Extracorporeal irradiation of an excised tumour-bearing segment of bone followed by its re-implantation is a technique used in bone sarcoma surgery for limb salvage when the bone is of reasonable quality. There is no agreement among previous studies about the dose of irradiation to be given: up to 300 Gy have been used.

We investigated the influence of extracorporeal irradiation on the elastic and viscoelastic properties of bone. Bone was harvested from mature cattle and subdivided into 13 groups: 12 were exposed to increasing levels of irradiation: one was not and was used as a control. The specimens, once irradiated, underwent mechanical testing in saline at 37°C.

The mechanical properties of each group, including Young’s modulus, storage modulus and loss modulus, were determined experimentally and compared with the control group.

There were insignificant changes in all of these mechanical properties with an increasing level of irradiation.

We conclude that the overall mechanical effect of high levels of extracorporeal irradiation (300 Gy) on bone is negligible. Consequently the dose can be maximised to reduce the risk of local tumour recurrence.

Cite this article: Bone Joint J 2015;97-B:1152–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 435 - 438
1 Apr 2011
Gilbody J

Aseptic loosening of the acetabular component continues to be the most common indication for revision of total hip replacements in younger patients. Early in the evolution of the cemented hip, arthroplasty surgeons switched from removal to retention of the acetabular subchondral bone plate, theorising that unfavourable mechanical forces were the cause of loosening at the bone-cement interface.

It is now known that the cause of aseptic loosening is probably biological rather than mechanical and removing the subchondral bone plate may enhance biological fixation of cement to bone. With this in mind, perhaps it is time to revive removal of the subchondral bone as a standard part of acetabular preparation.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 616 - 621
1 May 2011
O’Neill F Condon F McGloughlin T Lenehan B Coffey JC Walsh M

We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability.

We concluded that the DHS blade was the superior implant in this study.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1561 - 1565
1 Dec 2009
Blakey CM Eswaramoorthy VK Hamilton LC Biant LC Field RE

We report the minimum five-year follow-up of 352 primary total hip replacements using the uncemented hydroxyapatite-coated ANCA-Fit femoral component with a modular neck and head. The series comprised 319 patients (212 men, 107 women) with a mean age at operation of 64.4 years (28 to 97). The principal diagnosis was osteoarthritis. A total of 18 patients (21 hips) died before their follow-up at five years, nine patients (11 hips) were lost to follow-up, and four (four hips) declined further follow-up. Patient-reported outcomes have been recorded for 288 patients (316 hips).

Their mean Oxford Hip Score improved significantly from 41 points (16 to 57) pre-operatively to 20 points (12 to 44) at five-year follow-up. Radiological assessment showed good bony stability in 98% of implants. There were two cases of aseptic loosening of the femoral component. There were no clinical or radiological complications related to modularity. In our series we did not see the high rate of intra-operative fracture previously reported for this implant.

This medium-term follow-up study demonstrates that the clinical outcome of the ANCA-Fit femoral component is, to date, comparable with that of other metaphyseal loading femoral components.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 188 - 193
1 Feb 2011
Rudol G Wilcox R Jin Z Tsiridis E

The mechanical performance of the cement-in-cement interface in revision surgery has not been fully investigated. The quantitative effect posed by interstitial fluids and roughening of the primary mantle remains unclear. We have analysed the strength of the bilaminar cement-bone interface after exposure of the surface of the primary mantle to roughening and fluid interference. The end surfaces of cylindrical blocks of cement were machined smooth (Ra = 200 nm) or rough (Ra = 5 μm) and exposed to either different volumes of water and carboxymethylcellulose (a bone-marrow equivalent) or left dry. Secondary blocks were cast against the modelled surface. Monoblocks of cement were used as a control group. The porosity of the samples was investigated using micro-CT. Samples were exposed to a single shearing force to failure.

The mean failure load of the monoblock control was 5.63 kN (95% confidence interval (CI) 5.17 to 6.08) with an estimated shear strength of 36 MPa. When small volumes of any fluid or large volumes were used, the respective values fell between 4.66 kN and 4.84 kN with no significant difference irrespective of roughening (p > 0.05). Large volumes of carboxymethylcellulose significantly weakened the interface. Roughening in this group significantly increased the strength with failure loads of 2.80 kN (95% CI 2.37 to 3.21) compared with 0.86 kN (95% CI 0.43 to 1.27) in the smooth variant. Roughening of the primary mantle may not therefore be as crucial as has been previously thought in clinically relevant circumstances.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 76 - 82
1 Jan 2015
Siebachmeyer M Boddu K Bilal A Hester TW Hardwick T Fox TP Edmonds M Kavarthapu V

We report the outcomes of 20 patients (12 men, 8 women, 21 feet) with Charcot neuro-arthropathy who underwent correction of deformities of the ankle and hindfoot using retrograde intramedullary nail arthrodesis. The mean age of the patients was 62.6 years (46 to 83); their mean BMI was 32.7 (15 to 47) and their median American Society of Anaesthetists score was 3 (2 to 4). All presented with severe deformities and 15 had chronic ulceration. All were treated with reconstructive surgery and seven underwent simultaneous midfoot fusion using a bolt, locking plate or a combination of both. At a mean follow-up of 26 months (8 to 54), limb salvage was achieved in all patients and 12 patients (80%) with ulceration achieved healing and all but one patient regained independent mobilisation. There was failure of fixation with a broken nail requiring revision surgery in one patient. Migration of distal locking screws occurred only when standard screws had been used but not with hydroxyapatite-coated screws. The mean American Academy of Orthopaedic Surgeons Foot and Ankle (AAOS-FAO) score improved from 50.7 (17 to 88) to 65.2 (22 to 88), (p = 0.015). The mean Short Form (SF)-36 Health Survey Physical Component Score improved from 25.2 (16.4 to 42.8) to 29.8 (17.7 to 44.2), (p = 0.003) and the mean Euroqol EQ‑5D‑5L score improved from 0.63 (0.51 to 0.78) to 0.67 (0.57 to 0.84), (p = 0.012).

Single-stage correction of deformity using an intramedullary hindfoot arthrodesis nail is a good form of treatment for patients with severe Charcot hindfoot deformity, ulceration and instability provided a multidisciplinary care plan is delivered.

Cite this article: Bone Joint J 2015;97-B:76–82.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 489 - 495
1 Apr 2010
Ramaswamy R Kosashvili Y Cameron H

The hip joint is commonly involved in multiple epiphyseal dysplasia and patients may require total hip replacement before the age of 30 years.

We retrospectively reviewed nine patients (16 hips) from four families. The diagnosis of multiple epiphyseal dysplasia was based on a family history, genetic counselling, clinical features and radiological findings. The mean age at surgery was 32 years (17 to 63), with a mean follow-up of 15.9 years (5.5 to 24).

Of the 16 hips, ten required revision at a mean of 12.5 years (5 to 15) consisting of complete revision of the acetabular component in three hips and isolated exchange of the liner in seven. No femoral component has loosened or required revision during the period of follow-up.

With revision for any reason, the 15-year survival was only 11.4% (95% confidence interval 1.4 to 21.4). However, when considering revision of the acetabular shell in isolation the survival at ten years was 93.7% (95% confidence interval 87.7 to 99.7), reducing to 76.7% (95% confidence interval 87.7 to 98.7) at 15 and 20 years, respectively.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 259 - 263
1 Feb 2009
Dimmen S Nordsletten L Engebretsen L Steen H Madsen JE

Conventional non-steroidal anti-inflammatory drugs (NSAIDs) and newer specific cyclo-oxygenase-2 (cox-2) inhibitors are commonly used in musculoskeletal trauma and orthopaedic surgery to reduce the inflammatory response and pain. These drugs have been reported to impair bone metabolism. In reconstruction of the anterior cruciate ligament the hamstring tendons are mainly used as the graft of choice, and a prerequisite for good results is healing of the tendons in the bone tunnel. Many of these patients are routinely given NSAIDs or cox-2 inhibitors, although no studies have elucidated the effects of these drugs on tendon healing in the bone tunnel.

In our study 60 female Wistar rats were randomly allocated into three groups of 20. One received parecoxib, one indometacin and one acted as a control. In all the rats the tendo-Achillis was released proximally from the calf muscles. It was then pulled through a drill hole in the distal tibia and sutured anteriorly. The rats were given parecoxib, indometacin or saline intraperitoneally twice daily for seven days. After 14 days the tendon/bone-tunnel interface was subjected to mechanical testing.

Significantly lower maximum pull-out strength (p < 0.001), energy absorption (p < 0.001) and stiffness (p = 0.035) were found in rats given parecoxib and indometacin compared with the control group, most pronounced with parecoxib.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 82 - 87
1 Jan 2009
Charalambous CP Stanley JK Siddique I Aster A Gagey O

The lateral ligament complex is the primary constraint to posterolateral rotatory laxity of the elbow, and if it is disrupted during surgery, posterolateral instability may ensue. The Wrightington approach to the head of the radius involves osteotomising the ulnar insertion of this ligament, rather than incising through it as in the classic posterolateral (Kocher) approach. In this biomechanical study of 17 human cadaver elbows, we demonstrate that the surgical approach to the head can influence posterolateral laxity, with the Wrightington approach producing less posterolateral rotatory laxity than the posterolateral approach.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 870 - 876
1 Jul 2009
Kosashvili Y Backstein D Safir O Lakstein D Gross AE

Pelvic discontinuity with associated bone loss is a complex challenge in acetabular revision surgery. Reconstruction using ilio-ischial cages combined with trabecular metal acetabular components and morsellised bone (the component-cage technique) is a relatively new method of treatment.

We reviewed a consecutive series of 26 cases of acetabular revision reconstructions in 24 patients with pelvic discontinuity who had been treated by the component-cage technique. The mean follow-up was 44.6 months (24 to 68). Failure was defined as migration of a component of > 5 mm.

In 23 hips (88.5%) there was no clinical or radiological evidence of loosening at the last follow-up. The mean Harris hip score improved significantly from 46.6 points (29.5 to 68.5) to 76.6 points (55.5 to 92.0) at two years (p < 0.001). In three hips (11.5%) the construct had migrated at one year after operation. The complications included two dislocations, one infection and one partial palsy of the peroneal nerve.

Our findings indicate that treatment of pelvic discontinuity using the component-cage construct is a reliable option.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1424 - 1430
1 Nov 2009
Corten K Vanrykel F Bellemans J Frederix PR Simon J Broos PLO

The use of plate-and-cable constructs to treat periprosthetic fractures around a well-fixed femoral component in total hip replacements has been reported to have high rates of failure. Our aim was to evaluate the results of a surgical treatment algorithm to use these lateral constructs reliably in Vancouver type-B1 and type-C fractures. The joint was dislocated and the stability of the femoral component was meticulously evaluated in 45 type-B1 fractures. This led to the identification of nine (20%) unstable components. The fracture was considered to be suitable for single plate-and-cable fixation by a direct reduction technique if the integrity of the medial cortex could be restored.

Union was achieved in 29 of 30 fractures (97%) at a mean of 6.4 months (3 to 30) in 29 type-B1 and five type-C fractures. Three patients developed an infection and one construct failed.

Using this algorithm plate-and-cable constructs can be used safely, but indirect reduction with minimal soft-tissue damage could lead to shorter times to union and lower rates of complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 11 | Pages 1528 - 1532
1 Nov 2006
Pun SY Teng MS Kim HT

Desiccation of articular cartilage during surgery is often unavoidable and may result in the death of chondrocytes, with subsequent joint degeneration. This study was undertaken to determine the extent of chondrocyte death caused by exposure to air and to ascertain whether regular rewetting of cartilage could decrease cell death.

Macroscopically normal human cartilage was exposed to air for 0, 30, 60 or 120 minutes. Selected samples were wetted in lactated Ringer’s solution for ten seconds every ten or 20 minutes. The viability of chondrocytes was measured after three days by Live/Dead staining.

Chondrocyte death correlated with the length of exposure to air and the depth of the cartilage. Drying for 120 minutes caused extensive cell death mainly in the superficial 500 μm of cartilage. Rewetting every ten or 20 minutes significantly decreased cell death.

The superficial zone is most susceptible to desiccation. Loss of superficial chondrocytes likely decreases the production of essential lubricating glycoproteins and contributes to subsequent degeneration. Frequent wetting of cartilage during arthrotomy is therefore essential.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 426 - 433
1 Apr 2009
Musahl V Tarkin I Kobbe P Tzioupis C Siska PA Pape H

The operative treatment of displaced fractures of the tibial plateau is challenging. Recent developments in the techniques of internal fixation, including the development of locked plating and minimal invasive techniques have changed the treatment of these fractures. We review current surgical approaches and techniques, improved devices for internal fixation and the clinical outcome after utilisation of new methods for locked plating.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 291 - 295
1 Mar 2007
Langdown AJ Pickard RJ Hobbs CM Clarke HJ Dalton DJN Grover ML

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1396 - 1401
1 Oct 2007
Hirpara KM Sullivan PJ Raheem O O’Sullivan ME

We compared the bulking and tensile strength of the Pennington modified Kessler, Cruciate and the Savage repairs in an ex vivo model. A total of 60 porcine tendons were randomised to three groups, half repaired using a core suture alone and the remainder employing a core and peripheral technique. The tendons were distracted to failure. The force required to produce a 3 mm gap, the ultimate strength, the mode of failure and bulking for each repair were assessed. We found that there was a significant increase in strength without an increase in bulk as the number of strands increased. The Cruciate repair was significantly more likely to fail by suture pullout than the Pennington modified Kessler or Savage repairs. We advise the use of the Savage repair, especially in the thumb, and a Cruciate when a Savage is not possible. The Pennington modified Kessler repair should be reserved for multiple tendon injuries.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1334 - 1340
1 Oct 2008
Flavin R Halpin T O’Sullivan R FitzPatrick D Ivankovic A Stephens MM

Hallux rigidus was first described in 1887. Many aetiological factors have been postulated, but none has been supported by scientific evidence. We have examined the static and dynamic imbalances in the first metatarsophalangeal joint which we postulated could be the cause of this condition. We performed a finite-element analysis study on a male subject and calculated a mathematical model of the joint when subjected to both normal and abnormal physiological loads.

The results gave statistically significant evidence for an increase in tension of the plantar fascia as the cause of abnormal stress on the articular cartilage rather than mismatch of the articular surfaces or subclinical muscle contractures. Our study indicated a clinical potential cause of hallux rigidus and challenged the many aetiological theories. It could influence the choice of surgical procedure for the treatment of early grades of hallux rigidus.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 172 - 179
1 Feb 2008
Pinczewski LA Salmon LJ Jackson WFM von Bormann RBP Haslam PG Tashiro S

There is little evidence examining the relationship between anatomical landmarks, radiological placement of the tunnels and long-term clinical outcomes following anterior cruciate ligament (ACL) reconstruction. The aim of this study was to investigate the reproducibility of intra-operative landmarks for placement of the tunnels in single-bundle reconstruction of the ACL using four-strand hamstring tendon autografts.

Isolated reconstruction of the ACL was performed in 200 patients, who were followed prospectively for seven years with use of the International Knee Documentation Committee forms and radiographs. Taking 0% as the anterior and 100% as the posterior extent, the femoral tunnel was a mean of 86% (sd 5) along Blumensaat’s line and the tibial tunnel was 48% (sd 5) along the tibial plateau. Taking 0% as the medial and 100% as the lateral extent, the tibial tunnel was 46% (sd 3) across the tibial plateau and the mean inclination of the graft in the coronal plane was 19° (sd 5.5).

The use of intra-operative landmarks resulted in reproducible placement of the tunnels and an excellent clinical outcome seven years after operation. Vertical inclination was associated with increased rotational instability and degenerative radiological changes, while rupture of the graft was associated with posterior placement of the tibial tunnel. If the osseous tunnels are correctly placed, single-bundle reconstruction of the ACL adequately controls both anteroposterior and rotational instability.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 635 - 639
1 May 2005
Ikeuchi M Kawakami T Kitaoka K Okanoue Y Tani T

We describe a new technique of reconstruction of the deficient acetabulum in cementless total hip arthroplasty. The outer iliac table just above the deficient acetabulum is osteotomised and slid downwards. We have termed this an iliac sliding graft. Between October 1997 and November 2001, cementless total hip arthroplasty with an iliac sliding graft was performed on 19 patients (19 hips) with acetabular dysplasia. The mean follow-up was 3.4 years (2 to 6).

The mean pre-operative Harris hip score was 45.1 which improved significantly to 85.3 at the time of the final follow-up. No patient had post-operative abductor dysfunction. Incorporation of the graft was seen after two to three months in all patients. Resorption of the graft and radiolucencies were infrequent. This technique is a useful alternative to femoral head autografting when the patient’s own femoral head cannot be used.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1298 - 1302
1 Sep 2005
Iesaka K Jaffe WL Jones CM Kummer FJ

We have investigated the role of the penetration of saline on the shear strength of the cement-stem interface for stems inserted at room temperature and those preheated to 37°C using a variety of commercial bone cements. Immersion in saline for two weeks at 37°C reduced interfacial strength by 56% to 88% after insertion at room temperature and by 28% to 49% after preheating of the stem. The reduction in porosity as a result of preheating ranged from 71% to 100%. Increased porosity correlated with a reduction in shear strength after immersion in saline (r = 0.839, p < 0.01) indicating that interfacial porosity may act as a fluid conduit.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 697 - 707
1 Jun 2008
Fraser-Moodie JA Shortt NL Robinson CM

Injuries to the acromioclavicular joint are common but underdiagnosed. Sprains and minor subluxations are best managed conservatively, but there is debate concerning the treatment of complete dislocations and the more complex combined injuries in which other elements of the shoulder girdle are damaged. Confusion has been caused by existing systems for classification of these injuries, the plethora of available operative techniques and the lack of well-designed clinical trials comparing alternative methods of management. Recent advances in arthroscopic surgery have produced an even greater variety of surgical options for which, as yet, there are no objective data on outcome of high quality. We review the current concepts of the treatment of these injuries.