We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up. Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips).Aims
Patients and Methods
Our primary aim was to describe migration of the Exeter stem with a 32 mm head on highly crosslinked polyethylene and whether this is influenced by age. Our secondary aims were to assess functional outcome, satisfaction, activity, and bone mineral density (BMD) according to age. A prospective cohort study was conducted. Patients were recruited into three age groups: less than 65 years (n = 65), 65 to 74 years (n = 68), and 75 years and older (n = 67). There were 200 patients enrolled in the study, of whom 115 were female and 85 were male, with a mean age of 69.9 years (sd 9.5, 42 to 92). They were assessed preoperatively, and at three, 12 and, 24 months postoperatively. Stem migration was assessed using Einzel-Bild-Röntgen-Analyse (EBRA). Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Harris Hip Score (HHS), Hip Disability and Osteoarthritis Outcome Score (HOOS), EuroQol-5 domains questionnaire (EQ-5D), short form-36 questionnaire (SF-36,) and patient satisfaction were used to assess outcome. The Lower Extremity Activity Scale (LEAS), Timed Up and Go (TUG) test, and activPAL monitor (energy expelled, time lying/standing/walking and step count) were used to assess activity. The BMD was assessed in Gruen and Charnley zones.Objectives
Patients and Methods
The aim of this study was to analyze the effect of a lateral rim mesh on the survival of primary total hip arthroplasty (THA) in young patients, aged 50 years or younger. We compared a study group of 235 patients (257 hips) who received a primary THA with the use of impaction bone grafting (IBG) with an additional lateral rim mesh with a group of 306 patients (343 hips) who received IBG in the absence of a lateral rim mesh during the same period from 1988 to 2015. In the mesh group, there were 74 male and 183 female patients, with a mean age of 35 years (13 to 50). In the no-mesh group, there were 173 male and 170 female patients, with a mean age of 38 years (12.6 to 50). Cox regression analyses were performed to study the effect of a lateral rim mesh on acetabular component survival. Kaplan–Meier analyses with 95% confidence intervals (CIs) were performed to estimate the survival of the acetabular implant.Aims
Patients and Methods
Loosening is a well-known complication in the fixation of fractures using devices such as locking plates or unilateral fixators. It is believed that high strains in the bone at the bone-screw interface can initiate loosening, which can result in infection, and further loosening. Here, we present a new theory of loosening of implants. The time-dependent response of bone subjected to loads results in interfacial deformations in the bone which accumulate with cyclical loading and thus accentuates loosening. We used an ‘ideal’ bone-screw system, in which the screw is subjected to cyclical lateral loads and trabecular bone is modelled as non-linear viscoelastic and non-linear viscoelastic-viscoplastic material, based on recent experiments, which we conducted.Aims
Methods
Aims
Patients and Methods
The primary stability of the cementless Oxford Unicompartmental Knee Replacement (OUKR) relies on interference fit (or press fit). Insufficient interference may cause implant loosening, whilst excessive interference could cause bone damage and fracture. The aim of this study was to identify the optimal interference fit by measuring the force required to seat the tibial component of the cementless OUKR (push-in force) and the force required to remove the component (pull-out force). Six cementless OUKR tibial components were implanted in 12 new slots prepared on blocks of solid polyurethane foam (20 pounds per cubic foot (PCF), Sawbones, Malmo, Sweden) with a range of interference of 0.1 mm to 1.9 mm using a Dartec materials testing machine HC10 (Zwick Ltd, Herefordshire, United Kingdom) . The experiment was repeated with cellular polyurethane foam (15 PCF), which is a more porous analogue for trabecular bone.Objectives
Materials and Methods
Post-operative migration of cemented acetabular components as
measured by radiostereometric analysis (RSA) has a strong predictive
power for late, aseptic loosening. Also, radiolucent lines predict
late loosening. Migration has been reduced by systemic bisphosphonate
treatment in randomised trials of hip and knee arthroplasty. Used
as a local treatment, a higher local dose of bisphosphonate can
be achieved without systemic exposure. We wished to see if this
principle could be applied usefully in total hip arthroplasty (THA). In this randomised placebo-controlled, double-blinded trial with
60 participants, we compressed gauze soaked in bisphosphonate solution
(ibandronate) or saline against the acetabular bone bed immediately
before cementing the acetabular component. RSA, classification of
radiolucent lines, the Harris Hip Score (HHS) and the Western Ontario McMasters
Universities Osteoarthritis Index (WOMAC) were carried out at three-,
six-, 12-, and 24-month follow-up.Aims
Patients and Methods
The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA). Hydroxyapatite coatings have been
applied to improve bone ingrowth in uncemented designs, but may
only coat the directly accessible surface. As peri-apatite (PA)
is solution deposited, this may increase the coverage of the implant
surface and thereby fixation. We assessed the tibial component fixation
of uncemented PA-coated TKAs Patients were randomised to PA-coated or cemented TKAs. In 60
patients (30 in each group), radiostereometric analysis of tibial
component migration was evaluated as the primary outcome at baseline,
three months post-operatively and at one, two and five years. A
linear mixed-effects model was used to analyse the repeated measurements.Aims
Patients and Methods
Reconstruction of the acetabulum after failed total hip arthroplasty
(THA) can be a surgical challenge in the presence of severe bone
loss. We report the long-term survival of a porous tantalum revision
acetabular component, its radiological appearance and quality of
life outcomes. We reviewed the results of 46 patients who had undergone revision
of a failed acetabular component with a Paprosky II or III bone
defect and reconstruction with a hemispherical, tantalum acetabular
component, supplementary screws and a cemented polyethylene liner.Aims
Patients and Methods
Promising medium-term results from total shoulder arthroplasty
(TSA) have been reported for the treatment of primary osteoarthritis
in young and middle-aged patients. The aim of this study was to
evaluate the long-term functional and radiological outcome of TSA
in the middle-aged patient. The data of all patients from the previous medium-term study
were available. At a mean follow-up of 13 years (8 to 17), we reviewed
21 patients (12 men, nine women, 21 shoulders) with a mean age of
55 years (37 to 60). The Constant-Murley score (CS) with its subgroups
and subjective satisfaction were measured. Radiological signs of implant
loosening were analysed.Aims
Patients and Methods
A contact patch to rim (CPR) distance of <
10 mm has been
associated with edge-loading and excessive wear. However, not all
arthroplasties with a low CPR distance show problems with wear.
Therefore, CPR distance may not be the only variable affecting the
post-operative metal ion concentrations. We used multiple logistic regression to determine what variables
differed between the patients who had high and low cobalt (CoS)
and chromium (CrS) serum ion concentrations within a cohort of patients
with low (<
10 mm) CPR distances. A total of 56 patients treated
with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS
ion studies performed more than one year after surgery. The mean
age of the patients at the time of surgery was 51.7 years (29 to
70), with 38 women (68%) and 18 men (32%).Aims
Patients and Methods
The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated tapered short stem femoral component and
53 patients with a standard conventional femoral component, at a
minimum follow-up of two years. The mean follow-up was 31.4 months
(24 to 39). Patients with poor proximal femoral bone quality were
excluded. The mean age of the patients was 66.6 years (59 to 77)
and the mean body mass index was 30.2 kg/m2 (24.1 to
41.0). Outcome was assessed using the Oxford Hip Score (OHS) and
the University College Hospital (UCH) hip score which is a validated
performance based instrument. HRQoL was assessed using the EuroQol
5D (EQ-5D).Aims
Patients and Methods
The aim of this study was to compare the survivorship and radiographic
outcomes at ten-year follow-up of three prospective consecutive
series of patients each of which received a different design of
cementless femoral components for total hip arthroplasty (THA). In Cohort 1, 91 consecutive patients (100 hips) underwent THA
with a cementless porous-coated anatomic femoral stem (PCA) between
October 1983 and January 1986. In Cohort 2, 86 consecutive patients
(100 hips) underwent THA with an extensively porous-coated cementless
femoral stem (Prodigy) between June 1994 and October 1997. In Cohort
3, 88 consecutive patients (100 hips) underwent THA with a proximally
porous-coated triple-tapered cementless stem (Summit) between April
2002 and October 2003. All three groups underwent prospective clinical
and radiographic evaluation.Aims
Patients and Methods
This review summarises the technique of impaction
grafting with mesh augmentation for the treatment of uncontained
acetabular defects in revision hip arthroplasty. The ideal acetabular revision should restore bone stock, use
a small socket in the near-anatomic position, and provide durable
fixation. Impaction bone grafting, which has been in use for over
40 years, offers the ability to achieve these goals in uncontained
defects. The precepts of modern, revision impaction grafting are
that the segmental or cavitary defects must be supported with a
mesh; the contained cavity is filled with vigorously impacted morselised
fresh-frozen allograft; and finally, acrylic cement is used to stabilise
the graft and provide rigid, long-lasting fixation of the revised
acetabular component. Favourable results have been published with this technique. While
having its limitations, it is a viable option to address large acetabular
defects in revision arthroplasty. Cite this article:
The purpose of this study was to determine the sensitivity, specificity
and predictive values of previously reported thresholds of proximal
translation and sagittal rotation of cementless acetabular components
used for revision total hip arthroplasty (THA) at various times
during early follow-up. Migration of cementless acetabular components was measured retrospectively
in 84 patients (94 components) using Ein-Bild-Rontgen-Analyse (EBRA-Cup)
in two groups of patients. In Group A, components were recorded
as not being loose intra-operatively at re-revision THA (52 components/48
patients) and Group B components were recorded to be loose at re-revision
(42 components/36 patients).Aims
Patients and Methods