Despite the worldwide usage of the cemented Contemporary
acetabular component (Stryker), no published data are available
regarding its use in patients aged <
50 years. We undertook a
mid- to long-term follow-up study, including all consecutive patients
aged
<
50 years who underwent a primary total hip replacement using
the Contemporary acetabular component with the Exeter cemented stem
between January 1999 and January 2006. There were 152 hips in 126
patients, 61 men and 65 women, mean age at surgery 37.6 years (16
to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to
12.0). All clinical questionnaire scores, including Harris hip score,
Oxford hip score and several visual analogue scales, were found
to have improved. The eight year survivorship of all acetabular
components for the endpoints revision for any reason or revision
for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2
to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up
was complete for 146 implants. The eight year survival for the endpoint
radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving
implants were considered radiologically loose but were asymptomatic.
The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent
lines (n = 20, 13.9%) in the 144 surviving cups indicates a need
for continued observation in the second decade of follow-up in order
to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival
rate >
90% in young patients support the use of the Contemporary
acetabular component in this specific patient group. Cite this article:
The purpose of this study was to evaluate the
change in sagittal tibiotalar alignment after total ankle arthroplasty (TAA)
for osteoarthritis and to investigate factors affecting the restoration
of alignment. This retrospective study included 119 patients (120 ankles) who
underwent three component TAA using the Hintegra prosthesis. A total
of 63 ankles had anterior displacement of the talus before surgery
(group A), 49 had alignment in the normal range (group B), and eight
had posterior displacement of the talus (group C). Ankles in group
A were further sub-divided into those in whom normal alignment was
restored following TAA (41 ankles) and those with persistent displacement
(22 ankles). Radiographic and clinical results were assessed. Pre-operatively, the alignment in group A was significantly more
varus than that in group B, and the posterior slope of the tibial
plafond was greater (p <
0.01 in both cases). The posterior slope
of the tibial component was strongly associated with restoration
of alignment: ankles in which the alignment was restored had significantly
less posterior slope (p <
0.001). An anteriorly translated talus was restored to a normal position
after TAA in most patients. We suggest that surgeons performing
TAA using the Hintegra prosthesis should aim to insert the tibial
component at close to 90° relative to the axis of the tibia, hence
reducing posterior soft-tissue tension and allowing restoration
of normal tibiotalar alignment following surgery. Cite this article:
Instability in flexion after total knee replacement
(TKR) typically occurs as a result of mismatched flexion and extension
gaps. The goals of this study were to identify factors leading to
instability in flexion, the degree of correction, determined radiologically,
required at revision surgery, and the subsequent clinical outcomes.
Between 2000 and 2010, 60 TKRs in 60 patients underwent revision
for instability in flexion associated with well-fixed components.
There were 33 women (55%) and 27 men (45%); their mean age was 65
years (43 to 82). Radiological measurements and the Knee Society
score (KSS) were used to assess outcome after revision surgery.
The mean follow-up was 3.6 years (2 to 9.8). Decreased condylar
offset (p <
0.001), distalisation of the joint line (p <
0.001)
and increased posterior tibial slope (p <
0.001) contributed
to instability in flexion and required correction at revision to regain
stability. The combined mean correction of posterior condylar offset
and joint line resection was 9.5 mm, and a mean of 5° of posterior
tibial slope was removed. At the most recent follow-up, there was
a significant improvement in the mean KSS for the knee and function
(both p <
0.001), no patient reported instability and no patient
underwent further surgery for instability. The following step-wise approach is recommended: reduction of
tibial slope, correction of malalignment, and improvement of condylar
offset. Additional joint line elevation is needed if the above steps
do not equalise the flexion and extension gaps. Cite this article:
We undertook a randomised controlled trial to
compare bipolar hemiarthroplasty (HA) with a novel total hip replacement
(THR) comprising a polycarbonate–urethane (PCU) acetabular component
coupled with a large-diameter metal femoral head for the treatment
of displaced fractures of the femoral neck in elderly patients. Functional
outcome, assessed with the Harris hip score (HHS) at three months
and then annually after surgery, was the primary endpoint. Rates
of revision and complication were secondary endpoints. Based on a power analysis, 96 consecutive patients aged >
70
years were randomised to receive either HA (49) or a PCU-THR (47).
The mean follow-up was 30.1 months (23 to 50) and 28.6 months (22
to 52) for the HA and the PCU group, respectively. The HHS showed no statistically significant difference between
the groups at every follow-up. Higher pain was recorded in the PCU
group at one and two years’ follow-up
(p = 0.006 and p = 0.019, respectively). In the HA group no revision
was performed. In the PCU-THR group six patients underwent revision
and one patient is currently awaiting
re-operation. The three-year survival rate of the PCU-THR group
was 0.841 (95% confidence interval 0.680 to 0.926). Based on our findings we do not recommend the use of the PCU
acetabular component as part of the treatment of patients with fractures
of the femoral neck. Cite this article:
Dysplasia of the hip, hypotonia, osteopenia,
ligamentous laxity, and mental retardation increase the complexity
of performing and managing patients with Down syndrome who require
total hip replacement (THR). We identified 14 patients (six males,
eight females, 21 hips) with Down syndrome and degenerative disease
of the hip who underwent THR, with a minimum follow-up of two years
from 1969 to 2009. In seven patients, bilateral THRs were performed
while the rest had unilateral THRs. The mean clinical follow-up
was 5.8 years (standard deviation ( Cite this article:
Symptomatic cobalt toxicity from a failed total
hip replacement is a rare but devastating complication. It has been reported
following revision of fractured ceramic components, as well as in
patients with failed metal-on-metal articulations. Potential clinical
findings include fatigue, weakness, hypothyroidism, cardiomyopathy,
polycythaemia, visual and hearing impairment, cognitive dysfunction,
and neuropathy. We report a case of an otherwise healthy 46-year-old
patient, who developed progressively worsening symptoms of cobalt
toxicity beginning approximately six months following synovectomy
and revision of a fractured ceramic-on-ceramic total hip replacement
to a metal-on-polyethylene bearing. The whole blood cobalt levels
peaked at 6521 µg/l. The patient died from cobalt-induced cardiomyopathy.
Implant retrieval analysis confirmed a loss of 28.3 g mass of the
cobalt–chromium femoral head as a result of severe abrasive wear
by ceramic particles embedded in the revision polyethylene liner.
Autopsy findings were consistent with heavy metal-induced cardiomyopathy. We recommend using new ceramics at revision to minimise the risk
of wear-related cobalt toxicity following breakage of ceramic components. Cite this article:
In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any adverse effects for the patients. Cite this article:
Hip replacement is a very successful operation and the outcome is usually excellent. There are recognised complications that seem increasingly to give rise to litigation. This paper briefly examines some common scenarios where litigation may be pursued against hip surgeons. With appropriate record keeping, consenting and surgical care, the claim can be successfully defended if not avoided. We hope this short summary will help to highlight some common pitfalls. There is extensive literature available for detailed study.
Previous classification systems of failure of
limb salvage focused primarily on endoprosthetic failures and lacked sufficient
depth for the effective study of the causes of failure. In order
to address these inadequacies, the International Society of Limb
Salvage (ISOLS) formed a committee to recommend revisions of the
previous systems. The purpose of this study was to report on their
recommendations. The modifications were prepared using an earlier,
evidence-based model with subclassification based on the existing
medical literature. Subclassification for all five primary types
of failure of limb salvage following endoprosthetic reconstruction
were formulated and a complementary system was derived for the failure
of biological reconstruction. An additional classification of failure
in paediatric patients was also described. Limb salvage surgery presents a complex array of potential mechanisms
of failure, and a complete and precise classification of types of
failure is required. Earlier classification systems lacked specificity,
and the evidence-based system outlined here is designed to correct
these weaknesses and to provide a means of reporting failures of
limb salvage in order to allow the interpretation of outcome following
reconstructive surgery. Cite this article:
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article:
Total knee arthroplasty (TKA) and total hip arthroplasty
(THA) are recognised and proven interventions for patients with
advanced arthritis. Studies to date have demonstrated a steady increase
in the requirement for primary and revision procedures. Projected
estimates made for the United States show that by 2030 the demand
for primary TKA will grow by 673% and for revision TKA by 601% from
the level in 2005. For THA the projected estimates are 174% and
137% for primary and revision surgery, respectively. The purpose
of this study was to see if those predictions were similar for England
and Wales using data from the National Joint Registry and the Office
of National Statistics. Analysis of data for England and Wales suggest that by 2030,
the volume of primary and revision TKAs will have increased by 117%
and 332%, respectively between 2012 and 2030. The data for the United
States translates to a 306% cumulative rate of increase between
2012 and 2030 for revision surgery, which is similar to our predictions
for England and Wales. The predictions from the United States for primary TKA were similar
to our upper limit projections. For THA, we predicted an increase
of 134% and 31% for primary and revision hip surgery, respectively. Our model has limitations, however, it highlights the economic
burden of arthroplasty in the future in England and Wales as a real
and unaddressed problem. This will have significant implications
for the provision of health care and the management of orthopaedic
services in the future. Cite this article:
Because there have been no standard methods to determine pre-operatively
the thickness of resection of the proximal tibia in unicompartmental
knee arthroplasty (UKA), information about the relationship between
the change of limb alignment and the joint line elevation would
be useful for pre-operative planning. The purpose of this study
was to clarify the correlation between the change of limb alignment
and the change of joint line height at the medial compartment after
UKA. A consecutive series of 42 medial UKAs was reviewed retrospectively.
These patients were assessed radiographically both pre- and post-operatively
with standing anteroposterior radiographs. The thickness of bone
resection at the proximal tibia and the distal femur was measured
radiographically. The relationship between the change of femorotibial
angle (δFTA) and the change of joint line height, was analysed.Objectives
Methods
This study compares the cost-effectiveness of
treating dorsally displaced distal radial fractures with a volar
locking plate and percutaneous fixation. It was performed from the
perspective of the National Health Service (NHS) using data from
a single-centre randomised controlled trial. In total 130 patients
(18 to 73 years of age) with a dorsally displaced distal radial
fracture were randomised to treatment with either a volar locking
plate (n = 66) or percutaneous fixation (n = 64). The methodology
was according to National Institute for Health and Care Excellence guidance
for technology appraisals. . There were no significant differences in quality of life scores
between groups at any time point in the study. Both groups returned
to baseline one year post-operatively. NHS costs for the plate group were significantly higher (p <
0.001, 95% confidence interval 497 to 930). For an additional £713,
fixation with a volar locking plate offered 0.0178 additional quality-adjusted
life years in the year after surgery. The incremental cost-effectiveness
ratio (ICER) for plate fixation relative to percutaneous fixation
at list price was £40 068. When adjusting the prices of the implants
for a 20% hospital discount, the ICER was £31 898. Patients who
underwent plate fixation did not return to work earlier. We found no evidence to support the cost-effectiveness, from
the perspective of the NHS, of fixation using a volar locking plate
over percutaneous fixation for the operative treatment of a dorsally
displaced radial fracture. Cite this article:
Symptomatic hip osteonecrosis is a disabling
condition with a poorly understood aetiology and pathogenesis. Numerous
treatment options for hip osteonecrosis are described, which include
non-operative management and joint preserving procedures, as well
as total hip replacement (THR). Non-operative or joint preserving
treatment may improve outcomes when an early diagnosis is made before
the lesion has become too large or there is radiographic evidence
of femoral head collapse. The presence of a crescent sign, femoral
head flattening, and acetabular involvement indicate a more advanced-stage
disease in which joint preserving options are less effective than
THR. Since many patients present after disease progression, primary
THR is often the only reliable treatment option available. Prior
to the 1990s, outcomes of THR for osteonecrosis were poor. However,
according to recent reports and systemic reviews, it is encouraging
that with the introduction of newer ceramic and/or highly cross-linked
polyethylene bearings as well as highly-porous fixation interfaces,
THR appears to be a reliable option in the management of end-stage
arthritis following hip osteonecrosis in this historically difficult
to treat patient population. Cite this article:
Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium, tapered, conical stem for PFF in a series
of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of
27 patients had an American Society of Anesthesiologists grade of
at least 3. At a mean follow-up of 35 months (4 to 66) the mean
Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly
associated with a poorer OHS. All fractures united and no stem needed
to be revised. Three hips in three patients required further surgery
for infection, recurrent PFF and recurrent dislocation and three
other patients required closed manipulation for a single dislocation.
One stem subsided more than 5 mm but then stabilised and required
no further intervention. In this series, a modular, tapered, conical stem provided a versatile
reconstruction solution with a low rate of complications. Cite this article:
PCL retaining fixed-bearing TKA is a highly successful
operation with the need for more surgery occurring at the rate of
approximately 0.4% per year over the first 27 years. The most common
cause for revision surgery is related to polyethylene insert failure
and accounts for approximately 50% of re-operations. Late metastatic
infection is the next most frequent cause followed by patellar problems,
late instability and component loosening in decreasing frequency.
A myriad of rare miscellaneous problems can also occur.
The aim of this study was to analyse the gait
pattern, muscle force and functional outcome of patients who had undergone
replacement of the proximal tibia for tumour and alloplastic reconstruction
of the extensor mechanism using the patellar-loop technique. Between February 1998 and December 2009, we carried out wide
local excision of a primary sarcoma of the proximal tibia, proximal
tibial replacement and reconstruction of the extensor mechanism
using the patellar-loop technique in 18 patients. Of these, nine
were available for evaluation after a mean of 11.6 years (0.5 to
21.6). The strength of the knee extensors was measured using an
Isobex machine and gait analysis was undertaken in our gait assessment
laboratory. Functional outcome was assessed using the American Knee
Society (AKS) and Musculoskeletal Tumor Society (MSTS) scores. The gait pattern of the patients differed in ground contact time,
flexion heel strike, maximal flexion loading response and total
sagittal plane excursion. The mean maximum active flexion was 91°
(30° to 110°). The overall mean extensor lag was 1° (0° to 5°).
The mean extensor muscle strength was 25.8% (8.3% to 90.3%) of that
in the non-operated leg (p <
0.001). The mean functional scores
were 68.7% (43.4% to 83.3%) (MSTS) and 71.1 (30 to 90) (AKS functional
score). In summary, the results show that reconstruction of the extensor
mechanism using this technique gives good biomechanical and functional
results. The patients’ gait pattern is close to normal, except for
a somewhat stiff knee gait pattern. The strength of the extensor
mechanism is reduced, but sufficient for walking. Cite this article:
There is no consensus on the benefit of arthroscopically
assisted reduction of the articular surface combined with fixation
using a volar locking plate for the treatment of intra-articular
distal radial fractures. In this study we compared the functional
and radiographic outcomes of fluoroscopically and arthroscopically
guided reduction of these fractures. Between February 2009 and May 2013, 74 patients with unilateral
unstable intra-articular distal radial fractures were randomised
equally into the two groups for treatment. The mean age of these
74 patients was 64 years (24 to 92). We compared functional outcomes
including active range of movement of the wrist, grip strength and Disabilities
of the Arm, Shoulder, and Hand scores at six and 48 weeks; and radiographic
outcomes that included gap, step, radial inclination, volar angulation
and ulnar variance. There were no significant differences between the techniques
with regard to functional outcomes or radiographic parameters. The
mean gap and step in the fluoroscopic and arthroscopic groups were
comparable at 0.9 mm (standard deviation Arthroscopic reduction conferred no advantage over conventional
fluoroscopic guidance in achieving anatomical reduction of intra-articular
distal radial fractures when using a volar locking plate. Cite this article:
The August 2014 Shoulder &
Elbow Roundup360 looks at: Myofibroblasts perhaps not implicated in post-traumatic elbow stiffness; olecranon tip biomechanically sound for coranoid reconstruction; obesity and elbow replacement don’t mix; single column plating successful for extra-articular distal humeral fractures; satisfaction not predictable in frozen shoulder; tenodesis and repair both acceptable in Grade II SLAP tears; glenoid bone grafting is effective and glenohumeral articular lesions best seen with an arthroscope.
We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article: