We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
The April 2015 Hip &
Pelvis Roundup360 looks at: Goal-directed fluid therapy in hip fracture; Liberal blood transfusion no benefit in the longer term; Repeated measures: increased accuracy or compounded errors?; Peri-acetabular osteotomy safer than perhaps thought?; Obesity and peri-acetabular osteotomy: poor bedfellows; Stress fracture in peri-acetabular osteotomy; Infection and tantalum implants; Highly crosslinked polyethylene really does work
Ideal placement of the acetabular component remains
elusive both in terms of defining and achieving a target. Our aim
is to help restore original anatomy by using the transverse acetabular
ligament (TAL) to control the height, depth and version of the component.
In the normal hip the TAL and labrum extend beyond the equator of
the femoral head and therefore, if the definitive acetabular component
is positioned such that it is cradled by and just deep to the plane
of the TAL and labrum and is no more than 4mm larger than the original
femoral head, the centre of the hip should be restored. If the face
of the component is positioned parallel to the TAL and psoas groove
the patient specific version should be restored. We still use the
TAL for controlling version in the dysplastic hip because we believe
that the TAL and labrum compensate for any underlying bony abnormality. The TAL should not be used as an aid to inclination. Worldwide,
>
75% of surgeons operate with the patient in the lateral decubitus
position and we have shown that errors in post-operative radiographic
inclination (RI) of >
50° are generally caused by errors in patient positioning.
Consequently, great care needs to be taken when positioning the
patient. We also recommend 35° of apparent operative inclination
(AOI) during surgery, as opposed to the traditional 45°. Cite this article:
The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded.Aims
Patients and Methods
The June 2015 Foot &
Ankle Roundup360 looks at: Syndesmosis and outcomes in ankle fracture; Ankle arthrodesis or arthroplasty: a complications-based analysis; Crosslinked polyethylene and ankle arthroplasty; Reducing screw removal in calcaneal osteotomies; Revisiting infection control policies; Chevron osteotomy: proximal or distal?; Ankle distraction for osteoarthritis
The purpose of this study was to determine the lifetime risk
of revision surgery for patients undergoing Charnley cemented total
hip arthroplasty (THA), with 40-year follow up, using death as a
competing risk. We retrospectively reviewed 2000 cemented Charnley THAs, with
51 living hips available at 40 years. Aims
Materials and Methods
Partial knee arthroplasty (PKA), either medial
or lateral unicompartmental knee artroplasty (UKA) or patellofemoral arthroplasty
(PFA) are a good option in suitable patients and have the advantages
of reduced operative trauma, preservation of both cruciate ligaments
and bone stock, and restoration of normal kinematics within the
knee joint. However, questions remain concerning long-term survival.
The goal of this review article was to present the long-term results
of medial and lateral UKA, PFA and combined compartmental arthroplasty
for multicompartmental disease. Medium- and long-term studies suggest
reasonable outcomes at ten years with survival greater than 95% in
UKA performed for medial osteoarthritis or osteonecrosis, and similarly
for lateral Cite this article:
Malrotation of the femoral component can result in post-operative complications in total knee arthroplasty (TKA), including patellar maltracking. Therefore, we used computational simulation to investigate the influence of femoral malrotation on contact stresses on the polyethylene (PE) insert and on the patellar button as well as on the forces on the collateral ligaments. Validated finite element (FE) models, for internal and external malrotations from 0° to 10° with regard to the neutral position, were developed to evaluate the effect of malrotation on the femoral component in TKA. Femoral malrotation in TKA on the knee joint was simulated in walking stance-phase gait and squat loading conditions.Objectives
Materials and Methods
This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for peri-prosthetic fracture after
21 years, and one patient (one hip) was lost to follow-up. Using
aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular
revision but none needed femoral revision, demonstrating 100% survival
(95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral
component gives excellent long-term results. Cite this article:
Posterior tilt of the pelvis with sitting provides biological
acetabular opening. Our goal was to study the post-operative interaction
of skeletal mobility and sagittal acetabular component position. This was a radiographic study of 160 hips (151 patients) who
prospectively had lateral spinopelvic hip radiographs for skeletal
and implant measurements. Intra-operative acetabular component position
was determined according to the pre-operative spinal mobility. Sagittal
implant measurements of ante-inclination and sacral acetabular angle were
used as surrogate measurements for the risk of impingement, and
intra-operative acetabular component angles were compared with these.Aims
Materials and Methods
Risk of revision following total knee arthroplasty (TKA) is higher
in patients under 55 years, but little data are reported regarding
non-revision outcomes. This study aims to identify predictors of
dissatisfaction in these patients. We prospectively assessed 177 TKAs (157 consecutive patients,
99 women, mean age 50 years; 17 to 54) from 2008 to 2013. Age, gender,
implant, indication, body mass index (BMI), social deprivation,
range of movement, Kellgren-Lawrence (KL) grade of osteoarthritis
(OA) and prior knee surgery were recorded. Pre- and post-operative
Oxford Knee Score (OKS) as well as Short Form-12 physical (PCS)
and mental component scores were obtained. Post-operative range
of movement, complications and satisfaction were measured at one
year.Aims
Patients and Methods
The aim of this study was to investigate differences in pain,
range of movement function and satisfaction at three months and
one year after total knee arthroplasty (TKA) in patients with an
oblique pattern of kinematic graph of the knee and those with a
varus pattern. A total of 91 patients who underwent TKA were included in this
retrospective study. Patients (59 women and 32 men with mean age
of 68.7 years; 38.6 to 88.4) were grouped according to kinematic
graphs which were generated during navigated TKA and the outcomes
between the groups were compared.Aims
Patients and Methods
Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset femoral stems, and occurs within a central concentric
zone of the femoral head bore taper. Further investigation is required
to determine the effect of increased head size, and variations in
head–neck taper design. Cite this article:
The June 2015 Knee Roundup360 looks at: Cruciate substituting
The aim of this to study was to compare the previously unreported
long-term survival outcome of the Oxford medial unicompartmental
knee arthroplasty (UKA) performed by trainee surgeons and consultants. We therefore identified a previously unreported cohort of 1084
knees in 947 patients who had a UKA inserted for anteromedial knee
arthritis by consultants and surgeons in training, at a tertiary
arthroplasty centre and performed survival analysis on the group
with revision as the endpoint.Aims
Patients and Methods
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article: