The aim of this study was to obtain detailed long-term data on the cement-bone interface in patients with cemented stems, implanted using the constrained fixation technique. A total of eight stems were removed together with adjacent bone during Our results confirm that a complete cement mantle is not essential for the survival of Müller straight stems into the mid term, and support our hypothesis that no benefit to long-term survival can be expected from modern cementing techniques.
We report ten-year clinical and radiological
follow-up data for the Sigma Press Fit Condylar total knee replacement system
(Sigma PFC TKR). Between October 1998 and October 1999 a total of
235 consecutive PFC Sigma TKRs were carried out in 203 patients.
Patients were seen at a specialist nurse-led clinic seven to ten
days before admission and at six and 18 months, three, five and
eight to ten years after surgery. Data were recorded prospectively
at each clinic visit. Radiographs were obtained at the five- and
eight- to ten-year follow-up appointments. Of the 203 patients,
147 (171 knees) were alive at ten years and 12 were lost to follow-up.
A total of eight knees (3.4%) were revised, five for infection and
three to change the polyethylene insert. The survival at ten years
with an endpoint of revision for any reason was 95.9%, and with
an endpoint of revision for aseptic failure was 98.7%. The mean
American Knee Society Score (AKSS) was 79 (10 to 99) at eight to
ten years, compared with 31 (2 to 62) pre-operatively. Of 109 knee
with radiographs reviewed, 47 knees had radiolucent lines but none
showed evidence of loosening. Cite this article:
Data on early morbidity and complications after
revision total hip replacement (THR) are limited. The aim of this nationwide
study was to describe and quantify early morbidity after aseptic
revision THR and relate the morbidity to the extent of the revision
surgical procedure. We analysed all aseptic revision THRs from 1st
October 2009 to 30th September 2011 using the Danish National Patient
Registry, with additional information from the Danish Hip Arthroplasty
Registry. There were 1553 procedures (1490 patients) performed in
40 centres and we divided them into total revisions, acetabular
component revisions, femoral stem revisions and partial revisions.
The mean age of the patients was 70.4 years (25 to 98) and the median
hospital stay was five days (interquartile range 3 to 7). Within 90
days of surgery, the readmission rate was 18.3%, mortality rate
1.4%, re-operation rate 6.1%, dislocation rate 7.0% and infection
rate 3.0%. There were no differences in these outcomes between high-
and low-volume centres. Of all readmissions, 255 (63.9%) were due
to ‘surgical’ complications Cite this article:
There is a North Atlantic divide, with cementless
femoral stems being used more frequently in the USA and cemented
stems being used more frequently in many countries in Europe. This
is primarily because different cemented stems have been used on
different sides of the Atlantic and the results of the cemented
stems in the US have often been poor, whereas the results of the
stems used in Europe have been good. In the National registers in
Europe, cemented stems have tended to achieve better results than
cementless. Cite this article:
Intact abductors of the hip play a crucial role
in preventing limping and are known to be damaged through the direct lateral
approach. The extent of trauma to the abductors after revision total
hip replacement (THR) is unknown. The aim of this prospective study
was to compare the pre- and post-operative status of the gluteus
medius muscle after revision THR. We prospectively compared changes
in the muscle and limping in 30 patients who were awaiting aseptic
revision THR and 15 patients undergoing primary THR. The direct
lateral approach as described by Hardinge was used for all patients.
MRI scans of the gluteus medius and functional analyses were recorded
pre-operatively and six months post-operatively. The overall mean
fatty degeneration of the gluteus medius increased from 35.8% (1.1
to 98.8) pre-operatively to 41% (1.5 to 99.8) after multiple revision
THRs (p = 0.03). There was a similar pattern after primary THR,
but with considerably less muscle damage (p = 0.001), indicating
progressive muscle damage. Despite an increased incidence of a positive
Trendelenburg sign following revision surgery (p = 0.03) there was
no relationship between the cumulative fatty degeneration in the
gluteus medius and a positive Trendelenburg sign (p = 0.26). The
changes associated with other surgical approaches to the hip warrant
investigation. Cite this article:
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
Plots are an elegant and effective way to represent
data. At their best they encourage the reader and promote comprehension.
A graphical representation can give a far more intuitive feel to
the pattern of results in the study than a list of numerical data,
or the result of a statistical calculation. The temptation to exaggerate differences or relationships between
variables by using broken axes, overlaid axes, or inconsistent scaling
between plots should be avoided. A plot should be self-explanatory and not complicated. It should
make good use of the available space. The axes should be scaled
appropriately and labelled with an appropriate dimension. Plots are recognised statistical methods of presenting data and
usually require specialised statistical software to create them.
The statistical analysis and methods to generate the plots are as
important as the methodology of the study itself. The software,
including dates and version numbers, as well as statistical tests
should be appropriately referenced. Following some of the guidance provided in this article will
enhance a manuscript. Cite this article:
The February 2014 Shoulder &
Elbow Roundup360 looks at: whether arthroscopic acromioplasty is a cost-effective intervention; shockwave therapy in cuff tear; whether microfracture relieves short-term pain in cuff repair; the promising early results from L-PRF augmented cuff repairs; rehabilitation following cuff repair; supination strength following biceps tendon rupture; whether longer is better in humeral components; fatty degeneration in a rodent model; and the controversial acromioclavicular joint dislocation.
Tapered, fluted, modular, titanium stems have
a long history in Europe and are increasing in popularity in North America.
We have reviewed the results at our institution looking at stem
survival and clinical outcomes. Radiological outcomes and quality
of life assessments have been performed and compared to cylindrical
non-modular cobalt chromium stems. Survival at five years was 94%.
This fell to 85% at ten years due to stem breakage with older designs.
Review of radiology showed maintenance or improvement of bone stock
in 87% of cases. Outcome scores were superior in tapered stems despite
worse pre-operative femoral deficiency. Tapered stems have proved
to be a useful alternative in revision total hip arthroplasty across
the spectrum of femoral bone deficiency.
We report the clinical and radiological outcome of total ankle replacement performed in conjunction with hindfoot fusion or in isolation. Between May 2003 and June 2008, 60 ankles were treated with total ankle replacement with either subtalar or triple fusion, and the results were compared with a control group of 288 ankles treated with total ankle replacement alone. After the mean follow-up of 39.5 months (12 to 73), the ankles with hindfoot fusion showed significant improvement in the mean visual analogue score for pain (p <
0.001), the mean American Orthopaedic Foot and Ankle Society score (p <
0.001), and the mean of a modified version of this score (p <
0.001). The mean visual analogue pain score (p = 0.304) and mean modified American Orthopaedic Foot and Ankle Society score (p = 0.119) were not significantly different between the hindfoot fusion and the control groups. However, the hindfoot fusion group had a significantly lower mean range of movement (p = 0.009) and a higher rate of posterior focal osteolysis (p = 0.04). Both groups showed various complications (p = 0.131) and failure occurring at a similar rate (p = 0.685). Subtalar or triple fusion is feasible and has minimal adverse effects on ankles treated with total ankle replacement up to midterm follow-up. The clinical outcome of total ankle replacement when combined with hindfoot fusion is comparable to that of ankle replacement alone. Thus, hindfoot fusion should be performed in conjunction with total ankle replacement when indicated.
When fracture of an extensively porous-coated
femoral component occurs, its removal at revision total hip arthroplasty
(THA) may require a femoral osteotomy and the use of a trephine.
The remaining cortical bone after using the trephine may develop
thermally induced necrosis. A retrospective review identified 11
fractured, well-fixed, uncemented, extensively porous-coated femoral
components requiring removal using a trephine with a minimum of
two years of follow-up. The mean time to failure was 4.6 years (1.7 to 9.1, standard
deviation ( A total of four patients (36.4%) required further revision: three
for instability and one for fracture of the revision component.
There was no statistically significant difference in the mean Harris
hip score before implant fracture (82.4; These findings suggest that removal of a fractured, well-fixed,
uncemented, extensively porous-coated femoral component using a
trephine does not compromise subsequent fixation at revision THA
and the patient’s pre-operative level of function can be restored.
However, the loss of proximal bone stock before revision may be associated
with a high rate of dislocation post-operatively. Cite this article:
We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231 knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196 patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1 mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the tibial component.
These results confirm that the early uncemented Oxford medial unicompartmental
compartmental knee replacements were reliable and the incidence
of radiolucent lines was significantly decreased compared with the
reported results of cemented versions of this implant. These independent
results confirm those of the designing centre.
We report our experience of performing an elbow
hemiarthroplasty in the treatment of comminuted distal humeral fractures
in the elderly patients. A cohort of 42 patients (three men and 39 women, mean age 72;
56 to 84) were reviewed at a mean of 34.3 months (24 to 61) after
surgery. Functional outcome was measured with the Mayo Elbow Performance
Score (MEPS) and range of movement. The disabilities of the arm,
shoulder and hand questionnaire (DASH) was used as a patient rated
evaluation. Complications and ulnar nerve function were recorded.
Plain radiographs were obtained to assess prosthetic loosening,
olecranon wear and heterotopic bone formation. The mean extension deficit was 23.5° (0° to 60°) and mean flexion
was 126.8° (90° to 145°) giving a mean arc of 105.5° (60° to 145°).
The mean MEPS was 90 (50 to 100) and a mean DASH score of 20 (0
to 63). Four patients had additional surgery for limited range of
movement and one for partial instability. One elbow was revised
due to loosening, two patients had sensory ulnar nerve symptoms,
and radiographic signs of mild olecranon wear was noted in five
patients. Elbow hemiarthroplasty for comminuted intra-articular distal
humeral fractures produces reliable medium-term results with functional
outcome and complication rates, comparable with open reduction and
internal fixation and total elbow arthroplasty. Cite this article:
We wanted to investigate regional variations in the organisms
reported to be causing peri-prosthetic infections and to report
on prophylaxis regimens currently in use across England. Analysis of data routinely collected by Public Health England’s
(PHE) national surgical site infection database on elective primary
hip and knee arthroplasty procedures between April 2010 and March
2013 to investigate regional variations in causative organisms.
A separate national survey of 145 hospital Trusts (groups of hospitals
under local management) in England routinely performing primary
hip and/or knee arthroplasty was carried out by standard email questionnaire.Objectives
Methods
Many different lengths of stem are available
for use in primary total hip replacement, and the morphology of
the proximal femur varies greatly. The more recently developed shortened
stems provide a distribution of stress which closely mimics that
of the native femur. Shortening the femoral component potentially
comes at the cost of decreased initial stability. Clinical studies
on the performance of shortened cemented and cementless stems are promising,
although long-term follow-up studies are lacking. We provide an
overview of the current literature on the anatomical features of
the proximal femur and the biomechanical aspects and clinical outcomes
associated with the length of the femoral component in primary hip
replacement, and suggest a classification system for the length
of femoral stems. Cite this article:
Hip replacement is a very successful operation and the outcome is usually excellent. There are recognised complications that seem increasingly to give rise to litigation. This paper briefly examines some common scenarios where litigation may be pursued against hip surgeons. With appropriate record keeping, consenting and surgical care, the claim can be successfully defended if not avoided. We hope this short summary will help to highlight some common pitfalls. There is extensive literature available for detailed study.
The June 2015 Foot &
Ankle Roundup360 looks at: Syndesmosis and outcomes in ankle fracture; Ankle arthrodesis or arthroplasty: a complications-based analysis; Crosslinked polyethylene and ankle arthroplasty; Reducing screw removal in calcaneal osteotomies; Revisiting infection control policies; Chevron osteotomy: proximal or distal?; Ankle distraction for osteoarthritis
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
The June 2015 Trauma Roundup360 looks at: HIV-related implant surgery in trauma; Major transfusion under the spotlight; Surgery and mortality in elderly acetabular fractures; Traction pin safety; Obesity and trauma; Salvage of acetabular fixation in the longer term