This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
We hypothesised that the removal of the subchondral
bone plate (SCBP) for cemented acetabular component fixation in
total hip arthroplasty (THA) offers advantages over retention by
improving the cement-bone interface, without jeopardising implant
stability. We have previously published two-year follow-up data
of a randomised controlled trial (RCT), in which 50 patients with
primary osteoarthritis were randomised to either retention or removal
of the SCBP. The mean age of the retention group (n = 25, 13 males)
was 70.0 years ( Cite this article:
A total of 445 consecutive primary total knee
replacements (TKRs) were followed up prospectively at six and 18 months
and three, six and nine years. Patients were divided into two groups:
non-obese (body mass index (BMI) <
30 kg/m2) and obese
(BMI ≥ 30 kg/m2). The obese group was subdivided into
mildly obese (BMI 30 to 35 kg/m2) and highly obese (BMI ≥ 35
kg/m2) in order to determine the effects of increasing
obesity on outcome. The clinical data analysed included the Knee
Society score, peri-operative complications and implant survival.
There was no difference in the overall complication rates or implant
survival between the two groups. Obesity appears to have a small but significant adverse effect
on clinical outcome, with highly obese patients showing lower function
scores than non-obese patients. However, significant improvements
in outcome are sustained in all groups nine years after TKR. Given
the substantial, sustainable relief of symptoms after TKR and the low
peri-operative complication and revision rates in these two groups,
we have found no reason to limit access to TKR in obese patients.
Inherent disadvantages of reverse shoulder arthroplasty
designs based on the Grammont concept have raised a renewed interest
in less-medialised designs and techniques. The aim of this study
was to evaluate the outcome of reverse shoulder arthroplasty (RSA)
with the fully-constrained, less-medialised, Bayley–Walker prosthesis performed
for the treatment of rotator-cuff-deficient shoulders with glenohumeral
arthritis. A total of 97 arthroplasties in 92 patients (53 women
and 44 men, mean age 67 years (standard deviation ( The Bayley–Walker prosthesis provides reliable pain relief and
reasonable functional improvement for patients with symptomatic
cuff-deficient shoulders. Compared with other designs of RSA, it
offers a modest improvement in forward elevation, but restores external
rotation to some extent and prevents scapular notching. A longer
follow-up is required to assess the survival of the prosthesis and
the clinical performance over time. Cite this article:
The term developmental dysplasia of the hip (DDH)
describes a spectrum of disorders that results in abnormal development
of the hip joint. If not treated successfully in childhood, these
patients may go on to develop hip symptoms and/or secondary osteoarthritis
in adulthood. In this review we describe the altered anatomy encountered
in adults with DDH along with the management options, and the challenges
associated with hip arthroscopy, osteotomies and arthroplasty for
the treatment of DDH in young adults. Cite this article:
We compared the clinical and radiographic results
of total ankle replacement (TAR) performed in non-diabetic and diabetic
patients. We identified 173 patients who underwent unilateral TAR
between 2004 and 2011 with a minimum of two years’ follow-up. There
were 88 male (50.9%) and 85 female (49.1%) patients with a mean
age of 66 years ( The mean AOS and AOFAS scores were significantly better in the
non-diabetic group (p = 0.018 and p = 0.038, respectively). In all,
nine TARs (21%) in the diabetic group had clinical failure at a
mean follow-up of five years (24 to 109), which was significantly
higher than the rate of failure of 15 (11.6%) in the non-diabetic
group (p = 0.004). The uncontrolled diabetic subgroup had a significantly
poorer outcome than the non-diabetic group (p = 0.02), and a higher
rate of delayed wound healing. The incidence of early-onset osteolysis was higher in the diabetic
group than in the non-diabetic group (p = 0.02). These results suggest
that diabetes mellitus, especially with poor glycaemic control,
negatively affects the short- to mid-term outcome after TAR. Cite this article:
The aim of this study was to assess a specific
protocol for the treatment of patients with a parosteal osteosarcoma of
the distal femur with limb salvage involving hemicortical resection
and reconstruction using recycled pasteurised autograft and internal
fixation. Between January 2000 and January 2010, 13 patients with
a mean age of 26.5 years (17 to 39) underwent this procedure. All
the tumours were staged according to Enneking’s criteria: there
were eight stage IA tumours and five stage IB tumours. The mean
follow-up was 101.6 months (58 to 142), and mean post-operative
Musculoskeletal Tumour Society functional score was 88.6% (80% to
100%) at the final follow-up. All the patients had achieved bony
union; the mean time to union was 11.2 months (6 to 18). Local recurrence
occurred in one patient 27 months post-operatively. No patient had
a pulmonary metastasis. A hemicortical procedure for the treatment of a parosteal osteosarcoma
is safe and effective. Precise pre-operative planning using MRI
is essential in order to define the margins of resection. Although
it is a technically demanding procedure, gratifying results make
it worthwhile for selected patients. Cite this article:
Hips with metal-on-metal total hip arthroplasty (MoM THA) have a high rate of adverse local tissue reactions (ALTR), often associated with hypersensitivity reactions. Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) measures tissue perfusion with the parameter Ktrans (volume transfer constant of contrast agent). Our purpose was 1) to evaluate the feasibility of DCE-MRI in patients with THA and 2) to compare DCE-MRI in patients with MoM bearings with metal-on-polyethylene (MoP) bearings, hypothesising that the perfusion index Ktrans in hips with MoM THA is higher than in hips with MoP THA. In this pilot study, 16 patients with primary THA were recruited (eight MoM, eight MoP). DCE-MRI of the hip was performed at 1.5 Tesla (T). For each patient, Ktrans was computed voxel-by-voxel in all tissue lateral to the bladder. The mean Ktrans for all voxels was then calculated. These values were compared with respect to implant type and gender, and further correlated with clinical parameters.Objectives
Methods
Pelvic discontinuity represents a rare but challenging
problem for orthopaedic surgeons. It is most commonly encountered
during revision total hip replacement, but can also result from
an iatrogentic acetabular fracture during hip replacement. The general
principles in management of pelvic discontinuity include restoration
of the continuity between the ilium and the ischium, typically with
some form of plating. Bone grafting is frequently required to restore
pelvic bone stock. The acetabular component is then impacted, typically
using an uncemented, trabecular metal component. Fixation with multiple
supplemental screws is performed. For larger defects, a so-called
‘cup–cage’ reconstruction, or a custom triflange implant may be
required. Pre-operative CT scanning can greatly assist in planning
and evaluating the remaining bone stock available for bony ingrowth.
Generally, good results have been reported for constructs that restore
stability to the pelvis and allow some form of biologic ingrowth. Cite this article:
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
Increasing demand for total hip and knee arthroplasty (THA/TKA)
and associated follow-up has placed huge demands on orthopaedic
services. Feasible follow-up mechanisms are therefore essential. We conducted an audit of clinical follow-up decision-making for
THA/TKA based on questionnaire/radiograph review compared with local
practice of Arthroplasty Care Practitioner (ACP)-led outpatient
follow-up. In all 599 patients attending an ACP-led THA/TKA follow-up
clinic had a pelvic/knee radiograph, completed a pain/function questionnaire
and were reviewed by an ACP. An experienced orthopaedic surgeon
reviewed the same radiographs and questionnaires, without patient
contact or knowledge of the ACP’s decision. Each pathway classified
patients into: urgent review, annual monitoring, routine follow-up
or discharge. Aims
Methods
The most frequent cause of failure after total
hip replacement in all reported arthroplasty registries is peri-prosthetic
osteolysis. Osteolysis is an active biological process initiated
in response to wear debris. The eventual response to this process
is the activation of macrophages and loss of bone. Activation of macrophages initiates a complex biological cascade
resulting in the final common pathway of an increase in osteolytic
activity. The biological initiators, mechanisms for and regulation
of this process are beginning to be understood. This article explores current
concepts in the causes of, and underlying biological mechanism resulting
in peri-prosthetic osteolysis, reviewing the current basic science
and clinical literature surrounding the topic.
Survival analysis is an important tool for assessing the outcome of total joint replacement. The Kaplan-Meier method is used to estimate the incidence of revision of a prosthesis over time, but does not account appropriately for competing events which preclude revision. In the presence of competing death, this method will lead to statistical bias and the curve will lose its interpretability. A valid comparison of survival results between studies using the method is impossible without accounting for different rates of competing events. An alternative and easily applicable approach, the cumulative incidence of competing risk, is proposed. Using three simulated data sets and realistic data from a cohort of 406 consecutive cementless total hip prostheses, followed up for a minimum of ten years, both approaches were compared and the magnitude of potential bias was highlighted. The Kaplan-Meier method overestimated the incidence of revision by almost 4% (60% relative difference) in the simulations and more than 1% (31.3% relative difference) in the realistic data set. The cumulative incidence of competing risk approach allows for appropriate accounting of competing risk and, as such, offers an improved ability to compare survival results across studies.
As adverse events related to metal on metal hip
arthroplasty have been better understood, there has been increased
interest in toxicity related to the high circulating levels of cobalt ions.
However, distinguishing true toxicity from benign elevations in
cobalt levels can be challenging. The purpose of this review is
to examine the use of cobalt alloys in total hip arthroplasty, to
review the methods of measuring circulating cobalt levels, to define
a level of cobalt which is considered pathological and to review
the pathophysiology, risk factors and treatment of cobalt toxicity.
To the best of our knowledge, there are 18 published cases where
cobalt metal ion toxicity has been attributed to the use of cobalt-chromium
alloys in hip arthroplasty. Of these cases, the great majority reported
systemic toxic reactions at serum cobalt levels more than 100 μg/L.
This review highlights some of the clinical features of cobalt toxicity,
with the goal that early awareness may decrease the risk factors
for the development of cobalt toxicity and/or reduce its severity. Take home message: Severe adverse events can arise from the release
of cobalt from metal-on-metal arthroplasties, and as such, orthopaedic
surgeons should not only be aware of the presenting problems, but
also have the knowledge to treat appropriately. Cite this article:
We report the five-year outcome of a randomised
controlled trial which used radiostereometric analysis (RSA) to assess
the influence of surface oxidised zirconium (OxZr, Oxinium) on polyethylene
wear A total of 120 patients, 85 women and 35 men with a mean age
of 70 years (59 to 80) who were scheduled for primary cemented total
hip arthroplasty were randomly allocated to four study groups. Patients
were blinded to their group assignment and received either a conventional
polyethylene (CPE) or a highly cross-linked (HXL) acetabular component
of identical design. On the femoral side patients received a 28
mm head made of either cobalt-chromium (CoCr) or OxZr. The proximal head penetration (wear) was measured with repeated
RSA examinations over five years. Clinical outcome was measured
using the Harris hip score. There was no difference in polyethylene wear between the two
head materials when used with either of the two types of acetabular
component (p = 0.3 to 0.6). When comparing the two types of polyethylene
there was a significant difference in favour of HXLPE, regardless
of the head material used (p <
0.001). In conclusion, we found no advantage of OxZr over CoCr in terms
of polyethylene wear after five years of follow-up. Our findings
do not support laboratory results which have shown a reduced rate
of wear with OxZr. They do however add to the evidence on the better
resistance to wear of HXLPE over CPE. Cite this article:
Polyethylene wear debris can cause osteolysis
and the failure of total hip arthroplasty. We present the five-year
wear rates of a highly cross-linked polyethylene (X3) bearing surface
when used in conjunction with a 36 mm ceramic femoral head. This was a prospective study of a cohort of 100 THAs in 93 patients.
Pain and activity scores were measured pre- and post-operatively.
Femoral head penetration was measured at two months, one year, two
years and at five years using validated edge-detecting software
(PolyWare Auto). At a mean of 5.08 years (3.93 to 6.01), 85 hips in 78 patients
were available for study. The mean age of these patients was 59.08
years (42 to 73, the mean age of males (n = 34) was 59.15 years,
and females (n = 44) was 59.02 years). All patients had significant
improvement in their functional scores (p <
0.001). The steady
state two-dimensional linear wear rate was 0.109 mm/year. The steady
state volumetric wear rate was 29.61 mm3/year. No significant
correlation was found between rate of wear and age (p = 0.34), acetabular
component size (p = 0.12) or clinical score (p = 0.74). Our study shows low steady state wear rates at five years in
X3 highly cross-linked polyethylene in conjunction with a 36 mm
ceramic femoral head. The linear wear rate was almost identical
to the osteolysis threshold of 0.1 mm/year recommended in the literature. Cite this article:
This systematic review and meta-analysis was conducted to determine
the mid- to long-term clinical outcomes for a medial-pivot total
knee replacement (TKR) system. The objectives were to synthesise
available survivorship, Knee Society Scores (KSS), and reasons for
revision for this system. A systematic search was conducted of two online databases to
identify sources of survivorship, KSS, and reasons for revision.
Survivorship results were compared with values in the National Joint
Registry of England, Wales, and Northern Ireland (NJR).Objectives
Methods
Peri-prosthetic femoral fracture after total hip replacement (THR) is associated with a poor outcome and high mortality. However, little is known about its long-term incidence after uncemented THR. We retrospectively reviewed a consecutive series of 326 patients (354 hips) who had received a CLS Spotorno replacement with an uncemented, straight, collarless tapered titanium stem between January 1985 and December 1989. The mean follow-up was 17 years (15 to 20). The occurrence of peri-prosthetic femoral fracture during follow-up was noted. Kaplan-Meier survival analysis was used to estimate the cumulative incidence of fracture. At the last follow-up, 86 patients (89 hips) had died and eight patients (eight hips) had been lost to follow-up. A total of 14 fractures in 14 patients had occurred. In ten hips, the femoral component had to be revised and in four the fracture was treated by open reduction and internal fixation. The cumulative incidence of peri-prosthetic femoral fracture was 1.6% (95% confidence interval 0.7 to 3.8) at ten years and 4.5% (95% confidence interval 2.6 to 8.0) at 17 years after the primary THR. There was no association between the occurrence of fracture and gender or age at the time of the primary replacement. Our findings indicate that peri-prosthetic femoral fracture is a significant mode of failure in the long term after the insertion of an uncemented CLS Spotorno stem. Revision rates for this fracture rise in the second decade. Further research is required to investigate the risk factors involved in the occurrence of late peri-prosthetic femoral fracture after the implantation of any uncemented stem, and to assess possible methods of prevention.
Between 1999 and 2005, 10 264 patients who had undergone total hip replacement (THR) for subcapital fracture of the hip were compared with 76 520 in whom THR had been performed for other reasons. All the cases were identified through the Swedish Hip Arthroplasty Register. The THRs performed as primary treatment for fracture were also compared with those done after failure of internal fixation. After seven years the rate of revision was higher in THR after fracture (4.4% vs 2.9%). Dislocation and periprosthetic fracture were the most common causes of revision. The risk was higher in men than in women. The type of femoral component and the surgical approach influenced the risk. After correction for gender, type of component and the surgical approach the revision rates were similar in the primary and secondary fracture THR groups. Total hip replacement is therefore a safe method for both the primary and secondary management of fracture of the hip.