Conventional uncemented femoral implants provide
dependable long-term fixation in patients with a wide range of functional
requirements. Yet challenges associated with proximal–distal femoral
dimensional mismatch, preservation of bone stock, and minimally
invasive approaches have led to exploration into alternative implant designs.
Short stem designs focusing on a stable metaphyseal fit have emerged
to address these issues in total hip replacement (THR). Uncemented
metaphyseal-engaging short stem implants are stable and are associated
with proximal bone remodeling closer to the metaphysis when compared
with conventional stems and they also have comparable clinical performances.
Short stem metaphyseal-engaging implants can meet the goals of a
successful THR, including tolerating a high level of patient function,
as well as durable fixation. Cite this article:
The restoration of knee alignment is an important
goal during total knee arthroplasty (TKA). In the past surgeons aimed
to restore neutral limb alignment during surgery. However, previous
studies have demonstrated alignment to be dynamic, varying depending
on the position of the limb and the degree of weight-bearing, and
between patients. We used a validated computer navigation system
to measure the femorotibial mechanical angle (FTMA) in 264 knees in
77 male and 55 female healthy volunteers aged 18 to 35 years (mean
26.2). We found the mean supine alignment to be a varus angle of
1.2° (standard deviation ( Knee alignment is different in different individuals and is dynamic
in nature, changing with different postures. This may have implications
for the assessment of alignment in TKA, which is achieved in non-weight-bearing conditions
and which may not represent the situation observed during weight-bearing. Cite this article:
The aim of this study was to investigate differences in pain,
range of movement function and satisfaction at three months and
one year after total knee arthroplasty (TKA) in patients with an
oblique pattern of kinematic graph of the knee and those with a
varus pattern. A total of 91 patients who underwent TKA were included in this
retrospective study. Patients (59 women and 32 men with mean age
of 68.7 years; 38.6 to 88.4) were grouped according to kinematic
graphs which were generated during navigated TKA and the outcomes
between the groups were compared.Aims
Patients and Methods
The purpose of this study was to develop an accurate, reliable and easily applicable method for determining the anatomical location of the joint line during revision knee arthroplasty. The transepicondylar width (TEW), the perpendicular distance between the medial and lateral epicondyles and the distal articular surfaces (DMAD, DLAD) and the distance between the medial and lateral epicondyles and the posterior articular surfaces (PMAD, DLAD) were measured in 40 knees from 20 formalin-fixed adult cadavers (11 male and nine female; mean age at death 56.9 years, Objectives
Methods
Hip resurfacing has been proposed as an alternative
to traditional total hip arthroplasty in young, active patients.
Much has been learned following the introduction of metal-on-metal resurfacing
devices in the 1990s. The triad of a well-designed device, implanted
accurately, in the correct patient has never been more critical
than with these implants. Following Food and Drug Administration approval in 2006, we studied
the safety and effectiveness of one hip resurfacing device (Birmingham
Hip Resurfacing) at our hospital in a large, single-surgeon series.
We report our early to mid–term results in 1333 cases followed for
a mean of 4.3 years (2 to 5.7) using a prospective, observational
registry. The mean patient age was 53.1 years (12 to 84); 70% were
male and 91% had osteoarthritis. Complications were few, including
no dislocations, no femoral component loosening, two femoral neck
fractures (0.15%), one socket loosening (0.08%), three deep infections
(0.23%), and three cases of metallosis (0.23%). There were no destructive
pseudotumours. Overall survivorship at up to 5.7 years was 99.2%. Aseptic survivorship
in males under the age of 50 was 100%. We believe this is the largest
United States series of a single surgeon using a single resurfacing
system. Cite this article:
A total of 20 pairs of fresh-frozen cadaver femurs were assigned to four alignment groups consisting of relative varus (10° and 20°) and relative valgus (10° and 20°), 75 composite femurs of two neck geometries were also used. In both the cadaver and the composite femurs, placing the component in 20° of valgus resulted in a significant increase in load to failure. Placing the component in 10° of valgus had no appreciable effect on increasing the load to failure except in the composite femurs with varus native femoral necks. Specimens in 10° of varus were significantly weaker than the neutrally-aligned specimens. The results suggest that retention of the intact proximal femoral strength occurs at an implant angulation of ≥ 142°. However, the benefit of extreme valgus alignment may be outweighed in clinical practice by the risk of superior femoral neck notching, which was avoided in this study.
The December 2015 Knee Roundup360 looks at: Albumin and complications in knee arthroplasty; Tantalum: a knee fixation for all seasons?; Dynamic knee alignment; Tibial component design in UKA; Managing the tidal wave of revision knee arthroplasty; Scoring pain in TKR; Does anyone have a ‘normal’ tibial slope?; XLPE in TKR? A five-year clinical study; Spacers and infected revision arthroplasties; Dialysis and arthroplasty
Analysis of the morphology of the distal femur, and by extension
of the femoral components in total knee arthroplasty (TKA), has
largely been related to the aspect ratio, which represents the width
of the femur. Little is known about variations in trapezoidicity
(i.e. whether the femur is more rectangular or more trapezoidal).
This study aimed to quantify additional morphological characteristics
of the distal femur and identify anatomical features associated
with higher risks of over- or under-sizing of components in TKA. We analysed the shape of 114 arthritic knees at the time of primary
TKA using the pre-operative CT scans. The aspect ratio and trapezoidicity
ratio were quantified, and the post-operative prosthetic overhang
was calculated. We compared the morphological characteristics with
those of 12 TKA models.Aims
Methods
Mechanical alignment has been a fundamental tenet of total knee arthroplasty (TKA) since modern knee replacement surgery was developed in the 1970s. The objective of mechanical alignment was to infer the greatest biomechanical advantage to the implant to prevent early loosening and failure. Over the last 40 years a great deal of innovation in TKA technology has been focusing on how to more accurately achieve mechanical alignment. Recently the concept of mechanical alignment has been challenged, and other alignment philosophies are being explored with the intention of trying to improve patient outcomes following TKA. This article examines the evolution of the mechanical alignment concept and whether there are any viable alternatives.
Stress shielding resulting in diminished bone
density following total knee replacement (TKR) may increase the
risk of migration and loosening of the prosthesis. This retrospective
study was designed to quantify the effects of the method of fixation
on peri-prosthetic tibial bone density beneath cemented and uncemented
tibial components of similar design and with similar long-term survival
rates. Standard radiographs taken between two months and 15 years
post-operatively were digitised from a matched group of TKRs using
cemented (n = 67) and uncemented (n = 67) AGC tibial prostheses.
Digital radiograph densitometry was used to quantify changes in
bone density over time. Age, length of follow-up, gender, body mass
index and alignment each significantly influenced the long-term
pattern of peri-prosthetic bone density. Similar long-term changes
in density irrespective of the method of fixation correlated well
with the high rate of survival of this TKR at 20 years, and suggest
that cemented and uncemented fixation are both equally viable. Cite this article:
The June 2015 Knee Roundup360 looks at: Cruciate substituting
We performed a retrospective review of a consecutive
series of 178 Mobility total ankle replacements (TARs) performed
by three surgeons between January 2004 and June 2009, and analysed
radiological parameters and clinical outcomes in a subgroup of 129
patients. The mean follow-up was 4 years (2 to 6.3). A total of
ten revision procedures (5.6%) were undertaken. The mean Ankle Osteoarthritis
Scale (AOS) pain score was 17 (0 to 88) and 86% of patients were
clinically improved at follow-up. However, 18 patients (18 TARs,
14%) had a poor outcome with an AOS pain score of >
30. A worse
outcome was associated with a pre-operative diagnosis of post-traumatic degenerative
arthritis. However, no pre- or post-operative radiological parameters
were significantly associated with a poor outcome. Of the patients
with persistent pain, eight had predominantly medial-sided pain.
Thirty TARs (29%) had a radiolucency in at least one zone. The outcome of the Mobility TAR at a mean of four years is satisfactory
in >
85% of patients. However, there is a significant incidence
of persistent pain, particularly on the medial side, for which we
were unable to establish a cause. Cite this article:
Large femoral heads have become popular in total
hip replacement (THR) as a method of reducing the risk of dislocation.
However, if large heads are used in ceramic-on-ceramic THR, the
liner must be thinner, which may increase the risk of fracture.
To compare the rates of ceramic fracture and dislocation between
28 mm and 32 mm ceramic heads, 120 hips in 109 patients (51 men
and 58 women, mean age 49.2 years) were randomised to THR with either
a 28 mm or a 32 mm ceramic articulation. A total of 57/60 hips assigned
to the 28 mm group and 55/60 hips assigned to the 32 mm group were
followed for at least five years. No ceramic component fractures
occured in any patient in either group. There was one dislocation
in the 32 mm group and none in the 28 mm group (p = 0.464). No hip
had detectable wear, focal osteolysis or prosthetic loosening. In
our small study the 32 mm ceramic articulation appeared to be safe
in terms of ceramic liner fracture. Cite this article:
A prospective, randomised, controlled trial compared two different techniques of high tibial osteotomy with a lateral closing wedge or a medial opening wedge, stabilised by a Puddu plate. The clinical outcome and radiological results were examined at one year. The primary outcome measure was the achievement of an overcorrection of valgus of 4°. Secondary outcome measures were the severity of pain (visual analogue scale), knee function (Hospital for Special Surgery score), and walking distance. Between January 2001 and April 2004, 92 patients were randomised to one or other of the techniques. At follow-up at one year the post-operative hip-knee-ankle angle was 3.4° (± 3.6° The severity of pain, knee score and walking ability improved in both groups, but the difference was not significant. Because of pain, the staples required removal in 11 (23%) patients in the closing-wedge group and a Puddu plate was removed in 27 (60%) patients in the opening-wedge group. This difference was significant (p <
0.001). We conclude that closing-wedge osteotomy achieves a more accurate correction with less morbidity, although both techniques had improved the function of the knee at one year after the procedure.
We retrospectively reviewed the records of 1150
computer-assisted total knee replacements and analysed the clinical
and radiological outcomes of 45 knees that had arthritis with a
pre-operative recurvatum deformity. The mean pre-operative hyperextension
deformity of 11° (6° to 15°), as measured by navigation at the start
of the operation, improved to a mean flexion deformity of 3.1° (0°
to 7°) post-operatively. A total of 41 knees (91%) were managed
using inserts ≤ 12.5 mm thick, and none had mediolateral laxity
>
2 mm from a mechanical axis of 0° at the end of the surgery. At
a mean follow-up of 26.4 months (13 to 48) there was significant
improvement in the mean Knee Society, Oxford knee and Western Ontario
and McMaster Universities Osteoarthritis Index scores compared with
the pre-operative values. The mean knee flexion improved from 105°
(80° to 125°) pre-operatively to 131° (120° to 145°), and none of
the limbs had recurrent recurvatum. These early results show that total knee replacement using computer
navigation and an algorithmic approach for arthritic knees with
a recurvatum deformity can give excellent radiological and functional
outcomes without recurrent deformity.
We report the outcome of 39 patients who underwent
a modified Pauwels’ intertrochanteric osteotomy for nonunion of
a femoral neck fracture following failed osteosynthesis. There were
31 men and eight women with a mean age of 47.2 years (34 to 59).
By Pauwels’ classification, there were 11 Type II fractures and
28 Type III fractures. The mean follow-up was 7.9 years (2 to 19).
In the 11 patients whose initial treatment had been osteotomy, union was
achieved in nine (81.8%). In 28 patients whose initial treatment
had been with a lag screw or a dynamic hip screw, union was achieved
in 27 (96.4%). Limb lengths were equalised in 14 of 16 patients
(87.5%) with pre-operative shortening. The mean neck-shaft angle
improved significantly from 100.5° (80° to 120°) to 131.6° (120°
to 155°) (p = 0.004). The mean modified Harris hip score was 85.6
points (70 to 97) and the mean modified Merle d’Aubigné score was
14.3 (11 to 18). Good to excellent functional outcomes were achieved
in 32 patients (88.8%). A modified Pauwels’ intertrochanteric osteotomy
is a reliable method of treating ununited fractures of the femoral neck
following failed osteosynthesis: coxa vara and shortening can also
simultaneously be addressed. Cite this article
Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement tibial components have been referred to as physiological radiolucencies. These should be distinguished from pathological radiolucencies, which are poorly defined, wide and progressive, and associated with loosening and infection. We studied the incidence and clinical significance of tibial radiolucent lines in 161 Oxford unicondylar knee replacements five years after surgery. All the radiographs were aligned with fluoroscopic control to obtain views parallel to the tibial tray to reveal the tibial bone-implant interface. We found that 49 knees (30%) had complete, 52 (32%) had partial and 60 (37%) had no radiolucent lines. There was no relationship between the incidence of radiolucent lines and patient factors such as gender, body mass index and activity, or operative factors including the status of the anterior cruciate ligament and residual varus deformity. Nor was any statistical relationship established between the presence of radiolucent lines and clinical outcome, particularly pain, assessed by the Oxford Knee score and the American Knee Society score. We conclude that radiolucent lines are common after Oxford unicompartmental knee replacement but that their aetiology remains unclear. Radiolucent lines were not a source of adverse symptoms or pain. Therefore, when attempting to identify a source of postoperative pain after Oxford unicompartmental knee replacement the presence of a physiological radiolucency should be ignored.
The aim of this study was to find anatomical
landmarks for rotational alignment of the tibial component in total knee
replacement (TKR) in a CT-based study. Pre-operative CT scanning
was performed on 94 South Korean patients (nine men, 85 women, 188
knees) with osteoarthritis of the knee joint prior to TKR. The tibial
anteroposterior (AP) axis was defined as a line perpendicular to
the femoral surgical transepicondylar axis and passing through the centre
of the posterior cruciate ligament (PCL). The angles between the
defined tibial AP axis and anatomical landmarks at various levels
of the tibia were measured. The mean values of the angles between
the defined tibial AP axis and the line connecting the anterior
border of the proximal third of the tibia to the centre of the PCL
was -0.2° (-17 to 14.1, Cite this article:
The use of plate-and-cable constructs to treat periprosthetic fractures around a well-fixed femoral component in total hip replacements has been reported to have high rates of failure. Our aim was to evaluate the results of a surgical treatment algorithm to use these lateral constructs reliably in Vancouver type-B1 and type-C fractures. The joint was dislocated and the stability of the femoral component was meticulously evaluated in 45 type-B1 fractures. This led to the identification of nine (20%) unstable components. The fracture was considered to be suitable for single plate-and-cable fixation by a direct reduction technique if the integrity of the medial cortex could be restored. Union was achieved in 29 of 30 fractures (97%) at a mean of 6.4 months (3 to 30) in 29 type-B1 and five type-C fractures. Three patients developed an infection and one construct failed. Using this algorithm plate-and-cable constructs can be used safely, but indirect reduction with minimal soft-tissue damage could lead to shorter times to union and lower rates of complications.
Metal-on-metal (MoM) hip resurfacing was developed in the 1990s by surgeons in Birmingham, UK, as a surgical solution to the problem of osteoarthritis in younger, more active patients. Early results were promising and the procedure gained in popularity. However, adverse reports of soft-tissue reaction and failure started to appear from 2008 onwards. Surgeons may be asked to write medico-legal reports on the surgical aspects of an individual case for claimant lawyers or in defence for the NHSLA or indemnity insurers. The purpose of this article is to cover some of the aspects of the operation that may be considered in such medico-legal reports.