Using human cadaver specimens, we investigated
the role of supplementary fibular plating in the treatment of distal
tibial fractures using an intramedullary nail. Fibular plating is
thought to improve stability in these situations, but has been reported
to have increased soft-tissue complications and to impair union
of the fracture. We proposed that multidirectional locking screws
provide adequate stability, making additional fibular plating unnecessary.
A distal tibiofibular osteotomy model performed on matched fresh-frozen
lower limb specimens was stabilised with reamed nails using conventional
biplanar distal locking (CDL) or multidirectional distal locking
(MDL) options with and without fibular plating. Rotational stiffness
was assessed under a constant axial force of 150 N and a superimposed
torque of ± 5 Nm. Total movement, and neutral zone and fracture
gap movement were analysed. In the CDL group, fibular plating improved stiffness at the tibial
fracture site, albeit to a small degree (p = 0.013). In the MDL
group additional fibular plating did not increase the stiffness.
The MDL nail without fibular plating was significantly more stable
than the CDL nail with an additional fibular plate (p = 0.008). These findings suggest that additional fibular plating does not
improve stability if a multidirectional distal locking intramedullary
nail is used, and is therefore unnecessary if not needed to aid
reduction. Cite this article:
This study tests the biomechanical properties of adjacent locked
plate constructs in a femur model using Sawbones. Previous studies
have described biomechanical behaviour related to inter-device distances.
We hypothesise that a smaller lateral inter-plate distance will
result in a biomechanically stronger construct, and that addition
of an anterior plate will increase the overall strength of the construct. Sawbones were plated laterally with two large-fragment locking
compression plates with inter-plate distances of 10 mm or 1 mm.
Small-fragment locking compression plates of 7-hole, 9-hole, and
11-hole sizes were placed anteriorly to span the inter-plate distance.
Four-point bend loading was applied, and the moment required to
displace the constructs by 10 mm was recorded.Objectives
Methods
The purpose of this study was to evaluate the
risk of late displacement after the treatment of distal radial fractures with
a locking volar plate, and to investigate the clinical and radiological
factors that might correlate with re-displacement. From March 2007
to October 2009, 120 of an original cohort of 132 female patients
with unstable fractures of the distal radius were treated with a
volar locking plate, and were studied over a follow-up period of
six months. In the immediate post-operative and final follow-up
radiographs, late displacement was evaluated as judged by ulnar
variance, radial inclination, and dorsal angulation. We also analysed
the correlation of a variety of clinical and radiological factors
with re-displacement. Ulnar variance was significantly overcorrected
(p <
0.001) while radial inclination and dorsal angulation were
undercorrected when compared statistically (p <
0.001) with the unaffected
side in the immediate post-operative stage. During follow-up, radial
shortening and dorsal angulation progressed statistically, but none
had a value beyond the acceptable range. Bone mineral density measured
at the proximal femur and the position of the screws in the subchondral
region, correlated with slight progressive radial shortening, which
was not clinically relevant. Volar locking plating of distal radial fractures is a reliable
form of treatment without substantial late displacement. Cite this article:
Treatment guidelines for atypical femoral fractures associated
with bisphosphonates have not been established. We conducted a systematic
review of the treatment of atypical femoral fractures first, to
evaluate the outcomes of surgical fixation of complete atypical fractures
and secondly, to assess whether prophylactic surgery is necessary
for incomplete atypical fractures. Case reports and series were identified from the PubMed database
and were included if they described the treatment of atypical femoral
fractures. In total, 77 publications met our inclusion criteria
and 733 patients with 834 atypical complete or incomplete femoral fractures
were identified.Aims
Materials and Methods
We report the long-term outcome of 33 patients
(37 knees) who underwent proximal tibial open-wedge osteotomy with
hemicallotasis (HCO) for medial osteoarthritis of the knee between
1995 and 2000. Among these, 29 patients with unilateral HCO were
enrolled and 19 were available for review at a mean of 14.2 years
(10 to 15.7) post-operatively. For these 19 patients, the mean Hospital
for Special Surgery knee score was 60 (57 to 62) pre-operatively
and 85 (82 to 87) at final follow-up (p <
0.001; paired Cite this article:
In approximately 20% of patients with ankle fractures, there
is an concomitant injury to the syndesmosis which requires stabilisation,
usually with one or more syndesmotic screws. The aim of this review
is to evaluate whether removal of the syndesmotic screw is required
in order for the patient to obtain optimal functional recovery. A literature search was conducted in Medline, Embase and the
Cochrane Library for articles in which the syndesmotic screw was
retained. Articles describing both removal and retaining of syndesmotic
screws were included. Excluded were biomechanical studies, studies
not providing patient related outcome measures, case reports, studies
on skeletally immature patients and reviews. No restrictions regarding
year of publication and language were applied.Aims
Materials and Methods
Patients with diabetes are at increased risk of wound complications
after open reduction and internal fixation of unstable ankle fractures.
A fibular nail avoids large surgical incisions and allows anatomical
reduction of the mortise. We retrospectively reviewed the results of fluoroscopy-guided
reduction and percutaneous fibular nail fixation for unstable Weber
type B or C fractures in 24 adult patients with type 1 or type 2
diabetes. The re-operation rate for wound dehiscence or other indications
such as amputation, mortality and functional outcomes was determined.Aims
Patients and Methods
A successful outcome following treatment of nonunion requires the correct identification of all of the underlying cause(s) and addressing them appropriately. The aim of this study was to assess the distribution and frequency of causative factors in a consecutive cohort of nonunion patients in order to optimise the management strategy for individual patients presenting with nonunion. Causes of the nonunion were divided into four categories: mechanical; infection; dead bone with a gap; and host. Prospective and retrospective data of 100 consecutive patients who had undergone surgery for long bone fracture nonunion were analysed.Objectives
Methods
Bone loss secondary to primary or metastatic lesions of the proximal humerus remains a challenging surgical problem. Options include preservation of the joint with stabilisation using internal fixation or resection of the tumour with prosthetic replacement. Resection of the proximal humerus often includes the greater tuberosity and adjacent diaphysis, which may result in poor function secondary to loss of the rotator cuff and/or deltoid function. Preservation of the joint with internal fixation may reduce the time in hospital and peri-operative morbidity compared with joint replacement, and result in a better functional outcome. We included 32 patients with pathological fractures of the proximal humerus in this study. Functional and radiological assessments were performed. At a mean follow-up of 17.6 months (8 to 61) there was no radiological evidence of failure of fixation. The mean revised musculoskeletal Tumour Society functional score was 94.6% (86% to 99%). There was recurrent tumour requiring further surgery in four patients (12.5%). Of the 22 patients who were employed prior to presentation all returned to work without restrictions. The use of a locking plate combined with augmentation with cement extends the indications for salvage of the proximal humerus with good function in patients with pathological and impending pathological fractures.
This study describes the Osseointegration Group of Australia’s
Accelerated Protocol two-stage strategy (OGAAP-1) for the osseointegrated
reconstruction of amputated limbs. We report clinical outcomes in 50 unilateral trans-femoral amputees
with a mean age of 49.4 years (24 to 73), with a minimum one-year
follow-up. Outcome measures included the Questionnaire for persons
with a Trans-Femoral Amputation, the health assessment questionnaire
Short-Form-36 Health Survey, the Amputation Mobility Predictor scores
presented as K-levels, 6 Minute Walk Test and timed up and go tests.
Adverse events included soft-tissue problems, infection, fractures
and failure of the implant.Aims
Patients and Methods
We compared the outcome of closed intramedullary nailing with minimally invasive plate osteosynthesis using a percutaneous locked compression plate in patients with a distal metaphyseal fracture in a prospective study. A total of 85 patients were randomised to operative stabilisation either by a closed intramedullary nail (44) or by minimally invasive osteosynthesis with a compression plate (41). Pre-operative variables included the patients’ age and the side and pattern of the fracture. Peri-operative variables were the operating time and the radiation time. Postoperative variables were wound problems, the time to union of the fracture, the functional American Orthopaedic Foot and Ankle surgery score and removal of hardware. We found no significant difference in the pre-operative variables or in the time to union in the two groups. However, the mean radiation time and operating time were significantly longer in the locked compression plate group (3.0 We conclude that both closed intramedullary nailing and a percutaneous locked compression plate can be used safely to treat Orthopaedic Trauma Association type-43A distal metaphyseal fractures of the tibia. However, closed intramedullary nailing has the advantage of a shorter operating and radiation time and easier removal of the implant. We therefore prefer closed intramedullary nailing for patients with these fractures.
We evaluated the biomechanical properties of two different methods of fixation for unstable fractures of the proximal humerus. Biomechanical testing of the two groups, locking plate alone (LP), and locking plate with a fibular strut graft (LPSG), was performed using seven pairs of human cadaveric humeri. Cyclical loads between 10 N and 80 N at 5 Hz were applied for 1 000 000 cycles. Immediately after cycling, an increasing axial load was applied at a rate of displacement of 5 mm/min. The displacement of the construct, maximum failure load, stiffness and mode of failure were compared. The displacement was significantly less in the LPSG group than in the LP group (p = 0.031). All maximum failure loads and measures of stiffness in the LPSG group were significantly higher than those in the LP group (p = 0.024 and p = 0.035, respectively). In the LP group, varus collapse and plate bending were seen. In the LPSG group, the humeral head cut out and the fibular strut grafts fractured. No broken plates or screws were seen in either group. We conclude that strut graft augmentation significantly increases both the maximum failure load and the initial stiffness of this construct compared with a locking plate alone.
Secure fixation of displaced proximal fractures of the humerus is a challenging problem. A total of 32 patients with acutely displaced three- or four-part proximal fractures of the humerus were treated by open reduction and internal fixation using the proximal humeral internal locking system (PHILOS) plate. There were 23 women and nine men with a mean age of 59.9 years (18 to 87). Data were collected prospectively and the outcomes were assessed using the Constant score. The mean follow-up was for 11 months (3 to 24). In 31 patients (97%) the fracture united clinically and radiologically at a mean of 10 weeks (8 to 24). The mean Constant score at final review was 66.5 (30 to 92). There was no significant difference in outcome when comparing patients aged more than 60 years (18 patients) with those aged less than 60 years (14 patients) ( This plate provides an alternative method of fixation for fractures of the proximal humerus. It provides a stable fixation in young patients with good-quality bone sufficient to permit early mobilisation. Failure of the screws to maintain fixation in the elderly remains a problem.
We present the medium-term clinical results of a reverse total
shoulder arthroplasty with a trabecular metal glenoid base plate. We reviewed 125 consecutive primary reverse total shoulder arthroplasties
(RTSA) implanted in 124 patients for rotator cuff arthropathy. There
were 100 women and 24 men in the study group with a mean age of
76 years (58 to 89). The mean follow-up was 32 months (24 to 60).
No patient was lost to follow-up.Aim
Patients and Methods
Medium-term results of the Discovery elbow replacement
are presented. We reviewed 51 consecutive primary Discovery total
elbow replacements (TERs) implanted in 48 patients. The mean age
of the patients was 69.2 years (49 to 92), there were 19 males and
32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69).
A total of six patients were lost to follow-up. Statistically significant
improvements in range movement and Oxford Elbow Score were found
(p <
0.001). Radiolucent lines were much more common in, and
aseptic loosening was exclusive to, the humeral component. Kaplan–Meier
survivorship at five years was 92.2% (95% CI 74.5% to 96.4%) for
aseptic loosening. In four TERs, periprosthetic infection occurred
resulting in failure. A statistically significant association between
infection and increased BMI was found (p = 0.0268). Triceps failure
was more frequent after the Mayo surgical approach and TER performed
after previous trauma surgery. No failures of the implant were noted. Our comparison shows that the Discovery has early clinical results
that are similar to other semi-constrained TERs. We found continued
radiological surveillance with particular focus on humeral lucency
is warranted and has not previously been reported. Despite advances
in the design of total elbow replacement prostheses, rates of complication
remain high. Cite this article:
We review our experience of Coonrad-Morrey total elbow arthroplasty
(TEA) for fractures of the distal humerus in non-rheumatoid patients
with a minimum of ten years follow-up. TEA through a triceps splitting approach was peformed in 37 non-rheumatoid
patients for a fracture of the distal humerus between 1996 and 2004.
One patient could not be traced and 17 had died before the tenth
anniversary of their surgery. This left 19 patients with a minimum
follow-up of ten years to form the study group. Of these, 13 patients
were alive at the time of final review. The other six had died,
but after the tenth anniversary of their elbow arthroplasty. Their
clinical and radiological data were included in the study. Aims
Patients and Methods
Induced membrane technique is a relatively new technique in the reconstruction of large bone defects. It involves the implantation of polymethylmethacrylate (PMMA) cement in the bone defects to induce the formation of membranes after radical debridement and reconstruction of bone defects using an autologous cancellous bone graft in a span of four to eight weeks. The purpose of this study was to explore the clinical outcomes of the induced membrane technique for the treatment of post-traumatic osteomyelitis in 32 patients. A total of 32 cases of post-traumatic osteomyelitis were admitted to our department between August 2011 and October 2012. This retrospective study included 22 men and ten women, with a mean age of 40 years (19 to 70). Within this group there were 20 tibias and 12 femurs with a mean defect of 5 cm (1.5 to 12.5). Antibiotic-loaded PMMA cement was inserted into the defects after radical debridement. After approximately eight weeks, the defects were implanted with bone graft.Objectives
Methods
A consecutive series of 320 patients with an
intracapsular fracture of the hip treated with a dynamic locking
plate (Targon Femoral Neck (TFN)) were reviewed. All surviving patients
were followed for a minimum of two years. During the follow-up period
109 patients died. There were 112 undisplaced fractures, of which three (2.7%) developed
nonunion or re-displacement and five (4.5%) developed avascular
necrosis of the femoral head. Revision to an arthroplasty was required
for five patients (4.5%). A further six patients (5.4%) had elective
removal of the plate and screws. There were 208 displaced fractures, of which 32 (15.4%) developed
nonunion or re-displacement and 23 (11.1%) developed avascular necrosis.
A further four patients (1.9%) developed a secondary fracture around
the TFN. Revision to a hip replacement was required for 43 patients
(20.7%) patients and a further seven (3.3%) had elective removal
of the plate and screws. It is suggested that the stronger distal fixation combined with
rotational stability may lead to a reduced incidence of complications
related to the healing of the fracture when compared with other
contemporary fixation devices but this needs to be confirmed in
further studies. Cite this article:
The ideal form of fixation for displaced, extra-articular
fractures of the distal tibia remains controversial. In the UK, open
reduction and internal fixation with locking-plates and intramedullary
nailing are the two most common forms of treatment. Both techniques
provide reliable fixation but both are associated with specific
complications. There is little information regarding the functional
recovery following either procedure. We performed a randomised pilot trial to determine the functional
outcome of 24 adult patients treated with either a locking-plate
(n = 12) or an intramedullary nailing (n = 12). At six months, there
was an adjusted difference of 13 points in the Disability Rating
Index in favour of the intramedullary nail. However, this was not
statistically significant in this pilot trial (p = 0.498). A total
of seven patients required further surgery in the locking-plate
group and one in the intramedullary nail group. This study suggests that there may be clinically relevant, functional
differences in patients treated with nail