Early failure associated with adverse reactions to metal debris is an emerging problem after hip resurfacing but the exact mechanism is unclear. We analysed our entire series of 660 metal-on-metal resurfacings (Articular Surface Replacement (ASR) and Birmingham Hip Resurfacing (BHR)) and large-bearing ASR total hip replacements, to establish associations with metal debris-related failures. Clinical and radiological outcomes, metal ion levels, explant studies and lymphocyte transformation tests were performed. A total of 17 patients (3.4%) were identified (all ASR bearings) with adverse reactions to metal debris, for which revision was required. This group had significantly smaller components, significantly higher acetabular component anteversion, and significantly higher whole concentrations of blood and joint chromium and cobalt ions than asymptomatic patients did (all p <
0.001). Post-revision lymphocyte transformation tests on this group showed no reactivity to chromium or cobalt ions. Explants from these revisions had greater surface wear than retrievals for uncomplicated fractures. The absence of adverse reactions to metal debris in patients with well-positioned implants usually implies high component wear. Surgeons must consider implant design, expected component size and acetabular component positioning in order to reduce early failures when performing large-bearing metal-on-metal hip resurfacing and replacement.
Complex regional pain syndrome is characterised by an exaggerated response to injury in a limb with intense prolonged pain, vasomotor disturbance, delayed functional recovery and trophic changes. This review describes the current knowledge of the condition and outlines the methods of treatment available with particular emphasis on the knee.
The long-term effects of metal-on-metal arthroplasty are currently under scrutiny because of the potential biological effects of metal wear debris. This review summarises data describing the release, dissemination, uptake, biological activity, and potential toxicity of metal wear debris released from alloys currently used in modern orthopaedics. The introduction of risk assessment for the evaluation of metal alloys and their use in arthroplasty patients is discussed and this should include potential harmful effects on immunity, reproduction, the kidney, developmental toxicity, the nervous system and carcinogenesis.
Achieving arthrodesis of the ankle can be difficult
in the presence of infection, deformity, poor soft tissues and bone loss.
We present a series of 48 patients with complex ankle pathology,
treated with the Ilizarov technique. Infection was present in 30
patients and 30 had significant deformity before surgery. Outcome
was assessed clinically and with patient-reported outcome measures
(Modified American Orthopaedic Foot and Ankle Society (MAOFAS) scale and
the Short-Form (SF-36)). Arthrodesis was achieved in 40 patients with the Ilizarov technique
alone and in six further patients with additional surgery. Infection
was eradicated in all patients at a mean follow-up of 46.6 months
(13 to 162). Successful arthrodesis was less likely in those with
comorbidities and in tibiocalcaneal fusion compared with tibiotalar
fusion. These patients had poor general health scores compared with the
normal population before surgery. The mean MAOFAS score improved
significantly from 24.3 (0 to 90) pre-operatively to 56.2 (30 to
90) post-operatively, but there was only a modest improvement in
general health; the mean SF-36 improved from 44.8 (19 to 66) to
50.1 (21 to 76). There was a major benefit in terms of pain relief. Arthrodesis using the Ilizarov technique is an effective treatment
for complex ankle pathology, with good clinical outcomes and eradication
of infection. However, even after successful arthrodesis general
health scores remain limited. Cite this article:
We have prospectively studied the outcome of infections associated with implants which were retained and treated using a standardised antimicrobial protocol. Over a period of four years, we studied 24 consecutive patients who had symptoms of infection for less than one year, a stable implant, no sinus tract and a known pathogen which was susceptible to recommended antimicrobial agents. The infections involved hip prostheses (14), knee prostheses (5), an internal fixation device (4), and an ankle prosthesis (1). Twenty patients had a successful outcome at a median follow-up of 3.7 years (1.8 to 4.7); four had failure of the implant after a median follow-up of 1.2 years (0.3 to 2.5). The probability of survival without failure of treatment was 96% at one year (95% confidence interval (CI) 88 to 100), 92% at two years (95% CI 80 to 100) and 86% at three years (95% CI 72 to 100). Patients with a short-term infection but with a stable implant, no sinus tract and a known pathogen may be successfully treated by retention of the implant and the use of a standardised regimen of antimicrobial treatment.
Local corticosteroid infiltration is a common practice of treatment
for lateral epicondylitis. In recent studies no statistically significant
or clinically relevant results in favour of corticosteroid injections
were found. The injection of autologous blood has been reported
to be effective for both intermediate and long-term outcomes. It
is hypothesised that blood contains growth factors, which induce
the healing cascade. A total of 60 patients were included in this prospective randomised
study: 30 patients received 2 ml autologous blood drawn from contralateral
upper limb vein + 1 ml 0.5% bupivacaine, and 30 patients received
2 ml local corticosteroid + 1 ml 0.5% bupivacaine at the lateral
epicondyle. Outcome was measured using a pain score and Nirschl
staging of lateral epicondylitis. Follow-up was continued for total
of six months, with assessment at one week, four weeks, 12 weeks
and six months.Objectives
Methods
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.
Peri-articular soft-tissue masses or ‘pseudotumours’
can occur after large-diameter metal-on-metal (MoM) resurfacing
of the hip and conventional total hip replacement (THR). Our aim
was to assess the incidence of pseudotumour formation and to identify
risk factors for their formation in a prospective cohort study. A total of 119 patients who underwent 120 MoM THRs with large-diameter
femoral heads between January 2005 and November 2007 were included
in the study. Outcome scores, serum metal ion levels, radiographs
and CT scans were obtained. Patients with symptoms or an identified
pseudotumour were offered MRI and an ultrasound-guided biopsy. There were 108 patients (109 hips) eligible for evaluation by
CT scan at a mean follow-up of 3.6 years (2.5 to 4.5); 42 patients
(39%) were diagnosed with a pseudotumour. The hips of 13 patients
(12%) were revised to a polyethylene acetabular component with small-diameter
metal head. Patients with elevated serum metal ion levels had a
four times increased risk of developing a pseudotumour. This study shows a substantially higher incidence of pseudotumour
formation and subsequent revisions in patients with MoM THRs than
previously reported. Because most revision cases were identified
only after an intensive screening protocol, we recommend close monitoring
of patients with MoM THR.
We report the findings of an independent review
of 230 consecutive Birmingham hip resurfacings (BHRs) in 213 patients
(230 hips) at a mean follow-up of 10.4 years (9.6 to 11.7). A total
of 11 hips underwent revision; six patients (six hips) died from
unrelated causes; and 13 patients (16 hips) were lost to follow-up.
The survival rate for the whole cohort was 94.5% (95% confidence
interval (CI) 90.1 to 96.9). The survival rate in women was 89.1%
(95% CI 79.2 to 94.4) and in men was 97.5% (95% CI 92.4 to 99.2).
Women were 1.4 times more likely to suffer failure than men. For
each millimetre increase in component size there was a 19% lower
chance of a failure. The mean Oxford hip score was 45.0 (median
47.0, 28 to 48); mean University of California, Los Angeles activity
score was 7.4 (median 8.0, 3 to 9); mean patient satisfaction score
was 1.4 (median 1.0, 0 to 9). A total of eight hips had lysis in
the femoral neck and two hips had acetabular lysis. One hip had
progressive radiological changes around the peg of the femoral component.
There was no evidence of progressive neck narrowing between five
and ten years. Our results confirm that BHR provides good functional outcome
and durability for men, at a mean follow-up of ten years. We are
now reluctant to undertake hip resurfacing in women with this implant.
Impingement syndrome in the shoulder has generally been considered to be a clinical condition of mechanical origin. However, anomalies exist between the pathology in the subacromial space and the degree of pain experienced. These may be explained by variations in the processing of nociceptive inputs between different patients. We investigated the evidence for augmented pain transmission (central sensitisation) in patients with impingement, and the relationship between pre-operative central sensitisation and the outcomes following arthroscopic subacromial decompression. We recruited 17 patients with unilateral impingement of the shoulder and 17 age- and gender-matched controls, all of whom underwent quantitative sensory testing to detect thresholds for mechanical stimuli, distinctions between sharp and blunt punctate stimuli, and heat pain. Additionally Oxford shoulder scores to assess pain and function, and PainDETECT questionnaires to identify ‘neuropathic’ and referred symptoms were completed. Patients completed these questionnaires pre-operatively and three months post-operatively. A significant proportion of patients awaiting subacromial decompression had referred pain radiating down the arm and had significant hyperalgesia to punctate stimulus of the skin compared with controls (unpaired The presence of either hyperalgesia or referred pain pre-operatively resulted in a significantly worse outcome from decompression three months after surgery (unpaired These observations confirm the presence of central sensitisation in a proportion of patients with shoulder pain associated with impingement. Also, if patients had relatively high levels of central sensitisation pre-operatively, as indicated by higher levels of punctate hyperalgesia and/or referred pain, the outcome three months after subacromial decompression was significantly worse.
Limb salvage involving wide resection and reconstruction is now well established for managing musculoskeletal sarcomas. However, involvement of major nerves and vessels with a large volume of muscle and skin may result in a useless limb, contributing to depression and a low quality of life. We have been studying alternative treatments for musculoskeletal sarcoma since 1990, and have recently established a regime using photodynamic surgery with cells labelled with acridine orange, photodynamic therapy with cells treated similarly and radiodynamic treatment using the effect of X-rays on such cells. These techniques have been used after marginal or intralesional resection of tumours since 1999 and have enabled maintenance of excellent limb function in patients with sarcomas.
Frozen shoulder is commonly encountered in general
orthopaedic practice. It may arise spontaneously without an obvious
predisposing cause, or be associated with a variety of local or
systemic disorders. Diagnosis is based upon the recognition of the
characteristic features of the pain, and selective limitation of
passive external rotation. The macroscopic and histological features
of the capsular contracture are well-defined, but the underlying
pathological processes remain poorly understood. It may cause protracted
disability, and imposes a considerable burden on health service
resources. Most patients are still managed by physiotherapy in primary
care, and only the more refractory cases are referred for specialist
intervention. Targeted therapy is not possible and treatment remains predominantly
symptomatic. However, over the last ten years, more active interventions
that may shorten the clinical course, such as capsular distension
arthrography and arthroscopic capsular release, have become more popular. This review describes the clinical and pathological features
of frozen shoulder. We also outline the current treatment options,
review the published results and present our own treatment algorithm.
We review the history and literature of hip resurfacing arthroplasty. Resurfacing and the science behind it continues to evolve. Recent results, particularly from the national arthroplasty registers, have spread disquiet among both surgeons and patients. A hip resurfacing arthroplasty is not a total hip replacement, but should perhaps be seen as a means of delaying it. The time when hip resurfacing is offered to a patient may be different from that for a total hip replacement. The same logic can apply to the timing of revision surgery. Consequently, the comparison of resurfacing with total hip replacement may be a false one. Nevertheless, the need for innovative solutions for young arthroplasty patients is clear. Total hip replacement can be usefully delayed in many of these patients by the use of hip resurfacing arthroplasty.
We report serum metal ion level data in patients with unilateral and bilateral hip resurfacing over a ten-year period. In these patients there is an increase in both cobalt and chromium levels above the accepted reference ranges during the first 18 months after operation. Metal ion levels remain elevated, but decline slowly for up to five years. However, the levels then appear to start rising again in some patients up to the ten-year mark. There was no significant difference in cobalt or chromium levels between men and women. These findings appear to differ from much of the current literature. The clinical significance of a raised metal ion level remains under investigation.
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p <
0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the
We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging. Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.
This conversation represents an attempt by several
arthroplasty surgeons to critique several abstracts presented over
the last year as well as to use them as a jumping off point for trying
to figure out where they fit in into our current understanding of
multiple issues in modern hip and knee arthroplasty.
We describe three patients with pre-ganglionic (avulsion) injuries of the brachial plexus which caused a partial Brown-Séquard syndrome.
We have previously shown that joint distraction and movement with a hinged external fixation device for 12 weeks was useful for repairing a large articular cartilage defect in a rabbit model. We have now investigated the results after six months and one year. The device was applied to 16 rabbits who underwent resection of the articular cartilage and subchondral bone from the entire tibial plateau. In group A (nine rabbits) the device was applied for six months. In group B (seven rabbits) it was in place for six months, after which it was removed and the animals were allowed to move freely for an additional six months. The cartilage remained sound in all rabbits. The areas of type II collagen-positive staining and repaired soft tissue were larger in group B than in group A. These findings provide evidence of long-term persistence of repaired cartilage with this technique and that weight-bearing has a positive effect on the quality of the cartilage.