Narrowing of the femoral neck after resurfacing arthroplasty of the hip has been described previously in both cemented and uncemented hip resurfacing. The natural history of narrowing of the femoral neck is unknown. We retrospectively measured the diameter of the femoral neck in a series of 163 Birmingham hip resurfacings in 163 patients up to a maximum of six years after operation to determine the extent and progression of narrowing. There were 105 men and 58 women with a mean age of 52 years (18 to 82). At a mean follow-up of five years, the mean Harris hip score was 94.8 (47 to 100) and the mean flexion of the hip 112.5° (80° to 160°). There was some narrowing of the femoral neck in 77% (125) of the patients reviewed, and in 27.6% (45) the narrowing exceeded 10% of the diameter of the neck. A multiple logistic regression analysis showed a significant association (chi-squared test (derived from logistic regression) p = 0.01) of narrowing with female gender and a valgus femoral neck/shaft angle. There was no significant association between the range of movement, position or size of the component or radiological lucent lines and narrowing of the neck (chi-squared test; p = 0.10 (flexion), p = 0.08 (size of femoral component), p = 0.09 (size of acetabular component), p = 0.71 (femoral component angulation), p = 0.99 (lucent lines)). There was no significant difference between the diameter of the neck at a mean of three years (2.5 to 3.5) and that at five years (4.5 to 5.5), indicating that any change in the diameter of the neck had stabilised by three years (sign rank test, p = 0.60). We conclude that narrowing of the femoral neck which is found with the Birmingham hip resurfacing arthroplasty is in most cases associated with no adverse clinical or radiological outcome up to a maximum of six years after the initial operation.
We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7). The groups differed significantly from each other in Larsen grade, cementing of components and patellar resurfacing, but no statistically significant difference between the survival of the components was found (Log rank test, p = 0.91). The cumulative success rate of the moulded group was 96.8% (95% confidence interval 93.6% to 98.4%) at five years and 94.4% (95% confidence interval 90.4% to 96.7%) at ten years, and of the modular group 96.2% (95% confidence interval 94% to 97.6%) and 93.6% (95% confidence interval 89.7% to 96%), respectively. Revision was required in 37 total knee replacements, the main causes were infection, pain, loosening of the tibial component and patellar problems. Survival rates for both components were satisfactory.
Polymethylmethacrylate remains one of the most enduring materials in orthopaedic surgery. It has a central role in the success of total joint replacement and is also used in newer techniques such as percutaneous vertebroplasty and kyphoplasty. This article describes the current uses and limitations of polymethylmethacrylate in orthopaedic surgery. It focuses on its mechanical and chemical properties and links these to its clinical performance. The behaviour of antibiotic-loaded bone cement are discussed, together with areas of research that are now shedding light upon the behaviour of this unique biomaterial.
We report the clinical and radiographic outcome of a consecutive series of 138 hydroxyapatite-coated total knee replacements with a mean follow-up of 11 years (10 to 13). The patients were entered into a prospective study and all living patients (76 knees) were evaluated. The Hospital for Special Surgery knee score was obtained for comparison with the pre-operative situation. No patient was lost to follow-up. Radiographic assessment revealed no loosening. Seven prostheses have been revised, giving a cumulative survival rate of 93% at 13 years. We believe this to be the longest follow-up report available for an hydroxyapatite-coated knee replacement and the first for this design of Insall-Burstein II knee.
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.
Ciprofloxacin hydrochloride-loaded microspheres were prepared by a spray-drying method using pectin and chitosan. The effects of different polymers and drug ratios were investigated. The most appropriate carriers were selected by The drug was released rapidly from the pectin carrier but this was more sustained in the chitosan formulation. Chitosan microspheres loaded with ciprofloxacin hydrochloride were more effective for the treatment of osteomyelitis than equivalent intramuscular antibiotics.
There have been no large comparative studies of the blood levels of metal ions after implantation of commercially available hip resurfacing devices which have taken into account the effects of femoral size and inclination and anteversion of the acetabular component. We present the results in 90 patients with unilateral articular surface replacement (ASR) hip resurfacings (mean time to blood sampling 26 months) and 70 patients with unilateral Birmingham Hip Resurfacing (BHR) implants (mean time 47 months). The whole blood and serum chromium (Cr) and cobalt (Co) concentrations were inversely related to the size of the femoral component in both groups (p <
0.05). Cr and Co were more strongly influenced by the position of the acetabular component in the case of the ASR, with an increase in metal ions observed at inclinations >
45° and anteversion angles of <
10° and >
20°. These levels were only increased in the BHR group when the acetabular component was implanted with an inclination >
55°. A significant relationship was identified between the anteversion of the BHR acetabular component and the levels of Cr and Co (p <
0.05 for Co), with an increase observed at anteversion angles <
10° and >
20°. The median whole blood and serum Cr concentrations of the male ASR patients were significantly lower than those of the BHR men (p <
0.001). This indicates that reduced diametral clearance may equate to a reduction in metal ion concentrations in larger joints with satisfactory orientation of the acetabular component.
In a series of 1304 patients (1867 knees), the results of simultaneous and staged bilateral total knee arthroplasty were compared with each other and with unilateral total knee arthroplasty. The bilateral procedures had a significantly higher rate of complications than unilateral procedures, almost entirely because of thromboembolic problems. However, this did not correspond to an increase in mortality. If a bilateral procedure was indicated, then a simultaneous procedure had no increased risk over a staged procedure. There was no increase in cardiovascular complications, the rate of deep-vein thrombosis or pulmonary embolism or mortality. The rate of infection was lower with a bilateral procedure and the overall revision rate was less than 1% in all groups. The prosthesis functioned as well in all groups in the medium and longer term periods. We feel that simultaneous bilateral total knee arthroplasty is a safe and successful procedure when compared with a staged bilateral procedure. It also has the added benefit of single anaesthetic, reduced costs and decreased total recovery time when compared to a staged bilateral procedure. For these reasons it should be considered as an option in the presence of bilateral knee joint disease.
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.
We carried out a clinical and radiological review of 103 cementless primary hip arthroplasties with a tapered rectangular grit-blasted titanium press-fit femoral component and a threaded conical titanium acetabular component at a mean follow-up of 14.4 years (10.2 to 17.1). The mean Harris hip score at the last follow-up was 89.2 (32 to 100). No early loosening and no fracture of the implant were found. One patient needed revision surgery because of a late deep infection. In 11 hips (10.7%), the reason for revision was progressive wear of the polyethylene liner. Exchange of the acetabular component because of aseptic loosening without detectable liner wear was carried out in three hips (2.9%). After 15 years the survivorship with aseptic loosening as the definition for failure was 95.6% for the acetabular component and 100% for the femoral component.
The menisci of the knee have an important role in load-bearing and shock absorption within the joint. They may also function as secondary stabilisers, have a proprioceptive role, and aid the lubrication and nutrition of the articular cartilage. Complete or partial loss of a meniscus can have damaging effects on a knee, leading to serious long-term sequelae. This paper reviews the consequences of meniscectomy and summarises the body of evidence in the literature regarding those factors most relevant to long-term outcome.
We investigated the wear characteristics and clinical performance of four different total hip joint articulations in 114 patients. Wear and migration was measured by roentgenstereophotogrammetric analysis at five years or at the last follow-up. The mean annual wear was 0.11 mm for a stainless steel/Enduron articulation, 0.34 mm for stainless steel/Hylamer cup, 0.17 mm for zirconium oxide ceramic/Enduron and 0.40 mm for zirconium oxide ceramic/Hylamer. The difference between the groups was significant (p <
0.008) except for stainless steel/Hylamer At present, 12 patients have undergone a revision procedure, four at five years and eight thereafter. No patient who received a stainless steel/Enduron articulation at their primary replacement required revision. Conflicting results have been reported about the performance of the zirconium oxide ceramic femoral head, but our findings suggest that it should not be used with a polymethylmethacrylate acetabular component. Hylamer has already been withdrawn from the market.