In a retrospective study we compared 32 HINTEGRA
total ankle replacements (TARs) and 35 Mobility TARs performed between
July 2005 and May 2010, with a minimum follow-up of two years. The
mean follow-up for the HINTEGRA group was 53 months (24 to 76) and
for the Mobility group was 34 months (24 to 45). All procedures
were performed by a single surgeon. There was no significant difference between the two groups with
regard to the mean AOFAS score, visual analogue score for pain or
range of movement of the ankle at the latest follow-up. Most radiological
measurements did not differ significantly between the two groups.
However, the most common grade of heterotopic ossification (HO)
was grade 3 in the HINTEGRA group (10 of 13 TARs, 76.9%) and grade
2 in the Mobility group (four of seven TARs, 57.1%) (p = 0.025).
Although HO was more frequent in the HINTEGRA group (40.6%) than
in the Mobility group (20.0%), this was not statistically significant
(p = 0.065).The difference in peri-operative complications between
the two groups was not significant, but intra-operative medial malleolar
fractures occurred in four (11.4%) in the Mobility group; four (12.5%)
in the HINTEGRA group and one TAR (2.9%) in the Mobility group failed
(p = 0.185). Cite this article:
We present our experience with a double-mobility
acetabular component in 155 consecutive revision total hip replacements
in 149 patients undertaken between 2005 and 2009, with particular
emphasis on the incidence of further dislocation. The mean age of
the patients was 77 years (42 to 89) with 59 males and 90 females.
In all, five patients died and seven were lost to follow-up. Indications
for revision were aseptic loosening in 113 hips, recurrent instability
in 29, peri-prosthetic fracture in 11 and sepsis in two. The mean
follow-up was 42 months (18 to 68). Three hips (2%) in three patients
dislocated within six weeks of surgery; one of these dislocated
again after one year. All three were managed successfully with closed
reduction. Two of the three dislocations occurred in patients who
had undergone revision for recurrent dislocation. All three were
found at revision to have abductor deficiency. There were no dislocations
in those revised for either aseptic loosening or sepsis. These results demonstrate a good mid-term outcome for this component.
In the 29 patients revised for instability, only two had a further
dislocation, both of which were managed by closed reduction.
We systematically reviewed all the evidence published
in the English language on proximal interphalangeal joint (PIPJ)
replacement, to determine its effectiveness on the function of the
hand and the associated post-operative complications. Original studies were selected if they reported clinical outcome
with a minimum of one year’s follow-up. Quality was assessed using
the Cowley systematic review criteria modified for finger-joint
replacements. Of 319 articles identified, only five were adequately
reported according to our quality criteria; there were no randomised
controlled trials. PIPJ replacements had a substantial effect size
on hand pain of -23.2 (95% confidence interval (CI) -27.3 to -19.1)
and grip strength 1.2 (95% CI -10.7 to 13.1), and a small effect
on range of movement 0.2 (95% CI -0.4 to 0.8). A dorsal approach
was most successful. Post-operative loosening occurred in 10% (95%
CI 3 to 30) of ceramic and 12.5% (95% CI 7 to 21) of pyrocarbon
replacements. Post-operative complications occurred in 27.8% (95%
CI 20 to 37). We conclude that the effectiveness of PIPJ replacement has not
been established. Small observational case studies and short-term
follow-up, together with insufficient reporting of patient data,
functional outcomes and complications, limit the value of current
evidence. We recommend that a defined core set of patients, surgical and
outcome data for this intervention be routinely and systematically
collected within the framework of a joint registry.
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.
The Oxford unicompartmental knee replacement gives good results in patients with symptomatic osteoarthritis of the medial compartment. Previous studies have suggested that the presence of radiolucent lines (RLLs) does not reflect a poor outcome in such patients. However, the reliability and validity of this assessment have not been determined. Our aim was to assess the intra- and interobserver reliability and the sensitivity and specificity of the assessment of RLLs around both tibial and femoral components using standard radiographs. Two reviewers assessed the radiographs of 45 patients who had loosening of the tibial or femoral component confirmed at revision surgery and compared them with those of a series of 45 asymptomatic patients matched for age and gender. The results suggested that, using standard radiographs, tibial RLLs were 63.6% sensitive and 94.4% specific and femoral RLLs 63.9% sensitive and 72.7% specific for loosening. Overall intra- and interobserver reliability was highly variable, but zonal analysis showed that lucency at the tip of the femoral peg was significantly associated with loosening of the femoral component. Fluoroscopically guided radiographs may improve the zonal reliability of the assessment of RLLs, but further independent and comparative studies are required. In the meantime, the innocence of the physiological RLLs detected by standard radiographs should be viewed with caution.
Frozen shoulder is commonly encountered in general
orthopaedic practice. It may arise spontaneously without an obvious
predisposing cause, or be associated with a variety of local or
systemic disorders. Diagnosis is based upon the recognition of the
characteristic features of the pain, and selective limitation of
passive external rotation. The macroscopic and histological features
of the capsular contracture are well-defined, but the underlying
pathological processes remain poorly understood. It may cause protracted
disability, and imposes a considerable burden on health service
resources. Most patients are still managed by physiotherapy in primary
care, and only the more refractory cases are referred for specialist
intervention. Targeted therapy is not possible and treatment remains predominantly
symptomatic. However, over the last ten years, more active interventions
that may shorten the clinical course, such as capsular distension
arthrography and arthroscopic capsular release, have become more popular. This review describes the clinical and pathological features
of frozen shoulder. We also outline the current treatment options,
review the published results and present our own treatment algorithm.
This is a retrospective review of the results of the Acclaim total elbow replacement in 11 older patients aged ≥ 65 years with primary osteoarthritis of the elbow, with a mean follow-up of 57.6 months (30 to 86.4). Significant reductions in pain and improvement in range of movement and function were recorded. Radiological review revealed two patients with 1 mm lucencies in a single zone, and one patient with 1 mm lucencies in two zones. No components required revision. There were no deep infections, dislocations or mechanical failures. Complications included one intra-operative medial condylar fracture and one post-operative transient ulnar neuropathy, which resolved. This study demonstrates that the Acclaim prosthesis provides good symptomatic relief and improvement of function in patients with primary osteoarthritis, with low rates of loosening or other complications. This prosthesis can therefore be considered for patients aged ≥ 65 years with primary osteoarthritis of the elbow.
We describe a technique for the diagnosis of loosening of the femoral component of the Oxford Unicompartmental Knee Replacement using accurately aligned lateral radiographs in extension and flexion. If gaps are present between the component and cement on one radiograph and not on the other, the component is loose.
The aim of this study was to compare the outcome of bilateral sequential total ankle replacement (TAR) with that of unilateral TAR. We reviewed 23 patients who had undergone sequential bilateral TAR under a single anaesthetic and 46 matched patients with a unilateral TAR. There were no significant pre-operative differences between the two groups in terms of age, gender, body mass index, American Society of Anaesthesiologists classification and aetiology of the osteoarthritis of the ankle. Clinical and radiological follow-up was carried out at four months, one and two years. After four months, patients with simultaneous bilateral TAR reported a significantly higher mean pain score than those with a unilateral TAR. The mean American Orthopaedic Foot and Ankle Society hindfoot score and short-form 36 physical component summary score were better in the unilateral group. However, this difference disappeared at the one-and two-year follow-ups. Bilateral sequential TAR under one anaesthetic can be offered to patients with bilateral severe ankle osteoarthritis. However, they should be informed of the long recovery period.
We investigated the role of ion release in the assessment of fixation of the implant after total knee replacement and hypothesised that ion monitoring could be a useful parameter in the diagnosis of prosthetic loosening. We enrolled 59 patients with unilateral procedures and measured their serum aluminium, titanium, chromium and cobalt ion levels, blinded to the clinical and radiological outcome which was considered to be the reference standard. The cut-off levels for detection of the ions were obtained by measuring the levels in 41 healthy blood donors who had no implants. Based on the clinical and radiological evaluation the patients were divided into two groups with either stable (n = 24) or loosened (n = 35) implants. A significant increase in the mean level of Cr ions was seen in the group with failed implants (p = 0.001). The diagnostic accuracy was 71% providing strong evidence of failure when the level of Cr ions exceeded the cut-off value. The possibility of distinguishing loosening from other causes of failure was demonstrated by the higher diagnostic accuracy of 83%, when considering only patients with failure attributable to loosening. Measurement of the serum level of Cr ions may be of value for detecting failure due to loosening when the diagnosis is in doubt. The other metal ions studies did not have any diagnostic value.
The contraindications for unicompartmental knee replacement (UKR) remain controversial. The views of many surgeons are based on Kozinn and Scott’s 1989 publication which stated that patients who weighed more than 82 kg, were younger than 60 years, undertook heavy labour, had exposed bone in the patellofemoral joint or chondrocalcinosis, were not ideal candidates for UKR. Our aim was to determine whether these potential contraindications should apply to patients with a mobile-bearing UKR. In order to do this the outcome of patients with these potential contraindications was compared with that of patients without the contraindications in a prospective series of 1000 UKRs. The outcome was assessed using the Oxford knee score, the American Knee Society score, the Tegner activity score, revision rate and survival. The clinical outcome of patients with each of the potential contraindications was similar to or better than those without each contraindication. Overall, 678 UKRs (68%) were performed in patients who had at least one potential contraindication and only 322 (32%) in patients deemed to be ideal. The survival at ten years was 97.0% (95% confidence interval 93.4 to 100.0) for those with potential contraindications and 93.6% (95% confidence interval 87.2 to 100.0) in the ideal patients. We conclude that the thresholds proposed by Kozinn and Scott using weight, age, activity, the state of the patellofemoral joint and chondrocalcinosis should not be considered to be contraindications for the use of the Oxford UKR.
Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities >
10 mm in diameter considered to be ‘marked’. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous coating. Care should be taken with ankle arthroplasty until more is known about the reasons for these severe osteolyses.
We carried out 123 consecutive total ankle replacements in 111 patients with a mean follow-up of four years (2 to 8). Patients with a hindfoot deformity of up to 10° (group A, 91 ankles) were compared with those with a deformity of 11° to 30° (group B, 32 ankles). There were 18 failures (14.6%), with no significant difference in survival between groups A and B. The clinical outcome as measured by the post-operative American Orthopaedic Foot and Ankle Surgeons score was significantly better in group B (p = 0.036). There was no difference between the groups regarding the post-operative range of movement and complications. Correction of the hindfoot deformity was achieved to within 5° of neutral in 27 ankles (84%) of group B patients. However, gross instability was the most common mode of failure in group B. This was not adequately corrected by reconstruction of the lateral ligament. Total ankle replacement can safely be performed in patients with a hindfoot deformity of up to 30°. The importance of adequate correction of alignment and instability is highlighted.
We retrospectively reviewed 35 cemented unicompartmental knee replacements performed for medial unicompartmental osteoarthritis of the knee in 31 patients ≤50 years old (mean 46, 31 to 49). Patients were assessed clinically and radiologically using the Knee Society scores at a mean follow-up of 9.7 years (5 to 16) and survival at 12 years was calculated. The mean Knee Society Function Score improved from 54 points (25 to 64) pre-operatively to 89 (80 to 100) post-operatively (p <
0.0001). Six knees required revision, four for polyethylene wear treated with an isolated exchange of the tibial insert, one for aseptic loosening and one for progression of osteoarthritis. The 12-year survival according to Kaplan-Meier was 80.6% with revision for any reason as the endpoint. Despite encouraging clinical results, polyethylene wear remains a major concern affecting the survival of unicompartmental knee replacement in patients younger than 50.
The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years ( These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.
Cervical spinal disc replacement is used in the management of degenerative cervical disc disease in an attempt to preserve cervical spinal movement and to prevent adjacent disc overload and subsequent degeneration. A large number of patients have undergone cervical spinal disc replacement, but the effectiveness of these implants is still uncertain. In most instances, degenerative change at adjacent levels represents the physiological progression of the natural history of the arthritic disc, and is unrelated to the surgeon. Complications of cervical disc replacement include loss of movement from periprosthetic ankylosis and ossification, neurological deficit, loosening and failure of the device, and worsening of any cervical kyphosis. Strict selection criteria and adherence to scientific evidence are necessary. Only prospective, randomised clinical trials with long-term follow-up will establish any real advantage of cervical spinal disc replacement over fusion.
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
There are no long-term published results on the survival of a third-generation cemented total shoulder replacement. We describe a clinical and radiological study of the Aequalis total shoulder replacement for a minimum of ten years. Between September 1996 and May 1998, 39 consecutive patients underwent a primary cemented total shoulder replacement using this prosthesis. Data were collected prospectively on all patients each year, for a minimum of ten years, or until death or failure of the prosthesis. At a follow-up of at least ten years, 12 patients had died with the prosthesis intact and two had emigrated, leaving 25 available for clinical review. Of these, 13 had rheumatoid arthritis and 12 osteoarthritis. One refused radiological review leaving 24 with fresh radiographs. Survivorship at ten years was 100% for the humeral component and 92% for the glenoid component. The incidence of lucent lines was low. No humeral component was thought to be at risk and only two glenoid components. The osteoarthritic group gained a mean 65° in forward flexion and their Constant score improved by a mean 41.4 points (13 to 55). The rheumatoid group gained a mean of 24° in flexion and their Constant score improved by 29.4 points. This difference may have been due to failure of the rotator cuff in 75% of the patients with rheumatoid arthritis. Thus a third-generation total shoulder replacement gives an excellent result in patients with osteoarthritis and an intact rotator cuff. Patients with rheumatoid arthritis have a 75% risk of failure of the rotator cuff at ten years.
While frequently discussed as a standard treatment for the management of an infected shoulder replacement, there is little information on the outcome of two-stage re-implantation. We examined the outcome of 17 consecutive patients (19 shoulders) who were treated between 1995 and 2004 with a two-stage re-implantation for the treatment of a deep-infection after shoulder replacement. All 19 shoulders were followed for a minimum of two years or until the time of further revision surgery. The mean clinical follow-up was for 35 months (24 to 80). The mean radiological follow-up was 27 months (7 to 80). There were two excellent results, four satisfactory and 13 unsatisfactory. In 12 of the 19 shoulders (63%) infection was considered to be eradicated. The mean pain score improved from 4.2 (3 to 5 (out of 5)) to 1.8 (1 to 4). The mean elevation improved from 42° (0° to 140°) to 89° (0° to 165°), mean external rotation from 30° (0° to 90°) to 43° (0° to 90°), and mean internal rotation from the sacrum to L5. There were 14 complications. Our study suggests that two-stage re-implantation for an infected shoulder replacement is associated with a high rate of unsatisfactory results, marginal success at eradicating infection and a high complication rate.