The cementless Oxford unicompartmental knee replacement
has been demonstrated to have superior fixation on radiographs and
a similar early complication rate compared with the cemented version.
However, a small number of cases have come to our attention where,
after an apparently successful procedure, the tibial component subsides into
a valgus position with an increased posterior slope, before becoming
well-fixed. We present the clinical and radiological findings of
these six patients and describe their natural history and the likely
causes. Two underwent revision in the early post-operative period,
and in four the implant stabilised and became well-fixed radiologically with
a good functional outcome. This situation appears to be avoidable by minor modifications
to the operative technique, and it appears that it can be treated
conservatively in most patients. Cite this article:
Mobile-bearing unicompartmental knee replacements
(UKRs) with a flat tibial plateau have not performed well in the
lateral compartment, owing to a high dislocation rate. This led
to the development of the Domed Lateral Oxford UKR (Domed OUKR)
with a biconcave bearing. The aim of this study was to assess the
survival and clinical outcomes of the Domed OUKR in a large patient
cohort in the medium term. We prospectively evaluated 265 consecutive knees with isolated
disease of the lateral compartment and a mean age at surgery of
64 years (32 to 90). At a mean follow-up of four years ( The Domed Lateral OUKR gives good clinical outcomes, low re-operation
and revision rates and a low dislocation rate in patients with isolated
lateral compartmental disease, in the hands of the designer surgeons. Cite this article:
Bone loss in the proximal tibia and distal femur
is frequently encountered in revision knee replacement surgery.
The various options for dealing with this depend on the extent of
any bone loss. We present our results with the use of cementless
metaphyseal metal sleeves in 103 patients (104 knees) with a mean
follow-up of 43 months (30 to 65). At final follow-up, sleeves in
102 knees had good osseointegration. Two tibial sleeves were revised
for loosening, possibly due to infection. The average pre-operative Oxford Knee Score was 23 (11 to 36)
and this improved to 32 (15 to 46) post-operatively. These early
results encourage us to continue with the technique and monitor
the outcomes in the long term. Cite this article:
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
The significance of weight in the indications
for unicompartmental knee replacement (UKR) is unclear. Our hypothesis was
that weight does not affect the long-term rate of survival of UKRs. We undertook a retrospective study of 212 UKRs at a mean follow-up
of 12 years (7 to 22). The patients were distributed according to
body mass index (BMI; <
The ten-year rates of survival were similar in the two weight
subgroups (≥ 82 kg: 93.5% (95% confidence interval (CI) 66.5 to
96.3); <
82 kg: 92.5% (95% CI 82.5 to 94.1)) and also in the
two BMI subgroups (≥ 30 kg/m2: 92% (95% CI 82.5 to 95.3);
<
30 kg/m2: 94% (95% CI 78.4 to 95.9)). Multimodal
regression analysis revealed that weight plays a part in reducing
the risk of revision with a relative risk of 0.387, although this
did not reach statistical significance (p = 0.662). The results
relating weight and BMI to the clinical outcome were not statistically
significant. Thus, this study confirms that weight does not influence
the long-term rate of survival of UKR. Cite this article:
The outcome of high tibial osteotomy (HTO) deteriorates
with time, and additional procedures may be required. The aim of
this study was to compare the clinical and radiological outcomes
between unicompartmental knee replacement (UKR) and total knee replacement
(TKR) after HTO as well as after primary UKR. A total of 63 patients (63
knees) were studied retrospectively and divided into three groups:
UKR after HTO (group A; n = 22), TKR after HTO (group B; n = 18)
and primary UKR (group C; n = 22). The Oxford knee score (OKS),
Knee Society score (KSS), hip–knee–ankle angles, mechanical axis
and patellar height were evaluated pre- and post-operatively. At
a mean of 64 months (19 to 180) post-operatively the mean OKS was
43.8 (33 to 49), 43.3 (30 to 48) and 42.5 (29 to 48) for groups
A, B and C, respectively (p = 0.73). The mean KSS knee score was
88.8 (54 to 100), 88.11 (51 to 100) and 85.3 (45 to 100) for groups
A, B and C, respectively (p = 0.65), and the mean KSS function score
was 85.0 (50 to 100) in group A, 85.8 (20 to 100) in group B and
79.3 (50 to 100) in group C (p = 0.48). Radiologically the results
were comparable for all groups except for patellar height, with
a higher incidence of patella infra following a previous HTO (p
= 0.02). Cite this article:
The December 2012 Knee Roundup360 looks at: the demand for knee replacement; a Japanese knee outcome score; smoking and TKR; coronal alignment as a determinant of outcome in TKR; fixed flexion; MRI detected knee lesions; and lateral domed Oxford unicompartmental knee replacements.
The Oxford mobile-bearing unicompartmental knee
replacement (UKR) is an effective and safe treatment for osteoarthritis
of the medial compartment. The results in the lateral compartment
have been disappointing due to a high early rate of dislocation
of the bearing. A series using a newly designed domed tibial component
is reported. The first 50 consecutive domed lateral Oxford UKRs in 50 patients
with a mean follow-up of three years (2.0 to 4.3) were included.
Clinical scores were obtained prospectively and Kaplan-Meier survival
analysis was performed for different endpoints. Radiological variables
related to the position and alignment of the components were measured. One patient died and none was lost to follow-up. The cumulative
incidence of dislocation was 6.2% (95% confidence interval (CI)
2.0 to 17.9) at three years. Survival using revision for any reason
and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85
to 99) at three years, respectively. Outcome scores, visual analogue
scale for pain and maximum knee flexion showed a significant improvement
(p <
0.001). The mean Oxford knee score was 43 ( Clinical results are excellent and short-term survival has improved
when compared with earlier series. The risk of dislocation remains
higher using a mobile-bearing UKR in the lateral compartment when
compared with the medial compartment. Patients should be informed
about this complication. To avoid dislocations, care must be taken
not to elevate the lateral joint line.
The purpose of this study was to evaluate the
change in sagittal tibiotalar alignment after total ankle arthroplasty (TAA)
for osteoarthritis and to investigate factors affecting the restoration
of alignment. This retrospective study included 119 patients (120 ankles) who
underwent three component TAA using the Hintegra prosthesis. A total
of 63 ankles had anterior displacement of the talus before surgery
(group A), 49 had alignment in the normal range (group B), and eight
had posterior displacement of the talus (group C). Ankles in group
A were further sub-divided into those in whom normal alignment was
restored following TAA (41 ankles) and those with persistent displacement
(22 ankles). Radiographic and clinical results were assessed. Pre-operatively, the alignment in group A was significantly more
varus than that in group B, and the posterior slope of the tibial
plafond was greater (p <
0.01 in both cases). The posterior slope
of the tibial component was strongly associated with restoration
of alignment: ankles in which the alignment was restored had significantly
less posterior slope (p <
0.001). An anteriorly translated talus was restored to a normal position
after TAA in most patients. We suggest that surgeons performing
TAA using the Hintegra prosthesis should aim to insert the tibial
component at close to 90° relative to the axis of the tibia, hence
reducing posterior soft-tissue tension and allowing restoration
of normal tibiotalar alignment following surgery. Cite this article:
The Oxford unicompartmental knee replacement
(UKR) is an established treatment option in the management of symptomatic
end-stage medial compartmental osteoarthritis (MCOA), which works
well in the young and active patient. However, previous studies
have shown that it is reliable only in the presence of a functionally
intact anterior cruciate ligament (ACL). This review reports the
outcomes, at a mean of five years and a maximum of ten years, of 52
consecutive patients with a mean age of 51 years (36 to 57) who
underwent staged or simultaneous ACL reconstruction and Oxford UKR.
At the last follow-up (with one patient lost to follow-up), the
mean Oxford knee score was 41 ( In summary, ACL reconstruction and Oxford UKR gives good results
in patients with end-stage MCOA secondary to ACL deficiency.
In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career.
We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p <
0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p <
0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
Current analysis of unicondylar knee replacements
(UKRs) by national registries is based on the pooled results of medial
and lateral implants. Consequently, little is known about the differential
performance of medial and lateral replacements and the influence
of each implant type within these pooled analyses. Using data from
the National Joint Registry for England and Wales (NJR) we aimed
to determine the proportion of UKRs implanted on the lateral side
of the knee, and their survival and reason for failure compared
with medial UKRs. By combining information on the side of operation
with component details held on the NJR, we were able to determine
implant laterality (medial
The Oxford unicompartmental knee replacement gives good results in patients with symptomatic osteoarthritis of the medial compartment. Previous studies have suggested that the presence of radiolucent lines (RLLs) does not reflect a poor outcome in such patients. However, the reliability and validity of this assessment have not been determined. Our aim was to assess the intra- and interobserver reliability and the sensitivity and specificity of the assessment of RLLs around both tibial and femoral components using standard radiographs. Two reviewers assessed the radiographs of 45 patients who had loosening of the tibial or femoral component confirmed at revision surgery and compared them with those of a series of 45 asymptomatic patients matched for age and gender. The results suggested that, using standard radiographs, tibial RLLs were 63.6% sensitive and 94.4% specific and femoral RLLs 63.9% sensitive and 72.7% specific for loosening. Overall intra- and interobserver reliability was highly variable, but zonal analysis showed that lucency at the tip of the femoral peg was significantly associated with loosening of the femoral component. Fluoroscopically guided radiographs may improve the zonal reliability of the assessment of RLLs, but further independent and comparative studies are required. In the meantime, the innocence of the physiological RLLs detected by standard radiographs should be viewed with caution.
Aseptic loosening of the femoral component is
an important indication for revision surgery in unicompartmental knee
replacement (UKR). A new design of femoral component with an additional
peg was introduced for the cemented Oxford UKR to increase its stability.
The purpose of this study was to compare the primary stability of
the two designs of component. Medial Oxford UKR was performed in 12 pairs of human cadaver
knees. In each pair, one knee received the single peg and one received
the twin peg design. Three dimensional micromotion and subsidence
of the component in relation to the bone was measured under cyclical
loading at flexion of 40° and 70° using an optical measuring system.
Wilcoxon matched pairs signed-rank test was performed to detect
differences between the two groups. There was no significant difference in the relative micromotion
(p = 0.791 and 0.380, respectively) and subsidence (p = 0.301 and
0.176, respectively) of the component between the two groups at
both angles of flexion. Both designs of component offered good strength
of fixation in this cadaver study. Cite this article:
We describe a technique for the diagnosis of loosening of the femoral component of the Oxford Unicompartmental Knee Replacement using accurately aligned lateral radiographs in extension and flexion. If gaps are present between the component and cement on one radiograph and not on the other, the component is loose.
The amount of anteroposterior laxity required for a good range of movement and knee function in a cruciate-retaining total knee replacement (TKR) continues to be debated. We undertook a retrospective study to evaluate the effects of anteroposterior laxity on the range of movement and knee function in 55 patients following the e-motion cruciate-retaining TKR with a minimum follow-up of two years. The knees were divided into stable (anteroposterior translation, ≤ 10 mm, 38 patients) and unstable (anteroposterior translation, >
10 mm, 17) groups based on the anteroposterior laxity, measured using stress radiographs. We compared the Hospital for Special Surgery (HSS) scores, the Western Ontario MacMasters University Osteoarthritis (WOMAC) index, weight-bearing flexion, non-weight-bearing flexion and the reduction of flexion under weight-bearing There were no differences between the stable and unstable groups with regard to the mean HHS and WOMAC total scores, as well as weight-bearing and non-weight-bearing flexion (p = 0.277, p = 0.082, p = 0.095 and p = 0.646, respectively). However, the stable group had a better WOMAC function score and less delta flexion than the unstable group (p = 0.011 and p = 0.005, respectively). Our results suggest that stable knees with laxity ≤ 10 mm have a good functional outcome and less reduction of flexion under weight-bearing conditions than unstable knees with laxity >
10 mm following an e-motion cruciate-retaining TKR.
We reviewed the rate of revision of unicompartmental knee replacements (UKR) from the New Zealand Joint Registry between 1999 and 2008. There were 4284 UKRs, of which 236 required revision, 205 to a total knee replacement (U2T) and 31 to a further unicompartmental knee replacement (U2U). We used these data to establish whether the survival and functional outcome for revised UKRs were comparable with those of primary total knee replacement (TKR). The rate of revision for the U2T cohort was four times higher than that for a primary TKR (1.97 The poor outcome of a UKR converted to a primary TKR compared with a primary TKR should contra-indicate the use of a UKR as a more conservative procedure in the younger patient.