The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1019 - 1024
1 Aug 2009
Grammatopolous G Pandit H Kwon Y Gundle R McLardy-Smith P Beard DJ Murray DW Gill HS

Inflammatory pseudotumours occasionally occur after metal-on-metal hip resurfacing and often lead to revision. Our aim was to determine the severity of this complication by assessing the outcome of revision in these circumstances and by comparing this with the outcome of other metal-on-metal hip resurfacing revisions as well as that of matched primary total hip replacements.

We identified 53 hips which had undergone metal-on-metal hip resurfacing and required revision at a mean of 1.59 years (0.01 to 6.69) after operation. Of these, 16 were revised for pseudotumours, 21 for fracture and 16 for other reasons. These were matched by age, gender and diagnosis with 103 patients undergoing primary total hip replacement with the Exeter implant.

At a mean follow-up of three years (0.8 to 7.2) the outcome of metal-on-metal hip resurfacing revision for pseudotumour was poor with a mean Oxford hip score of 20.9 (sd 9.3) and was significantly worse (p < 0.001) than the outcome for fracture with a mean Oxford hip score of 40.2 (sd 9.2) or that for other causes with a mean Oxford hip score of 37.8 (sd 9.4). The clinical outcome of revision for pseudotumour was also significantly worse (p < 0.001) than the outcome of matched primary total hip replacements. By contrast, the outcome for fracture and other causes was not significantly different from that of matched primary total hip replacements (p = 0.065). After revision for pseudotumour there were three cases of recurrent dislocation, three of palsy of the femoral nerve, one of stenosis of the femoral artery and two of loosening of the component. Five hips required further revision. In three of these there was evidence of recurrent pseudotumour, and one is currently awaiting further revision. The incidence of major complications after revision for pseudotumour (50%) was significantly higher (p = 0.018) than that after revision for other causes (14%).

The outcome of revision for pseudotumour is poor and consideration should be given to early revision to limit the extent of the soft-tissue destruction. The outcome of resurfacing revision for other causes is good.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 12 | Pages 1566 - 1574
1 Dec 2009
Glyn-Jones S Pandit H Kwon Y Doll H Gill HS Murray DW

Metal-on-metal hip resurfacing is commonly performed for osteoarthritis in young active patients. We have observed cystic or solid masses, which we have called inflammatory pseudotumours, arising around these devices. They may cause soft-tissue destruction with severe symptoms and a poor outcome after revision surgery. The aim of this study was to determine the incidence of and risk factors for pseudotumours that are serious enough to require revision surgery.

Since 1999, 1419 metal-on-metal hip resurfacings have been implanted by our group in 1224 patients; 1.8% of the patients had a revision for pseudotumour. In this series the Kaplan-Meier cumulative revision rate for pseudotumour increased progressively with time. At eight years, in all patients, it was 4% (95% confidence interval (CI) 2.2 to 5.8). Factors significantly associated with an increase in revision rate were female gender (p < 0.001), age under 40 (p = 0.003), small components (p = 0.003), and dysplasia (p = 0.019), whereas implant type was not (p = 0.156). These factors were inter-related, however, and on fitting a Cox proportional hazard model only gender (p = 0.002) and age (p = 0.024) had a significant independent influence on revision rate; size nearly reached significance (p = 0.08). Subdividing the cohort according to significant factors, we found that the revision rate for pseudotumours in men was 0.5% (95% CI 0 to 1.1) at eight years wheras in women over 40 years old it was 6% (95% CI 2.3 to 10.1) at eight years and in women under 40 years it was 13.1% at six years (95% CI 0 to 27) (p < 0.001).

We recommend that resurfacings are undertaken with caution in women, particularly those under 40 years of age but they remain a good option in young men. Further work is required to understand the aetiology of pseudotumours so that this complication can be avoided.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1356 - 1362
1 Oct 2010
Simpson DJ Kendrick BJL Hughes M Glyn-Jones S Gill HS Rushforth GF Murray DW

We have evaluated the difference in the migration patterns over two years of two cementless stems in a randomised, controlled trial using radiostereophotogrammetric analysis (RSA). The implants studied were the Furlong HAC stem, which has good long-term results and the Furlong Active stem, which is a modified version of the former designed to minimise stress concentrations between the implant and bone, and thus to improve fixation.

A total of 23 Furlong HAC and 20 Furlong Active stems were implanted in 43 patients. RSA examinations were carried out immediately post-operatively and at six, 12 and 24 months post-operatively.

The subsidence during the first year in the Furlong HAC stem, was approximately one-third that of the Furlong Active stem, the measured mean subsidence of the femoral head at six months being 0.27 mm (95% confidence interval (CI) 0.03 to 0.51) and 0.99 mm (95% CI 0.38 to 1.60), respectively (p = 0.03). One Active stem continued to subside during the second year. All hips, regardless of the type of stem were clinically successful as judged by the Oxford hip score and a derived pain score without any distinction between the two types of stem.

The initial stability of the Furlong Active stem was not as good as the established stem which might compromise osseo-integration to the detriment of long-term success. The changes in the geometry of the stem, to minimise stress have affected the attainment of initial stability.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 430 - 435
1 Apr 2008
Biant LC Bruce WJM Assini JB Walker PM Walsh WR

This prospective study presents the ten-year (5 to 16) clinical and radiological results of 55 primary total hip replacements (THR) using a cementless modular femoral component (S-ROM). All patients had a significant anatomical abnormality which rendered the primary THR difficult.

The mean Harris hip score was 36 (12 to 72) pre-operatively, 83 (44 to 100) at five years, and 85 (45 to 99) at ten years. The Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and short-form (SF)-12 scores were recorded from the year 2000. The mean SF-12 score at five years after surgery was 45.24 (22.74 to 56.58) for the physical component and 54.14 (29.20 to 66.61) for the mental component. By ten years the SF-12 scores were 42.86 (21.59 to 58.95) and 51.03 (33.78 to 61.40), respectively. The mean WOMAC score at five years post-operatively was 25 (0 to 59), and at ten years was 27 (2 to 70).

No femoral components were radiologically loose, although five had osteolysis in Gruen zone 1, three had osteolysis in zone 7, and two showed osteolysis in both zones 1 and 7. No osteolysis was observed around or distal to the prosthetic sleeve. No femoral components were revised, although three hips underwent an acetabular revision and two required a liner exchange. At a mean of ten years’ follow-up the S-ROM femoral component implanted for an anatomically difficult primary THR has excellent clinical and radiological results.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 27 - 30
1 Jan 2008
Rajaratnam SS Jack C Tavakkolizadeh A George MD Fletcher RJ Hankins M Shepperd JAN

Between 1986 and 1991 we implanted 331 consecutive Furlong hydroxyapatite-coated femoral components of a total hip replacement in 291 patients. A cemented acetabular prosthesis was used in 217 hips and a hydroxyapatite-coated component in 114. We describe the long-term clinical and radiological survival of the femoral component at a mean follow-up of 17.5 years (15 to 21). Only two patients (0.68%) were lost to follow-up. With revision of the femoral component for any reason as the endpoint, the survival at a mean of 17 years was 97.4% (95% confidence interval 94.1 to 99.5), and with revision for aseptic loosening as the endpoint it was 100%. The survival at a maximum of 21 years with revision of the femoral component for any reason as the endpoint was 97.4% (95% confidence interval 81.0 or 99.5). These results compare favourably with the best long-term results of cemented or uncemented femoral components used in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 12 | Pages 1574 - 1580
1 Dec 2007
Hallan G Lie SA Furnes O Engesaeter LB Vollset SE Havelin LI

Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data.

With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well.

Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1176 - 1181
1 Aug 2010
Tayton E Evans S O’Doherty D

We implanted titanium and carbon fibre-reinforced plastic (CFRP) femoral prostheses of the same dimensions into five prosthetic femora. An abductor jig was attached and a 1 kN load applied. This was repeated with five control femora. Digital image correlation was used to give a detailed two-dimensional strain map of the medial cortex of the proximal femur. Both implants caused stress shielding around the calcar. Distally, the titanium implant showed stress shielding, whereas the CFRP prosthesis did not produce a strain pattern which was statistically different from the controls. There was a reduction in strain beyond the tip of both the implants.

This investigation indicates that use of the CFRP stem should avoid stress shielding in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1066 - 1071
1 Aug 2010
Chee YH Teoh KH Sabnis BM Ballantyne JA Brenkel IJ

We compared 55 consecutive total hip replacements performed on 53 morbidly obese patients with osteoarthritis with a matched group of 55 total hip replacements in 53 non-obese patients. The groups were matched for age, gender, prosthesis type, laterality and preoperative Harris Hip Score. They were followed prospectively for five years and the outcomes were assessed using the Harris Hip Score, the Short-form 36 score and radiological findings.

Survival at five years using revision surgery as an endpoint, was 90.9% (95% confidence interval 82.9 to 98.9) for the morbidly obese and 100% for the non-obese patients. The Harris Hip and the Short-form 36 scores were significantly better in the non-obese group (p < 0.001). The morbidly obese patients had a higher rate of complications (22% vs 5%, p = 0.012), which included dislocation and both superficial and deep infection.

In light of these inferior results, morbidly obese patients should be advised to lose weight before undergoing a total hip replacement, and counselled regarding the complications. Despite these poorer results, however, the patients have improved function and quality of life.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 436 - 441
1 Apr 2008
Steffen RT Pandit HP Palan J Beard DJ Gundle R McLardy-Smith P Murray DW Gill HS

Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series.

A total of 610 Birmingham Hip Resurfacing arthroplasties were performed in 532 patients with a mean age of 51.8 years (16.5 to 81.6). They were followed for between two and eight years; 107 patients (120 hips) had been followed up for more than five years. Two patients were lost to follow-up. At a minimum of five years’ follow-up, 79 of 85 hips (93%) had an excellent or good outcome according to the Harris hip score. The mean Oxford hip score was 16.1 points (sd 7.7) and the mean University of California Los Angeles activity score was 6.6 points (sd 1.9). There were no patients with definite radiological evidence of loosening or of narrowing of the femoral neck exceeding 10% of its width. There were 23 revisions (3.8%), giving an overall survival of 95% (95% confidence interval 85.3 to 99.2) at seven years. Fractured neck of femur in 12 hips was the most common indication for revision, followed by aseptic loosening in four. In three hips (three patients) (0.5%), failure was possibly related to metal debris.

Considering that these patients are young and active these results are good, and support the use of resurfacing. Further study is needed to address the early failures, particularly those related to fracture and metal debris.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1068 - 1072
1 Aug 2005
Morag G Zalzal P Liberman B Safir O Flint M Gross AE

Our aim was to determine if the height of the cup, lateralisation or the abduction angle correlated with functional outcome or survivorship in revision total hip replacement in patients with a previous diagnosis of developmental dysplasia of the hip. A retrospective investigation of 51 patients (63 hips) who had undergone revision total hip replacement was performed. The mean duration of follow-up was 119 months. Forty-one patients (52 hips) were available for both determination of functional outcome and survivorship analysis. Ten patients (11 hips) were only available for survivorship analysis.

The height of the cup was found to have a statistically significant correlation with functional outcome and a high hip centre correlated with a worse outcome score. Patients with a hip centre of less than 3.5 cm above the anatomical level had a statistically better survivorship of the cup than those with centres higher than this. Restoration of the height of the centre of the hip to as near the anatomical position as possible improved functional outcome and survivorship of the cup.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1299 - 1302
1 Oct 2007
Ramiah RD Ashmore AM Whitley E Bannister GC

We determined the ten-year life expectancy of 5831 patients who had undergone 6653 elective primary total hip replacements at a regional orthopaedic centre between April 1993 and October 2004. Using hospital, general practitioner and the local health authority records, we recorded the dates of death for those who died following surgery.

The mean age at operation was 67 years (13 to 96) with a male:female ratio of 2:3. Of 1154 patients with a ten-year follow-up 340 (29.5%) had died a mean of 5.6 years (0 to 10) after surgery. Using Kaplan-Meier curves, the ten-year survival was 89% in patients under 65 years at surgery, 75% in patients aged between 65 and 74 years, and 51% in patients over 75.

The standardised mortality rates were considerably higher for patients under 45 years, 20% higher for those between 45 and 64 years, and steadily reduced in patients aged 65 and over.

The survival of cemented hip replacement derived from the Swedish Hip Arthroplasty Register Annual Report 2004 exceeds the life expectancy of patients over the age of 60 in our area, suggesting that cemented hip replacement is the procedure of choice in this population.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 500 - 501
1 Apr 2008
Rose DM Guryel E Acton KJ Clark DW

We describe a case of symptomatic focal femoral osteolysis around a screw hole distal to the hydroxyapatite-coated portion of a cannulated femoral component in a revision hip replacement. No locking screw had been inserted into this, the most proximal of the three distal holes for locking screws. The presence of polyethylene wear debris in the tissue excised from the lesion suggested that it had passed through the cannulated portion of the stem and out of the proximal unfilled distal locking hole, initiating an osteolytic reaction in an otherwise well-fixed stem. This case highlights an important design characteristic of such cannulated, uncemented femoral components. We recommend that the proximal aperture of these cannulated stems be occluded at implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 730 - 735
1 Jun 2007
Derbyshire B Porter ML

We performed a three-year radiostereometric analysis (RSA) study of the Elite Plus femoral component on 25 patients undergoing primary total hip replacement. Additional assessments and measurements from standard radiographs were also made.

Subsidence of the stem occurred at the cement-stem interface. At 36 months the subsidence of the stem centroid was a mean of 0.30 mm (0.02 to 1.28), and was continuing at a slow rate. At the same time point, internal rotation and posterior migration of the femoral head had ceased. One stem migrated excessively and additional assessments suggested that this was probably due to high patient demand.

The failure rate of 4% in our study is consistent with data from arthroplasty registers but contrasts with poor results from another RSA study, and from some clinical studies. We believe that the surgical technique, particularly the use of high-viscosity cement, may have been an important factor contributing to our results.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 838 - 838
1 Jun 2006
BEAULÉ PE


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 8 | Pages 1045 - 1048
1 Aug 2011
Avery PP Baker RP Walton MJ Rooker JC Squires B Gargan MF Bannister GC

We reviewed the seven- to ten-year results of our previously reported prospective randomised controlled trial comparing total hip replacement and hemiarthroplasty for the treatment of displaced intracapsular fracture of the femoral neck. Of our original study group of 81 patients, 47 were still alive.

After a mean follow up of nine years (7 to 10) overall mortality was 32.5% and 51.2% after total hip replacement and hemiarthroplasty, respectively (p = 0.09). At 100 months postoperatively a significantly greater proportion of hemiarthroplasty patients had died (p = 0.026). Three hips dislocated following total hip replacement and none after hemiarthroplasty. In both the total hip replacement and hemiarthroplasty groups a deterioration had occurred in walking distance (p = 0.02 and p < 0.001, respectively). One total hip replacement required revision compared with four hemiarthroplasties which were revised to total hip replacements. All surviving patients with a total hip replacement demonstrated wear of the cemented polyethylene component and all hemiarthroplasties had produced acetabular erosion.

There was lower mortality (p = 0.013) and a trend towards superior function in patients with a total hip replacement in the medium term.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 124 - 130
1 Jan 2009
Deuel CR Jamali AA Stover SM Hazelwood SJ

Bone surface strains were measured in cadaver femora during loading prior to and after resurfacing of the hip and total hip replacement using an uncemented, tapered femoral component. In vitro loading simulated the single-leg stance phase during walking. Strains were measured on the medial and the lateral sides of the proximal aspect and the mid-diaphysis of the femur. Bone surface strains following femoral resurfacing were similar to those in the native femur, except for proximal shear strains, which were significantly less than those in the native femur. Proximomedial strains following total hip replacement were significantly less than those in the native and the resurfaced femur.

These results are consistent with previous clinical evidence of bone loss after total hip replacement, and provide support for claims of bone preservation after resurfacing arthroplasty of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1019 - 1024
1 Aug 2007
Hing CB Young DA Dalziel RE Bailey M Back DL Shimmin AJ

Narrowing of the femoral neck after resurfacing arthroplasty of the hip has been described previously in both cemented and uncemented hip resurfacing. The natural history of narrowing of the femoral neck is unknown. We retrospectively measured the diameter of the femoral neck in a series of 163 Birmingham hip resurfacings in 163 patients up to a maximum of six years after operation to determine the extent and progression of narrowing.

There were 105 men and 58 women with a mean age of 52 years (18 to 82). At a mean follow-up of five years, the mean Harris hip score was 94.8 (47 to 100) and the mean flexion of the hip 112.5° (80° to 160°). There was some narrowing of the femoral neck in 77% (125) of the patients reviewed, and in 27.6% (45) the narrowing exceeded 10% of the diameter of the neck. A multiple logistic regression analysis showed a significant association (chi-squared test (derived from logistic regression) p = 0.01) of narrowing with female gender and a valgus femoral neck/shaft angle. There was no significant association between the range of movement, position or size of the component or radiological lucent lines and narrowing of the neck (chi-squared test; p = 0.10 (flexion), p = 0.08 (size of femoral component), p = 0.09 (size of acetabular component), p = 0.71 (femoral component angulation), p = 0.99 (lucent lines)). There was no significant difference between the diameter of the neck at a mean of three years (2.5 to 3.5) and that at five years (4.5 to 5.5), indicating that any change in the diameter of the neck had stabilised by three years (sign rank test, p = 0.60).

We conclude that narrowing of the femoral neck which is found with the Birmingham hip resurfacing arthroplasty is in most cases associated with no adverse clinical or radiological outcome up to a maximum of six years after the initial operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 586 - 590
1 May 2007
Sauvé P Mountney J Khan T De Beer J Higgins B Grover M

Metal-on-metal bearings for total hip replacement (THR) are becoming increasingly popular. Improved wear characteristics mean that these articulations are being inserted into younger patients in the form of THR and resurfacing procedures. This has led to concerns regarding potential carcinogenicity because of the increased exposure to metal ions that the procedure brings.

We have studied the serum cobalt and chromium concentrations in patients who had primary, well-fixed Ring metal-on-metal THRs for more than 30 years. The levels of cobalt and chromium were elevated by five and three times, respectively compared with those in our reference groups. Metal-on-metal articulations appear to be the source of metal ions throughout the life of the prosthesis. In three patients who had undergone revision of a previous metal-on-metal THR to a metal-on-polyethylene replacement the levels of metal ions were within the normal range. The elevations of cobalt and chromium ions seen in our study were comparable with those in patients with modern metal-on-metal THRs.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1050 - 1054
1 Aug 2005
Shetty AA Slack R Tindall A James KD Rand C

We describe the survival of 134 consecutive JRI Furlong hydroxyapatite-coated uncemented total hip replacements. The mean follow-up was for 14.2 years (13 to 15).

Patients were assessed clinically, using the Merle d’Aubigné and Postel score. Radiographs were evaluated using Gruen zones for the stem and DeLee and Charnley zones for the cup. Signs of subsidence, radiolucent lines, endosteal bone formation (spot welds) and pedestal formation were used to assess fixation and stability of the stem according to Engh’s criteria. Cup angle, migration and radiolucency were used to assess loosening of the cup. The criteria for failure were revision, or impending revision because of pain or loosening. Survival analysis was performed using a life table and the Kaplan-Meier curve.

The mean total Merle d’Aubigné and Postel score was 7.4 pre-operatively and 15.9 at follow-up. During the study period 22 patients died and six were lost to follow-up. None of the cups was revised. One stem was revised for a periprosthetic fracture following a fall but none was revised for loosening, giving a 99% survival at 13 years. Our findings suggest that the long-term results of these hydroxyapatite-coated prostheses are more than satisfactory.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 291 - 295
1 Mar 2007
Langdown AJ Pickard RJ Hobbs CM Clarke HJ Dalton DJN Grover ML

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.