The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 917 - 924
1 Jul 2016
Whittaker RK Hothi HS Meswania JM Berber R Blunn GW Skinner JA Hart AJ

Aims

Surgeons have commonly used modular femoral heads and stems from different manufacturers, although this is not recommended by orthopaedic companies due to the different manufacturing processes.

We compared the rate of corrosion and rate of wear at the trunnion/head taper junction in two groups of retrieved hips; those with mixed manufacturers (MM) and those from the same manufacturer (SM).

Materials and Methods

We identified 151 retrieved hips with large-diameter cobalt-chromium heads; 51 of two designs that had been paired with stems from different manufacturers (MM) and 100 of seven designs paired with stems from the same manufacturer (SM). We determined the severity of corrosion with the Goldberg corrosion score and the volume of material loss at the head/stem junction. We used multivariable statistical analysis to determine if there was a significant difference between the two groups.


Bone & Joint Research
Vol. 5, Issue 7 | Pages 294 - 300
1 Jul 2016
Nishioka H Nakamura E Hirose J Okamoto N Yamabe S Mizuta H

Objectives

The purpose of this study was to clarify the appearance of the reparative tissue on the articular surface and to analyse the properties of the reparative tissue after hemicallotasis osteotomy (HCO) using MRI T1ρ and T2 mapping.

Methods

Coronal T1ρ and T2 mapping and three-dimensional gradient-echo images were obtained from 20 subjects with medial knee osteoarthritis. We set the regions of interest (ROIs) on the full-thickness cartilage of the medial femoral condyle (MFC) and medial tibial plateau (MTP) of the knee and measured the cartilage thickness (mm) and T1ρ and T2 relaxation times (ms). Statistical analysis of time-dependent changes in the cartilage thickness and the T1ρ and T2 relaxation times was performed using one-way analysis of variance, and Scheffe’s test was employed for post hoc multiple comparison.


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1112 - 1118
1 Aug 2016
Pedersen AB Christiansen CF Gammelager H Kahlert J Sørensen HT

Aims

We examined risk of developing acute renal failure and the associated mortality among patients aged > 65 years undergoing surgery for a fracture of the hip.

Patients and Methods

We used medical databases to identify patients who underwent surgical treatment for a fracture of the hip in Northern Denmark between 2005 and 2011. Acute renal failure was classified as stage 1, 2 and 3 according to the Kidney Disease Improving Global Outcome criteria. We computed the risk of developing acute renal failure within five days after surgery with death as a competing risk, and the short-term (six to 30 days post-operatively) and long-term mortality (31 days to 365 days post-operatively). We calculated adjusted hazard ratios (HRs) for death with 95% confidence intervals (CIs).


Bone & Joint Research
Vol. 5, Issue 5 | Pages 175 - 177
1 May 2016
Rubin G Rinott M Wolovelsky A Rosenberg L Shoham Y Rozen N

Objectives

Injectable Bromelain Solution (IBS) is a modified investigational derivate of the medical grade bromelain-debriding pharmaceutical agent (NexoBrid) studied and approved for a rapid (four-hour single application), eschar-specific, deep burn debridement. We conducted an ex vivo study to determine the ability of IBS to dissolve-disrupt (enzymatic fasciotomy) Dupuytren’s cords.

Materials and Methods

Specially prepared medical grade IBS was injected into fresh Dupuytren’s cords excised from patients undergoing surgical fasciectomy. These cords were tested by tension-loading them to failure with the Zwick 1445 (Zwick GmbH & Co. KG, Ulm, Germany) tension testing system.


Bone & Joint 360
Vol. 5, Issue 3 | Pages 26 - 28
1 Jun 2016


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 867 - 873
1 Jul 2016
Dalury DF

As the number of younger and more active patients treated with total knee arthroplasty (TKA) continues to increase, consideration of better fixation as a means of improving implant longevity is required. Cemented TKA remains the reference standard with the largest body of evidence and the longest follow-up to support its use. However, cementless TKA, may offer the opportunity of a more bone-sparing procedure with long lasting biological fixation to the bone. We undertook a review of the literature examining advances of cementless TKA and the reported results.

Cite this article: Bone Joint J 2016;98-B:867–73.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 874 - 883
1 Jul 2016
Ballal MS Pearce CJ Calder JDF

Sporting injuries around the ankle vary from simple sprains that will resolve spontaneously within a few days to severe injuries which may never fully recover and may threaten the career of a professional athlete. Some of these injuries can be easily overlooked altogether or misdiagnosed with potentially devastating effects on future performance. In this review article, we cover some of the common and important sporting injuries involving the ankle including updates on their management and outcomes.

Cite this article: Bone Joint J 2016;98-B:874–83.


Bone & Joint Research
Vol. 5, Issue 6 | Pages 247 - 252
1 Jun 2016
Tabuchi K Soejima T Murakami H Noguchi K Shiba N Nagata K

Objectives

The objective of this study was to determine if the use of fascia lata as a tendon regeneration guide (placed into the tendon canal following harvesting the semitendinosus tendon) would improve the incidence of tissue regeneration and prevent fatty degeneration of the semitendinosus muscle.

Materials and Methods

Bilateral semitendinosus tendons were harvested from rabbits using a tendon stripper. On the inducing graft (IG) side, the tendon canal and semitendinosus tibial attachment site were connected by the fascia lata, which was harvested at the same width as the semitendinosus tendon. On the control side, no special procedures were performed. Two groups of six rabbits were killed at post-operative weeks 4 and 8, respectively. In addition, three healthy rabbits were killed to obtain normal tissue. We evaluated the incidence of tendon tissue regeneration, cross-sectional area of the regenerated tendon tissue and proportion of fatty tissue in the semitendinosus muscle.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 786 - 792
1 Jun 2016
Schotanus MGM Sollie R van Haaren EH Hendrickx RPM Jansen EJP Kort NP

Aims

This prospective randomised controlled trial was designed to evaluate the outcome of both the MRI- and CT-based patient-specific matched guides (PSG) from the same manufacturer.

Patients and Methods

A total of 137 knees in 137 patients (50 men, 87 women) were included, 67 in the MRI- and 70 in the CT-based PSG group. Their mean age was 68.4 years (47.0 to 88.9). Outcome was expressed as the biomechanical limb alignment (centre hip-knee-ankle: HKA-axis) achieved post-operatively, the position of the individual components within 3° of the pre-operatively planned alignment, correct planned implant size and operative data (e.g. operating time and blood loss).


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1578 - 1585
1 Dec 2014
Rankin KS Sprowson AP McNamara I Akiyama T Buchbinder R Costa ML Rasmussen S Nathan SS Kumta S Rangan A

Trauma and orthopaedics is the largest of the surgical specialties and yet attracts a disproportionately small fraction of available national and international funding for health research. With the burden of musculoskeletal disease increasing, high-quality research is required to improve the evidence base for orthopaedic practice. Using the current research landscape in the United Kingdom as an example, but also addressing the international perspective, we highlight the issues surrounding poor levels of research funding in trauma and orthopaedics and indicate avenues for improving the impact and success of surgical musculoskeletal research.

Cite this article: Bone Joint J 2014; 96-B:1578–85.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 64 - 70
1 Jan 2015
Hamilton DF Burnett R Patton JT Howie CR Moran M Simpson AHRW Gaston P

Total knee arthroplasty (TKA) is an established and successful procedure. However, the design of prostheses continues to be modified in an attempt to optimise the functional outcome of the patient.

The aim of this study was to determine if patient outcome after TKA was influenced by the design of the prosthesis used.

A total of 212 patients (mean age 69; 43 to 92; 131 female (62%), 81 male (32%)) were enrolled in a single centre double-blind trial and randomised to receive either a Kinemax (group 1) or a Triathlon (group 2) TKA.

Patients were assessed pre-operatively, at six weeks, six months, one year and three years after surgery. The outcome assessments used were the Oxford Knee Score; range of movement; pain numerical rating scales; lower limb power output; timed functional assessment battery and a satisfaction survey. Data were assessed incorporating change over all assessment time points, using repeated measures analysis of variance longitudinal mixed models. Implant group 2 showed a significantly greater range of movement (p = 0.009), greater lower limb power output (p = 0.026) and reduced report of ‘worst daily pain’ (p = 0.003) over the three years of follow-up. Differences in Oxford Knee Score (p = 0.09), report of ‘average daily pain’ (p = 0.57) and timed functional performance tasks (p = 0.23) did not reach statistical significance. Satisfaction with outcome was significantly better in group 2 (p = 0.001).

These results suggest that patient outcome after TKA can be influenced by the prosthesis used.

Cite this article: Bone Joint J 2015;97-B:64–70.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 579 - 584
1 May 2016
Osman K Panagiotidou AP Khan M Blunn G Haddad FS

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface.

We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice.

Cite this article: Bone Joint J 2016;98-B:579–84.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 616 - 621
1 May 2016
Teeter MG Thoren J Yuan X McCalden RW MacDonald SJ Lanting BA Naudie DDR

Aims

The purpose of the present study was to examine the long-term fixation of a cemented fixed-bearing polished titanium tibial baseplate (Genesis ll).

Patients and Methods

Patients enrolled in a previous two-year prospective trial (n = 35) were recalled at ten years. Available patients (n = 15) underwent radiostereometric analysis (RSA) imaging in a supine position using a conventional RSA protocol. Migration of the tibial component in all planes was compared between initial and ten-year follow-up. Outcome scores including the Knee Society Score, Western Ontario and McMaster Universities Arthritis Index, 12-item Short Form Health Survey, Forgotten Joint Score, and University of California, Los Angeles Activity Score were recorded.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 169 - 174
1 May 2016
Wang Y Chu M Rong J Xing B Zhu L Zhao Y Zhuang X Jiang L

Objectives

Previous genome-wide association studies (GWAS) have reported significant association of the single nucleotide polymorphism (SNP) rs8044769 in the fat mass and obesity-associated gene (FTO) with osteoarthritis (OA) risk in European populations. However, these findings have not been confirmed in Chinese populations.

Methods

We systematically genotyped rs8044769 and evaluated the association between the genetic variants and OA risk in a case-controlled study including 196 OA cases and 442 controls in a northern Chinese population. Genotyping was performed using the Sequenom MassARRAY iPLEX platform.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 564 - 568
1 Apr 2016
Kothari A Bhuva S Stebbins J Zavatsky AB Theologis T

Aims

There is increasing evidence that flexible flatfoot (FF) can lead to symptoms and impairment in health-related quality of life. As such we undertook an observational study investigating the aetiology of this condition, to help inform management. The hypothesis was that as well as increased body mass index (BMI) and increased flexibility of the lower limb, an absent anterior subtalar articulation would be associated with a flatter foot posture.

Patients and Methods

A total of 84 children aged between eight and 15 years old were prospectively recruited. The BMI for each child was calculated, flexibility was assessed using the lower limb assessment scale (LLAS) and foot posture was quantified using the arch height index (AHI). Each child underwent a sagittal T1-weighted MRI scan of at least one foot.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1264 - 1270
1 Sep 2015
Karantana A Scammell BE Davis TRC Whynes DK

This study compares the cost-effectiveness of treating dorsally displaced distal radial fractures with a volar locking plate and percutaneous fixation. It was performed from the perspective of the National Health Service (NHS) using data from a single-centre randomised controlled trial. In total 130 patients (18 to 73 years of age) with a dorsally displaced distal radial fracture were randomised to treatment with either a volar locking plate (n = 66) or percutaneous fixation (n = 64). The methodology was according to National Institute for Health and Care Excellence guidance for technology appraisals. .

There were no significant differences in quality of life scores between groups at any time point in the study. Both groups returned to baseline one year post-operatively.

NHS costs for the plate group were significantly higher (p < 0.001, 95% confidence interval 497 to 930). For an additional £713, fixation with a volar locking plate offered 0.0178 additional quality-adjusted life years in the year after surgery. The incremental cost-effectiveness ratio (ICER) for plate fixation relative to percutaneous fixation at list price was £40 068. When adjusting the prices of the implants for a 20% hospital discount, the ICER was £31 898. Patients who underwent plate fixation did not return to work earlier.

We found no evidence to support the cost-effectiveness, from the perspective of the NHS, of fixation using a volar locking plate over percutaneous fixation for the operative treatment of a dorsally displaced radial fracture.

Cite this article: Bone Joint J 2015;97-B:1264–70.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 116 - 121
1 Apr 2016
Leow JM Clement ND Tawonsawatruk T Simpson CJ Simpson AHRW

Objectives

The radiographic union score for tibial (RUST) fractures was developed by Whelan et al to assess the healing of tibial fractures following intramedullary nailing. In the current study, the repeatability and reliability of the RUST score was evaluated in an independent centre (a) using the original description, (b) after further interpretation of the description of the score, and (c) with the immediate post-operative radiograph available for comparison.

Methods

A total of 15 radiographs of tibial shaft fractures treated by intramedullary nailing (IM) were scored by three observers using the RUST system. Following discussion on how the criteria of the RUST system should be implemented, 45 sets (i.e. AP and lateral) of radiographs of IM nailed tibial fractures were scored by five observers. Finally, these 45 sets of radiographs were rescored with the baseline post-operative radiograph available for comparison.


The Bone & Joint Journal
Vol. 98-B, Issue 2 | Pages 145 - 146
1 Feb 2016
Haddad FS Zagra L


Bone & Joint 360
Vol. 3, Issue 5 | Pages 21 - 22
1 Oct 2014

The October 2014 Shoulder & Elbow Roundup360 looks at: PRP is not effective in tennis elbow; eccentric physiotherapy effective in subacromial pain; dexamethasone in shoulder surgery; arthroscopic remplissage for engaging Hill-Sach’s lesions; a consistent approach to subacromial impingement; delay in fixation of proximal humeral fractures detrimental to outcomes.


The Bone & Joint Journal
Vol. 98-B, Issue 4 | Pages 519 - 525
1 Apr 2016
Lees DA Penny JB Baker P

Aims

The aim of this study was to compare the pain caused by the application of a tourniquet after exsanguination of the upper limb with that occurring after simple elevation.

Patients and Methods

We used 26 healthy volunteers (52 arms), each of whom acted as their own matched control.

The primary outcome measure was the total pain experienced by each volunteer while the tourniquet was inflated for 20 minutes. This was calculated as the area under the pain curve for each individual subject. Secondary outcomes were pain at each time point; the total pain experienced during the recovery phase; the ability to tolerate the tourniquet and the time for full recovery after deflation of the tourniquet.