Hook S Moulder E Yates PJ Burston BJ Whitley E Bannister GC

We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up.

The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips.

All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure.

Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p < 0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p < 0.001).

The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis.

Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1071 - 1076
1 Aug 2012
Lim H Bae J Song S Kim S

Medium-term survivorship of the Oxford phase 3 unicompartmental knee replacement (UKR) has not yet been established in an Asian population. We prospectively evaluated the outcome of 400 phase 3 Oxford UKRs in 320 Korean patients with a mean age at the time of operation of 69 years (48 to 82). The mean follow-up was 5.2 years (1 to 10). Clinical and radiological assessment was carried out pre- and post-operatively. At five years, the mean Knee Society knee and functional scores had increased significantly from 56.2 (30 to 91) pre-operatively to 87.2 (59 to 98) (p = 0.034) and from 59.2 (30 to 93) to 88.3 (50 to 100) (p = 0.021), respectively. The Oxford knee score increased from a mean of 25.8 (12 to 39) pre-operatively to 39.8 (25 to 58) at five years (p = 0.038). The ten-year survival rate was 94% (95% confidence interval 90.1 to 98.0). A total of 14 UKRs (3.5%) required revision. The most common reason for revision was dislocation of the bearing in 12 (3%). Conversion to a total knee replacement was required in two patients who developed osteoarthritis of the lateral compartment.

This is the largest published series of UKR in Korean patients. It shows that the mid-term results after a minimally invasive Oxford phase 3 UKR can yield satisfactory clinical and functional results in this group of patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 937 - 940
1 Jul 2012
Manopoulos P Havet E Pearce O Lardanchet JF Mertl P

This was a retrospective analysis of the medium- to long-term results of 46 TC3 Sigma revision total knee replacements using long uncemented stems in press-fit mode.

Clinical and radiological analysis took place pre-operatively, at two years post-operatively, and at a mean follow-up of 8.5 years (4 to 12). The mean pre-operative International Knee Society (IKS) clinical score was 42 points (0 to 74), improving to 83.7 (52 to 100) by the final follow-up. The mean IKS score for function improved from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up. At the final follow-up 30 knees (65.2%) had an excellent result, seven (15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor result. There were two failures, one with anteroposterior instability and one with aseptic loosening.

The TC3 revision knee system, when used with press-fit for long intramedullary stems and cemented femoral and tibial components, in both septic and aseptic revisions, results in a satisfactory clinical and radiological outcome, and has a good medium- to long-term survival rate.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 583 - 597
1 May 2013
Kurien T Pearson RG Scammell BE

We reviewed 59 bone graft substitutes marketed by 17 companies currently available for implantation in the United Kingdom, with the aim of assessing the peer-reviewed literature to facilitate informed decision-making regarding their use in clinical practice. After critical analysis of the literature, only 22 products (37%) had any clinical data. Norian SRS (Synthes), Vitoss (Orthovita), Cortoss (Orthovita) and Alpha-BSM (Etex) had Level I evidence. We question the need for so many different products, especially with limited published clinical evidence for their efficacy, and conclude that there is a considerable need for further prospective randomised trials to facilitate informed decision-making with regard to the use of current and future bone graft substitutes in clinical practice.

Cite this article: Bone Joint J 2013;95-B:583–97.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1475 - 1481
1 Nov 2012
Berton C Puskas GJ Christofilopoulos P Stern R Hoffmeyer P Lübbeke A

There are no recent studies comparing cable with wire for the fixation of osteotomies or fractures in total hip replacement (THR). Our objective was to evaluate the five-year clinical and radiological outcomes and complication rates of the two techniques. We undertook a review including all primary and revision THRs performed in one hospital between 1996 and 2005 using cable or wire fixation. Clinical and radiological evaluation was performed five years post-operatively. Cables were used in 51 THRs and wires in 126, and of these, 36 THRs with cable (71%) and 101 with wire (80%) were evaluated at follow-up. The five-year radiographs available for 33 cable and 91 wire THRs revealed rates of breakage of fixation of 12 of 33 (36%) and 42 of 91 (46%), respectively. With cable there was a significantly higher risk of metal debris (68% vs 9%; adjusted relative risk (RR) 6.6; 95% confidence interval (CI) 3.0 to 14.1), nonunion (36% vs 21%; adjusted RR 2.0; 95% CI 1.0 to 3.9) and osteolysis around the material, acetabulum or femur (61% vs 19%; adjusted RR 3.9; 95% CI 2.3 to 6.5). Cable breakage increased the risk of osteolysis to 83%. There was a trend towards foreign-body reaction and increased infection with cables. Clinical results did not differ between the groups.

In conclusion, we found a higher incidence of complications and a trend towards increased infection and foreign-body reaction with the use of cables.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 7 | Pages 1006 - 1012
1 Jul 2010
Davids JR Hydorn C Dillingham C Hardin JW Pugh LI

We have reviewed our experience of the removal of deep extremity orthopaedic implants in children to establish the nature, rate and risk of complications associated with this procedure. A retrospective review was performed of 801 children who had 1223 implants inserted and subsequently removed over a period of 17 years. Bivariate analysis of possible predictors including clinical factors, complications associated with implant insertion and indications for removal and the complications encountered at removal was performed. A logistical regression model was then constructed using those predictors which were significantly associated with surgical complications from the bivariate analyses. Odds ratios estimated in the logistical regression models were converted to risk ratios.

The overall rate of complications after removal of the implant was 12.5% (100 complications in 801 patients), with 48 (6.0%) major and 52 (6.5%) minor. Children with a complication after insertion of the initial implant or with a non-elective indication for removal, a neuromuscular disease associated with a seizure disorder or a neuromuscular disease in those unable to walk, had a significantly greater chance of having a major complication after removal of the implant. Children with all four of these predictors were 14.6 times more likely to have a major complication.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 310 - 313
1 Mar 2005
Buckley SC Stockley I Hamer AJ Kerry RM

We report the results of the revision of 123 acetabular components for aseptic loosening treated by impaction bone grafting using frozen, morsellised, irradiated femoral heads and cemented sockets. This is the first large series using this technique to be reported. A survivorship of 88% with revision as the end-point after a mean of five years is comparable with that of other series.


Bone & Joint 360
Vol. 1, Issue 2 | Pages 23 - 25
1 Apr 2012

The April 2012 Spine Roundup360 looks at yoga for lower back pain, spinal tuberculosis, complications of spinal surgery, fusing the subaxial cervical spine, minimally invasive surgery and osteoporotic vertebral fractures, spinal surgery in the over 65s, and pain relief after spinal surgery


Bone & Joint 360
Vol. 1, Issue 1 | Pages 23 - 24
1 Feb 2012


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 889 - 894
1 Jul 2012
Burston BJ Barnett AJ Amirfeyz R Yates PJ Bannister GC

We prospectively followed 191 consecutive collarless polished tapered (CPT) femoral stems, implanted in 175 patients who had a mean age at operation of 64.5 years (21 to 85). At a mean follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were still alive. The fate of all original stems is known. The 16-year survivorship with re-operation for any reason was 80.7% (95% confidence interval 72 to 89.4). There was no loss to follow-up, with clinical data available on all 95 hips and radiological assessment performed on 90 hips (95%). At latest follow-up, the mean Harris hip score was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48). Stems subsided within the cement mantle, with a mean subsidence of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem (0.5%) has been revised due to aseptic loosening. In total seven stems were revised for any cause, of which four revisions were required for infection following revision of the acetabular component. A total of 21 patients (11%) required some sort of revision procedure; all except three of these resulted from failure of the acetabular component. Cemented acetabular components had a significantly lower revision burden (three hips, 2.7%) than Harris Galante uncemented components (17 hips, 21.8%) (p < 0.001).

The CPT stem continues to provide excellent radiological and clinical outcomes at 15 years following implantation. Its results are consistent with other polished tapered stem designs.


Bone & Joint 360
Vol. 1, Issue 3 | Pages 30 - 33
1 Jun 2012

The June 2012 Research Roundup360 looks at: platelet-rich plasma; ageing, bone and mesenchymal stem cells; cytokines and the herniated intervertebral disc; ulcerative colitis, Crohn’s disease and anti-inflammatories; the effect of NSAIDs on bone healing; osteoporosis of the fractured hip; herbal medicine and recovery after acute muscle injury; and ultrasound and the time to fracture union.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1656 - 1658
1 Dec 2006
Hatano H Morita T Kobayashi H Otsuka H

Despite extensive experience with prosthetic replacement for the reconstruction of limbs following juxta-articular resection of tumours, there are few reports of prosthetic replacement of the distal radius. We present two cases of massive bone defects of the distal radius in which alumina ceramic prosthetic replacements were used. We evaluated the patients more than ten years after the procedure. Both patients had degenerative changes to the wrist. This, however, was not associated with pain or decreased function, and both had returned to their previous occupation after surgery.

When a patient has a massive defect of the distal radius, reconstruction using a ceramic prosthesis is a reasonable alternative to using autograft. This method of treatment results in little pain, a moderate range of movement and satisfactory function.


Bone & Joint Research
Vol. 1, Issue 10 | Pages 272 - 280
1 Oct 2012
De Mattos CBR Binitie O Dormans JP

Pathological fractures in children can occur as a result of a variety of conditions, ranging from metabolic diseases and infection to tumours. Fractures through benign and malignant bone tumours should be recognised and managed appropriately by the treating orthopaedic surgeon. The most common benign bone tumours that cause pathological fractures in children are unicameral bone cysts, aneurysmal bone cysts, non-ossifying fibromas and fibrous dysplasia. Although pathological fractures through a primary bone malignancy are rare, these should be recognised quickly in order to achieve better outcomes. A thorough history, physical examination and review of plain radiographs are crucial to determine the cause and guide treatment. In most benign cases the fracture will heal and the lesion can be addressed at the time of the fracture, or after the fracture is healed. A step-wise and multidisciplinary approach is necessary in caring for paediatric patients with malignancies. Pathological fractures do not have to be treated by amputation; these fractures can heal and limb salvage can be performed when indicated.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 623 - 627
1 May 2005
Kim Y

Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20).

The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 776 - 780
1 Jun 2005
Kim Y Kim S Kim S Park B Kim P Ihn J

We performed 114 consecutive primary total hip arthroplasties with a cementless expansion acetabular component in 101 patients for advanced osteonecrosis of the femoral head. The mean age of the patients at surgery was 51 years (36 to 62) and the mean length of follow-up was 110 months (84 to 129).

The mean pre-operative Harris hip score of 47 points improved to 93 points at final follow-up. The polyethylene liner was exchanged in two hips during this period and one broken acetabular component was revised. The mean linear wear rate of polyethylene was 0.07 mm/year and peri-acetabular osteolysis was seen in two hips (1.9%). Kaplan-Meier analysis indicated that the survival of the acetabular component without revision was 97.8% (95% confidence interval 0.956 to 1.000) at ten years.

Our study has shown that the results of THA with a cementless expansion acetabular component and an alumina-polyethylene bearing surface are good.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1068 - 1072
1 Aug 2005
Morag G Zalzal P Liberman B Safir O Flint M Gross AE

Our aim was to determine if the height of the cup, lateralisation or the abduction angle correlated with functional outcome or survivorship in revision total hip replacement in patients with a previous diagnosis of developmental dysplasia of the hip. A retrospective investigation of 51 patients (63 hips) who had undergone revision total hip replacement was performed. The mean duration of follow-up was 119 months. Forty-one patients (52 hips) were available for both determination of functional outcome and survivorship analysis. Ten patients (11 hips) were only available for survivorship analysis.

The height of the cup was found to have a statistically significant correlation with functional outcome and a high hip centre correlated with a worse outcome score. Patients with a hip centre of less than 3.5 cm above the anatomical level had a statistically better survivorship of the cup than those with centres higher than this. Restoration of the height of the centre of the hip to as near the anatomical position as possible improved functional outcome and survivorship of the cup.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 6 | Pages 768 - 773
1 Jun 2012
Wang Q Zhang XL Chen YS Shen H Shao JJ

In this prospective study a total of 80 consecutive Chinese patients with Crowe type I or II developmental dysplasia of the hip were randomly assigned for hip resurfacing arthroplasty (HRA) or total hip replacement (THR).

Three patients assigned to HRA were converted to THR, and three HRA patients and two THR patients were lost to follow-up. This left a total of 34 patients (37 hips) who underwent HRA and 38 (39 hips) who underwent THR. The mean follow-up was 59.4 months (52 to 70) in the HRA group and 60.6 months (50 to 72) in the THR group. There was no failure of the prosthesis in either group. Flexion of the hip was significantly better after HRA, but there was no difference in the mean post-operative Harris hip scores between the groups. The mean size of the acetabular component in the HRA group was significantly larger than in the THR group (49.5 mm vs 46.1 mm, p = 0.001). There was no difference in the mean abduction angle of the acetabular component between the two groups.

Although the patients in this series had risk factors for failure after HRA, such as low body weight, small femoral heads and dysplasia, the clinical results of resurfacing in those with Crowe type I or II hip dysplasia were satisfactory. Patients in the HRA group had a better range of movement, although neck-cup impingement was observed. However, more acetabular bone was sacrificed in HRA patients, and it is unclear whether this will have an adverse effect in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 179 - 185
1 Jan 2010
Väänänen P Pajamäki I Paakkala A Nurmi JT Pajamäki J

We used a biodegradable mesh to convert an acetabular defect into a contained defect in six patients at total hip replacement. Their mean age was 61 years (46 to 69). The mean follow-up was 32 months (19 to 50). Before clinical use, the strength retention and hydrolytic in vitro degradation properties of the implants were studied in the laboratory over a two-year period. A successful clinical outcome was determined by the radiological findings and the Harris hip score.

All the patients had a satisfactory outcome and no mechanical failures or other complications were observed. No protrusion of any of the impacted grafts was observed beyond the mesh. According to our preliminary laboratory and clinical results the biodegradable mesh is suitable for augmenting uncontained acetabular defects in which the primary stability of the implanted acetabular component is provided by the host bone. In the case of defects of the acetabular floor this new application provides a safe method of preventing graft material from protruding excessively into the pelvis and the mesh seems to tolerate bone-impaction grafting in selected patients with primary and revision total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1261 - 1267
1 Sep 2007
Tohyama H Yasuda K Uchida H Nishihira J

In order to clarify the role of cytokines in the remodelling of the grafted tendon for ligament reconstruction we compared the responses to interleukin (IL)-1β, platelet-derived growth factor (PDGF)-BB and transforming growth factor (TGF)-β1 of extrinsic fibroblasts infiltrating the frozen-thawed patellar tendon in rats with that of the normal tendon fibroblasts, in regard to the gene expression of matrix metalloproteinase (MMP)-13, using Northern blot analysis. We also examined, immunohistologically, the local expression of IL-1β, PDGF-BB, and TGF-β1 in fibroblasts infiltrating the frozen-thawed patellar tendon.

Northern blot analysis showed that fibroblasts derived from the patellar tendon six weeks after the freeze-thaw procedure in situ showed less response to IL-1β than normal tendon fibroblasts with respect to MMP-13 mRNA gene expression. The immunohistological findings revealed that IL-1β was over-expressed in extrinsic fibroblasts which infiltrated the patellar tendon two and six weeks after the freeze-thaw procedure in situ, but neither PDGF-BB nor TGF-β1 was over-expressed in these extrinsic fibroblasts. Our findings indicated that IL-1β had a close relationship to matrix remodelling of the grafted tendon for ligament reconstruction, in addition to the commencement of inflammation during the tissue-healing process.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 116 - 120
1 Jan 2007
Laing AJ Dillon JP Condon E Coffey JC Street JT Wang JH McGuinness AJ Redmond HP

Post-natal vasculogenesis, the process by which vascular committed bone marrow stem cells or endothelial precursor cells migrate, differentiate and incorporate into the nacent endothelium and thereby contribute to physiological and pathological neurovascularisation, has stimulated much interest. Its contribution to neovascularisation of tumours, wound healing and revascularisation associated with ischaemia of skeletal and cardiac muscles is well established. We evaluated the responses of endothelial precursor cells in bone marrow to musculoskeletal trauma in mice.

Bone marrow from six C57 Black 6 mice subjected to a standardised, closed fracture of the femur, was analysed for the combined expression of cell-surface markers stem cell antigen 1 (sca-1+) and stem cell factor receptor, CD117 (c-kit+) in order to identify the endothelial precursor cell population. Immunomagnetically-enriched sca-1+ mononuclear cell (MNCsca-1+) populations were then cultured and examined for functional vascular endothelial differentiation. Bone marrow MNCsca-1+,c-kit+ counts increased almost twofold within 48 hours of the event, compared with baseline levels, before decreasing by 72 hours.

Sca-1+ mononuclear cell populations in culture from samples of bone marrow at 48 hours bound together Ulex Europus-1, and incorporated fluorescent 1,1′-dioctadecyl- 3,3,3,’3′-tetramethylindocarbocyanine perchlorate-labelled acetylated low-density lipoprotein intracellularily, both characteristics of mature endothelium.

Our findings suggest that a systemic provascular response of bone marrow is initiated by musculoskeletal trauma. Its therapeutic manipulation may have implications for the potential enhancement of neovascularisation and the healing of fractures.