Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


The Bone & Joint Journal
Vol. 100-B, Issue 3 | Pages 331 - 337
1 Mar 2018
Inui H Nobuhara K

Aims

We report the clinical results of glenoid osteotomy in patients with atraumatic posteroinferior instability associated with glenoid dysplasia.

Patients and Methods

The study reports results in 211 patients (249 shoulders) with atraumatic posteroinferior instability. The patients comprised 63 men and 148 women with a mean age of 20 years. The posteroinferior glenoid surface was elevated by osteotomy at the scapular neck. A body spica was applied to maintain the arm perpendicular to the glenoid for two weeks postoperatively. Clinical results were evaluated using the Rowe score and Japan Shoulder Society Shoulder Instability Score (JSS-SIS); bone union, osteoarthrosis, and articular congruity were examined on plain radiographs.


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1596 - 1602
1 Dec 2017
Dunbar MJ Laende EK Collopy D Richardson CG

Aims

Hydroxyapatite coatings for uncemented fixation in total knee arthroplasty can theoretically provide a long-lasting biological interface with the host bone. The objective of this study was to test this hypothesis with propriety hydroxyapatite, peri-apatite, coated tibial components using component migration measured with radiostereometric analysis over two years as an indicator of long-term fixation.

Patients and Methods

A total of 29 patients at two centres received uncemented PA-coated tibial components and were followed for two years with radiostereometric analysis exams to quantify the migration of the component.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 116 - 119
1 Jan 2016
Petrie JR Haidukewych GJ

Instability is a common indication for early revision after both primary and revision total knee arthroplasty (TKA), accounting for up to 20% in the literature. The number of TKAs performed annually continues to climb exponentially, thus having an effective algorithm for treatment is essential. This relies on a thorough pre- and intra-operative assessment of the patient. The underlying cause of the instability must be identified initially and subsequently, the surgeon must be able to balance the flexion and extension gaps and be comfortable using a variety of constrained implants.

This review describes the assessment of the unstable TKA, and the authors’ preferred form of treatment for these difficult cases where the source of instability is often multifactorial.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):116–19.


Bone & Joint 360
Vol. 5, Issue 1 | Pages 14 - 16
1 Feb 2016


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 622 - 627
1 May 2016
Donaldson JR Tudor F Gollish J

Aims

The aim of this study was to examine the results of revision total knee arthroplasty (TKA) undertaken for stiffness in the absence of sepsis or loosening.

Patients and Methods

We present the results of revision surgery for stiff TKA in 48 cases (35 (72.9%) women and 13 (27.1%) men). The mean age at revision surgery was 65.5 years (42 to 83). All surgeries were performed by a single surgeon. Stiffness was defined as an arc of flexion of < 70° or a flexion contracture of > 15°. The changes in the range of movement (ROM) and the Western Ontario and McMasters Osteoarthritis index scores (WOMAC) were recorded.


Objectives

Preservation of both anterior and posterior cruciate ligaments in total knee arthroplasty (TKA) can lead to near-normal post-operative joint mechanics and improved knee function. We hypothesised that a patient-specific bicruciate-retaining prosthesis preserves near-normal kinematics better than standard off-the-shelf posterior cruciate-retaining and bicruciate-retaining prostheses in TKA.

Methods

We developed the validated models to evaluate the post-operative kinematics in patient-specific bicruciate-retaining, standard off-the-shelf bicruciate-retaining and posterior cruciate-retaining TKA under gait and deep knee bend loading conditions using numerical simulation.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 793 - 798
1 Jun 2017
Anderson FL Koch CN Elpers ME Wright TM Haas SB Heyse TJ

Aims

We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty.

Materials and Methods

A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1640 - 1644
1 Dec 2015
Dzaja I Vasarhelyi EM Lanting BA Naudie DD Howard JL Somerville L McCalden RW MacDonald SJ

The purpose of this study was to compare clinical outcomes of total knee arthroplasty (TKA) after manipulation under anaesthesia (MUA) for post-operative stiffness with a matched cohort of TKA patients who did not requre MUA.

In total 72 patients (mean age 59.8 years, 42 to 83) who underwent MUA following TKA were identified from our prospective database and compared with a matched cohort of patients who had undergone TKA without subsequent MUA. Patients were evaluated for range of movement (ROM) and clinical outcome scores (Western Ontario and McMaster Universities Arthritis Index, Short-Form Health Survey, and Knee Society Clinical Rating System) at a mean follow-up of 36.4 months (12 to 120). MUA took place at a mean of nine weeks (5 to 18) after TKA. In patients who required MUA, mean flexion deformity improved from 10° (0° to 25°) to 4.4° (0° to 15°) (p < 0.001), and mean range of flexion improved from 79.8° (65° to 95°) to 116° (80° to 130°) (p < 0.001). There were no statistically significant differences in ROM or functional outcome scores at three months, one year, or two years between those who required MUA and those who did not. There were no complications associated with manipulation.

At most recent follow-up, patients requiring MUA achieved equivalent ROM and clinical outcome scores when compared with a matched control group. While other studies have focused on ROM after manipulation, the current study adds to current literature by supplementing this with functional outcome scores.

Cite this article: Bone Joint J 2015;97-B:1640–4.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 794 - 798
1 Jun 2010
Minoda Y Ikebuchi M Kobayashi A Iwaki H Inori F Nakamura H

Bone mineral density (BMD) around the femoral component has been reported to decrease after total knee replacement (TKR) because of stress shielding. Our aim was to determine whether a cemented mobile-bearing component reduced the post-operative loss of BMD. In our study 28 knees receiving a cemented fixed-bearing TKR were matched with 28 receiving a cemented mobile-bearing TKR. They underwent dual-energy x-ray absorptiometry, pre-operatively and at three weeks and at three, six, 12, 18 and 24 months post-operatively. The patients were not taking medication to improve the BMD.

The pre-operative differences in the BMD of the femoral neck, wrist, lumbar spine and knee in the two groups were not significant. The BMD of the femur decreased postoperatively in the fixed-bearing group, but not the mobile-bearing group. The difference in the post-operative change in the BMD in the two groups was statistically significant (p < 0.05) at 18 and 24 months.

Our findings show that a cemented mobile-bearing TKR has a favourable effect on the BMD of the distal femur after TKR in the short term. Further study is required to determine the long-term effects.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 105 - 112
1 Jan 2016
Khan M Osman K Green G Haddad FS

Total knee arthroplasty (TKA) is a cost effective and extremely successful operation. As longevity increases, the demand for primary TKA will continue to rise. The success and survivorship of TKAs are dependent on the demographics of the patient, surgical technique and implant-related factors.

Currently the risk of failure of a TKA requiring revision surgery ten years post-operatively is 5%.

The most common indications for revision include aseptic loosening (29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses considerable clinical burdens on patients and financial burdens on healthcare systems.

We present a current concepts review on the epidemiology of failed TKAs using data from worldwide National Joint Registries.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):105–12.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 475 - 480
1 Apr 2009
Matsumoto T Kuroda R Kubo S Muratsu H Mizuno K Kurosaka M

We have developed a new tensor for total knee replacements which is designed to assist with soft-tissue balancing throughout the full range of movement with a reduced patellofemoral joint. Using this tensor in 40 patients with osteoarthritis we compared the intra-operative joint gap in cruciate-retaining and posterior-stabilised total knee replacements at 0°, 10°, 45°, 90° and 135° of flexion, with the patella both everted and reduced.

While the measurement of the joint gap with a reduced patella in posterior-stabilised knees increased from extension to flexion, it remained constant for cruciate-retaining joints throughout a full range of movement. The joint gaps at deep knee flexion were significantly smaller for both types of prosthetic knee when the patellofemoral joint was reduced (p < 0.05).


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 217 - 223
1 Feb 2014
Namba RS Inacio MCS Cafri G

The outcome of total knee replacement (TKR) using components designed to increase the range of flexion is not fully understood. The short- to mid-term risk of aseptic revision in high flexion TKR was evaluated. The endpoint of the study was aseptic revision and the following variables were investigated: implant design (high flexion vs non-high flexion), the thickness of the tibial insert (≤ 14 mm vs > 14 mm), cruciate ligament (posterior stabilised (PS) vs cruciate retaining), mobility (fixed vs rotating), and the manufacturer (Zimmer, Smith & Nephew and DePuy). Covariates included patient, implant, surgeon and hospital factors. Marginal Cox proportional hazard models were used.

In a cohort of 64 000 TKRs, high flexion components were used in 8035 (12.5%). The high flexion knees with tibial liners of thickness > 14 mm had a density of revision of 1.45/100 years of observation, compared with 0.37/100 in non-high flexion TKR with liners ≤ 14 mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer, Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence interval (CI) 1.48 to 3.50)), an effect that was magnified when a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)).

Surgeons should be cautious when choosing high flexion TKRs, particularly when thicker tibial liners might be required.

Cite this article: Bone Joint J 2014;96-B:217–23.


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1185 - 1188
1 Sep 2016
Hommel H Perka C Kopf S

Aims

It is widely held that most Baker’s cysts resolve after treatment of the intra-articular knee pathology. The present study aimed to evaluate the fate of Baker’s cysts and their associated symptoms after total knee arthroplasty (TKA).

Patients and Methods

In this prospective cohort study, 102 patients with (105 were included, however three were lost to follow-up) an MRI-verified Baker’s cyst, primary osteoarthritis and scheduled for TKA were included. Ultrasound was performed to evaluate the existence and the gross size of the cyst before and at one year after TKA. Additionally, associated symptoms of Baker's cyst were recorded pre- and post-operatively.


Bone & Joint 360
Vol. 5, Issue 5 | Pages 27 - 29
1 Oct 2016


Bone & Joint 360
Vol. 3, Issue 4 | Pages 14 - 16
1 Aug 2014

The August 2014 Knee Roundup360 looks at: re-admission following total knee replacement; out with the old and in with the new? computer navigation revisited; approach less important in knee replacement; is obesity driving a rise in knee replacements?; knee replacement isn’t cheap in the obese; cruciate substitution doesn’t increase knee flexion; and sonication useful diagnostic aid in two-stage revision.


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1050 - 1055
1 Aug 2016
Karachalios T Varitimidis S Bargiotas K Hantes M Roidis N Malizos KN

Aims

The Advance Medial-Pivot total knee arthroplasty (TKA) was designed to reflect contemporary data regarding the kinematics of the knee. We wished to examine the long-term results obtained with this prosthesis by extending a previous evaluation.

Patients and Methods

We retrospectively evaluated prospectively collected data from 225 consecutive patients (41 men and 184 women; mean age at surgery 71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up of 13.4 years (11 to 15). Implant failure, complication rate, clinical (both subjective and objective) and radiological outcome were assessed. Pre- and post-operative clinical and radiographic data were available at regular intervals for all patients. A total of ten patients (4.4%; ten TKAs) were lost to follow-up.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 867 - 873
1 Jul 2016
Dalury DF

As the number of younger and more active patients treated with total knee arthroplasty (TKA) continues to increase, consideration of better fixation as a means of improving implant longevity is required. Cemented TKA remains the reference standard with the largest body of evidence and the longest follow-up to support its use. However, cementless TKA, may offer the opportunity of a more bone-sparing procedure with long lasting biological fixation to the bone. We undertook a review of the literature examining advances of cementless TKA and the reported results.

Cite this article: Bone Joint J 2016;98-B:867–73.


Bone & Joint Research
Vol. 5, Issue 5 | Pages 185 - 190
1 May 2016
Yuenyongviwat V Iamthanaporn K Hongnaparak T Tangtrakulwanich B

Objectives

Nylon sutures and skin staples are used commonly in total knee arthroplasty (TKA) surgical wound closure. However, there is no study that compares the wound healing efficacy and patient satisfaction scores of both techniques in the same knee.

Methods

We randomised 70 patients who underwent primary TKA into two groups. In one group of 34 patients, the skin at the upper half of the wound was closed with skin staples and the lower half of the wound was closed with simple interrupted nylon sutures. In the other group of 36 patients, the skin at the upper half of the wound was closed with nylon stitches and the lower half of the wound was closed with skin staples. We recorded the wound closure time, pain score at the time of stitch removal, wound complication rate, patient satisfaction score, and the Hollander wound evaluation score at the post-operative periods of five days, 14 days, six weeks, three months, and six months. Each half wound was analysed separately.