Advertisement for orthosearch.org.uk
Results 61 - 80 of 116
Results per page:
Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


The Bone & Joint Journal
Vol. 98-B, Issue 12 | Pages 1589 - 1596
1 Dec 2016
Magill P Blaney J Hill JC Bonnin MP Beverland DE

Aims

Our aim was to report survivorship data and lessons learned with the Corail/Pinnacle cementless total hip arthroplasty (THA) system.

Patients and Methods

Between August 2005 and March 2015, a total of 4802 primary cementless Corail/Pinnacle THAs were performed in 4309 patients. In March 2016, we reviewed these hips from a prospectively maintained database.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 749 - 758
1 Jun 2017
García-Rey E Cruz-Pardos A García-Cimbrelo E

Aims

To determine the effect of a change in design of a cementless ceramic acetabular component in fixation and clinical outcome after total hip arthroplasty

Patients and Methods

We compared 342 hips (302 patients) operated between 1999 and 2005 with a relatively smooth hydroxyapatite coated acetabular component (group 1), and 337 hips (310 patients) operated between 2006 and 2011 using a similar acetabular component with a macrotexture on the entire outer surface of the component (group 2). The mean age of the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70) in group 2. The mean follow-up was 12.7 years (10 to 17) for group 1 and 7.2 years (4 to 10) for group 2.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 317 - 324
1 Mar 2017
Schilcher J Palm L Ivarsson I Aspenberg P

Aims

Post-operative migration of cemented acetabular components as measured by radiostereometric analysis (RSA) has a strong predictive power for late, aseptic loosening. Also, radiolucent lines predict late loosening. Migration has been reduced by systemic bisphosphonate treatment in randomised trials of hip and knee arthroplasty. Used as a local treatment, a higher local dose of bisphosphonate can be achieved without systemic exposure. We wished to see if this principle could be applied usefully in total hip arthroplasty (THA).

Patients and Methods

In this randomised placebo-controlled, double-blinded trial with 60 participants, we compressed gauze soaked in bisphosphonate solution (ibandronate) or saline against the acetabular bone bed immediately before cementing the acetabular component. RSA, classification of radiolucent lines, the Harris Hip Score (HHS) and the Western Ontario McMasters Universities Osteoarthritis Index (WOMAC) were carried out at three-, six-, 12-, and 24-month follow-up.


The Bone & Joint Journal
Vol. 99-B, Issue 1 | Pages 44 - 50
1 Jan 2017
Salo PP Honkanen PB Ivanova I Reito A Pajamäki J Eskelinen A

Aims

We evaluated the short-term functional outcome and prevalence of bearing-specific generation of audible noise in 301 patients (336 hips) operated on with fourth generation (Delta) medium diameter head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs).

Patients and Methods

There were 191 female (63%) and 110 male patients (37%) with a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years (1.3 to 3.4). Patients completed three questionnaires: Oxford Hip Score (OHS), Research and Development 36-item health survey (RAND-36) and a noise-specific symptom questionnaire. Plain radiographs were also analysed. A total of three hips (0.9%) were revised.


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 50 - 59
1 Jan 2017
Carli AV Negus JJ Haddad FS

Aims

Periprosthetic femoral fractures (PFF) following total hip arthroplasty (THA) are devastating complications that are associated with functional limitations and increased overall mortality. Although cementless implants have been associated with an increased risk of PFF, the precise contribution of implant geometry and design on the risk of both intra-operative and post-operative PFF remains poorly investigated. A systematic review was performed to aggregate all of the PFF literature with specific attention to the femoral implant used.

Patients and Methods

A systematic search strategy of several journal databases and recent proceedings from the American Academy of Orthopaedic Surgeons was performed. Clinical articles were included for analysis if sufficient implant description was provided. All articles were reviewed by two reviewers. A review of fundamental investigations of implant load-to-failure was performed, with the intent of identifying similar conclusions from the clinical and fundamental literature.


Bone & Joint Research
Vol. 5, Issue 11 | Pages 531 - 537
1 Nov 2016
Burgo FJ Mengelle DE Ozols A Fernandez C Autorino CM

Objectives

Studies reporting specifically on squeaking in total hip arthroplasty have focused on cementless, and not on hybrid, fixation. We hypothesised that the cement mantle of the femur might have a damping effect on the sound transmitted through the metal stem. The objective of this study was to test the effect of cement on sound propagation along different stem designs and under different fixation conditions.

Methods

An in vitro model for sound detection, composed of a mechanical suspension structure and a sound-registering electronic assembly, was designed. A pulse of sound in the audible range was propagated along bare stems and stems implanted in cadaveric bone femurs with and without cement. Two stems of different alloy and geometry were compared.


Bone & Joint 360
Vol. 5, Issue 4 | Pages 29 - 31
1 Aug 2016


The Bone & Joint Journal
Vol. 98-B, Issue 8 | Pages 1050 - 1055
1 Aug 2016
Karachalios T Varitimidis S Bargiotas K Hantes M Roidis N Malizos KN

Aims

The Advance Medial-Pivot total knee arthroplasty (TKA) was designed to reflect contemporary data regarding the kinematics of the knee. We wished to examine the long-term results obtained with this prosthesis by extending a previous evaluation.

Patients and Methods

We retrospectively evaluated prospectively collected data from 225 consecutive patients (41 men and 184 women; mean age at surgery 71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up of 13.4 years (11 to 15). Implant failure, complication rate, clinical (both subjective and objective) and radiological outcome were assessed. Pre- and post-operative clinical and radiographic data were available at regular intervals for all patients. A total of ten patients (4.4%; ten TKAs) were lost to follow-up.


Bone & Joint 360
Vol. 4, Issue 4 | Pages 8 - 11
1 Aug 2015
McBride A Nicol S Monsell F


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 75 - 80
1 Jan 2016
Merolla G Chin P Sasyniuk TM Paladini P Porcellini G

Aims

We evaluated clinical and radiographic outcomes of total shoulder arthroplasty (TSA) using the second-generation Trabecular Metal (TM) Glenoid component. The first generation component was withdrawn in 2005 after a series of failures were reported. Between 2009 and 2012, 40 consecutive patients with unilateral TSA using the second-generation component were enrolled in this clinical study. The mean age of the patients was 63.8 years (40 to 75) and the mean follow-up was 38 months (24 to 42).

Methods

Patients were evaluated using the Constant score (CS), the American Shoulder and Elbow Surgeons (ASES) score and routine radiographs.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1611 - 1617
1 Dec 2012
Jameson SS Baker PN Mason J Gregg PJ Brewster N Deehan DJ Reed MR

Despite excellent results, the use of cemented total hip replacement (THR) is declining. This retrospective cohort study records survival time to revision following primary cemented THR using the most common combination of components that accounted for almost a quarter of all cemented THRs, exploring risk factors independently associated with failure. All patients with osteoarthritis who had an Exeter V40/Contemporary THR (Stryker) implanted before 31 December 2010 and recorded in the National Joint Registry for England and Wales were included in the analysis. Cox’s proportional hazard models were used to analyse the extent to which risk of revision was related to patient, surgeon and implant covariates, with a significance threshold of p < 0.01. A total of 34 721 THRs were included in the study. The overall seven-year rate of revision for any reason was 1.70% (99% confidence interval (CI) 1.28 to 2.12). In the final adjusted model the risk of revision was significantly higher in THRs with the Contemporary hooded component (hazard ratio (HR) 1.88, p < 0.001) than with the flanged version, and in smaller head sizes (< 28 mm) compared with 28 mm diameter heads (HR 1.50, p = 0.005). The seven-year revision rate was 1.16% (99% CI 0.69 to 1.63) with a 28 mm diameter head and flanged component. The overall risk of revision was independent of age, gender, American Society of Anesthesiologists grade, body mass index, surgeon volume, surgical approach, brand of cement/presence of antibiotic, femoral head material (stainless steel/alumina) and stem taper size/offset. However, the risk of revision for dislocation was significantly higher with a ‘plus’ offset head (HR 2.05, p = 0.003) and a hooded acetabular component (HR 2.34, p < 0.001).

In summary, we found that there were significant differences in failure between different designs of acetabular component and sizes of femoral head after adjustment for a range of covariates.


Bone & Joint Research
Vol. 3, Issue 6 | Pages 183 - 186
1 Jun 2014
Wyatt MC Jesani S Frampton C Devane P Horne JG

Objectives

Our study aimed to examine not only the incidence but also the impact of noise from two types of total hip replacement articulations: ceramic-on-ceramic and ceramic-on-polyethylene.

Methods

We performed a case-controlled study comparing subjective and objective questionnaire scores of patients receiving a ceramic-on-ceramic or a ceramic-on-polyethylene total hip replacement by a single surgeon.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 367 - 370
1 Mar 2013
Bini SA Chen Y Khatod M Paxton EW

We evaluated the impact of pre-coating the tibial component with polymethylmethacrylate (PMMA) on implant survival in a cohort of 16 548 primary NexGen total knee replacements (TKRs) in 14 113 patients. In 13 835 TKRs a pre-coated tray was used while in 2713 TKRs the non-pre-coated version of the same tray was used. All the TKRs were performed between 2001 and 2009 and were cemented. TKRs implanted with a pre-coated tibial component had a lower cumulative survival than those with a non-pre-coated tibial component (p = 0.01). After adjusting for diagnosis, age, gender, body mass index, American Society of Anesthesiologists grade, femoral coupling design, surgeon volume and hospital volume, pre-coating was an independent risk factor for all-cause aseptic revision (hazard ratio 2.75, p = 0.006). Revision for aseptic loosening was uncommon for both pre-coated and non-pre-coated trays (rates of 0.12% and 0%, respectively). Pre-coating with PMMA does not appear to be protective of revision for this tibial tray design at short-term follow-up.

Cite this article: Bone Joint J 2013;95-B:367–70.


Bone & Joint Research
Vol. 3, Issue 4 | Pages 95 - 100
1 Apr 2014
Kaneyama R Otsuka M Shiratsuchi H Oinuma K Miura Y Tamaki T

Objectives

Because posterior cruciate ligament (PCL) resection makes flexion gaps wider in total knee replacement (TKR), preserving or sacrificing a PCL affects the gap equivalence; however, there are no criteria for the PCL resection that consider gap situations of each knee. This study aims to investigate gap characteristics of knees and to consider the criteria for PCL resection.

Methods

The extension and flexion gaps were measured, first with the PCL preserved and subsequently with the PCL removed (in cases in which posterior substitute components were selected). The PCL preservation or sacrifice was solely determined by the gap measurement results, without considering other functions of the PCL such as ‘roll back.’


Bone & Joint 360
Vol. 2, Issue 3 | Pages 38 - 39
1 Jun 2013

The June 2013 Research Roundup360 looks at: a contact patch to rim distance and metal ions; the matrix of hypoxic cartilage; CT assessment of early fracture healing; Hawthornes and radiographs; cardiovascular mortality and fragility fractures; and muscle strength decline preceding OA changes.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 3 | Pages 307 - 314
1 Mar 2011
Matthies A Underwood R Cann P Ilo K Nawaz Z Skinner J Hart AJ

This study compared component wear rates and pre-revision blood metal ions levels in two groups of failed metal-on-metal hip arthroplasties: hip resurfacing and modular total hip replacement (THR).

There was no significant difference in the median rate of linear wear between the groups for both acetabular (p = 0.4633) and femoral (p = 0.0872) components. There was also no significant difference in the median linear wear rates when failed hip resurfacing and modular THR hips of the same type (ASR and Birmingham hip resurfacing (BHR)) were compared.

Unlike other studies of well-functioning hips, there was no significant difference in pre-revision blood metal ion levels between hip resurfacing and modular THR.

Edge loading was common in both groups, but more common in the resurfacing group (67%) than in the modular group (57%). However, this was not significant (p = 0.3479). We attribute this difference to retention of the neck in resurfacing of the hip, leading to impingement-type edge loading. This was supported by visual evidence of impingement on the femur.

These findings show that failed metal-on-metal hip resurfacing and modular THRs have similar component wear rates and are both associated with raised pre-revision blood levels of metal ions.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 6 - 14
1 Jun 2013
Wallace WA

In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 160 - 165
1 Feb 2013
McDonnell SM Boyce G Baré J Young D Shimmin AJ

Noise generation has been reported with ceramic-on-ceramic articulations in total hip replacement (THR). This study evaluated 208 consecutive Delta Motion THRs at a mean follow-up of 21 months (12 to 35). There were 141 women and 67 men with a mean age of 59 years (22 to 84). Patients were reviewed clinically and radiologically, and the incidence of noise was determined using a newly described assessment method. Noise production was examined against range of movement, ligamentous laxity, patient-reported outcome scores, activity level and orientation of the acetabular component. There were 143 silent hips (69%), 22 (11%) with noises other than squeaking, 17 (8%) with unreproducible squeaking and 26 (13%) with reproducible squeaking. Hips with reproducible squeaking had a greater mean range of movement (p < 0.001) and mean ligament laxity (p = 0.004), smaller median head size (p = 0.01) and decreased mean acetabular component inclination (p = 0.02) and anteversion angle (p = 0.02) compared with the other groups. There was no relationship between squeaking and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body mass index (p = 0.643), gender (p = 0.07) or patient outcome score (p = 0.422). There were no revisions during follow-up. Despite the surprisingly high incidence of squeaking, all patients remain satisfied with their hip replacement.

Cite this article: Bone Joint J 2013;95-B:160–5.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1482 - 1486
1 Nov 2012
Brogan K Charity J Sheeraz A Whitehouse SL Timperley AJ Howell JR Hubble MJW

The technique of femoral cement-in-cement revision is well established, but there are no previous series reporting its use on the acetabular side at the time of revision total hip replacement. We describe the technique and report the outcome of 60 consecutive acetabular cement-in-cement revisions in 59 patients at a mean follow-up of 8.5 years (5 to 12). All had a radiologically and clinically well-fixed acetabular cement mantle at the time of revision. During the follow-up 29 patients died, but no hips were lost to follow-up. The two most common indications for acetabular revision were recurrent dislocation (46, 77%) and to complement femoral revision (12, 20%).

Of the 60 hips, there were two cases of aseptic loosening of the acetabular component (3.3%) requiring re-revision. No other hip was clinically or radiologically loose (96.7%) at the latest follow-up. One hip was re-revised for infection, four for recurrent dislocation and one for disarticulation of a constrained component. At five years the Kaplan-Meier survival rate was 100% for aseptic loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause as the endpoint.

These results support the use of cement-in-cement revision on the acetabular side in appropriate cases. Theoretical advantages include preservation of bone stock, reduced operating time, reduced risk of complications and durable fixation.