The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 83 - 87
1 Jan 1992
Kim Y Kim V

We reviewed 82 primary arthroplasties (in 71 patients) in which cementless porous-coated hip prostheses were used. The mean age of the patients at operation was 52 years (24 to 86); they were followed up for an average of 62.1 months (60 to 66). The diagnosis was avascular necrosis of the femoral head in 35%, fracture of the femoral neck in 24%, primary osteoarthritis in 16% and miscellaneous in 25%. The average preoperative Harris hip score was 56.7 points and the average postoperative score was 83.3 points. Eight hips (10%) had component loosening; four had been revised and four were awaiting revision. In 27 hips (33%) there was a radiolucent line wider than 2 mm in zones 1 and 7. In 55 hips (67%) there was calcar resorption of more than 10 mm. Twenty patients (28%) complained of thigh pain although they had no radiographic evidence of loosening of a component. Factors that may have contributed to the poor clinical and radiographic results were: 1) inadequate surface area for bone ingrowth, particularly on the lateral aspect of the upper part of the prosthesis, 2) poor initial fit of the stem in the metaphysis, which resulted in cantilever motion of the proximal part of the stem about the well-fixed distal stem, and 3) the collar of the prosthesis, which prevented it from subsiding to a naturally stable position and caused damage to the calcar


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 4 | Pages 603 - 606
1 Aug 1988
Eskola A Santavirta S Konttinen Y Tallroth K Hoikka V Lindholm S

We report the results of cementless total joint replacement in 18 patients with old tuberculosis of the hip, performed, on average, 34 years after the onset of infection. Mean follow-up was 3.5 years. Only seven of the patients had antituberculous drugs during or after the operation. Using the Mayo hip score, 15 patients had excellent or good results and two had a fair rating. One patient had the prosthesis removed more than one year postoperatively for late haematogenous staphylococcal infection and had a poor rating. All the patients had relief of hip-related pain. Despite the absence of any reactivation of tuberculosis in our series, we recommend the use of specific prophylaxis


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 4 | Pages 675 - 679
1 Jul 1997
Ang KC Das De S Goh JCH Low SL Bose K

In a prospective study of 14 patients undergoing total hip replacement we have used dual-energy X-ray absorptiometry (DEXA) to investigate remodelling of the bone around two different designs of cementless femoral prosthesis. The bone mineral density (BMD) was measured at 12-weekly intervals for a year. Eight patients (group A) had a stiff, collarless implant and six (group B) a flexible isoelastic implant. Patients in group A showed a decrease in BMD from 14 weeks after operation. By 12 months, the mean loss in BMD was 27%, both medially and laterally to the proximal part of the implant. Those in group B showed an overall increase in BMD which reached a mean of 12.6% on the lateral side of the distal portion of the implant. Our results support the current concepts of the effects of stem stiffness and flexibility on periprosthetic remodelling


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 2 | Pages 209 - 214
1 Mar 2003
Aldinger PR Breusch SJ Lukoschek M Mau H Ewerbeck V Thomsen M

We followed the first 354 consecutive implantations of a cementless, double-tapered straight femoral stem in 326 patients. Follow-up was at a mean of 12 years (10 to 15). The mean age of the patients was 57 years (13 to 81). At follow-up, 56 patients (59 hips) had died, and eight (eight hips) had been lost to follow-up. Twenty-five hips underwent femoral revision, eight for infection, three for periprosthetic fracture and 14 for aseptic loosening. The overall survival was 92% at 12 years (95% CI 88 to 95). Survival with femoral revision for aseptic loosening as an endpoint was 95% (95% CI 92 to 98). The median Harris hip score at follow-up was 84 points (23 to 100). Radiolucent lines (< 2 mm) in Gruen zones 1 and 7 were present in 38 (16%) and 34 hips (14%), respectively. Radiolucencies in zones 2 to 6 were found in five hips (2%). The results for mid- to long-term survival with this femoral component are encouraging and compare with those achieved in primary cemented total hip arthroplasty. The high rate of loosening of the cup and the high rate of pain are, however, a source of concern


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 2 | Pages 236 - 241
1 Mar 1995
Kim Y Oh J Oh S

We report the incidence of osteolysis in the femur, tibia, and patella of 44 consecutive patients (60 knees) who were followed for more than seven years after cementless knee arthroplasty with a Porous-Coated Anatomic prosthesis. The average age of the patients was 56.5 years (17 to 73); the operative diagnosis was osteoarthritis (33 knees), rheumatoid arthritis (17), tuberculous arthritis (7) and post-traumatic arthritis (3). All patellae were resurfaced. No femoral or tibial component was loose at the final follow-up examination. Thirty patellar components were loose of which six had been revised. Radiographs revealed osteolysis in 90% of the tibial plateaux and in 80% of the 30 intact patellar prostheses. No osteolysis was seen around any femoral component. In 50 knees (83%) the average wear of the polyethylene liner was 2.5 mm in the medial compartment and 1.7 mm in the lateral compartment. Four of 60 knees (6.7%) were revised for complete wear of the polyethylene liner of the tibial component. Fixation of the tibial and patellar components without cement fails to seal the interface between bone and prosthesis and allows the migration of polyethylene particulate debris which causes osteolysis


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 366 - 368
1 May 1995
Sweetnam D Lavelle J Allwood W Cohen B

We reviewed 12 patients six years after they had undergone total hip replacement with a cementless prosthesis, the Ribbed Hip System (Waldemar Link GmbH & Co, Hamburg, Germany). Aseptic loosening of one or both components had necessitated revision surgery in seven patients, in five within two years of operation. In view of our experience we question the wisdom of allowing the uncontrolled use of new prosthesis without postoperative surveillance


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 2 | Pages 205 - 211
1 Mar 1990
Paavilainen T Hoikka V Solonen K

We report the short-term results of 100 cementless total hip replacements in 52 severely dysplastic and 48 totally dislocated hips, with some new technical solutions to the problems involved. In cases with a very narrow iliac bone, the acetabular screw ring is seated below the true cotyloid area. In hips with tight flexor and abductor muscles or with deformities of the proximal femur, various osteotomies were performed. Special attention was paid to careful pre-operative planning and precise operative technique. In spite of a high complication rate the results were generally good and even patients who required reoperation were satisfied with the final result


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 455 - 460
1 Apr 2007
Sugano N Nishii T Miki H Yoshikawa H Sato Y Tamura S

We have developed a CT-based navigation system using infrared light-emitting diode markers and an optical camera. We used this system to perform cementless total hip replacement using a ceramic-on-ceramic bearing couple in 53 patients (60 hips) between 1998 and 2001. We reviewed 52 patients (59 hips) at a mean of six years (5 to 8) postoperatively. The mid-term results of total hip replacement using navigation were compared with those of 91 patients (111 hips) who underwent this procedure using the same implants, during the same period, without navigation. There were no significant differences in age, gender, diagnosis, height, weight, body mass index, or pre-operative clinical score between the two groups. The operation time was significantly longer where navigation was used, but there was no significant difference in blood loss or navigation-related complications. With navigation, the acetabular components were placed within the safe zone defined by Lewinnek, while without, 31 of the 111 components were placed outside this zone. There was no significant difference in the Merle d’Aubigne and Postel hip score at the final follow-up. However, hips treated without navigation had a higher rate of dislocation. Revision was performed in two cases undertaken without navigation, one for aseptic acetabular loosening and one for fracture of a ceramic liner, both of which showed evidence of neck impingement on the liner. A further five cases undertaken without navigation showed erosion of the posterior aspect of the neck of the femoral component on the lateral radiographs. These seven impingement-related mechanical problems correlated with malorientation of the acetabular component. There were no such mechanical problems in the navigated group. We conclude that CT-based navigation increased the precision of orientation of the acetabular component and control of limb length in total hip replacement, without navigation-related complications. It also reduced the rate of dislocation and mechanical problems related to impingement


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 3 | Pages 419 - 422
1 May 1994
Hooten J Engh C Engh C

We investigated the radiographic and clinical course of 31 patients in whom a bulk acetabular allograft had been used during the cementless revision of a total hip replacement. Two patients died and two were lost to follow-up within 24 months, but of the remaining 27 acetabular components, 12 (44%) showed radiographic evidence of instability at a mean of 46 months. Five of these have been revised. In the 12 failures, signs of instability had been noted at an average of 29 months (1 to 60). Failures during the first 24 months were usually due to technical errors, later failures to gradual migration of the cup into the graft. The cups with the greatest amount of their surface supported by grafts were most likely to migrate, but this migration was usually asymptomatic. Screw fixation of the cup, used in 24 cases, appeared to control the mechanism of failure. Femoral head allografts and distal femoral allografts had been used, with failure in 6 of 16, and 6 of 11 respectively; distal femoral allografts were used only for large defects. The insidious course of late cup migration and graft failure necessitates close radiographic follow-up of patients treated with bulk allografts


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 2 | Pages 200 - 204
1 Mar 2004
Dixon P Parish EN Chan B Chitnavis J Cross MJ

We have carried out a prospective study comparing the results at five years in patients older than 75 years of age undergoing hydroxyapatite-coated, cementless total knee replacement (TKR) with those who were younger. The Knee Society clinical rating scores were recorded before and after operation. Of 559 patients undergoing TKR, 135 were in the elderly age group. The knee scores at five years or more showed comparable results, with patients under 75 years of age reaching a score of 183 and those aged 75 years or over reaching 174. These differences were predominantly due to the functional component of the score. We conclude that elderly patients do just as well as the younger group using this prosthesis


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 3 | Pages 431 - 434
1 May 1990
Phillips T Messieh S McDonald P

A leading problem with cementless hip replacement is thigh pain, probably due to some degree of loosening. This is most pronounced during activities such as stair climbing or getting up from a chair. Our study compares the immediate fixation of three stems in widespread clinical use, using a cadaver test simulation in which the femur was horizontal, and cyclic forces acted vertically downward on the prosthetic head with gradual increments of load. We implanted and tested 18 pairs of femurs, making analyses between stem types by right to left comparisons. We found no statistically significant differences in loosening between cementless AML and PCA prostheses, but cemented stems withstood greater loads than uncemented stems by a factor of five or more


The Bone & Joint Journal
Vol. 101-B, Issue 5 | Pages 502 - 511
1 May 2019
Lidder S Epstein DJ Scott G

Aims

Short-stemmed femoral implants have been used for total hip arthroplasty (THA) in young and active patients to conserve bone, provide physiological loading, and reduce the incidence of thigh pain. Only short- to mid-term results have been presented and there have been concerns regarding component malalignment, incorrect sizing, and subsidence. This systematic review reports clinical and radiological outcomes, complications, revision rates, and implant survival in THA using short-stemmed femoral components.

Materials and Methods

A literature review was performed using the EMBASE, Medline, and Cochrane databases. Strict inclusion and exclusion criteria were used to identify studies reporting clinical and radiological follow-up for short-stemmed hip arthroplasties.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 53 - 56
1 Nov 2013
Su EP Barrack RL

Cementless femoral stems are currently preferred for total hip replacement (THR) in the United States. Improvements in stem design, instrumentation and surgical technique have made this technology highly successful, reproducible, and applicable to the vast majority of patients requiring a THR. However, there are ongoing developments in some aspects of stem design that influence clinical results, the incidence of complications and their inherent adaptability in accommodating the needs of individual patients. Here we examine some of these design features.

Cite this article: Bone Joint J 2013;95-B, Supple A:53–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 85 - 89
1 Nov 2012
Drexler M Dwyer T Marmor M Abolghasemian M Sternheim A Cameron HU

In this study we present our experience with four generations of uncemented total knee arthroplasty (TKA) from Smith & Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing on the failure rates correlating with each design change. Beginning in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix were implanted by the senior author. The rate of revision for loosening was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the Tricon 2 with a HA coated tibial component, and 1.3% for the Profix TKA. No loosening of the femoral component was seen with the Tricon M, Tricon LS or Tricon 2, with no loosening seen of the tibial component with the Profix TKA. Regarding revision for wear, the incidence was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the Tricon 2, and 0% for the Profix. These results demonstrate that improvements in the design of uncemented components, including increased polyethylene thickness, improved polyethylene quality, and the introduction of hydroxyapatite coating, has improved the outcomes of uncemented TKA over time.


The Journal of Bone & Joint Surgery British Volume
Vol. 72-B, Issue 3 | Pages 423 - 430
1 May 1990
Wilson-MacDonald J Morscher E Masar Z

We reviewed the results of 545 consecutive total hip replacements using a cementless non-coated high-density polyethylene acetabular component combined with a cemented Muller stem at five to 10 years. In all, 421 patients (445 hips) were available for review, 118 by questionnaire and 303 by examination and radiography. Of these, 86% had a good or excellent result. We found a high rate of radiological loosening of the cup after the sixth year, and a high rate of clinical loosening after the eighth year. Loosening was commoner in women, in younger patients and where a smaller size of acetabulum had been used. Calcar resorption was significantly related to loosening of the acetabulum. Loosening appeared to be mainly due to polyethylene debris produced by micro-movement of the acetabulum against the bone, which had resulted in a giant cell foreign body reaction and subsequent bone erosion. We have abandoned the use of this prosthesis and suggest that direct contact between bone and polyethylene should be prevented by a coating of metal or some other material


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 484 - 489
1 May 1995
Won C Hearn T Tile M

Adult human cadaver pelves were tested to determine micromotion at the prosthesis-bone interface in cementless hemispherical acetabular components during simulated single-limb stance. The micromotion of non-press-fit components with screw fixation in response to cyclic loads to a maximum of 1500N was compressive (interface closing) at the superior iliac rim and distractive (interface opening) at the inferior ischial rim; that of press-fit components was compressive all around the acetabular rim regardless of screw fixation. Adding screws to the component decreased the micromotion at the site of the screw, but sometimes increased it at the opposite side. Two dome screws with the press-fit component decreased the micromotion at the superior iliac rim but at the inferior ischial rim there was either no change or increased movement. A press-fit cup shows less micromotion than a non-press-fit cup with screw fixation. The addition of screws to a press-fit cup does not necessarily increase the initial stability


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 2 | Pages 279 - 287
1 Mar 1993
Kilgus D Shimaoka E Tipton J Eberle R

The measurement of bone mineral density in defined areas around metal implants has improved with the development of dual-energy X-ray absorptiometry. We used this technique to compare the bone mineral density adjacent to metal cementless femoral implants with that of identical regions of bone in normal proximal femora. We studied the anteroposterior views only of 72 femora which contained total hip implants and 34 non-operated femora. We compared the regional bone mineral density of bone adjacent to proximally porous-coated and distally porous-coated implants of one design, to measure the relative differences in the remodelling changes induced by different amounts of porous coating. We also measured differences in bone density with time and with variations in implant size (and therefore stiffness). The greatest decrease in bone mineral density (34.8%) occurred in the most proximal 1 cm of the medial femoral cortex around relatively stiff, extensively porous-coated implants. The next most severe decrease (20% to 25%) was in the next most proximal 6 cm of the medial femoral cortex. Small, progressive decreases in bone mineral density continued for five to seven years after implantation


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 595 - 602
1 May 2015
McCalden RW Korczak A Somerville L Yuan X Naudie DD

This was a randomised controlled trial studying the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised that it would have similar early clinical results and micromovement to those of a standard-length tapered Synergy metaphyseal fixation stem. Using radiostereometric analysis (RSA) we compared the two stems in 43 patients. A short metaphyseal fixation stem was used in 22 patients and a Synergy stem in 21 patients. No difference was found in the clinical outcomes pre- or post-operatively between groups. RSA showed no significant differences two years post-operatively in mean micromovement between the two stems (except for varus/valgus tilt at p = 0.05) (subsidence 0.94 mm (sd 1.71) vs 0.32 mm (sd 0.45), p = 0.66; rotation 0.96° (sd 1.49) vs 1.41° (sd 2.95), p = 0.88; and total migration 1.09 mm (sd 1.74) vs 0.73 mm (sd 0.72), p = 0.51). A few stems (four SMF and three Synergy) had initial migration > 1.0 mm but stabilised by three to six months, with the exception of one SMF stem which required revision three years post-operatively. For most stems, total micromovement was very low at two years (subsidence < 0.5 mm, rotation < 1.0°, total migration < 0.5 mm), which was consistent with osseous ingrowth. The small sample makes it difficult to confirm the universal applicability of or elucidate the potential contraindications to the use of this particular new design of stem.

Cite this article: Bone Joint J 2015; 97-B:595–602.


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 65 - 74
1 Jan 2016
Phadnis J Huang T Watts A Krishnan J Bain GI

Aims

To date, there is insufficient evidence available to compare the outcome of cemented and uncemented fixation of the humeral stem in reverse shoulder arthroplasty (RSA).

Methods

A systemic review comprising 41 clinical studies was performed to compare the functional outcome and rate of complications of cemented and uncemented stems in RSA. These included 1455 cemented and 329 uncemented shoulders. The clinical characteristics of the two groups were similar.

Variables were compared using pooled frequency-weighted means and relative risk ratios (RR).


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 856 - 861
1 Jun 2010
Emms NW Stockley I Hamer AJ Wilkinson JM

Between 1988 and 1998 we implanted 318 total hip replacements (THRs) in 287 patients using the Plasmacup (B. Braun Ltd, Sheffield, United Kingdom) and a conventional metal-on-polyethylene articulation. The main indications for THR were primary or secondary osteoarthritis.

At follow-up after a mean 11.6 years (7.6 to 18.4) 17 patients had died and 20 could not be traced leaving a final series of 280 THRs in 250 patients. There were 62 revisions (22.1%) in 59 patients. A total of 43 acetabular shells (15.4%) had been revised and 13 (4.6%) had undergone exchange of the liner. The most frequent indications for revision were osteolysis and aseptic loosening, followed by polyethylene wear. The mean Kaplan-Meier survival of the Plasmacup was 91% at ten years and 58% at 14 years. Osteolysis was found around 36 (17.1%) of the 211 surviving shells. The median annual rate of linear wear in the surviving shells was 0.12 mm/year and 0.25 mm/year in those which had been revised (p < 0.001). Polyethylene wear was a strong independent risk factor for osteolysis and aseptic loosening. The percentage of patients with osteolysis increased proportionately with each quintile of wear-rate.

There is a high late rate of failure of the Plasmacup. Patients with the combination of this prosthesis and bearing should be closely monitored after ten years.