Rarely, the extent of a malignant bone tumour
may necessitate resection of the complete humerus to achieve adequate
oncological clearance. We present our experience with reconstruction
in such cases using a total humeral endoprosthesis (THER) in 20
patients (12 male and eight female) with a mean age of 22 years
(6 to 59). We assessed the complications, the oncological and functional
outcomes and implant survival. Surgery was performed between June
2001 and October 2009. The diagnosis included osteosarcoma in nine,
Ewing’s sarcoma in eight and chondrosarcoma in three. One patient
was lost to follow-up. The mean follow-up was 41 months (10 to 120)
for all patients and 56 months (25 to 120) in survivors. There were
five local recurrences (26.3%) and 11 patients were alive at time
of last follow-up, with overall survival for all patients being
52% (95% confidence interval (CI) 23.8 to 74) at five years. The
mean Musculoskeletal Tumor Society score for the survivors was 22
(73%; 16 to 23). The implant survival was 95% (95% CI 69.5 to 99.3)
at five years. The use of a THER in the treatment of malignant tumours of bone
is oncologically safe; it gives consistent and predictable results
with low rates of complication.
Amputation was once widely practised for primary bone tumours of the limbs. Yet this situation has changed with limb salvage surgery becoming increasingly popular in the last 30 years. Many different techniques are now available. These include allografts, autografts, endoprostheses and allograft-prosthesis composites. This article reviews these methods, concentrating on the functional outcomes and complications that have been reported.
Numerous complications following total knee replacement (TKR)
relate to the patellofemoral (PF) joint, including pain and patellar
maltracking, yet the options for A total of three knees with end-stage osteoarthritis and three
knees that had undergone TKR at more than one year’s follow-up were
investigated. In each knee, sequential biplane radiological images
were acquired from the sagittal direction (i.e. horizontal X-ray
source and 10° below horizontal) for a sequence of eight flexion
angles. Three-dimensional implant or bone models were matched to
the biplane images to compute the six degrees of freedom of PF tracking
and TF kinematics, and other clinical measures.Objectives
Methods
The Oxford hip score (OHS) is a 12-item questionnaire designed
and developed to assess function and pain from the perspective of
patients who are undergoing total hip replacement (THR). The OHS
has been shown to be consistent, reliable, valid and sensitive to
clinical change following THR. It has been translated into different
languages, but no adequately translated, adapted and validated Danish
language version exists. The OHS was translated and cross-culturally adapted into Danish
from the original English version, using methods based on best-practice
guidelines. The translation was tested for psychometric quality
in patients drawn from a cohort from the Danish Hip Arthroplasty
Register (DHR).Objectives
Methods
We report a case of spontaneous physeal growth arrest of the distal femur in a nine-year-old child with Ewing’s sarcoma of the proximal femur treated with chemotherapy and endoprosthetic replacement. Owing to the extent of disuse osteoporosis at the time of surgery, the entire intramedullary canal up to the distal femoral physis was filled with cement. Three years later, the femur remained at its pre-operative length of 19 cm. Pre-operative calculations of further growth failed to account for the growth arrest, and the initial expandable growing prosthesis inserted has been revised to a longer one in order to address the leg-length discrepancy. To our knowledge, this is the only reported case of distal femoral physeal growth arrest following cemented endoprosthetic replacement of the proximal femur.
A clinical and radiological study was conducted on 97 total hip replacements performed for congenital hip dislocation in 79 patients between 1989 and 1998 using a three-dimensional custom-made cementless stem. The mean age at operation was 48 years (17 to 72) and the mean follow-up was for 123 months (83 to 182). According to the Crowe classification, there were 37 class I, 28 class II, 13 class III and 19 class IV hips. The mean leg lengthening was 25 mm (5 to 58), the mean pre-operative femoral anteversion was 38.6° (2° to 86°) and the mean correction in the prosthetic neck was −23.6° (−71° to 13°). The mean Harris hip score improved from 58 (15 to 84) to 93 (40 to 100) points. A revision was required in six hips (6.2%). The overall survival rate was 89.5% (95% confidence interval 89.2 to 89.8) at 13 years when two hips were at risk. This custom-made cementless femoral component, which can be accommodated in the abnormal proximal femur and will correct the anteversion and frontal offset, provided good results without recourse to proximal femoral corrective osteotomy.
Endoprosthetic replacement of the pelvis is one of the most challenging types of limb-salvage surgery, with a high rate of complications. In an attempt to reduce this and build greater versatility into the reconstruction process, a new type of pelvic endoprosthesis was developed in 2003, based on the old McKee-Farrar prosthesis. This study reviews the outcomes in 27 patients who had an ice-cream cone pelvic prosthesis inserted at two different specialist bone tumour centres in the United Kingdom over the past six years. The indications for treatment included primary bone tumours in 19 patients and metastatic disease in two, and six implants were inserted following failure of a previous pelvic reconstruction. Most of the patients had a P2+P3 resection as classified by Enneking, and most had resection of the ilium above the sciatic notch. The mean age of the patients at operation was 49 years (13 to 81). Complications occurred in ten patients (37.0%), of which dislocation was the most common, affecting four patients (14.8%). A total of three patients (11.1%) developed a deep infection around the prosthesis but all were successfully controlled by early intervention and two patients (7.4%) developed a local recurrence, at the same time as widespread metastases appeared. In one patient the prosthesis was removed for severe pain. This method of treatment is still associated with high morbidity, but early results are promising. Complications are diminishing with increasing experience.
We report the use of the reverse shoulder prosthesis in the revision of a failed shoulder hemiarthroplasty in 19 shoulders in 18 patients (7 men, 11 women) with severe pain and loss of function. The primary procedure had been undertaken for glenohumeral arthritis associated with severe rotator cuff deficiency. Statistically significant improvements were seen in pain and functional outcome. After a mean follow-up of 44 months (24 to 89), mean forward flexion improved by 26.4° and mean abduction improved by 35°. There were six prosthesis-related complications in six shoulders (32%), five of which had severe bone loss of the glenoid, proximal humerus or both. Three shoulders (16%) had non-prosthesis related complications. The use of the reverse shoulder prosthesis provides improvement in pain and function for patients with failure of a hemiarthroplasty for glenohumeral arthritis and rotator cuff deficiency. However, high rates of complications were associated with glenoid and proximal humeral bone loss.
We studied 33 third generation, alumina ceramic-on-ceramic bearings retrieved from cementless total hip replacements after more than six months in situ. Wear volume was measured with a Roundtest machine, and acetabular orientation from the anteroposterior pelvic radiograph. The overall median early wear rate was 0.1 mm3/yr for the femoral heads, and 0.04 mm3/yr for the acetabular liners. We then excluded hips where the components had migrated. In this stable subgroup of 22 bearings, those with an acetabular anteversion of <
15° (seven femoral heads) had a median femoral head wear rate of 1.2 mm3/yr, compared with 0 mm3/yr for those with an anteversion of ≥15° (15 femoral heads, p <
0.001). Even under edge loading, wear volumes with ceramic-on-ceramic bearings are small in comparison to other bearing materials. Low acetabular anteversion is associated with greater wear.
Peri-prosthetic bone loss caused by stress shielding may be associated with aseptic loosening of femoral components. In order to increase primary stability and to reduce stress shielding, a three-dimensional, cementless individual femoral (Evolution K) component was manufactured using pre-operative CT scans. Using dual energy x-ray absorptiometry, peri-prosthetic bone density was measured in 43 patients, three months, six months, 3.6 and 4.6 years after surgery. At final follow-up there was a significant reduction in mean bone density in the proximal Gruen zones of −30.3% (zone 7) and −22.8% (zone 1). The density in the other zones declined by a mean of between −4% and −16%. We conclude that the manufacture of a three-dimensional, custom-made femoral component could not prevent a reduction in peri-prosthetic bone density.
We reviewed 39 patients with displaced three- and four-part fractures of the humerus. In 21 patients (group A) we had used an anatomical prosthesis for the humeral head and in 18 (group B) an implant designed for fractures. When followed up at a mean of 29.3 months after surgery the overall Constant score was 51.9 points; in group A it was 51.5 and in group B 52.4 points. The subjective satisfaction of the patients was assessed using a numerical rating scale and was similar in both groups. In group A complete healing of the tuberosities was found in 29% and 50% in group B. Partial integration was seen in 29% of group A and in only one patient in group B, while resorption was noted in 43% of group A and 44% of group B. The functional outcome was significantly better in patients with complete or partial healing of the tuberosities (p = 0.022). The specific trauma prosthesis did not lead to better healing of the tuberosities. The difference in clinical outcome obtained by the two designs did not reach statistical significance.
The use of a navigation system in musculoskeletal tumour surgery enables the integration of pre-operative CT and MRI images to generate a precise three-dimensional anatomical model of the site and the extent of the tumour. We carried out six consecutive resections of musculoskeletal tumour in five patients using an existing commercial computer navigation system. There were three women and two men with a mean age of 41 years (24 to 47). Reconstruction was performed using a tumour prosthesis in three lesions and a vascularised fibular graft in one. No reconstruction was needed in two cases. The mean follow-up was 6.9 months (3.5 to 10). The mean duration of surgery was 28 minutes (13 to 50). Examination of the resected specimens showed clear margins in all the tumour lesions and a resection that was exactly as planned.
We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter. The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection. The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s We found no evidence of the superiority of one design over the other at long-term follow-up.
When performing the Scandinavian Total Ankle Replacement (STAR), the positioning of the talar component and the selection of mobile-bearing thickness are critical. A biomechanical experiment was undertaken to establish the effects of these variables on the range of movement (ROM) of the ankle. Six cadaver ankles containing a specially-modified STAR prosthesis were subjected to ROM determination, under weight-bearing conditions, while monitoring the strain in the peri-ankle ligaments. Each specimen was tested with the talar component positions in neutral, as well as 3 and 6 mm of anterior and posterior displacement. The sequence was repeated with an anatomical bearing thickness, as well as at 2 mm reduced and increased thicknesses. The movement limits were defined as 10% strain in any ligament, bearing lift-off from the talar component or limitations of the hardware. Both anterior talar component displacement and bearing thickness reduction caused a decrease in plantar flexion, which was associated with bearing lift-off. With increased bearing thickness, posterior displacement of the talar component decreased plantar flexion, whereas anterior displacement decreased dorsiflexion.