We retrospectively reviewed 44 consecutive patients
(50 hips) who underwent acetabular re-revision after a failed previous
revision that had been performed using structural or morcellised
allograft bone, with a cage or ring for uncontained defects. Of
the 50 previous revisions, 41 cages and nine rings were used with
allografts for 14 minor-column and 36 major-column defects. We routinely
assessed the size of the acetabular bone defect at the time of revision
and re-revision surgery. This allowed us to assess whether host
bone stock was restored. We also assessed the outcome of re-revision
surgery in these circumstances by means of radiological characteristics,
rates of failure and modes of failure. We subsequently investigated
the factors that may affect the potential for the restoration of bone
stock and the durability of the re-revision reconstruction using
multivariate analysis. At the time of re-revision, there were ten host acetabula with
no significant defects, 14 with contained defects, nine with minor-column,
seven with major-column defects and ten with pelvic discontinuity.
When bone defects at re-revision were compared with those at the
previous revision, there was restoration of bone stock in 31 hips, deterioration
of bone stock in nine and remained unchanged in ten. This was a
significant improvement (p <
0.001). Morselised allografting
at the index revision was not associated with the restoration of
bone stock. In 17 hips (34%), re-revision was possible using a simple acetabular
component without allograft, augments, rings or cages. There were
47 patients with a mean follow-up of 70 months (6 to 146) available
for survival analysis. Within this group, the successful cases had
a minimum follow-up of two years after re-revision. There were 22 clinical
or radiological failures (46.7%), 18 of which were due to aseptic
loosening. The five and ten year Kaplan–Meier survival rate was
75% (95% CI, 60 to 86) and 56% (95% CI, 40 to 70) respectively with
aseptic loosening as the endpoint. The rate of aseptic loosening
was higher for hips with pelvic discontinuity (p = 0.049) and less
when the allograft had been in place for longer periods (p = 0.040). The use of a cage or ring over structural allograft bone for
massive uncontained defects in acetabular revision can restore host
bone stock and facilitate subsequent re-revision surgery to a certain
extent. Cite this article:
Dysplasia of the hip, hypotonia, osteopenia,
ligamentous laxity, and mental retardation increase the complexity
of performing and managing patients with Down syndrome who require
total hip replacement (THR). We identified 14 patients (six males,
eight females, 21 hips) with Down syndrome and degenerative disease
of the hip who underwent THR, with a minimum follow-up of two years
from 1969 to 2009. In seven patients, bilateral THRs were performed
while the rest had unilateral THRs. The mean clinical follow-up
was 5.8 years (standard deviation ( Cite this article:
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
In 2005, we demonstrated that the polished triple-tapered
C-stem at two years had migrated distally and rotated internally.
From that series, 33 patients have now been followed radiologically,
clinically and by radiostereometric analysis (RSA) for up to ten
years. The distal migration within the cement mantle had continued
and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation,
also within the cement mantle, was a mean 3.8° (external 1.6° to
internal 6.6°) The cement mantle did not show any sign of migration
or loosening in relation to the femoral bone. There were no clinical
or radiological signs indicating that the migration or rotation
within the cement mantle had had any adverse effects for the patients. Cite this article:
We compared the outcome of total hip arthroplasty (THA) in obese
patients who previously underwent bariatric surgery and those who
did not, in a matched cohort study. There were 47 THAs in the bariatric group (42 patients), and
94 THAs in the comparison group (92 patients). The mean age of the
patients was 57 years in both groups (24 to 79) and 57% of the patients
in both groups were women. The mean time between bariatric surgery
and THA was five years (four months to 12 years) in the bariatric group.
The mean follow-up after THA was three years (2 to 9). Aims
Patients and Methods
We report a 12- to 15-year implant survival assessment
of a prospective single-surgeon series of Birmingham Hip Resurfacings
(BHRs). The earliest 1000 consecutive BHRs including 288 women (335
hips) and 598 men (665 hips) of all ages and diagnoses with no exclusions
were prospectively followed-up with postal questionnaires, of whom
the first 402 BHRs (350 patients) also had clinical and radiological
review. Mean follow-up was 13.7 years (12.3 to 15.3). In total, 59 patients
(68 hips) died 0.7 to 12.6 years following surgery from unrelated
causes. There were 38 revisions, 0.1 to 13.9 years (median 8.7)
following operation, including 17 femoral failures (1.7%) and seven
each of infections, soft-tissue reactions and other causes. With
revision for any reason as the end-point Kaplan–Meier survival analysis
showed 97.4% (95% confidence interval (CI) 96.9 to 97.9) and 95.8%
(95% CI 95.1 to 96.5) survival at ten and 15 years, respectively.
Radiological assessment showed 11 (3.5%) femoral and 13 (4.1%) acetabular
radiolucencies which were not deemed failures and one radiological
femoral failure (0.3%). Our study shows that the performance of the BHR continues to
be good at 12- to 15-year follow-up. Men have better implant survival
(98.0%; 95% CI 97.4 to 98.6) at 15 years than women (91.5%; 95%
CI 89.8 to 93.2), and women <
60 years (90.5%; 95% CI 88.3 to
92.7) fare worse than others. Hip dysplasia and osteonecrosis are
risk factors for failure. Patients under 50 years with osteoarthritis
fare best (99.4%; 95% CI 98.8 to 100 survival at 15 years), with
no failures in men in this group. Cite this article:
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
Vascular injuries during total hip arthroplasty
(THA) are rare but when they occur, have serious consequences. These have
traditionally been managed with open exploration and repair, but
more recently there has been a trend towards percutaneous endovascular
management. We performed a systematic review of the literature to assess
if this change in trend has led to an improvement in the overall
reported rates of morbidity and mortality during the last 22 years
in comparison with the reviews of the literature published previously. We found a total of 61 articles describing 138 vascular injuries
in 124 patients. Injuries because of a laceration were the most
prevalent (n = 51, 44%) and the most common presenting feature,
when recorded, was bleeding (n = 41, 53.3%). Delay in diagnosis
was associated with the type of vascular lesion (p <
0.001) and
the clinical presentation (p = 0.002). Open exploration and repair was the most common form of management,
however percutaneous endovascular intervention was used in one third
of the injuries and more constantly during the last 13 years. The main overall reported complications included death (n = 9,
7.3%), amputation (n = 2, 1.6%), and persistent ischaemia (n = 9,
7.3%). When compared with previous reviews there was a similar rate
of mortality but lower rates of amputation and permanent disability,
especially in patients managed by endovascular strategies. Cite this article:
We investigated whether blood metal ion levels could effectively
identify patients with bilateral Birmingham Hip Resurfacing (BHR)
implants who have adverse reactions to metal debris (ARMD). Metal ion levels in whole blood were measured in 185 patients
with bilateral BHRs. Patients were divided into those with ARMD
who either had undergone a revision for ARMD or had ARMD on imaging
(n = 30), and those without ARMD (n = 155). Receiver operating characteristic
analysis was used to determine the optimal thresholds of blood metal
ion levels for identifying patients with ARMD.Aims
Patients and Methods
We report the clinical and radiological outcomes
of a series of contemporary cementless ceramic-on-ceramic total hip
replacements (THRs) at ten years in patients aged ≤ 55 years of
age. Pre- and post-operative activity levels are described. A total
of 120 consecutive ceramic cementless THRs were performed at a single
centre in 110 patients from 1997 to 1999. The mean age of the patients
at operation was 45 years (20 to 55). At ten years, four patients
had died and six were lost to follow-up, comprising ten hips. The
mean post-operative Harris hip score was 94.7 (55 to 100). Radiological
analysis was undertaken in 90 available THRs of the surviving 106
hips at final review: all had evidence of stable bony ingrowth,
with no cases of osteolysis. Wear was undetectable. There were four
revisions. The survival for both components with revision for any
cause as an endpoint was 96.5% (95% confidence interval 94.5 to
98.7). The mean modified University of California, Los Angeles activity
level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6
to 10) at the ten-year post-operative period. Alumina ceramic-on-ceramic bearings in cementless primary THR
in this series have resulted in good clinical and radiological outcomes
with undetectable rates of wear and excellent function in the demanding
younger patient group at ten years. Cite this article:
Femoral stem version has a major influence on
impingement and early post-operative stability after total hip arthroplasty
(THA). The main objective of this study was to evaluate the validity
of a novel radiological method for measuring stem version. Anteroposterior
(AP) radiographs and three-dimensional CT scans were obtained for
115 patients (female/male 63/72, mean age 62.5 years (50 to 75))
who had undergone minimally invasive, cementless THA. Stem version was
calculated from the AP hip radiograph by rotation-based change in
the projected prosthetic neck–shaft (NSA*) angle using the mathematical
formula ST = arcos [tan (NSA*) / tan (135)]. We used two independent
observers who repeated the analysis after a six-week interval. Radiological
measurements were compared with 3D-CT measurements by an independent,
blinded external institute. We found a mean difference of 1.2° ( We found that femoral tilt was associated with the mean radiological
measurement error (r = 0.22, p = 0.02). The projected neck–shaft angle is a reliable method for measuring
stem version on AP radiographs of the hip after a THA. However,
a highly standardised radiological technique is required for its
precise measurement. Cite this article:
The use of ilioischial cage reconstruction for
pelvic discontinuity has been replaced by the Trabecular Metal (Zimmer,
Warsaw, Indiana) cup-cage technique in our institution, due to the
unsatisfactory outcome of using a cage alone in this situation.
We report the outcome of 26 pelvic discontinuities in 24 patients
(20 women and four men, mean age 65 years (44 to 84)) treated by
the cup-cage technique at a mean follow-up of 82 months (12 to 113)
and compared them with a series of 19 pelvic discontinuities in
19 patients (18 women and one man, mean age 70 years (42 to 86))
treated with a cage at a mean follow-up of 69 months (1 to 170).
The clinical and radiological outcomes as well as the survivorship
of the groups were compared. In all, four of the cup-cage group
(15%) and 13 (68%) of the cage group failed due to septic or aseptic
loosening. The seven-year survivorship was 87.2% (95% confidence interval
(CI) 71 to 103) for the cup-cage group and 49.9% (95% CI 15 to 84)
for the cage-alone group (p = 0.009). There were four major complications
in the cup-cage group and nine in the cage group. Radiological union
of the discontinuity was found in all successful cases in the cup-cage
group and three of the successful cage cases. Three hips in the
cup-cage group developed early radiological migration of the components,
which stabilised with a successful outcome. Cup-cage reconstruction is a reliable technique for treating
pelvic discontinuity in mid-term follow-up and is preferred to ilioischial
cage reconstruction. If the continuity of the bone graft at the
discontinuity site is not disrupted, early migration of the components
does not necessarily result in failure. Cite this article:
There are few reports describing the technique
of managing acetabular protrusio in primary total hip replacement. Most
are small series with different methods of addressing the challenges
of significant medial and proximal migration of the joint centre,
deficient medial bone and reduced peripheral bony support to the
acetabular component. We describe our technique and the clinical
and radiological outcome of using impacted morsellised autograft
with a porous-coated cementless cup in 30 primary THRs with mild
(n = 8), moderate (n = 10) and severe (n = 12) grades of acetabular
protrusio. The mean Harris hip score had improved from 52 pre-operatively
to 85 at a mean follow-up of 4.2 years (2 to 10). At final follow-up,
27 hips (90%) had a good or excellent result, two (7%) had a fair
result and one (3%) had a poor result. All bone grafts had united
by the sixth post-operative month and none of the hips showed any
radiological evidence of recurrence of protrusio, osteolysis or
loosening. By using impacted morsellised autograft and cementless
acetabular components it was possible to achieve restoration of
hip mechanics, provide a biological solution to bone deficiency
and ensure long-term fixation without recurrence in arthritic hips
with protrusio undergoing THR. Cite this article:
Squeaking arising from a ceramic-on-ceramic (CoC)
total hip replacement (THR) may cause patient concern and in some
cases causes patients to seek revision surgery. We performed a meta-analysis
to determine the incidence of squeaking and the incidence of revision
surgery for squeaking. A total of 43 studies including 16 828 CoC
THR that reported squeaking, or revision for squeaking, were entered
into the analysis. The incidence of squeaking was 4.2% and the incidence
of revision for squeaking was 0.2%. The incidence of squeaking in
patients receiving the Accolade femoral stem was 8.3%, and the incidence
of revision for squeaking in these patients was 1.3%. Cite this article:
We report patient-reported outcomes and complications
associated with contemporary periacetabular osteotomy (PAO) surgery
in treating symptomatic acetabular dysplasia and compare these outcomes
with total hip arthroplasty (THA) in patients with similar demographic
details. Two consecutive cohorts included patients between aged
18 to 40 years who had undergone either PAO (100 hips; 24 male,
76 female) or THA (55 hips; 18 male, 37 female). At a mean follow-up
of 5.9 years (2 to 13), there was significant improvement in the
modified Harris hip pain (p <
0.001, PAO and p <
0.001, THA),
function (p <
0.001, PAO and p = 0.001, THA), and total scores
(p <
0.001, PAO and p <
0.001, THA) within each cohort. There
were no significant differences in the clinical outcome scores between
the groups. Complication rates were low and similar in each cohort
(p = 0.68). Similar to THA, contemporary PAO surgery is a clinically
effective procedure that improves function and activity levels,
provides pain relief and is associated with an acceptable complication
rate. Cite this article:
When fracture of an extensively porous-coated
femoral component occurs, its removal at revision total hip arthroplasty
(THA) may require a femoral osteotomy and the use of a trephine.
The remaining cortical bone after using the trephine may develop
thermally induced necrosis. A retrospective review identified 11
fractured, well-fixed, uncemented, extensively porous-coated femoral
components requiring removal using a trephine with a minimum of
two years of follow-up. The mean time to failure was 4.6 years (1.7 to 9.1, standard
deviation ( A total of four patients (36.4%) required further revision: three
for instability and one for fracture of the revision component.
There was no statistically significant difference in the mean Harris
hip score before implant fracture (82.4; These findings suggest that removal of a fractured, well-fixed,
uncemented, extensively porous-coated femoral component using a
trephine does not compromise subsequent fixation at revision THA
and the patient’s pre-operative level of function can be restored.
However, the loss of proximal bone stock before revision may be associated
with a high rate of dislocation post-operatively. Cite this article:
The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos
R) bone cement is no longer commercially available and was superseded
by Refobacin bone cement R and Palacos R + G in 2005. However, the
performance of these newly introduced bone cements have not been
tested in a phased evidence-based manner, including roentgen stereophotogrammetric
analysis (RSA). In this blinded, randomised, clinical RSA study, the migration
of the Stanmore femoral component was compared between Refobacin
bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties.
The primary outcome measure was femoral component migration measured
using RSA and secondary outcomes were Harris hip score (HHS), Hip
disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D)
and Short Form 36 (SF-36).Aims
Patients and Methods
We evaluated the outcome of 41 consecutive Charnley
low-friction arthroplasties (LFAs) performed by a single surgeon
in 28 patients aged ≤ 35 years at operation between 23 and 36 years
previously. There were 20 women and eight men with a mean age of
32 years (23 to 35) at surgery. Two patients (three hips) were lost
to follow-up at 12 and 17 years post-operatively, respectively,
and one patient (one hip) died at 13 years post-operatively. These patients
were excluded from the final evaluation. The survival rate of the
acetabular components was 92.7% (95% confidence interval (CI) 88.7
to 96.7) at ten years, 67.1% (95% CI 59.75 to 74.45) at 20 years
and 53.2% (95% CI 45.3 to 61.1) at 25 years. For the femoral component
the survival was 95.1% (95% CI 91.8 to 98.5) at ten years, 77.1%
(95% CI 73.9 to 80.3) at 20 years and 68.2% (95% CI 60.7 to 75.8)
at 25 years. The results indicate that the Charnley LFA remains
a reasonable choice in the treatment of young patients and can serve
for comparison with newer techniques and implants. Cite this article:
The aims of this study were to determine the
functional impact and financial burden of isolated and recurrent dislocation
after total hip arthroplasty (THA). Our secondary goal was to determine
whether there was a difference between patients who were treated
non-operatively and those who were treated operatively. We retrospectively reviewed 71 patients who had suffered dislocation
of a primary THA. Their mean age was 67 years (41 to 92) and the
mean follow-up was 3.8 years (2.1 to 8.2). Because patients with recurrent dislocation were three times
more likely to undergo operative treatment (p <
0.0001), they
ultimately had a significantly higher mean Harris Hip Score (HHS)
(p = 0.0001), lower mean Western Ontario and McMaster Universities
Arthritis Index (WOMAC) scores (p = 0.001) and a higher mean SF-12
score (p <
0.0001) than patients with a single dislocation. Likewise,
those who underwent operative treatment had a higher mean HHS (p
<
0.0001), lower mean WOMAC score (p <
0.0001) and a higher
mean SF-12 score (p <
0.0001) than those who were treated non-operatively. Recurrent dislocation and operative treatment increased costs
by 300% (£11 456;
p <
0.0001) and 40% (£5217; p <
0.0001), respectively. The operative treatment of recurrent dislocation results in significantly
better function than non-operative management. Moreover, the increase
in costs for operative treatment is modest compared with that of
non-operative measures. Cite this article:
We report a multicentre prospective consecutive
study assessing the long-term outcome of the proximally hydroxyapatite
(HA)-coated ABG II monobloc femoral component in a series of 1148
hips in 1053 patients with a mean age at surgery of 64.77 years
(22 to 80) at a mean follow-up of 10.84 years (10 to 15.25). At
latest follow-up, the mean total Harris hip score was 94.7 points
( Cite this article: