Advertisement for orthosearch.org.uk
Results 721 - 740 of 1140
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 780 - 785
1 Jun 2015
Baauw M van Hellemondt GG van Hooff ML Spruit M

We evaluated the accuracy with which a custom-made acetabular component could be positioned at revision arthroplasty of the hip in patients with a Paprosky type 3 acetabular defect.

A total of 16 patients with a Paprosky type 3 defect underwent revision surgery using a custom-made trabecular titanium implant. There were four men and 12 women with a median age of 67 years (48 to 79). The planned inclination (INCL), anteversion (AV), rotation and centre of rotation (COR) of the implant were compared with the post-operative position using CT scans.

A total of seven implants were malpositioned in one or more parameters: one with respect to INCL, three with respect to AV, four with respect to rotation and five with respect to the COR.

To the best of our knowledge, this is the first study in which CT data acquired for the pre-operative planning of a custom-made revision acetabular implant have been compared with CT data on the post-operative position. The results are encouraging.

Cite this article: Bone Joint J 2015; 97-B:780–5.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 724 - 731
1 Jun 2017
Mei-Dan O Jewell D Garabekyan T Brockwell J Young DA McBryde CW O’Hara JN

Aims

The aim of this study was to evaluate the long-term clinical and radiographic outcomes of the Birmingham Interlocking Pelvic Osteotomy (BIPO).

Patients and Methods

In this prospective study, we report the mid- to long-term clinical outcomes of the first 100 consecutive patients (116 hips; 88 in women, 28 in men) undergoing BIPO, reflecting the surgeon’s learning curve. Failure was defined as conversion to hip arthroplasty. The mean age at operation was 31 years (7 to 57). Three patients (three hips) were lost to follow-up.


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1024 - 1030
1 Aug 2015
Whitehouse MR Endo M Zachara S Nielsen TO Greidanus NV Masri BA Garbuz DS Duncan CP

Adverse reaction to wear and corrosion debris is a cause for concern in total hip arthroplasty (THA). Modular junctions are a potential source of such wear products and are associated with secondary pseudotumour formation.

We present a consecutive series of 17 patients treated at our unit for this complication following metal-on-highly cross-linked polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as infection, and present the aggregate laboratory results and pathological findings in this series.

The clinical presentation was pain, swelling or instability. Solid, cystic and mixed soft-tissue lesions were noted on imaging and confirmed intra-operatively. Corrosion at the head–neck junction was noted in all cases. No bacteria were isolated on multiple pre- and intra-operative samples yet the mean erythrocyte sedimentation rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and stromal polymorphonuclear cell counts were noted in nine cases.

Adverse soft–tissue reactions can occur in MoP THA owing to corrosion products released from the head–neck junction. The diagnosis should be carefully considered when investigating pain after THA. This may avoid the misdiagnosis of periprosthetic infection with an unidentified organism and mitigate the unnecessary management of these cases with complete single- or two-stage exchange.

Cite this article: Bone Joint J 2015;97-B:1024–1030.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1459 - 1463
1 Nov 2014
Lee YK Ha YC Koo KH

Large femoral heads have become popular in total hip replacement (THR) as a method of reducing the risk of dislocation. However, if large heads are used in ceramic-on-ceramic THR, the liner must be thinner, which may increase the risk of fracture. To compare the rates of ceramic fracture and dislocation between 28 mm and 32 mm ceramic heads, 120 hips in 109 patients (51 men and 58 women, mean age 49.2 years) were randomised to THR with either a 28 mm or a 32 mm ceramic articulation. A total of 57/60 hips assigned to the 28 mm group and 55/60 hips assigned to the 32 mm group were followed for at least five years. No ceramic component fractures occured in any patient in either group. There was one dislocation in the 32 mm group and none in the 28 mm group (p = 0.464). No hip had detectable wear, focal osteolysis or prosthetic loosening. In our small study the 32 mm ceramic articulation appeared to be safe in terms of ceramic liner fracture.

Cite this article: Bone Joint J 2014;96-B:1459–63.


Bone & Joint Research
Vol. 6, Issue 6 | Pages 391 - 398
1 Jun 2017
Lenguerrand E Whitehouse MR Beswick AD Jones SA Porter ML Blom* AW

Objectives

We used the National Joint Registry for England, Wales, Northern Ireland and the Isle of Man (NJR) to investigate the risk of revision due to prosthetic joint infection (PJI) for patients undergoing primary and revision hip arthroplasty, the changes in risk over time, and the overall burden created by PJI.

Methods

We analysed revision total hip arthroplasties (THAs) performed due to a diagnosis of PJI and the linked index procedures recorded in the NJR between 2003 and 2014. The cohort analysed consisted of 623 253 index primary hip arthroplasties, 63 222 index revision hip arthroplasties and 7585 revision THAs performed due to a diagnosis of PJI. The prevalence, cumulative incidence functions and the burden of PJI (total procedures) were calculated. Overall linear trends were investigated with log-linear regression.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 732 - 740
1 Jun 2017
Meermans G Konan S Das R Volpin A Haddad FS

Aims

The most effective surgical approach for total hip arthroplasty (THA) remains controversial. The direct anterior approach may be associated with a reduced risk of dislocation, faster recovery, reduced pain and fewer surgical complications. This systematic review aims to evaluate the current evidence for the use of this approach in THA.

Materials and Methods

Following the Cochrane collaboration, an extensive literature search of PubMed, Medline, Embase and OvidSP was conducted. Randomised controlled trials, comparative studies, and cohort studies were included. Outcomes included the length of the incision, blood loss, operating time, length of stay, complications, and gait analysis.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 771 - 779
1 Jun 2015
te Stroet MAJ Rijnen WHC Gardeniers JWM van Kampen A Schreurs BW

We report the clinical and radiographic outcomes of 208 consecutive femoral revision arthroplasties performed in 202 patients (119 women, 83 men) between March 1991 and December 2007 using the X-change Femoral Revision System, fresh-frozen morcellised allograft and a cemented polished Exeter stem. All patients were followed prospectively. The mean age of the patients at revision was 65 years (30 to 86). At final review in December 2013 a total of 130 patients with 135 reconstructions (64.9%) were alive and had a non re-revised femoral component after a mean follow-up of 10.6 years (4.7 to 20.9). One patient was lost to follow-up at six years, and their data were included up to this point. Re-operation for any reason was performed in 33 hips (15.9%), in 13 of which the femoral component was re-revised (6.3%). The mean pre-operative Harris hip score was 52 (19 to 95) (n = 73) and improved to 80 (22 to 100) (n = 161) by the last follow-up. Kaplan–Meier survival with femoral re-revision for any reason as the endpoint was 94.9% (95% confidence intervals (CI) 90.2 to 97.4) at ten years; with femoral re-revision for aseptic loosening as the endpoint it was 99.4% (95% CI 95.7 to 99.9); with femoral re-operation for any reason as the endpoint it was 84.5% (95% CI 78.3 to 89.1); and with subsidence ≥ 5 mm it was 87.3% (95% CI 80.5 to 91.8). Femoral revision with the use of impaction allograft bone grafting and a cemented polished stem results in a satisfying survival rate at a mean of ten years’ follow-up.

Cite this article: Bone Joint J 2015; 97-B:771–9.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 762 - 770
1 Jun 2015
Pennington MW Grieve R van der Meulen JH

There is little evidence on the cost effectiveness of different brands of hip prostheses. We compared lifetime cost effectiveness of frequently used brands within types of prosthesis including cemented (Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed) and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy). We used data from three linked English national databases to estimate the lifetime risk of revision, quality-adjusted life years (QALYs) and cost.

For women with osteoarthritis aged 70 years, the Exeter V40 Elite Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0 QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision risk, 9.22 QALYs), the most commonly used brand, and assuming a willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is most cost effective, with an incremental net monetary benefit of £876. Differences in cost effectiveness between the hybrid CPT Trilogy and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc Exceed were small, and a cautious interpretation is required, given the limitations of the available information.

However, it is unlikely that cemented brands are among the most cost effective. Similar patterns of results were observed for men and other ages. The gain in quality of life after total hip arthroplasty, rather than the risk of revision, was the main driver of cost effectiveness.

Cite this article: Bone Joint J 2015;97-B:762–70.


The Bone & Joint Journal
Vol. 99-B, Issue 3 | Pages 325 - 329
1 Mar 2017
Viste A Perry KI Taunton MJ Hanssen AD Abdel MP

Aims

Loss or absence of proximal femoral bone in revision total hip arthroplasty (THA) remains a significant challenge. While the main indication for the use of proximal femoral replacements (PFRs) is in the treatment of malignant disease, they have a valuable role in revision THA for loosening, fracture and infection in patients with bone loss. Our aim was to determine the clinical outcomes, implant survivorship, and complications of PFRs used in revision THA for indications other than malignancy.

Patients and Methods

A retrospective review of 44 patients who underwent revision THA using a PFR between 2000 and 2013 was undertaken. Their mean age was 79 years (53 to 97); 31 (70%) were women. The bone loss was classified as Paprosky IIIB or IV in all patients. The mean follow-up was six years (2 to 12), at which time 22 patients had died and five were lost to follow-up.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 754 - 760
1 Jun 2016
Malek IA Royce G Bhatti SU Whittaker JP Phillips SP Wilson IRB Wootton JR Starks I

Aims

We assessed the difference in hospital based and early clinical outcomes between the direct anterior approach and the posterior approach in patients who undergo total hip arthroplasty (THA).

Patients and Methods

The outcome was assessed in 448 (203 males, 245 females) consecutive patients undergoing unilateral primary THA after the implementation of an ‘Enhanced Recovery’ pathway. In all, 265 patients (mean age: 71 years (49 to 89); 117 males and 148 females) had surgery using the direct anterior approach (DAA) and 183 patients (mean age: 70 years (26 to 100); 86 males and 97 females) using a posterior approach. The groups were compared for age, gender, American Society of Anesthesiologists grade, body mass index, the side of the operation, pre-operative Oxford Hip Score (OHS) and attendance at ‘Joint school’. Mean follow-up was 18.1 months (one to 50).


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 614 - 622
1 May 2017
Grammatopoulos G Bolduc M Atkins BL Kendrick BJL McLardy-Smith P Murray DW Gundle R Taylor AH

Aims

Advocates of debridement, antibiotics and implant retention (DAIR) in hip periprosthetic joint infection (PJI) argue that a procedure not disturbing a sound prosthesis-bone interface is likely to lead to better survival and functional outcome compared with revision. This case-control study aims were to compare outcome of DAIRs for infected primary total hip arthroplasty (THA) with outcomes following primary THA and two-stage revision of infected primary THAs.

Patients and Methods

We retrospectively reviewed all DAIRs, performed for confirmed infected primary hip arthropasty (n = 82) at out institution, between 1997 and 2013. Data recorded included full patient information and type of surgery. Outcome measures included complications, mortality, implant survivorship and functional outcome. Outcome was compared with two control groups matched for gender and age; a cohort of primary THAs (n = 120) and a cohort of two-stage revisions for infection (n = 66).


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1472 - 1477
1 Nov 2014
Vioreanu MH Parry MC Haddad FS Duncan CP

The Unified Classification System (UCS) emphasises the key principles in the assessment and management of peri-prosthetic fractures complicating partial or total joint replacement.

We tested the inter- and intra-observer agreement for the UCS as applied to the pelvis and femur using 20 examples of peri-prosthetic fracture in 17 patients. Each subtype of the UCS was represented by at least one case. Specialist orthopaedic surgeons (experts) and orthopaedic residents (pre-experts) assessed reliability on two separate occasions.

For the pelvis, the UCS showed inter-observer agreement of 0.837 (95% confidence intervals (CI) 0.798 to 0.876) for the experts and 0.728 (95% CI 0.689 to 0.767) for the pre-experts. The intra-observer agreement for the experts was 0.861 (95% CI 0.760 to 0.963) and 0.803 (95% 0.688 to 0.918) for the pre-experts. For the femur, the UCS showed an inter-observer kappa value of 0.805 (95% CI 0.765 to 0.845) for the experts and a value of 0.732 (95% CI 0.690 to 0.773) for the pre-experts. The intra-observer agreement was 0.920 (95% CI 0.867 to 0.973) for the experts, and 0.772 (95% CI 0.652 to 0.892) for the pre-experts. This corresponds to a substantial and ‘almost perfect’ inter- and intra-observer agreement for the UCS for peri-prosthetic fractures of the pelvis and femur.

We hope that unifying the terminology of these injuries will assist in their assessment, treatment and outcome.

Cite this article: Bone Joint J 2014;96-B:1472–7.


The Bone & Joint Journal
Vol. 99-B, Issue 4_Supple_B | Pages 33 - 40
1 Apr 2017
Galea VP Laaksonen I Matuszak SJ Connelly JW Muratoglu O Malchau H

Aims

Our first aim was to determine whether there are significant changes in the level of metal ions in the blood at mid-term follow-up, in patients with an Articular Surface Replacement (ASR) arthroplasty. Secondly, we sought to identify risk factors for any increases.

Patients and Methods

The study involved 435 patients who underwent unilateral, metal-on-metal (MoM) hip resurfacing (HRA) or total hip arthroplasty (THA). These patients all had one measurement of the level of metal ions in the blood before seven years had passed post-operatively (early evaluation) and one after seven years had passed post-operatively (mid-term evaluation). Changes in ion levels were tested using a Wilcoxon signed-rank test. We identified subgroups at the highest risk of increase using a multivariable linear logistic regression model.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 451 - 457
1 Apr 2017
Gromov K Bersang A Nielsen CS Kallemose T Husted H Troelsen A

Aims

The aim of this study was to identify patient- and surgery-related risk factors for sustaining an early periprosthetic fracture following primary total hip arthroplasty (THA) performed using a double-tapered cementless femoral component (Bi-Metric femoral stem; Biomet Inc., Warsaw, Indiana).

Patients and Methods

A total of 1598 consecutive hips, in 1441 patients receiving primary THA between January 2010 and June 2015, were retrospectively identified. Level of pre-operative osteoarthritis, femoral Dorr type and cortical index were recorded. Varus/valgus placement of the stem and canal fill ratio were recorded post-operatively. Periprosthetic fractures were identified and classified according to the Vancouver classification. Regression analysis was performed to identify risk factors for early periprosthetic fracture.


The Bone & Joint Journal
Vol. 98-B, Issue 5 | Pages 579 - 584
1 May 2016
Osman K Panagiotidou AP Khan M Blunn G Haddad FS

There is increasing global awareness of adverse reactions to metal debris and elevated serum metal ion concentrations following the use of second generation metal-on-metal total hip arthroplasties. The high incidence of these complications can be largely attributed to corrosion at the head-neck interface. Severe corrosion of the taper is identified most commonly in association with larger diameter femoral heads. However, there is emerging evidence of varying levels of corrosion observed in retrieved components with smaller diameter femoral heads. This same mechanism of galvanic and mechanically-assisted crevice corrosion has been observed in metal-on-polyethylene and ceramic components, suggesting an inherent biomechanical problem with current designs of the head-neck interface.

We provide a review of the fundamental questions and answers clinicians and researchers must understand regarding corrosion of the taper, and its relevance to current orthopaedic practice.

Cite this article: Bone Joint J 2016;98-B:579–84.


Aims

To investigate the longevity of uncemented fixation of a femoral component in total hip arthroplasty (THA) in patients with Dorr type C proximal femoral morphology.

Patients and Methods

A total of 350 consecutive uncemented THA in 320 patients were performed between 1983 and 1987, by a single surgeon using the Taperloc femoral component. The 63 patients (68 hips) with Dorr type C proximal femoral morphology were the focus of this review. The mean age of the patients was 69 years (24 to 88) and mean follow-up was 16.6 years (ten to 29). Survival analysis included eight patients (eight hips) who died without undergoing revision surgery prior to obtaining ten years follow-up. All 55 surviving patients (60 hips) were available for clinical assessment and radiographic review. As a comparator group, the survival and implant fixation in the remaining 282 THAs (257 patients) with Dorr type A and B morphology were evaluated. The mean age of these patients was 52 years (20 to 82).


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1600 - 1609
1 Dec 2014
Matharu GS Pynsent PB Sumathi VP Mittal S Buckley CD Dunlop DJ Revell PA Revell MP

We undertook a retrospective cohort study to determine clinical outcomes following the revision of metal-on-metal (MoM) hip replacements for adverse reaction to metal debris (ARMD), and to identify predictors of time to revision and outcomes following revision. Between 1998 and 2012 a total of 64 MoM hips (mean age at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings and 18 (28%) total hip replacements) were revised for ARMD at one specialist centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision for ARMD there were 13 hips (20.3%) with post-operative complications and eight (12.5%) requiring re-revision.

The Kaplan–Meier five-year survival rate for ARMD revision was 87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding re-revisions, the median absolute Oxford hip score (OHS) following ARMD revision using the percentage method (0% best outcome and 100% worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%), which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using the modified OHS. Histopathological response did not affect time to revision for ARMD (p = 0.334) or the subsequent risk of re-revision (p = 0.879). Similarly, the presence or absence of a contralateral MoM hip bearing did not affect time to revision for ARMD (p = 0.066) or the subsequent risk of re-revision (p = 0.178).

Patients revised to MoM bearings had higher rates of re-revision (five of 16 MoM hips re-revised; p = 0.046), but those not requiring re-revision had good functional results (median absolute OHS 14.6% or 41.0/48). Short-term morbidity following revision for ARMD was comparable with previous reports. Caution should be exercised when choosing bearing surfaces for ARMD revisions.

Cite this article: Bone Joint J 2014;96-B:1600–9.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 925 - 933
1 Jul 2016
Sidaginamale RP Joyce TJ Bowsher JG Lord JK Avery PJ Natu S Nargol AVF Langton DJ

Aims

We wished to investigate the influence of metal debris exposure on the subsequent immune response and resulting soft-tissue injury following metal-on-metal (MoM) hip arthroplasty. Some reports have suggested that debris generated from the head-neck taper junction is more destructive than equivalent doses from metal bearing surfaces.

Patients and Methods

We investigated the influence of the source and volume of metal debris on chromium (Cr) and cobalt (Co) concentrations in corresponding blood and hip synovial fluid samples and the observed agglomerated particle sizes in excised tissues using multiple regression analysis of prospectively collected data. A total of 199 explanted MoM hips (177 patients; 132 hips female) were analysed to determine rates of volumetric wear at the bearing surfaces and taper junctions.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 883 - 889
1 Jul 2015
Jassim SS Patel S Wardle N Tahmassebi J Middleton R Shardlow DL Stephen A Hutchinson J Haddad FS

Oxidised zirconium (OxZi) has been developed as an alternative bearing surface for femoral heads in total hip arthroplasty (THA). This study has investigated polyethylene wear, functional outcomes and complications, comparing OxZi and cobalt–chrome (CoCr) as part of a three-arm, multicentre randomised controlled trial. Patients undergoing THA from four institutions were prospectively randomised into three groups. Group A received a CoCr femoral head and highly cross-linked polyethylene (XLPE) liner; Group B received an OxZi femoral head and XLPE liner; Group C received an OxZi femoral head and ultra-high molecular weight polyethylene (UHMWPE) liner. At five years, 368 patients had no statistically significant differences in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western Ontario and McMaster Universities Osteoarthritis Index (p = 0.847), pain scores (p = 0.458) or complications. The mean rate of linear wear was 0.028 mm/year (standard deviation (sd) 0.010) for Group A, 0.023 mm/year (sd 0.010) for Group B, and 0.09 mm/year (sd 0.045) for Group C. Penetration was significantly higher in the UHMWPE liner group compared with both XLPE liner groups (p < 0.001) but no significant difference was noted between CoCr and OxZi when articulating with XLPE (p = 0.153). In this, the largest randomised study of this bearing surface, it appears that using a XLPE acetabular liner is more important in reducing THA component wear than the choice of femoral head bearing, at mid-term follow-up. There is a non-significant trend towards lower wear, coupling OxZi rather than CoCr with XLPE but long-term analysis is required to see if this observation changes with time and becomes significant.

Cite this article: Bone Joint J 2015;97-B:883–9.


The Bone & Joint Journal
Vol. 97-B, Issue 12 | Pages 1615 - 1622
1 Dec 2015
Müller M Abdel MP Wassilew GI Duda G Perka C

The accurate reconstruction of hip anatomy and biomechanics is thought to be important in achieveing good clinical outcomes following total hip arthroplasty (THA). To this end some newer hip designs have introduced further modularity into the design of the femoral component such that neck­shaft angle and anteversion, which can be adjusted intra-operatively. The clinical effect of this increased modularity is unknown. We have investigated the changes in these anatomical parameters following conventional THA with a prosthesis of predetermined neck–shaft angle and assessed the effect of changes in the hip anatomy on clinical outcomes.

In total, 44 patients (mean age 65.3 years (standard deviation (sd) 7); 17 male/27 female; mean body mass index 26.9 (kg/m²) (sd 3.1)) underwent a pre- and post-operative three-dimensional CT scanning of the hip. The pre- and post-operative neck–shaft angle, offset, hip centre of rotation, femoral anteversion, and stem alignment were measured. Additionally, a functional assessment and pain score were evaluated before surgery and at one year post-operatively and related to the post-operative anatomical changes.

The mean pre-operative neck–shaft angle was significantly increased by 2.8° from 128° (sd 6.2; 119° to 147°) to 131° (sd 2.1; 127° to 136°) (p = 0.009). The mean pre-operative anteversion was 24.9° (sd 8; 7.9 to 39.1) and reduced to 7.4° (sd 7.3; -11.6° to 25.9°) post-operatively (p < 0.001). The post-operative changes had no influence on function and pain. Using a standard uncemented femoral component, high pre- and post-operative variability of femoral anteversion and neck–shaft angles was found with a significant decrease of the post-operative anteversion and slight increase of the neck–shaft angles, but without any impact on clinical outcome.

Cite this article: Bone Joint J 2015;97-B:1615–22.