Femoral stem version has a major influence on
impingement and early post-operative stability after total hip arthroplasty
(THA). The main objective of this study was to evaluate the validity
of a novel radiological method for measuring stem version. Anteroposterior
(AP) radiographs and three-dimensional CT scans were obtained for
115 patients (female/male 63/72, mean age 62.5 years (50 to 75))
who had undergone minimally invasive, cementless THA. Stem version was
calculated from the AP hip radiograph by rotation-based change in
the projected prosthetic neck–shaft (NSA*) angle using the mathematical
formula ST = arcos [tan (NSA*) / tan (135)]. We used two independent
observers who repeated the analysis after a six-week interval. Radiological
measurements were compared with 3D-CT measurements by an independent,
blinded external institute. We found a mean difference of 1.2° ( We found that femoral tilt was associated with the mean radiological
measurement error (r = 0.22, p = 0.02). The projected neck–shaft angle is a reliable method for measuring
stem version on AP radiographs of the hip after a THA. However,
a highly standardised radiological technique is required for its
precise measurement. Cite this article:
All-suture anchors are increasingly used in rotator cuff repair procedures. Potential benefits include decreased bone damage. However, there is limited published evidence for the relative strength of fixation for all-suture anchors compared with traditional anchors. A total of four commercially available all-suture anchors, the ‘Y-Knot’ (ConMed), Q-FIX (Smith & Nephew), ICONIX (Stryker) and JuggerKnot (Zimmer Biomet) and a traditional anchor control TWINFIX Ultra PK Suture Anchor (Smith & Nephew) were tested in cadaveric human humeral head rotator cuff repair models (n = 24). This construct underwent cyclic loading applied by a mechanical testing rig (Zwick/Roell). Ultimate load to failure, gap formation at 50, 100, 150 and 200 cycles, and failure mechanism were recorded. Significance was set at p < 0.05.Objectives
Materials and Methods
Whiplash injury is surrounded by controversy in both the medical and legal world. The debate on whether it is either a potentially serious medical condition or a social problem is ongoing. This paper briefly examines a selection of studies on low velocity whiplash injury (LVWI) and whiplash associated disorder (WAD) and touches upon the pathophysiological and epidemiological considerations, cultural and geographical differences and the effect of litigation on chronicity. The study concludes that the evidence for significant physical injury after LVWI is poor, and if significant disability is present after such injury, it will have to be explained in terms of psychosocial factors.
To assess the extent of osteointegration in two designs of shoulder
resurfacing implants. Bony integration to the Copeland cylindrical
central stem design and the Epoca RH conical-crown design were compared. Implants retrieved from six patients in each group were pair-matched.
Mean time to revision surgery of Copeland implants was 37 months
(standard deviation (Aims
Patients and Methods
Myxofibrosarcomas (MFSs) are malignant soft-tissue sarcomas characteristically
presenting as painless slowly growing masses in the extremities.
Locally infiltrative growth means that the risk of local recurrence
is high. We reviewed our experience to make recommendations about
resection strategies and the role of the multidisciplinary team
in the management of these tumours. Patients with a primary or recurrent MFS who were treated surgically
in our unit between 1997 and 2012 were included in the study. Clinical
records and imaging were reviewed. A total of 50 patients with a
median age of 68.4 years (interquartile range 61.6 to 81.8) were
included. There were 35 men; 49 underwent surgery in our unit.Aims
Patients and Methods
Hip resurfacing has been proposed as an alternative
to traditional total hip arthroplasty in young, active patients.
Much has been learned following the introduction of metal-on-metal resurfacing
devices in the 1990s. The triad of a well-designed device, implanted
accurately, in the correct patient has never been more critical
than with these implants. Following Food and Drug Administration approval in 2006, we studied
the safety and effectiveness of one hip resurfacing device (Birmingham
Hip Resurfacing) at our hospital in a large, single-surgeon series.
We report our early to mid–term results in 1333 cases followed for
a mean of 4.3 years (2 to 5.7) using a prospective, observational
registry. The mean patient age was 53.1 years (12 to 84); 70% were
male and 91% had osteoarthritis. Complications were few, including
no dislocations, no femoral component loosening, two femoral neck
fractures (0.15%), one socket loosening (0.08%), three deep infections
(0.23%), and three cases of metallosis (0.23%). There were no destructive
pseudotumours. Overall survivorship at up to 5.7 years was 99.2%. Aseptic survivorship
in males under the age of 50 was 100%. We believe this is the largest
United States series of a single surgeon using a single resurfacing
system. Cite this article:
The goals of this study were to define the risk factors, nature,
chronology, and treatment strategies adopted for periprosthetic
femoral fractures in 32 644 primary total hip arthroplasties (THAs). There were 564 intra-operative fractures (1.7%); 529 during uncemented
stem placement (3.0%) and 35 during cemented stem placement (0.23%).
Intra-operative fractures were more common in females and patients
over 65 years (p <
0.001). The majority occurred during placement
of the femoral component (60%), and involved the calcar (69%). There
were 557 post-operative fractures (20-year probability: 3.5%; 95%
confidence interval (CI) 3.2 to 3.9); 335 fractures after placement
of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to
9.1) and 222 after placement of a cemented stem (20-year probability:
2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture
within 30 days after an uncemented stem was ten times higher than
a cemented stem. The most common post-operative fracture type was
a Vancouver AG (32%; n = 135), with 67% occurring after
a fall. In all, 36% (n = 152) were treated with revision arthroplasty. Aim and Methods
Results
Our aim was to report survivorship data and lessons learned with
the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Between August 2005 and March 2015, a total of 4802 primary cementless
Corail/Pinnacle THAs were performed in 4309 patients. In March 2016,
we reviewed these hips from a prospectively maintained database.Aims
Patients and Methods
Diabetes mellitus is the most common co-morbidity associated
with necrotising fasciitis. This study aims to compare the clinical
presentation, investigations, Laboratory Risk Indicator for Necrotising
Fasciitis (LRINEC) score, microbiology and outcome of management
of this condition in diabetic and non-diabetic patients. The medical records of all patients with surgically proven necrotising
fasciitis treated at our institution between 2005 and 2014 were
reviewed. Diagnosis of necrotising fasciitis was made on findings
of ‘dishwater’ fluid, presence of greyish necrotic deep fascia and
lack of bleeding on muscle dissection found intra-operatively. Information
on patients’ demographics, presenting symptoms, clinical signs,
investigations, treatment and outcome were recorded and analysed.Aims
Patients and Methods
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
Smoking is associated with post-operative complications but smokers
often under-report the amount they smoke. Our objective was to determine
whether a urine dipstick test could be used as a substitute for
quantitative cotinine assays to determine smoking status in patients. Between September 2013 and July 2014 we conducted a prospective
cohort study in which 127 consecutive patients undergoing a planned
foot and ankle arthrodesis or osteotomy were included. Patients
self-reported their smoking status and were classified as: ‘never
smoked’ (61 patients), ‘ex-smoker’ (46 patients), or ‘current smoker’ (20
patients). Urine samples were analysed with cotinine assays and
cotinine dipstick tests.Aims
Patients and Methods
Rib fracture fixation by orthopaedic and cardiothoracic surgeons has become increasingly popular for the treatment of chest injuries in trauma. The literature, though mainly limited to Level II and III evidence, shows favourable results for operative fixation. In this paper we review the literature and discuss the indications for rib fracture fixation, surgical approaches, choice of implants and the future direction for management. With the advent of NICE guidance and new British Orthopaedic Association Standards for Trauma (BOAST) guidelines in production, the management of rib fractures is going to become more and more commonplace.
A trial-based comparison of the use of resources, costs and quality
of life outcomes of arthroscopic and open surgical management for
rotator cuff tears in the United Kingdom NHS was performed using
data from the United Kingdom Rotator Cuff Study (UKUFF) randomised
controlled trial. Using data from 273 patients, healthcare-related use of resources,
costs and quality-adjusted life years (QALYs) were estimated at
12 months and 24 months after surgery on an intention-to-treat basis
with adjustment for covariates. Uncertainty about the incremental
cost-effectiveness ratio for arthroscopic Aims
Patients and Methods
The surgical management of ankle arthritis with tibiotalar arthrodesis
is known to alter gait, as compared with normal ankles. The purpose
of this study was to assess post-operative gait function with gait
before arthrodesis. We prospectively studied 20 patients who underwent three-dimensional
gait analysis before and after tibiotalar arthrodesis. Cadence,
step length, walking velocity and total support time were assessed.
Kinetic parameters, including the moment and power of the ankle
in the sagittal plane and hip power were also recorded.Aims
Patients and Methods
We performed a retrospective, comparative study of elderly patients
with an increased risk from anaesthesia who had undergone either
anterior screw fixation (ASF) or halo vest immobilisation (HVI)
for a type II odontoid fracture. A total of 80 patients aged 65 years or more who had undergone
either ASF or HVI for a type II odontoid fracture between 1988 and
2013 were reviewed. There were 47 women and 33 men with a mean age
of 73 (65 to 96; standard deviation 7). All had an American Society
of Anesthesiologists score of 2 or more.Aims
Patients and Methods
We report the kinematic and early clinical results
of a patient- and observer-blinded randomised controlled trial in which
CT scans were used to compare potential impingement-free range of
movement (ROM) and acetabular component cover between patients treated
with either the navigated ‘femur-first’ total hip arthroplasty (THA) method
(n = 66; male/female 29/37, mean age 62.5 years; 50 to 74) or conventional
THA (n = 69; male/female 35/34, mean age 62.9 years; 50 to 75).
The Hip Osteoarthritis Outcome Score, the Harris hip score, the
Euro-Qol-5D and the Mancuso THA patient expectations score were
assessed at six weeks, six months and one year after surgery. A
total of 48 of the patients (84%) in the navigated ‘femur-first’
group and 43 (65%) in the conventional group reached all the desirable
potential ROM boundaries without prosthetic impingement for activities
of daily living (ADL) in flexion, extension, abduction, adduction
and rotation (p = 0.016). Acetabular component cover and surface
contact with the host bone were >
87% in both groups. There was
a significant difference between the navigated and the conventional
groups’ Harris hip scores six weeks after surgery (p = 0.010). There
were no significant differences with respect to any clinical outcome
at six months and one year of follow-up. The navigated ‘femur-first’
technique improves the potential ROM for ADL without prosthetic
impingement, although there was no observed clinical difference
between the two treatment groups. Cite this article: