Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated in
vitro to that seen in explanted bearings. The aim of this
study was to determine whether second-generation ceramic bearings
exhibited a better pattern of wear than those reported in the literature
for first-generation bearings. We considered both macro- and microscopic
findings. We found that long-term alumina wear in association with a loose
acetabular component could be categorised into three groups. Of
20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight
‘severe wear’, which was characterised by a change in the physical
shape of the bearing and a loss of volume. This suggests that the
wear in alumina–alumina bearings in association with a loose acetabular
component may be variable in pattern, and may explain, in part,
why the wear of a ceramic
Our retrospective analysis reports the outcome of patients operated for slipped capital femoral epiphysis using the modified Dunn procedure. Results, complications, and the need for revision surgery are compared with the recent literature. We retrospectively evaluated 17 patients (18 hips) who underwent the modified Dunn procedure for the treatment of slipped capital femoral epiphysis. Outcome measurement included standardized scores. Clinical assessment included ambulation, leg length discrepancy, and hip mobility. Radiographically, the quality of epiphyseal reduction was evaluated using the Southwick and Alpha-angles. Avascular necrosis, heterotopic ossifications, and osteoarthritis were documented at follow-up.Aims
Methods
We aimed to establish the short- and long-term efficacy of corticosteroid injection for coccydynia, and to determine if betamethasone or triamcinolone has the best effect. During 2009 to 2016, we treated 277 patients with chronic coccydynia with either one 6 mg betamethasone or one 20 mg triamcinolone cortisone injection. A susequent injection was given to 62 (26%) of the patients. All were reviewed three to four months after injection, and 241 replied to a questionnaire a mean of 36 months (12 to 88) after the last injection. No pain at the early review was considered early success. When the patient had not been subsequently operated on, and indicated on the questionnaire that they were either well or much better, it was considered a long-term success.Aims
Methods
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6) in situ. These
implants were obtained from 54 patients (16 men and 38 women) with
a mean age of 67 years (33 to 88) who underwent revision for a variety
of reasons. Posterior edge loading was found in the majority of
these retrievals (32 out of 54). Anterosuperior edge loading occurred
less often but produced a higher rate of wear. Stripe wear on the
femoral
We assessed the intermediate functional results of eight patients after wide resection of the proximal humerus for malignant bone tumour. We used a free vascularised fibular graft as a functional spacer and a sling procedure to preserve passive scapulohumeral movement. Scapulohumeral arthrodesis was not carried out. Five patients had osteosarcoma, two achondrosarcoma and one a malignant fibrous histiocytoma of the bone. The mean duration of follow-up was 70 months (median, 76) for the seven patients who were still alive at the time of the latest follow-up. One patient died from the disease 12 months after surgery. There were no local recurrences. The functional results were described and graded quantitatively according to the rating system of the Musculoskeletal Tumour Society. Our results were satisfactory with regard to pain, emotional acceptance and manual dexterity. Function and lifting ability were unsatisfactory in two patients. One patient had delayed union between host and graft, but this united after six months without further surgery. Radiographs of the shoulder showed absorption or collapse of the
We adduce the following conclusions from our experience of using this spring-loaded compression screw on completely displaced medial fractures of the femoral neck:. 1. That this method probably eliminates non-union when the
Despite advances in the treatment of paediatric hip disease, adolescent and young adult patients can develop early onset end-stage osteoarthritis. The aims of this study were to address the indications and medium-term outcomes for total hip arthroplasty (THA) with ceramic bearings for teenage patients. Surgery was performed by a single surgeon working in the paediatric orthopaedic unit of a tertiary referral hospital. Databases were interrogated from 2003 to 2017 for all teenage patients undergoing THA with a minimum 2.3 year follow-up. Data capture included patient demographics, the underlying hip pathology, number of previous surgeries, and THA prostheses used. Institutional ethical approval was granted to contact patients for prospective clinical outcomes and obtain up-to-date radiographs. In total, 60 primary hips were implanted in 51 patients (35 female, 16 male) with nine bilateral cases. The mean age was 16.7 years (12 to 19) and mean follow-up was 9.3 years (2.3 to 16.8).Aims
Methods
The primary source for the blood supply of the
Latissimus dorsi tendon transfer (LDTT) is technically
challenging. In order to clarify the local structural anatomy, we undertook
a morphometric study using six complete cadavers (12 shoulders).
Measurements were made from the tendon to the nearby neurovascular
structures with the arm in two positions: flexed and internally
rotated, and adducted in neutral rotation. The tendon was then transferred
and measurements were taken from the edge of the tendon to a reference
point on the humeral
Due to the overwhelming demand for trauma services, resulting from increasing emergency department attendances over the past decade, virtual fracture clinics (VFCs) have become the fashion to keep up with the demand and help comply with the BOA Standards for Trauma and Orthopaedics (BOAST) guidelines. In this article, we perform a systematic review asking, “How useful are VFCs?”, and what injuries and conditions can be treated safely and effectively, to help decrease patient face to face consultations. Our primary outcomes were patient satisfaction, clinical efficiency and cost analysis, and clinical outcomes. We performed a systematic literature search of all papers pertaining to VFCs, using the search engines PubMed, MEDLINE, and the Cochrane Database, according to the Preferred Reporting Items for Systematic review and Meta-Analysis (PRISMA) checklist. Searches were carried out and screened by two authors, with final study eligibility confirmed by the senior author.Background
Methods
Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median
We present a case of early retrieval of an Oxinium femoral head and corresponding polyethylene liner where there was significant surface damage to the
Osteolysis remains a common reason for revision
after total hip arthroplasty (THA). For osteolysis associated with loose
cups, revision is indicated. For osteolysis around a well-fixed
cup, the decision is more controversial. The data available data
support retention of the cupwith lesional treatment, working through
screw holes and access channels for debridement and grafting. The
choice of graft material to fill defects, if any, remains controversial. Several
studies demonstrate good survivorship with cup retention strategies.
Complete revision allows more complete debridement of the lesion
and better graft fill, and allows implantation of a modern cup,
typically with a full line of liners and bearing surfaces available.
Additionally, revision allows fine tuning of the orientation of
the cup, which may be advantageous for optimising hip stability.
The author prefers to retain a well-fixed cup if it meets the following
criteria: it is well-fixed to intra-operative testing, it is well-positioned,
it is of sufficient size to allow insertion of a new liner with
a reasonable
Intravenous dexamethasone has been shown to reduce immediate postoperative pain after total hip arthroplasty (THA), though the effects are short-lived. We aimed to assess whether two equivalent perioperative split doses were more effective than a single preoperative dose. A total of 165 patients were randomly assigned into three groups: two perioperative saline injections (Group A, placebo), a single preoperative dose of 20 mg dexamethasone and a postoperative saline injection (Group B), and two perioperative doses of 10 mg dexamethasone (Group C). Patients, surgeons, and staff collecting outcome data were blinded to allocation. The primary outcome was postoperative pain level reported on a ten-point Numerical Rating Scale (NRS) at rest and during activity. The use of analgesic and antiemetic rescue, incidence of postoperative nausea and vomiting (PONV), CRP and interleukin-6 (IL-6) levels, range of motion (ROM), length of stay (LOS), patient satisfaction, and the incidence of surgical site infection (SSI) and gastrointestinal bleeding (GIB) in the three months postoperatively, were also compared.Aims
Methods
1. The anatomy of the forefoot in hallux valgus is compared with the normal, with a review of the literature and descriptions of anatomical preparations, observations at operation and radiographs. 2. The early and essential lesions are stretching of the ligaments on the medial side of the metatarso-phalangeal joint that attach the medial sesamoid and basal phalanx to the metatarsal, and erosion of the ridge that separates the grooves for the sesamoids on the metatarsal
The efficacy and safety of intrawound vancomycin for preventing surgical site infection in primary hip and knee arthroplasty is uncertain. A systematic review of the literature was conducted, indexed from inception to March 2020 in PubMed, Web of Science, Cochrane Library, Embase, and Google Scholar databases. All studies evaluating the efficacy and/or safety of intrawound vancomycin in patients who underwent primary hip and knee arthroplasty were included. Incidence of periprosthetic joint infection (PJI), superficial infection, aseptic wound complications, acute kidney injury, anaphylactic reaction, and ototoxicity were meta-analyzed. Results were reported as odds ratios (ORs) and 95% confidence intervals (CIs). The quality of included studies was assessed using the risk of bias in non-randomized studies of interventions (ROBINS-I) assessment tool.Aims
Methods
This study is a prospective, non-randomized trial for the treatment of fractures of the medial malleolus using lean, bioabsorbable, rare-earth element (REE)-free, magnesium (Mg)-based biodegradable screws in the adult skeleton. A total of 20 patients with isolated, bimalleolar, or trimalleolar ankle fractures were recruited between July 2018 and October 2019. Fracture reduction was achieved through bioabsorbable Mg-based screws composed of pure Mg alloyed with zinc (Zn) and calcium (Ca) ( Mg-Zn0.45-Ca0.45, in wt.%; ZX00). Visual analogue scale (VAS) and the presence of complications (adverse events) during follow-up (12 weeks) were used to evaluate the clinical outcomes. The functional outcomes were analyzed through the range of motion (ROM) of the ankle joint and the American Orthopaedic Foot and Ankle Society (AOFAS) score. Fracture reduction and gas formation were assessed using several plane radiographs.Aims
Methods
We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital. All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with National Joint Registry (NJR) data and general practitioner (GP) records to assess whether they had undergone revision for any reason.Aims
Methods
There are few reports of the surgical treatment for late stage Freiburg’s disease with flattening of the metatarsal