The aim of this study was to analyse the gait
pattern, muscle force and functional outcome of patients who had undergone
replacement of the proximal tibia for tumour and alloplastic reconstruction
of the extensor mechanism using the patellar-loop technique. Between February 1998 and December 2009, we carried out wide
local excision of a primary sarcoma of the proximal tibia, proximal
tibial replacement and reconstruction of the extensor mechanism
using the patellar-loop technique in 18 patients. Of these, nine
were available for evaluation after a mean of 11.6 years (0.5 to
21.6). The strength of the knee extensors was measured using an
Isobex machine and gait analysis was undertaken in our gait assessment
laboratory. Functional outcome was assessed using the American Knee
Society (AKS) and Musculoskeletal Tumor Society (MSTS) scores. The gait pattern of the patients differed in ground contact time,
flexion heel strike, maximal flexion loading response and total
sagittal plane excursion. The mean maximum active flexion was 91°
(30° to 110°). The overall mean extensor lag was 1° (0° to 5°).
The mean extensor muscle strength was 25.8% (8.3% to 90.3%) of that
in the non-operated leg (p <
0.001). The mean functional scores
were 68.7% (43.4% to 83.3%) (MSTS) and 71.1 (30 to 90) (AKS functional
score). In summary, the results show that reconstruction of the extensor
mechanism using this technique gives good biomechanical and functional
results. The patients’ gait pattern is close to normal, except for
a somewhat stiff knee gait pattern. The strength of the extensor
mechanism is reduced, but sufficient for walking. Cite this article:
We retrospectively reviewed the outcomes of 33
consecutive patients who had undergone an extra-articular, total or
partial scapulectomy for a malignant tumour of the shoulder girdle
between 1 July 2001 and 30 September 2013. Of these, 26 had tumours
which originated in the scapula or the adjacent soft tissue and
underwent a classic Tikhoff–Linberg procedure, while seven with
tumours arising from the proximal humerus were treated with a modified
Tikhoff-Linberg operation. We used a Ligament Advanced Reinforcement
System for soft-tissue reconstruction in nine patients, but not
in the other 24. The mean Musculoskeletal Tumor Society score (MSTS) was 17.6
(95% confidence interval (CI) 15.9 to 19.4); 17.6 (95% CI 15.5 to
19.6) after the classic Tikhoff–Linberg procedure and 18.1 (95%
CI 13.8 to 22.3) after the modified Tikhoff–Linberg procedure. Patients
who had undergone a LARS soft-tissue reconstruction had a mean score
of 18.6 (95% (CI) 13.9 to 22.4) compared with 17.2 (95% CI 15.5
to 19.0) for those who did not. The Tikhoff–Linberg procedure is a useful method for wide resection
of a malignant tumour of the shoulder girdle which helps to preserve
hand and elbow function. The method of soft-tissue reconstruction
has no effect on functional outcome. Cite this article:
Periprosthetic joint infection (PJI) is one of
the most feared and challenging complications following total knee arthroplasty.
We provide a detailed description of our current understanding regarding
the management of PJI of the knee, including diagnostic aids,
pre-operative planning, surgical treatment, and outcome. Cite this article:
The February 2014 Children’s orthopaedics Roundup360 looks at: flexible plasters; dual 8-plate or ablation for knee epiphysiodesis; ultrasounds for pulled elbow; leg length without the radiation; Boyd amputation in limb deficiencies; gold standard club foot treatment; quadrupled semitendinosis graft effective in paediatric ACL reconstruction; and predicting complications following cerebral palsy hip reconstruction
Dislocation is one of the most common causes
of patient and surgeon dissatisfaction following hip replacement
and to treat it, the causes must first be understood. Patient factors
include age greater than 70 years, medical comorbidities, female
gender, ligamentous laxity, revision surgery, issues with the abductors,
and patient education. Surgeon factors include the annual quantity
of procedures and experience, the surgical approach, adequate restoration
of femoral offset and leg length, component position, and soft-tissue
or bony impingement. Implant factors include the design of the head
and neck region, and so-called skirts on longer neck lengths. There
should be offset choices available in order to restore soft-tissue
tension. Lipped liners aid in gaining stability, yet if improperly placed
may result in impingement and dislocation. Late dislocation may
result from polyethylene wear, soft-tissue destruction, trochanteric
or abductor disruption and weakness, or infection. Understanding
the causes of hip dislocation facilitates prevention in a majority
of instances. Proper pre-operative planning includes the identification
of patients with a high offset in whom inadequate restoration of
offset will reduce soft-tissue tension and abductor efficiency.
Component position must be accurate to achieve stability without impingement.
Finally, patient education cannot be over-emphasised, as most dislocations occur
early, and are preventable with proper instructions. Cite this article:
The August 2012 Hip &
Pelvis Roundup360 looks at: whether cemented hip replacement might be bad for your health; highly cross-linked polyethylene; iHOT-33 - a new hip outcome measure; hamstring injuries; total hip replacement; stemmed metal-on-metal THR; bipolar hemiarthroplasty, neuromuscular disease and dislocation; the high risk of secondary hemiarthroplasty; and whether we have to repair the labrum after all?
The purpose of this study was to undertake a
meta-analysis to determine whether there is lower polyethylene wear and
longer survival when using mobile-bearing implants in total knee
replacement when compared with fixed-bearing implants. Of 975 papers
identified, 34 trials were eligible for data extraction and meta-analysis
comprising 4754 patients (6861 knees). We found no statistically
significant differences between the two designs in terms of the incidence
of radiolucent lines, osteolysis, aseptic loosening or survival.
There is thus currently no evidence to suggest that the use of mobile-bearing
designs reduce polyethylene wear and prolong survival after total
knee replacement. Cite this article:
The management of spinal deformity in children
with univentricular cardiac pathology poses significant challenges to
the surgical and anaesthetic teams. To date, only posterior instrumented
fusion techniques have been used in these children and these are
associated with a high rate of complications. We reviewed our experience
of both growing rod instrumentation and posterior instrumented fusion
in children with a univentricular circulation. Six children underwent spinal corrective surgery, two with cavopulmonary
shunts and four following completion of a Fontan procedure. Three
underwent growing rod instrumentation, two had a posterior fusion
and one had spinal growth arrest. There were no complications following
surgery, and the children undergoing growing rod instrumentation
were successfully lengthened. We noted a trend for greater blood
loss and haemodynamic instability in those whose surgery was undertaken
following completion of a Fontan procedure. At a median follow-up
of 87.6 months (interquartile range (IQR) 62.9 to 96.5) the median
correction of deformity was 24.2% (64.5° (IQR 46° to 80°) We believe that early surgical intervention with growing rod
instrumentation systems allows staged correction of the spinal deformity
and reduces the haemodynamic insult to these physiologically compromised
children. Due to the haemodynamic changes that occur with the completed
Fontan circulation, the initial scoliosis surgery should ideally
be undertaken when in the cavopulmonary shunt stage. Cite this article:
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
Isolated patellofemoral arthritis is a common
condition and there are varying opinions on the most effective treatments.
Non-operative and operative treatments have failed to demonstrate
effective long-term treatment for those in an advanced stage of
the condition. Newer designs and increased technology in patellofemoral replacement
(PFR) have produced more consistent outcomes. This has led to a
renewed enthusiasm for this procedure. Newer PFR prostheses have
addressed the patellar maltracking issues plaguing some of the older designs.
Short-term results with contemporary prostheses and new technology
are described here. Cite this article:
In this paper we make the case for the use of
single-stage revision for infected knee arthroplasty.
The February 2015 Oncology Roundup360 looks at: Achieving global collaboration; A new standard for limb salvage; Inoperable chondrosarcoma and chemotherapy; Soft-tissue sarcoma and adjuvant chemotherapy; Missed diagnoses and malpractice in sarcoma; Radiofrequency and cartilage tumours
Trabecular metal (TM) augments are a relatively
new option for reconstructing segmental bone loss during acetabular
revision. We studied 34 failed hip replacements in 34 patients that
were revised between October 2003 and March 2010 using a TM acetabular
shell and one or two augments. The mean age of the patients at the
time of surgery was 69.3 years (46 to 86) and the mean follow-up
was 64.5 months (27 to 107). In all, 18 patients had a minor column
defect, 14 had a major column defect, and two were associated with
pelvic discontinuity. The hip centre of rotation was restored in
27 patients (79.4%). The Oxford hip score increased from a mean
of 15.4 points (6 to 25) before revision to a mean of 37.7 (29 to
47) at the final follow-up. There were three aseptic loosenings
of the construct, two of them in the patients with pelvic discontinuity.
One septic loosening also occurred in a patient who had previously
had an infected hip replacement. The augments remained stable in
two of the failed hips. Whenever there was a loose acetabular component
in contact with a stable augment, progressive metal debris shedding
was evident on the serial radiographs. Complications included another
deep infection treated without revision surgery. Good clinical and
radiological results can be expected for bone-deficient acetabula
treated by a TM cup and augment, but for pelvic discontinuities
this might not be a reliable option. Cite this article:
Bone loss in the proximal tibia and distal femur
is frequently encountered in revision knee replacement surgery.
The various options for dealing with this depend on the extent of
any bone loss. We present our results with the use of cementless
metaphyseal metal sleeves in 103 patients (104 knees) with a mean
follow-up of 43 months (30 to 65). At final follow-up, sleeves in
102 knees had good osseointegration. Two tibial sleeves were revised
for loosening, possibly due to infection. The average pre-operative Oxford Knee Score was 23 (11 to 36)
and this improved to 32 (15 to 46) post-operatively. These early
results encourage us to continue with the technique and monitor
the outcomes in the long term. Cite this article:
The aim of this study was to assess the role
of synovial C-reactive protein (CRP) in the diagnosis of chronic periprosthetic
hip infection. We prospectively collected synovial fluid from 89
patients undergoing revision hip arthroplasty and measured synovial
CRP, serum CRP, erythrocyte sedimentation rate (ESR), synovial white
blood cell (WBC) count and synovial percentages of polymorphonuclear
neutrophils (PMN). Patients were classified as septic or aseptic
by means of clinical, microbiological, serum and synovial fluid
findings. The high viscosity of the synovial fluid precluded the
analyses in nine patients permitting the results in 80 patients
to be studied. There was a significant difference in synovial CRP
levels between the septic (n = 21) and the aseptic (n = 59) cohort.
According to the receiver operating characteristic curve, a synovial
CRP threshold of 2.5 mg/l had a sensitivity of 95.5% and specificity
of 93.3%. The area under the curve was 0.96. Compared with serum
CRP and ESR, synovial CRP showed a high diagnostic value. According
to these preliminary results, synovial CRP may be a useful parameter
in diagnosing chronic periprosthetic hip infection. Cite this article:
Reported rates of dislocation in hip hemiarthroplasty
(HA) for the treatment of intra-capsular fractures of the hip, range
between 1% and 10%. HA is frequently performed through a direct
lateral surgical approach. The aim of this study is to determine
the contribution of the anterior capsule to the stability of a cemented
HA through a direct lateral approach. A total of five whole-body cadavers were thawed at room temperature,
providing ten hip joints for investigation. A Thompson HA was cemented
in place via a direct lateral approach. The cadavers were then positioned
supine, both knee joints were disarticulated and a digital torque
wrench was attached to the femur using a circular frame with three
half pins. The wrench applied an external rotation force with the
hip in extension to allow the hip to dislocate anteriorly. Each
hip was dislocated twice; once with a capsular repair and once without
repairing the capsule. Stratified sampling ensured the order in
which this was performed was alternated for the paired hips on each
cadaver. Comparing peak torque force in hips with the capsule repaired
and peak torque force in hips without repair of the capsule, revealed
a significant difference between the ‘capsule repaired’ (mean 22.96
Nm, standard deviation ( Cite this article:
Drug therapy forms an integral part of the management
of many orthopaedic conditions. However, many medicines can produce
serious adverse reactions if prescribed inappropriately, either
alone or in combination with other drugs. Often these hazards are
not appreciated. In response to this, the European Union recently
issued legislation regarding safety measures which member states
must adopt to minimise the risk of errors of medication. In March 2014 the Medicines and Healthcare products Regulatory
Agency and NHS England released a Patient Safety Alert initiative
focussed on errors of medication. There have been similar initiatives
in the United States under the auspices of The National Coordinating
Council for Medication Error and The Joint Commission on the Accreditation
of Healthcare Organizations. These initiatives have highlighted
the importance of informing and educating clinicians. Here, we discuss common drug interactions and contra-indications
in orthopaedic practice. This is germane to safe and effective clinical
care. Cite this article:
We evaluated an operative technique, described
by the Exeter Hip Unit, to assist accurate introduction of the femoral
component. We assessed whether it led to a reduction in the rate
of leg-length discrepancy after total hip arthroplasty (THA). A total of 100 patients undergoing THA were studied retrospectively;
50 were undertaken using the test method and 50 using conventional
methods as a control group. The groups were matched with respect
to patient demographics and the grade of surgeon. Three observers
measured the depth of placement of the femoral component on post-operative
radiographs and measured the length of the legs. There was a strong correlation between the depth of insertion
of the femoral component and the templated depth in the test group
(R = 0.92), suggesting accuracy of the technique. The mean leg-length
discrepancy was 5.1 mm (0.6 to 21.4) pre-operatively and 1.3 mm
(0.2 to 9.3) post-operatively. There was no difference between Consultants
and Registrars as primary surgeons. Agreement between the templated
and post-operative depth of insertion was associated with reduced
post-operative leg-length discrepancy. The intra-class coefficient
was R ≥ 0.88 for all measurements, indicating high observer agreement.
The post-operative leg-length discrepancy was significantly lower
in the test group (1.3 mm) compared with the control group (6.3
mm, p <
0.001). The Exeter technique is reproducible and leads to a lower incidence
of leg-length discrepancy after THA. Cite this article:
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score.Objectives
Methods