Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
In the native hip, the hip capsular ligaments tighten at the limits of range of hip motion and may provide a passive stabilizing force to protect the hip against edge loading. In this study we quantified the stabilizing force vectors generated by capsular ligaments at extreme range of motion (ROM), and examined their ability to prevent edge loading. Torque-rotation curves were obtained from nine cadaveric hips to define the rotational restraint contributions of the capsular ligaments in 36 positions. A ligament model was developed to determine the line-of-action and effective moment arms of the medial/lateral iliofemoral, ischiofemoral, and pubofemoral ligaments in all positions. The functioning ligament forces and stiffness were determined at 5 Nm rotational restraint. In each position, the contribution of engaged capsular ligaments to the joint reaction force was used to evaluate the net force vector generated by the capsule.Aims
Methods
Elevated levels of circulating cobalt ions have been linked with a wide range of systemic complications including neurological, endocrine, and cardiovascular symptoms. Case reports of patients with elevated blood cobalt ions have described significant cardiovascular complications including cardiomyopathy. However, correlation between the actual level of circulating cobalt and extent of cardiovascular injury has not previously been performed. This review examines evidence from the literature for a link between elevated blood cobalt levels secondary to metal-on-metal (MoM) hip arthroplasties and cardiomyopathy. Correlation between low, moderate, and high blood cobalt with cardiovascular complications has been considered. Elevated blood cobalt at levels over 250 µg/l have been shown to be a risk factor for developing systemic complications and published case reports document cardiomyopathy, cardiac transplantation, and death in patients with severely elevated blood cobalt ions. However, it is not clear that there is a hard cut-off value and cardiac dysfunction may occur at lower levels. Clinical and laboratory research has found conflicting evidence of cobalt-induced cardiomyopathy in patients with MoM hips. Further work needs to be done to clarify the link between severely elevated blood cobalt ions and cardiomyopathy. Cite this article:
Metal allergy in knee arthroplasty patients is a controversial topic. We aimed to conduct a scoping review to clarify the management of metal allergy in primary and revision total knee arthroplasty (TKA). Studies were identified by searching electronic databases: Cochrane Central Register of Controlled Trials, Ovid MEDLINE, and Embase, from their inception to November 2020, for studies evaluating TKA patients with metal hypersensitivity/allergy. All studies reporting on diagnosing or managing metal hypersensitivity in TKA were included. Data were extracted and summarized based on study design, study population, interventions and outcomes. A practical guide is then formulated based on the available evidence.Aims
Methods
This study assessed the impact of COVID-19 on hip and distal femur fracture patient outcomes across three successive UK lockdown periods over one year. A single-centre retrospective cohort study was performed at an acute NHS Trust. Hip and distal femur fracture patients admitted within the first month from each of the three starting dates of each national lockdown were included and compared to a control group in March 2019. Data were collected as per the best practice tariff outcomes including additional outcomes as required. Data collection included COVID-19 status, time to theatre, 30-day mortality, presence of acute kidney injury (AKI) and pneumonia, and do not attempt cardiopulmonary resuscitation (DNACPR) status. Data were analyzed using an independent-samples Aims
Methods
Aims. The aim of this study was to examine the efficacy and safety
of multiple boluses of intravenous (IV) tranexamic acid (TXA) on
the hidden blood loss (HBL) and inflammatory response following
primary total hip arthroplasty (THA). Patients and Methods. A total of 150 patients were allocated randomly to receive a
single bolus of 20 mg/kg IV TXA before the incision (group A), a
single bolus followed by a second bolus of 1 g IV-TXA three hours
later (group B) or a single bolus followed by two boluses of 1 g
IV-TXA three and six hours later (group C). All patients were treated
using a standard peri-operative enhanced recovery protocol. Primary
outcomes were HBL and the level of haemoglobin (Hb) as well as the
levels of C-reactive protein (CRP) and interleukin-6 (IL-6) as markers
of inflammation. Secondary outcomes included the length of stay
in hospital and the incidence of venous thromboembolism (VTE). Results. The mean HBL was significantly lower in group C (402.13 ml standard
deviation (. sd). 225.97) than group A (679.28 ml. sd. 277.16,
p <
0.001) or B (560.62 ml . sd. 295.22, p = 0.010). The
decrease in the level of Hb between the pre-operative baseline and
the level on the third post-operative day was 30.82 g/L (. sd. 6.31
g/L) in group A, 27.16 g/L (. sd. 6.83) in group B and 21.98
g/L (. sd. 3.72) in group C. This decrease differed significantly
among the three groups (p <
0.01). The mean level of CRP was
significantly lower in group C than in the other two groups on the
second (p ≤ 0.034) and third post-operative days (p ≤ 0.014). The
levels of IL-6 were significantly lower in group C than group A
on the first three post-operative days (p = 0.023). The mean length
of stay was significantly lower in group C than group A (p = 0.023).
No VTE or other adverse events occurred. Conclusion. Multiple boluses of IV-TXA can effectively reduce HBL following
primary
The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach. A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.Aims
Methods
Current guidelines consider analyses of joint aspirates, including leucocyte cell count (LC) and polymorphonuclear percentage (PMN%) as a diagnostic mainstay of periprosthetic joint infection (PJI). It is unclear if these parameters are subject to a certain degree of variability over time. Therefore, the aim of this study was to evaluate the variation of LC and PMN% in patients with aseptic revision total knee arthroplasty (TKA). We conducted a prospective, double-centre study of 40 patients with 40 knee joints. Patients underwent joint aspiration at two different time points with a maximum period of 120 days in between these interventions and without any events such as other joint aspirations or surgeries. The main indications for TKA revision surgery were aseptic implant loosening (n = 24) and joint instability (n = 11).Aims
Methods
Aims. We report the incidence of radiolucent lines (RLLs) using two
flanged acetabular components at total hip arthroplasty (THA) and
the effect of the Rim Cutter. Patients and Methods. We performed a retrospective review of 300 hips in 292 patients
who underwent primary cemented
The aim of the study was to compare two methods of calculating pelvic incidence (PI) and pelvic tilt (PT), either by using the femoral heads or acetabular domes to determine the bicoxofemoral axis, in patients with unilateral or bilateral primary hip osteoarthritis (OA). PI and PT were measured on standing lateral radiographs of the spine in two groups: 50 patients with unilateral (Group I) and 50 patients with bilateral hip OA (Group II), using the femoral heads or acetabular domes to define the bicoxofemoral axis. Agreement between the methods was determined by intraclass correlation coefficient (ICC) and the standard error of measurement (SEm). The intraobserver reproducibility and interobserver reliability of the two methods were analyzed on 31 radiographs in both groups to calculate ICC and SEm.Aims
Methods
Aims. To demonstrate, with concrete examples, the value of in-depth
exploration and comparison of data published in National Joint Arthroplasty
registry reports. Patients and Methods. The author reviewed published current reports of National Joint
Arthroplasty registries for findings of current significance to
current orthopaedic practice. Results. A total of six observations that demonstrate actionable or unexpected
findings from joint registries are described. These include: one
third to one half of all arthroplasty failures in the first decade
occur in the first one to two years; infection rates after arthroplasty
have not declined in the last three decades; infection after TKA
is more common in men than women; outcomes of TKA are more variable
in young compared with older patients; new technologies (uncemented
implants and crosslinked polyethylene) have improved results of
THA and a real-time shift in use of ceramic femoral heads is occurring
in
Aims. There is no consensus about the best method of achieving equal
leg lengths at total hip arthroplasty (THA) in patients with Crowe
type-IV developmental dysplasia of the hip (DDH). We reviewed our
experience of a consecutive series of patients who underwent THA
for this indication. Patients and Methods. We retrospectively reviewed 78 patients (86 THAs) with Crowe
type-IV DDH, including 64 women and 14 men, with a minimum follow-up
of two years. The mean age at the time of surgery was 52.2 years
(34 to 82). We subdivided Crowe type-IV DDH into two major types
according to the number of dislocated hips, and further categorised
them into three groups according to the occurrence of pelvic obliquity
or spinal curvature. Leg length discrepancy (LLD) and functional
scores were analysed. Results. Type-I included 53 patients with unilateral dislocation, in which
25 (category A) had no pelvic obliquity or spinal deformity, 19
(category B) had pelvic obliquity with a compensated spinal curvature
and nine (category C) had pelvic obliquity and decompensated spinal
degenerative changes. Type-II included 25 patients with one dislocated and one dysplastic
hip, in which there were eight of category A, 15 of category B and
two of category C. Pre-operatively, there were significant differences between the
anatomical and functional LLD in type-IB (p = 0.005) and -IC (p
<
0.001), but not in type-IA, -IIA or -IIB. Post-operatively, bony LLD increased significantly in types-IB,
-IC and -IIB, whereas functional LLD decreased significantly in
each type except for IIA. The mean functional LLD decreased from
30.7 mm (standard deviation (. sd. ) 18.5) pre-operatively
to 6.2 mm (. sd. 4.4) post-operatively and the mean anatomical
LLD improved from 35.8 mm (. sd. 19.7) pre-operatively to
12.4 mm (. sd. 8.3) post-operatively. Conclusion. Pelvic and spinal changes are common in patients with Crowe type-IV
DDH and need to be taken into consideration when planning
The aims of this systematic review were to assess the learning curve of semi-active robotic arm-assisted total hip arthroplasty (rTHA), and to compare the accuracy, patient-reported functional outcomes, complications, and survivorship between rTHA and manual total hip arthroplasty (mTHA). Searches of PubMed, Medline, and Google Scholar were performed in April 2020 in line with the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “hip”, and “arthroplasty”. The criteria for inclusion were published clinical research articles reporting the learning curve for rTHA (robotic arm-assisted only) and those comparing the implantation accuracy, functional outcomes, survivorship, or complications with mTHA.Aims
Methods
Aims. We conducted a prospective study of a delta ceramic total hip
arthroplasty (THA) to determine the rate of ceramic fracture, to
characterise post-operative noise, and to evaluate the mid-term
results and survivorship. Patients and Methods. Between March 2009 and March 2011, 274 patients (310 hips) underwent
cementless
Aims. We evaluated the short-term functional outcome and prevalence
of bearing-specific generation of audible noise in 301 patients
(336 hips) operated on with fourth generation (Delta) medium diameter
head, ceramic-on-ceramic (CoC) total hip arthroplasties (THAs). Patients and Methods. There were 191 female (63%) and 110 male patients (37%) with
a mean age of 61 years (29 to 78) and mean follow-up of 2.1 years
(1.3 to 3.4). Patients completed three questionnaires: Oxford Hip
Score (OHS), Research and Development 36-item health survey (RAND-36)
and a noise-specific symptom questionnaire. Plain radiographs were also
analysed. A total of three hips (0.9%) were revised. . Results. There were 52 patients (54 hips, 17%) who reported noise, and
in 25 (48%) of them the noise was frequently heard. In the multiple
regression analysis, the only independent risk factor for noise
was a specific
This study evaluates risk factors influencing fracture characteristics for postoperative periprosthetic femoral fractures (PFFs) around cemented stems in total hip arthroplasty. Data were collected for PFF patients admitted to eight UK centres between 25 May 2006 and 1 March 2020. Radiographs were assessed for Unified Classification System (UCS) grade and AO/OTA type. Statistical comparisons investigated relationships by age, gender, and stem fixation philosophy (polished taper-slip (PTS) vs composite beam (CB)). The effect of multiple variables was estimated using multinomial logistic regression to estimate odds ratios (ORs) with 95% confidence intervals (CIs). Surgical treatment (revision vs fixation) was compared by UCS grade and AO/OTA type.Aims
Methods
The aim of our study was to investigate the effect of asymmetric crosslinked polyethylene liner use on the risk of revision of cementless and hybrid total hip arthroplasties (THAs). We undertook a registry study combining the National Joint Registry dataset with polyethylene manufacturing characteristics as supplied by the manufacturers. The primary endpoint was revision for any reason. We performed further analyses on other reasons including instability, aseptic loosening, wear, and liner dissociation. The primary analytic approach was Cox proportional hazard regression.Aims
Methods
Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.Aims
Methods
Aims. The aims of the study were to review and analyse the reported
series of debridement, antibiotics and implant retention (DAIR)
in the management of infected total hip arthroplasties (THAs) to
establish the overall success and the influencing factors. Patients and methods. Using a standardised recognised study protocol, meta-analysis
of observational studies in epidemiology guidelines, a comprehensive
review and analysis of the literature was performed. The primary
outcome measure was the success of treatment. The search strategy
and inclusion criteria which involved an assessment of quality yielded
39 articles for analysis, which included 1296 patients. Results. The proportion of success following DAIR in the management of
an infected