Aims . The primary aim of this independent prospective randomised trial
was to compare serum metal ion levels for ceramic-on-metal (CoM)
and metal-on-metal (MoM) bearing surfaces in total hip arthroplasty
(THA). Our one-year results demonstrated elevation in metal ion
levels above baseline with no significant difference between the
CoM and MoM groups. This paper reviews the five-year data. Patients and Methods. The implants used in each patient differed only in respect to
the type of femoral head (ceramic or metal). At five-year follow-up
of the 83 enrolled patients, data from 67 (36 CoM, 31 MoM) was available
for comparison. Results. The mean serum cobalt (Co) and chromium (Cr) ion levels remained
above baseline in both groups (CoM: Co 1.16 μg/l (0.41 to 14.67),
Cr 1.05 μg/l (0.16 to 12.58); MoM: Co 2.93 μg/l (0.35 to 30.29),
Cr 1.85 μg/l (0.36 to 17.00)) but the increase was significantly
less in the CoM cohort (Co difference p = 0.001, Cr difference p
= 0.002). These medium-term results, coupled with lower revision
rates from national joint registries, suggest that the performance
of CoM
In 2020, the COVID-19 pandemic meant that proceeding with elective surgery was restricted to minimize exposure on wards. In order to maintain throughput of elective cases, our hospital (St Michaels Hospital, Toronto, Canada) was forced to convert as many cases as possible to same-day procedures rather than overnight admission. In this retrospective analysis, we review the cases performed as same-day arthroplasty surgeries compared to the same period in the previous 12 months. We conducted a retrospective analysis of patients undergoing total hip and knee arthroplasties over a three-month period between October and December in 2019, and again in 2020, in the middle of the COVID-19 pandemic. Patient demographics, number of outpatient primary arthroplasty cases, length of stay for admissions, 30-day readmission, and complications were collated.Aims
Methods
We studied the outcomes of hip and knee arthroplasties in a high-volume arthroplasty centre to determine if patients with morbid obesity (BMI ≥ 40 kg/m2) had unacceptably worse outcomes as compared to those with BMI < 40 kg/m2. In a two-year period, 4,711 patients had either total hip arthroplasty (THA; n = 2,370), total knee arthroplasty (TKA; n = 2,109), or unicompartmental knee arthroplasty (UKA; n = 232). Of these patients, 392 (8.3%) had morbid obesity. We compared duration of operation, anaesthetic time, length of stay (LOS), LOS > three days, out of hours attendance, emergency department attendance, readmission to hospital, return to theatre, and venous thromboembolism up to 90 days. Readmission for wound infection was recorded to one year. Oxford scores were recorded preoperatively and at one year postoperatively.Aims
Methods
To investigate whether idiopathic osteonecrosis of the femoral head (ONFH) is related to impaired osteoblast activities. We cultured osteoblasts isolated from trabecular bone explants taken from the femoral head and the intertrochanteric region of patients with idiopathic ONFH, or from the intertrochanteric region of patients with osteoarthritis (OA), and compared their viability, mineralization capacity, and secretion of paracrine factors.Aims
Methods
Aims. This paper describes the methodology, validation and reliability
of a new computer-assisted method which uses models of the patient’s
bones and the components to measure their migration and polyethylene
wear from radiographs after total hip arthroplasty (THA). Materials and Methods. Models of the patient’s acetabular and femoral component obtained
from the manufacturer and models of the patient’s pelvis and femur
built from a single computed tomography (CT) scan, are used by a
computer program to measure the migration of the components and
the penetration of the femoral head from anteroposterior and lateral radiographs
taken at follow-up visits. The program simulates the radiographic
setup and matches the position and orientation of the models to
outlines of the pelvis, the acetabular and femoral component, and
femur on radiographs. Changes in position and orientation reflect
the migration of the components and the penetration of the femoral
head. Validation was performed using radiographs of phantoms simulating
known migration and penetration, and the clinical feasibility of
measuring migration was assessed in two patients. Results. Migration of the acetabular and femoral components can be measured
with limits of agreement (LOA) of 0.37 mm and 0.33 mm, respectively.
Penetration of the femoral head can be measured with LOA of 0.161
mm. Conclusion. The migration of components and polyethylene wear can be measured
without needing specialised radiographs. Accurate measurement may
allow earlier prediction of failure after
The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes. A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).Aims
Methods
Due to the opioid epidemic in the USA, our service progressively decreased the number of opioid tablets prescribed at discharge after primary hip (THA) and knee (TKA) arthroplasty. The goal of this study was to analyze the effect on total morphine milligram equivalents (MMEs) prescribed and post-discharge opioid repeat prescriptions. We retrospectively reviewed 19,428 patients undergoing a primary THA or TKA between 1 February 2016 and 31 December 2019. Two reductions in the number of opioid tablets prescribed at discharge were implemented over this time; as such, we analyzed three periods (P1, P2, and P3) with different routine discharge MME (750, 520, and 320 MMEs, respectively). We investigated 90-day refill rates, refill MMEs, and whether discharge MMEs were associated with represcribing in a multivariate model.Aims
Methods
Aims. Increasing demand for total hip and knee arthroplasty (THA/TKA)
and associated follow-up has placed huge demands on orthopaedic
services. Feasible follow-up mechanisms are therefore essential. . Methods. We conducted an audit of clinical follow-up decision-making for
THA/TKA based on questionnaire/radiograph review compared with local
practice of Arthroplasty Care Practitioner (ACP)-led outpatient
follow-up. In all 599 patients attending an ACP-led
This study investigates head-neck taper corrosion with varying head size in a novel hip simulator instrumented to measure corrosion related electrical activity under torsional loads. In all, six 28 mm and six 36 mm titanium stem-cobalt chrome head pairs with polyethylene sockets were tested in a novel instrumented hip simulator. Samples were tested using simulated gait data with incremental increasing loads to determine corrosion onset load and electrochemical activity. Half of each head size group were then cycled with simulated gait and the other half with gait compression only. Damage was measured by area and maximum linear wear depth.Aims
Methods
The proportion of arthroplasties performed in the ambulatory setting has increased considerably. However, there are concerns whether same-day discharge may increase the risk of complications. The aim of this study was to compare 90-day outcomes between inpatient arthroplasties and outpatient arthroplasties performed at an ambulatory surgery centre (ASC), and determine whether there is a learning curve associated with performing athroplasties in an ASC. Among a single-surgeon cohort of 970 patients who underwent arthroplasty at an ASC, 854 (88.0%) were matched one-to-one with inpatients based on age, sex, American Society of Anesthesiologists (ASA) grade, BMI, and procedure (105 could not be adequately matched and 11 lacked 90-day follow-up). The cohort included 281 total hip arthroplasties (THAs) (32.9%), 267 unicompartmental knee arthroplasties (31.3%), 242 primary total knee arthroplasties (TKAs) (28.3%), 60 hip resurfacings (7.0%), two revision THAs (0.3%), and two revision TKAs (0.3%). Outcomes included readmissions, reoperations, visits to the emergency department, unplanned clinic visits, and complications.Aims
Methods
Accumulated evidence indicates that local cell origins may ingrain differences in the phenotypic activity of human osteoblasts. We hypothesized that these differences may also exist in osteoblasts harvested from the same bone type at periarticular sites, including those adjacent to the fixation sites for total joint implant components. Human osteoblasts were obtained from the acetabulum and femoral neck of seven patients undergoing total hip arthroplasty (THA) and from the femoral and tibial cuts of six patients undergoing total knee arthroplasty (TKA). Osteoblasts were extracted from the usually discarded bone via enzyme digestion, characterized by flow cytometry, and cultured to passage three before measurement of metabolic activity, collagen production, alkaline phosphatase (ALP) expression, and mineralization.Aims
Methods
The purpose of this study is to evaluate early outcomes with the use of a smartphone-based exercise and educational care management system after total hip arthroplasty (THA) and demonstrate decreased use of in-person physiotherapy (PT). A multicentre, prospective randomized controlled trial was conducted to evaluate a smartphone-based care platform for primary THA. Patients randomized to the control group (198) received the institution’s standard of care. Those randomized to the treatment group (167) were provided with a smartwatch and smartphone application. PT use, THA complications, readmissions, emergency department/urgent care visits, and physician office visits were evaluated. Outcome scores include the Hip disability and Osteoarthritis Outcome Score (HOOS, JR), health-related quality-of-life EuroQol five-dimension five-level score (EQ-5D-5L), single leg stance (SLS) test, and the Timed Up and Go (TUG) test.Aims
Methods
To establish the survivorship, function, and metal ion levels in an unselected series of metal-on-metal hip resurfacing arthroplasties (HRAs) performed by a non-designer surgeon. We reviewed 105 consecutive HRAs in 83 patients, performed by a single surgeon, at a mean follow-up of 14.9 years (9.3 to 19.1). The cohort included 45 male and 38 female patients, with a mean age of 49.5 years (SD 12.5)Aims
Methods
We aimed to investigate if the use of the largest possible cobalt-chromium head articulating with polyethylene acetabular inserts would increase the in vivo wear rate in total hip arthroplasty. In a single-blinded randomized controlled trial, 96 patients (43 females), at a median age of 63 years (interquartile range (IQR) 57 to 69), were allocated to receive either the largest possible modular femoral head (36 mm to 44 mm) in the thinnest possible insert or a standard 32 mm head. All patients received a vitamin E-doped cross-linked polyethylene insert and a cobalt-chromium head. The primary outcome was proximal head penetration measured with radiostereometric analysis (RSA) at two years. Secondary outcomes were volumetric wear, periacetabular radiolucencies, and patient-reported outcomes.Aims
Methods
Adverse local tissue reactions associated with abnormal wear considerably slowed down the general use of metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), now limited to a few specialized centres. In this study, we provide the clinical results of 400 consecutive MoM HRAs implanted more than 20 years ago in one such centre. A total of 355 patients (400 hips) were treated with Conserve Plus HRA between November 1996 and November 2000. There were 96 female (27%) and 259 male patients (73%). Their mean age was 48.2 years (SD 10.9). The University of California, Los Angeles (UCLA) hip scores and 12-item Short Form Survey (SF-12) quality of life scores were reported. Survivorship was assessed using Kaplan-Meier analyses.Aims
Methods
Aims. The aim of this study was to assess the current available evidence
about when patients might resume driving after elective, primary
total hip (THA) or total knee arthroplasty (TKA) undertaken for
osteoarthritis (OA). Materials and Methods. In February 2016, EMBASE, MEDLINE, Web of Science, Scopus, Cochrane,
PubMed Publisher, CINAHL, EBSCO and Google Scholar were searched
for clinical studies reporting on ‘THA’, ‘TKA’, ‘car driving’, ‘reaction
time’ and ‘brake response time’. Two researchers (CAV and JJT) independently
screened the titles and abstracts for eligibility and assessed the
risk of bias. Both fixed and random effects were used to pool data
and calculate mean differences (MD) and 95% confidence intervals
(CI) between pre- and post-operative total brake response time (TBRT). Results. A total of 19 studies were included. The assessment of the risk
of bias showed that one study was at high risk, six studies at moderate
risk and 12 studies at low risk. Meta-analysis of TBRT showed a
MD decrease of 25.54 ms (95% CI -32.02 to 83.09) two weeks after
right-sided
To find out if there is an inverse association between estimated glomerular filtration rate (eGFR) and whole blood cobalt (Co) and chromium (Cr) levels in patients with metal-on-metal (MoM) hip arthroplasties and renal insufficiency, suggesting that renal insufficiency could cause accumulation of Co and Cr in blood. Out of 2,520 patients with 3,013 MoM hip arthroplasties, we identified 1,244 patients with whole blood Co, Cr, and creatinine measured within no more than a one-year interval. We analyzed the correlation of blood metal ion levels and eGFR to identify a potential trend of accumulating Co or Cr with decreasing eGFR.Aims
Methods
The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems. A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.Aims
Methods
Hip resurfacing is being performed more frequently in the United Kingdom. The possible benefits include more accurate restoration of leg length, femoral offset and femoral anteversion than occurs after total hip arthroplasty (THA). We compared anteroposterior radiographs from 26 patients who had undergone hybrid