Advances in polyethylene (PE) in total hip arthroplasty
have led to interest and increased use of highly crosslinked PE
(HXLPE) in total knee arthroplasty (TKA). Biomechanical data suggest
improved wear characteristics for HXLPE inserts over conventional
PE in TKA. Short-term results from registry data and few clinical
trials are promising. Our aim is to present a review of the history
of HXLPEs, the use of HXLPE inserts in TKA, concerns regarding potential mechanical
complications, and a thorough review of the available biomechanical
and clinical data. Cite this article:
Patellofemoral complications are common after
total knee replacement (TKR). Leaving the patellar unsurfaced after
TKR may lead to complications such as anterior knee pain, and re-operation
to surface it. Complications after patellar resurfacing include
patellar fracture, aseptic loosening, patellar instability, polyethylene
wear, patellar clunk and osteonecrosis. Historically, patellar complications
account for one of the larger proportions of causes of failure in
TKR, however, with contemporary implant designs, complication rates
have decreased. Most remaining failures relate to patellofemoral
tracking. Understanding the causes of patellofemoral maltracking
is essential to prevent these complications as well as manage them
when they occur. Cite this article:
Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments. Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.Objectives
Materials and Methods
Fractures around total knee arthroplasties pose
a significant surgical challenge. Most can be managed with osteosynthesis
and salvage of the replacement. The techniques of fixation of these
fractures and revision surgery have evolved and so has the assessment
of outcome. This specialty update summarises the current evidence
for the classification, methods of fixation, revision surgery and
outcomes of the management of periprosthetic fractures associated
with total knee arthroplasty. Cite this article:
Isolated patellofemoral arthritis is a common
condition and there are varying opinions on the most effective treatments.
Non-operative and operative treatments have failed to demonstrate
effective long-term treatment for those in an advanced stage of
the condition. Newer designs and increased technology in patellofemoral replacement
(PFR) have produced more consistent outcomes. This has led to a
renewed enthusiasm for this procedure. Newer PFR prostheses have
addressed the patellar maltracking issues plaguing some of the older designs.
Short-term results with contemporary prostheses and new technology
are described here. Cite this article:
A key to the success of revision total knee arthroplasty
(TKA) is a safe surgical approach using an exposure that minimises
complications. In most patients, a medial parapatellar arthrotomy
with complete synovectomy is sufficient. If additional exposure
is needed, a quadriceps snip performed through the quadriceps tendon
often provides the additional exposure required. It is simple to
perform and does not alter the post-operative rehabilitative protocol.
In rare cases, in which additional exposure is needed, or when removal
of a cemented long-stemmed tibial component is required, a tibial
tubercle osteotomy (TTO) may be used. Given the risk of post-operative
extensor lag, a V-Y quadricepsplasty is rarely indicated and usually
considered only if TTO is not possible. Cite this article:
Disruption of the extensor mechanism in total
knee arthroplasty may occur by tubercle avulsion, patellar or quadriceps
tendon rupture, or patella fracture, and whether occurring intra-operatively
or post-operatively can be difficult to manage and is associated
with a significant rate of failure and associated complications.
This surgery is frequently performed in compromised tissues, and
repairs must frequently be protected with cerclage wiring and/or
augmentation with local tendon (semi-tendinosis, gracilis) which
may also be used to treat soft-tissue loss in the face of chronic
disruption. Quadriceps rupture may be treated with conservative
therapy if the patient retains active extension. Component loosening
or loss of active extension of 20° or greater are clear indications
for surgical treatment of patellar fracture. Acute patellar tendon
disruption may be treated by primary repair. Chronic extensor failure
is often complicated by tissue loss and retraction can be treated
with medial gastrocnemius flaps, achilles tendon allografts, and
complete extensor mechanism allografts. Attention to fixing the
graft in full extension is mandatory to prevent severe extensor
lag as the graft stretches out over time.
Total knee arthroplasty (TKA) is a cost effective
and extremely successful operation. As longevity increases, the demand
for primary TKA will continue to rise. The success and survivorship
of TKAs are dependent on the demographics of the patient, surgical
technique and implant-related factors. Currently the risk of failure of a TKA requiring revision surgery
ten years post-operatively is 5%. The most common indications for revision include aseptic loosening
(29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses
considerable clinical burdens on patients and financial burdens
on healthcare systems. We present a current concepts review on the epidemiology of failed
TKAs using data from worldwide National Joint Registries. Cite this article:
There are many reasons why a total knee replacement
(TKR) may fail and qualify for revision. Successful revision surgery
depends as much on accurate assessment of the problem TKR as it
does on revision implant design and surgical technique. Specific
modes of failure require specific surgical solutions. Causes of
failure are often presented as a list or catalogue, without a system
or process for making a decision. In addition, strict definitions
and consensus on modes of failure are lacking in published series
and registry data. How we approach the problem TKR is an essential
but neglected aspect of understanding knee replacement surgery.
It must be carried out systematically, comprehensively and efficiently.
Eight modes of failure are described: 1) sepsis; 2) extensor discontinuity;
3) stiffness; 4) tibial- femoral instability; 5) patellar tracking;
6) aseptic loosening and osteolysis; 7) periprosthetic fracture
and 8) component breakage. A ninth ‘category’, unexplained pain
is an indication for further investigation but not surgery. Cite this article:
The long-term success of total knee replacement is multifactorial, including factors relating to the patient, the operation and the implant. The purpose of this study was to examine the 20-year survival of the cemented Anatomical Graduated Component (AGC) total knee replacement. Between 1983 and 2004, 7760 of these were carried out at our institution. Of these, 6726 knees which received the non-modular metal-backed tibial component with compression-moulded polyethylene and had a minimum two-year follow-up were available for study. In all, 36 knees were followed over 20 years with a survival of the tibial and femoral components together of 97.8% (95% confidence interval (CI) 0.9851 to 0.9677), with no implants being revised for polyethylene wear or osteolysis. Age >
70 was associated with increased survival (99.6%, 95% CI 99.0 to 99.8) (p <
0.0001) but pre-operative valgus alignment reduced survival (95.1%, 95% CI 90.0 to 97.6) (p = 0.0056). Age <
55 (p = 0.129), pre-operative varus alignment (p = 0.707), osteonecrosis (p = 0.06), rheumatoid arthritis (p = 0.247), and gender (p = 0.666) were not statistically associated with failure. We attribute the success of the AGC implant to its relatively unconstrained articular geometry and the durability of a non-modular metal-backed tibial component with compression moulded polyethylene.
We have examined the differences in clinical outcome of total knee replacement (TKR) with and without patellar resurfacing in a prospective, randomised study of 181 osteoarthritic knees in 142 patients using the Profix total knee system which has a femoral component with features considered to be anatomical and a domed patellar implant. The procedures were carried out between February 1998 and November 2002. A total of 159 TKRs in 142 patients were available for review at a mean of four years (3 to 7). The patients and the clinical evaluator were blinded in this prospective study. Evaluation was undertaken annually by an independent observer using the knee pain scale and the Knee Society clinical rating system. Specific evaluation of anterior knee pain, stair-climbing and rising from a seated to a standing position was also undertaken. No benefit was shown of TKR with patellar resurfacing over that without resurfacing with respect to any of the measured outcomes. In 22 of 73 knees (30.1%) with and 18 of 86 knees (20.9%) without patellar resurfacing there was some degree of anterior knee pain (p = 0.183). No revisions related to the patellofemoral joint were performed in either group. Only one TKR in each group underwent a re-operation related to the patellofemoral joint. A significant association between knee flexion contracture and anterior knee pain was observed in those knees with patellar resurfacing (p = 0.006).
We report on the long-term results of 163 bicruciate-retaining
Hermes 2C total knee replacements in 130 patients at a mean follow-up
of 22.4 years (20.3 to 23.5). Even when the anterior cruciate ligament
had a partially degenerative appearance it was preserved as long
as the knee had a normal anterior drawer and Lachman’s test pre-operatively. The
description and surgical technique of this minimally constrained
prosthesis were published in 1983 and the ten-year clinical results
in 1999. A total of 12% of the knees (20 of 163) in this study were revised
because of wear of the polyethylene tibial insert. Excellent stability
was achieved and the incidence of aseptic component loosening was
4.3% (seven of 163). The survival rate using revision for any reason
as the endpoint was 82% (95% confidence interval 76.2 to 88.0). Although this series included a relatively small number of replacements,
it demonstrated that the anterior cruciate ligament, even when partially
degenerated at the time of TKR, remained functional and provided
adequate stability at a long-term follow-up. Cite this article:
The aim of our study was to describe the characteristics,
treatment, and outcomes of patients with periprosthetic joint infection
(PJI) and normal inflammatory markers after total knee arthroplasty
(TKA) and total hip arthroplasty (THA). In total 538 TKAs and 414 THAs underwent surgical treatment for
PJI and met the inclusion criteria. Pre-operative erythrocyte sedimentation
rate (ESR) and C-reactive protein level (CRP) were reviewed to identify
the seronegative cohort. An age- and gender-matched cohort was identified
from the remaining patients for comparison. Overall, 4% of confirmed
infections were seronegative (21 TKA and 17 THA). Of those who underwent
pre-operative aspiration, cultures were positive in 76% of TKAs
(n = 13) and 64% of THAs (n = 7). Cell count and differential were
suggestive of infection in 85% of TKA (n = 11) and all THA aspirates
(n = 5). The most common organism was coagulase-negative Cite this article:
PCL retaining fixed-bearing TKA is a highly successful
operation with the need for more surgery occurring at the rate of
approximately 0.4% per year over the first 27 years. The most common
cause for revision surgery is related to polyethylene insert failure
and accounts for approximately 50% of re-operations. Late metastatic
infection is the next most frequent cause followed by patellar problems,
late instability and component loosening in decreasing frequency.
A myriad of rare miscellaneous problems can also occur.
We describe the survivorship of the Medial Rotation total knee replacement (TKR) at ten years in 228 cemented primary replacements implanted between October 1994 and October 2006, with their clinical and radiological outcome. This implant has a highly congruent medial compartment, with the femoral component represented by a portion of a sphere which articulates with a matched concave surface on the medial side of the tibial insert. There were 78 men (17 bilateral TKRs) and 111 women (22 bilateral TKRs) with a mean age of 67.9 years (28 to 90). All the patients were assessed clinically and radiologically using the American Knee Society scoring systems. The mean follow-up was for six years (1 to 13) with only two patients lost to follow-up and 34 dying during the period of study, one of whom had required revision for infection. There were 11 revisions performed in total, three for aseptic loosening, six for infection, one for a periprosthetic fracture and one for a painful but well-fixed replacement performed at another centre. With revision for any cause as the endpoint, the survival at ten years was 94.5% (95% CI 85.1 to 100), and with aseptic loosening as the endpoint 98.4% (95% CI 93 to 100). The mean American Knee Society score improved from 47.6 (0 to 88) to 72.2 (26 to 100) and for function from 45.1 (0 to 100) to 93.1 (45 to 100). Radiological review failed to detect migration in any of the surviving knees. The clinical and radiological results of the Medial Rotation TKR are satisfactory at ten years. The increased congruence of the medial compartment has not led to an increased rate of loosening and continued use can be supported.
We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article:
We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up.
Malrotation of the femoral component is a cause of patellofemoral maltracking after total knee arthroplasty. Its precise effect on the patellofemoral mechanics has not been well quantified. We have developed an in vitro method to measure the influence of patellar maltracking on contact. Maltracking was induced by progressively rotating the femoral component either internally or externally. The contact mechanics were analysed using Tekscan. The results showed that excessive malrotation of the femoral component, both internally and externally, had a significant influence on the mechanics of contact. The contact area decreased with progressive maltracking, with a concomitant increase in contact pressure. The amount of contact area that carries more than the yield stress of ultra-high molecular weight polyethylene significantly increases with progressive maltracking. It is likely that the elevated pressures noted in malrotation could cause accelerated and excessive wear of the patellar button.
Peri-prosthetic patellar fracture following resurfacing
as part of total knee replacement (TKR) is an infrequent yet challenging
complication. This case-control study was performed to identify
clinical, radiological and surgical factors that increase the risk
of developing a spontaneous patellar fracture after TKR. Patellar
fractures were identified in 74 patients (88 knees) from a series
of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding
those with a previous history of extensor mechanism realignment
or a clear traumatic event, a metal-backed patella, any uncemented
component or subsequent infection, the remaining 64 fractures were
compared with a matched group of TKRs with an excellent outcome
defined by the Knee Society score. The mean age of patients with
a fracture was 70 years (51 to 81) at the time of TKR. Patellar
fractures were detected at a mean of 13.4 months (2 to 84) after
surgery. The incidence of patellar fracture was found to be strongly
associated with the number of previous knee operations, greater
pre-operative mechanical malalignment, smaller post-operative patellar
tendon length, thinner post-resection patellar thickness, and a
lower post-operative Insall-Salvati ratio. An understanding of the risk factors associated with spontaneous
patellar fracture following TKR provides a valuable insight into
prevention of this challenging complication.
In posterior stabilised total knee replacement
(TKR) a larger femoral component is sometimes selected to manage the
increased flexion gap caused by resection of the posterior cruciate
ligament. However, concerns remain regarding the adverse effect
of the increased anteroposterior dimensions of the femoral component
on the patellofemoral (PF) joint. Meanwhile, the gender-specific
femoral component has a narrower and thinner anterior flange and
is expected to reduce the PF contact force. PF contact forces were
measured at 90°, 120°, 130° and 140° of flexion using the NexGen
Legacy Posterior Stabilized (LPS)-Flex Fixed Bearing Knee system
using Standard, Upsized and Gender femoral components during TKR.
Increasing the size of the femoral component significantly increased
mean PF forces at 120°, 130° and 140° of flexion (p = 0.005, p <
0.001 and p <
0.001, respectively). No difference was found in
contact force between the Gender and the Standard components. Among
the patients who had overhang of the Standard component, mean contact
forces with the Gender component were slightly lower than those
of the Standard component, but no statistical difference was found
at 90°, 120°, 130° or 140° of flexion (p = 0.689, 0.615, 0.253 and
0.248, respectively). Upsized femoral components would increase PF forces in deep knee
flexion. Gender-specific implants would not reduce PF forces.