Advertisement for orthosearch.org.uk
Results 41 - 60 of 82
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1011 - 1021
1 Aug 2013
Krishnan H Krishnan SP Blunn G Skinner JA Hart AJ

Following the recall of modular neck hip stems in July 2012, research into femoral modularity will intensify over the next few years. This review aims to provide surgeons with an up-to-date summary of the clinically relevant evidence. The development of femoral modularity, and a classification system, is described. The theoretical rationale for modularity is summarised and the clinical outcomes are explored. The review also examines the clinically relevant problems reported following the use of femoral stems with a modular neck.

Joint replacement registries in the United Kingdom and Australia have provided data on the failure rates of modular devices but cannot identify the mechanism of failure. This information is needed to determine whether modular neck femoral stems will be used in the future, and how we should monitor patients who already have them implanted.

Cite this article: Bone Joint J 2013;95-B:1011–21.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1192 - 1196
1 Sep 2015
Amanatullah DF Siman H Pallante GD Haber DB Sierra RJ Trousdale RT

When fracture of an extensively porous-coated femoral component occurs, its removal at revision total hip arthroplasty (THA) may require a femoral osteotomy and the use of a trephine. The remaining cortical bone after using the trephine may develop thermally induced necrosis. A retrospective review identified 11 fractured, well-fixed, uncemented, extensively porous-coated femoral components requiring removal using a trephine with a minimum of two years of follow-up.

The mean time to failure was 4.6 years (1.7 to 9.1, standard deviation (sd) 2.3). These were revised using a larger extensively porous coated component, fluted tapered modular component, a proximally coated modular component, or a proximal femoral replacement. The mean clinical follow-up after revision THA was 4.9 years (2 to 22, sd 3.1). The mean diameter of the femoral component increased from 12.7 mm (sd 1.9) to 16.2 mm (sd 3.4; p >  0.001). Two revision components had radiographic evidence of subsidence that remained radiographically stable at final follow-up. The most common post-operative complication was instability affecting six patients (54.5%) on at least one occasion.

A total of four patients (36.4%) required further revision: three for instability and one for fracture of the revision component. There was no statistically significant difference in the mean Harris hip score before implant fracture (82.4; sd 18.3) and after trephine removal and revision THA (81.2; sd 14.8, p = 0.918).

These findings suggest that removal of a fractured, well-fixed, uncemented, extensively porous-coated femoral component using a trephine does not compromise subsequent fixation at revision THA and the patient’s pre-operative level of function can be restored. However, the loss of proximal bone stock before revision may be associated with a high rate of dislocation post-operatively.

Cite this article: Bone Joint J 2015;97-B:1192–6.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 950 - 956
1 Jul 2015
Tsitsilonis S Schaser KD Wichlas F Haas NP Manegold S

The incidence of periprosthetic fractures of the ankle is increasing. However, little is known about the outcome of treatment and their management remains controversial. The aim of this study was to assess the impact of periprosthetic fractures on the functional and radiological outcome of patients with a total ankle arthroplasty (TAA).

A total of 505 TAAs (488 patients) who underwent TAA were retrospectively evaluated for periprosthetic ankle fracture: these were then classified according to a recent classification which is orientated towards treatment. The outcome was evaluated clinically using the American Orthopedic Foot and Ankle Society (AOFAS) score and a visual analogue scale for pain, and radiologically.

A total of 21 patients with a periprosthetic fracture of the ankle were identified. There were 13 women and eight men. The mean age of the patients was 63 years (48 to 74). Thus, the incidence of fracture was 4.17%.

There were 11 intra-operative and ten post-operative fractures, of which eight were stress fractures and two were traumatic. The prosthesis was stable in all patients. Five stress fractures were treated conservatively and the remaining three were treated operatively.

A total of 17 patients (81%) were examined clinically and radiologically at a mean follow-up of 53.5 months (12 to 112). The mean AOFAS score at follow-up was 79.5 (21 to 100). The mean AOFAS score in those with an intra-operative fracture was 87.6 (80 to 100) and for those with a stress fracture, which were mainly because of varus malpositioning, was 67.3 (21 to 93). Periprosthetic fractures of the ankle do not necessarily adversely affect the clinical outcome, provided that a treatment algorithm is implemented with the help of a new classification system.

Cite this article: Bone Joint J 2015;97-B:950–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1610 - 1616
1 Dec 2011
Pegg E Pandit H Gill HS Keys GW Svard UGC O’Connor JJ Murray DW

Since the Oxford knee was first used unicompartmentally in 1982, a small number of bearings have fractured. Of 14 retrieved bearings, we examined ten samples with known durations in situ (four Phase 1, four Phase 2 and two Phase 3). Evidence of impingement and associated abnormally high wear (> 0.05 mm per year) as well as oxidation was observed in all bearings. In four samples the fracture was associated with the posterior radio-opaque wire. Fracture surfaces indicated fatigue failure, and scanning electron microscopy suggested that the crack initiated in the thinnest region. The estimated incidence of fracture was 3.20% for Phase 1, 0.74% for Phase 2, 0.35% for Phase 3, and 0% for Phase 3 without the posterior marker wire. The important aetiological factors for bearing fracture are impingement leading to high wear, oxidation, and the posterior marker wire. With improved surgical technique, impingement and high wear should be prevented and modern polyethylene may reduce the oxidation risk. A posterior marker wire is no longer used in the polyethylene meniscus. Therefore, the rate of fracture, which is now very low, should be reduced to a negligible level.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 10 | Pages 1385 - 1390
1 Oct 2005
Niva MH Kiuru MJ Haataja R Pihlajamäki HK

The purpose of this study was to describe the anatomical distribution and incidence of fatigue injuries of the femur in physically-active young adults, based upon MRI studies. During a period of 70 months, 1857 patients with exercise-induced pain in the femur underwent MRI of the pelvis, hips, femora, and/or knees.

Of these, 170 patients had a total of 185 fatigue injuries, giving an incidence of 199 per 100 000 person-years. Bilateral injuries occurred in 9% of patients. The three most common sites affected were the femoral neck (50%), the condylar area (24%) and the proximal shaft (18%). A fatigue reaction was seen in 57%, and a fracture line in 22%. There was a statistical correlation between the severity of the fatigue injury and the duration of pain (p = 0.001). The location of the pain was normally at the site of the fatigue injury. Fatigue injuries of the femur appear to be relatively common in physically-active patients.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 36 - 42
1 Nov 2014
Sheth NP Melnic CM Paprosky WG

Acetabular bone loss is a challenging problem facing the revision total hip replacement surgeon. Reconstruction of the acetabulum depends on the presence of anterosuperior and posteroinferior pelvic column support for component fixation and stability. The Paprosky classification is most commonly used when determining the location and degree of acetabular bone loss. Augments serve the function of either providing primary construct stability or supplementary fixation.

When a pelvic discontinuity is encountered we advocate the use of an acetabular distraction technique with a jumbo cup and modular porous metal acetabular augments for the treatment of severe acetabular bone loss and associated chronic pelvic discontinuity.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):36–42.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 863 - 867
1 Jul 2014
Aitken SA Hutchison JD McQueen MM Court-Brown CM

Epidemiological studies enhance clinical practice in a number of ways. However, there are many methodological difficulties that need to be addressed in designing a study aimed at the collection and analysis of data concerning fractures and other injuries. Most can be managed and errors minimised if careful attention is given to the design and implementation of the research.

Cite this article: Bone Joint J 2014;96-B:863–7.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 806 - 809
1 Jun 2008
Burkhart KJ Rommens PM

We describe a patient with insufficient bone regeneration of the tibia after bone transport over an intramedullary nail, in whom union was ultimately achieved after exchange nailing and intramedullary application of rh-bone morphogenetic protein-7 at the site of distraction.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 366 - 371
1 Mar 2007
Krieg AH Davidson AW Stalley PD

Between 1996 and 2003, 16 patients (nine female, seven male) were treated for a primary bone sarcoma of the femur by wide local excision of the tumour, extracorporeal irradiation and re-implantation. An additional vascularised fibular graft was used in 13 patients (81%). All patients were free from disease when reviewed at a minimum of two years postoperatively (mean 49.7 months (24 to 96).

There were no cases of infection. Primary union was achieved after a median of nine months (interquartile range 7 to 11). Five host-donor junctions (16%) united only after a second procedure. Primary union recurred faster at metaphyseal junctions (94% (15) at a median of 7.5 months (interquartile range 4 to 12)) than at diaphyseal junctions (75% (12) at a median of 11.1 months (interquartile range 5 to 18)).

Post-operatively, the median Musculoskeletal Tumour Society score was 85% (interquartile range 75 to 96) and the median Toronto Extremity Salvage score 94% (interquartile range 82 to 99). The Mankin score gave a good or excellent result in 14 patients (88%).

The range of movement of the knee was significantly worse when the extracorporeally irradiated autografts were fixed by plates rather than by nails (p = 0.035).

A total of 16 (62%) of the junctions of the vascularised fibular grafts underwent hypertrophy, indicating union and loading.

Extracorporeal irradiation autografting with supplementary vascularised fibular grafting is a promising biological alternative for intercalary reconstruction after wide resection of malignant bone tumours of the femur.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 114 - 121
1 Jan 2014
Pekmezci M McDonald E Buckley J Kandemir U

We investigated a new intramedullary locking nail that allows the distal interlocking screws to be locked to the nail. We compared fixation using this new implant with fixation using either a conventional nail or a locking plate in a laboratory simulation of an osteoporotic fracture of the distal femur. A total of 15 human cadaver femora were used to simulate an AO 33-A3 fracture pattern. Paired specimens compared fixation using either a locking or non-locking retrograde nail, and using either a locking retrograde nail or a locking plate. The constructs underwent cyclical loading to simulate single-leg stance up to 125 000 cycles. Axial and torsional stiffness and displacement, cycles to failure and modes of failure were recorded for each specimen. When compared with locking plate constructs, locking nail constructs had significantly longer mean fatigue life (75 800 cycles (sd 33 900) vs 12 800 cycles (sd 6100); p = 0.007) and mean axial stiffness (220 N/mm (sd 80) vs 70 N/mm (sd 18); p = 0.005), but lower mean torsional stiffness (2.5 Nm/° (sd 0.9) vs 5.1 Nm/° (sd 1.5); p = 0.008). In addition, in the nail group the mode of failure was either cut-out of the distal screws or breakage of nails, and in the locking plate group breakage of the plate was always the mode of failure. Locking nail constructs had significantly longer mean fatigue life than non-locking nail constructs (78 900 cycles (sd 25 600) vs 52 400 cycles (sd 22 500); p = 0.04).

The new locking retrograde femoral nail showed better stiffness and fatigue life than locking plates, and superior fatigue life to non-locking nails, which may be advantageous in elderly patients.

Cite this article: Bone Joint J 2014;96-B:114–21.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 751 - 759
1 Jun 2010
Tsirikos AI Garrido EG

A review of the current literature shows that there is a lack of consensus regarding the treatment of spondylolysis and spondylolisthesis in children and adolescents. Most of the views and recommendations provided in various reports are weakly supported by evidence. There is a limited amount of information about the natural history of the condition, making it difficult to compare the effectiveness of various conservative and operative treatments. This systematic review summarises the current knowledge on spondylolysis and spondylolisthesis and attempts to present a rational approach to the evaluation and management of this condition in children and adolescents.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 61 - 64
1 Nov 2012
Mayle Jr RE Paprosky WG

Revision total hip arthroplasty (THA) is projected to increase by 137% from the years 2005 to 2030. Reconstruction of the femur with massive bone loss can be a formidable undertaking. The goals of revision surgery are to create a stable construct, preserve bone and soft tissues, augment deficient host bone, improve function, provide a foundation for future surgery, and create a biomechanically restored hip. Options for treatment of the compromised femur include: resection arthroplasty, allograft prosthetic composite (APC), proximal femoral replacement, cementless fixation with a modular tapered fluted stem, and impaction grafting. The purpose of this article is to review the treatment options along with their associated outcomes in the more severe femoral defects (Paprosky types IIIb and IV) in revision THA.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 9 | Pages 1288 - 1291
1 Sep 2012
Sanghrajka AP Hill RA Murnaghan CF Simpson AHRW Bellemore MC

We describe three cases of infantile tibia vara resulting from an atraumatic slip of the proximal tibial epiphysis upon the metaphysis. There appears to be an association between this condition and severe obesity. Radiologically, the condition is characterised by a dome-shaped metaphysis, an open growth plate and disruption of the continuity between the lateral borders of the epiphysis and metaphysis, with inferomedial translation of the proximal tibial epiphysis. All patients were treated by realignment of the proximal tibia by distraction osteogenesis with an external circulator fixator, and it is suggested that this is the optimal method for correction of this complex deformity. There are differences in the radiological features and management between conventional infantile Blount’s disease and this ‘slipped upper tibial epiphysis’ variant.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1367 - 1372
1 Oct 2006
Gupta A Pollock R Cannon SR Briggs TWR Skinner J Blunn G

We used a knee-sparing distal femoral endoprosthesis in young patients with malignant bone tumours of the distal femur in whom it was possible to resect the tumour and to preserve the distal femoral condyles. The proximal shaft of the endoprosthesis had a coated hydroxyapatite collar, while the distal end had hydroxyapatite-coated extracortical plates to secure it to the small residual femoral condylar fragment. We reviewed the preliminary results of this endoprosthesis in eight patients with primary bone tumours of the distal femur. Their mean age at surgery was 17.years (14 to 21). The mean follow-up was 24 months (20 to 31). At final follow-up the mean flexion at the knee was 102° (20° to 120°) and the mean Musculoskeletal Tumour Society score was 80% (57% to 96.7%).

There was excellent osteointegration at the prosthesis-proximal bone interface with formation of new bone around the hydroxyapatite collar. The prosthesis allowed preservation of the knee and achieved a good functional result. Formation of new bone and remodelling at the interface make the implant more secure. Further follow-up is required to determine the long-term structural integrity of the prosthesis.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1110 - 1115
1 Aug 2006
Ong KL Kurtz SM Manley MT Rushton N Mohammed NA Field RE

The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented versus uncemented fixation showed no advantage of either with regard to bone loading. When the Birmingham hip resurfacing femoral component was fixed to bone, proximal femoral stresses and strains were non-physiological. Bone resorption was predicted in the inferomedial and superolateral bone within the Birmingham hip resurfacing shell. Resorption was limited to the superolateral region when the stem was not fixed. The increased bone strain observed adjacent to the distal stem should stimulate an increase in bone density at that location. The remodelling of bone seen during revision of failed Birmingham hip resurfacing implants appears to be consistent with the predictions of our finite element analysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 671 - 677
1 May 2012
Raiss P Sowa B Bruckner T Eck S Woerz S Rohr K Rickert M Kasten P

The aim of this study was to compare a third-generation cementing procedure for glenoid components with a new technique for cement pressurisation. In 20 pairs of scapulae, 20 keeled and 20 pegged glenoid components were implanted using either a third-generation cementing technique (group 1) or a new pressuriser (group 2). Cement penetration was measured by three-dimensional (3D) analysis of micro-CT scans. The mean 3D depth of penetration of the cement was significantly greater in group 2 (p < 0.001). The mean thickness of the cement mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and 5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm (0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement mantle < 2 mm was detected less frequently in group 2 (p < 0.001). Using the cement pressuriser the proportion of cement mantles < 2 mm was significantly reduced compared with the third-generation cementing technique.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 717 - 723
1 Jun 2014
Altaf F Heran MKS Wilson LF

Back pain is a common symptom in children and adolescents. Here we review the important causes, of which defects and stress reactions of the pars interarticularis are the most common identifiable problems. More serious pathology, including malignancy and infection, needs to be excluded when there is associated systemic illness. Clinical evaluation and management may be difficult and always requires a thorough history and physical examination. Diagnostic imaging is obtained when symptoms are persistent or severe. Imaging is used to reassure the patient, relatives and carers, and to guide management.

Cite this article: Bone Joint J 2014;96-B:717–23.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design.

At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%).

In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1690 - 1694
1 Dec 2010
Scoccianti G Campanacci DA Beltrami G Caldora P Capanna R

Several techniques have been described to reconstruct a mobile wrist joint after resection of the distal radius for tumour. We reviewed our experience of using an osteo-articular allograft to do this in 17 patients with a mean follow-up of 58.9 months (28 to 119).

The mean range of movement at the wrist was 56° flexion, 58° extension, 84° supination and 80° pronation. The mean ISOLS-MSTS score was 86% (63% to 97%) and the mean patient-rated wrist evaluation score was 16.5 (3 to 34). There was no local recurrence or distant metastases. The procedure failed in one patient with a fracture of the graft and an arthrodesis was finally required. Union was achieved at the host-graft interface in all except two cases. No patient reported more than modest non-disabling pain and six reported no pain at all. Radiographs showed early degenerative changes at the radiocarpal joint in every patient.

A functional pain-free wrist can be restored with an osteo-articular allograft after resection of the distal radius for bone tumour, thereby avoiding the donor site morbidity associated with an autograft. These results may deteriorate with time.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 760 - 768
1 Jun 2011
ten Broeke RHM Alves A Baumann A Arts JJC Geesink RGT

Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface.

Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer.

The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no adverse effects identified from either the BONIT-HA coating or the DOTIZE surface treatment.