We studied the tapered interface between the head and the neck of 139 modular femoral components of hip prostheses which had been removed for a variety of reasons. In 91 the same alloy had been used for the head and the stem; none of them showed evidence of corrosion. In contrast, there was definite corrosion in 25 of the 48 prostheses in which the stem was of titanium alloy and the head of
In a post-mortem study, we compared subjects with metal implants with and without visible wear with an age-matched control group to determine the extent and effects of dissemination of wear debris. In subjects with stainless-steel and
We implanted cylinders of
We aimed to evaluate the long-term outcome of highly cross-linked polyethylene (HXLPE) cemented acetabular components and assess whether any radiolucent lines (RLLs) which arose were progressive. We retrospectively reviewed 170 patients who underwent 187 total hip arthroplasties at two hospitals with a minimum follow-up of ten years. All interventions were performed using the same combination of HXLPE cemented acetabular components with femoral stems made of titanium alloy. Kaplan-Meier survival analysis was performed for the primary endpoint of acetabular component revision surgery for any reason and secondary endpoint of the appearance of RLLs. RLLs that had appeared once were observed over time. We statistically assessed potential relationships between RLLs and a number of factors, including the technique of femoral head autografting and the Japanese Orthopaedic Association score.Aims
Methods
We investigated the fixation of fully-threaded cementless acetabular prostheses in 20 patients with osteoarthritis, measuring the migration of the cup using roentgen stereophotogrammetric analysis (RSA). All the cups migrated proximally, 13 moved laterally or medially, and nine moved anteriorly or posteriorly in the first two postoperative years, the average migration being 1.1 to 1.4 mm in either direction. Rotatory movements of up to 5.7 degrees were found in nine of the 13 hips where this analysis could be performed. Movements of
We have studied 27 tibial prostheses retrieved from knee replacements after 1 to 9 years. In 22 the femoral components were of
Ten uncemented total hip replacements were performed in 1975 using an implant in which the
The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncemented metal-backed acetabular components. A total of 9113 primary uncemented acetabular components were implanted in 7937 patients between 1987 and 2007. These were included in a prospective, population-based observational study. All the implants were modular and metal-backed with ultra-high-molecular-weight polyethylene liners. The femoral heads were made of stainless steel,
We examined stainless-steel,
Total hip replacement using porous-coated
Orthopaedic and reconstructive surgeons are faced with large defects after the resection of malignant tumours of the sacrum. Spinopelvic reconstruction is advocated for resections above the level of the S1 neural foramina or involving the sacroiliac joint. Fixation may be augmented with either free vascularized fibular flaps (FVFs) or allograft fibular struts (AFSs) in a cathedral style. However, there are no studies comparing these reconstructive techniques. We reviewed 44 patients (23 female, 21 male) with a mean age of 40 years (SD 17), who underwent en bloc sacrectomy for a malignant tumour of the sacrum with a reconstruction using a total (n = 20), subtotal (n = 2), or hemicathedral (n = 25) technique. The reconstructions were supplemented with a FVF in 25 patients (57%) and an AFS in 19 patients (43%). The mean length of the strut graft was 13 cm (SD 4). The mean follow-up was seven years (SD 5).Aims
Methods
We reviewed 264 consecutive primary total hip replacements in 244 patients in which the Harris-Galante I porous-coated acetabular component had been used. The mean follow-up was 95 months (69 to 132). In 221 arthroplasties a 32 mm ceramic head had been used, and in the other 43 one of
To establish our early clinical results of a new total knee arthroplasty (TKA) tibial component introduced in 2013 and compare it to other designs in use at our hospital during the same period. This is a retrospective study of 166 (154 patients) consecutive cemented, fixed bearing, posterior-stabilized (PS) TKAs (ATTUNE) at one hospital performed by five surgeons. These were compared with a reference cohort of 511 knees (470 patients) of other designs (seven manufacturers) performed at the same hospital by the same surgeons. There were no significant differences in age, sex, BMI, or follow-up times between the two cohorts. The primary outcome was revision performed or pending.Aims
Methods
1. The probable greatest bending moment applied to a plated or nailed fracture of the tibia during restricted weight-bearing is estimated to be, in men, up to about 79 Newton metres (58 poundsforce feet). The maximum twisting moment is estimated to be about 29 Newton metres (22 poundsforce feet). 2. Twenty-two human tibiae were loaded in three-point bending and broke at bending moments of from 57·9 to 294 Newton metres (42·7 to 216 poundsforce feet) if they had not previously been drilled; tibiae which had holes made through both cortices with a c. 3-millimetre (⅛-inch) drill broke at from 32·4 to 144 Newton metres (23·8 to 106 poundsforce feet). Tibiae loaded in torsion broke at twisting moments of from 27·5 to 892 Newton metres (20·2 to 65·8 poundsforce feet) when not drilled, 23·6 to 77·5 Newton metres (l7·3 to 57·1 poundsforce feet) when drilled. 3. When bent so as to open the fracture site, the 14-centimetre Stamm was the strongest of all the single plates tested (reaching its elastic limit at a bending moment of 17·6 Newton metres (13 poundsforce feet) and 5 degrees total angulation at 22·6 Newton metres (16·6 poundsforce feet)), while the Venable was the weakest (elastic limit 4·9 Newton metres (3·6 poundsforce feet) and 5 degrees at 7·9 Newton metres (5·8 poundsforce feet)). A 13-millimetre Küntscher nail reached its elastic limit at 42·2 Newton metres (31·1 poundsforce feet) and 5 degrees total angulation at 49 Newton metres (36 poundsforce feet). 4. In torsion the 15-centimetre Hicks was the strongest ofthe plates (elastic limit 27·5 Newton metres (20·2 poundsforce feet) and 5 degrees rotation at 16·7 Newton metres (l2·3 poundsforce feet)). 5. Küntscher nails in bones provided no dependable strength in torsion. 6. In both bending and torsion, a preparation of one Venable plate on each of the two anterior surfaces was stronger than any single plate, and was as strong as the weaker drilled tibiae. 7. The three currently available metallic materials (stainless steel,
To report the mid-term results of a modified self-growing rod (SGR) technique for the treatment of idiopathic and neuromuscular early-onset scoliosis (EOS). We carried out a retrospective analysis of 16 consecutive patients with EOS treated with an SGR construct at a single hospital between September 2008 and December 2014. General demographics and deformity variables (i.e. major Cobb angle, T1 to T12 length, T1 to S1 length, pelvic obliquity, shoulder obliquity, and C7 plumb line) were recorded preoperatively, and postoperatively at yearly follow-up. Complications and revision procedures were also recorded. Only patients with a minimum follow-up of five years after surgery were included.Aims
Methods
To evaluate the histopathological examination of peri-implant tissue samples as a technique in the diagnosis of postoperative spinal implant infection (PSII). This was a retrospective analysis. Patients who underwent revision spinal surgery at our institution were recruited for this study. PSII was diagnosed by clinical signs, histopathology, and microbiological examination of intraoperatively collected samples. Histopathology was defined as the gold standard. The sensitivity for histopathology was calculated. A total of 47 patients with PSII and at least one microbiological and histopathological sample were included in the study.Aims
Methods
We have previously demonstrated raised cobalt and chromium levels in patients with larger diameter femoral heads, following metal-on-polyethylene uncemented total hip arthroplasty. Further data have been collected, to see whether these associations have altered with time and to determine the long-term implications for these patients and our practice. Patients from our previous study who underwent Trident-Accolade primary total hip arthroplasties using a metal-on-polyethylene bearing in 2009 were reviewed. Patients were invited to have their cobalt and chromium levels retested, and were provided an Oxford Hip Score. Serum ion levels were then compared between groups (28 mm, 36 mm, and 40 mm heads) and over time.Aims
Methods
Biofilm formation is intrinsic to prosthetic joint infection (PJI). In the current study, we evaluated the effects of silver-containing hydroxyapatite (Ag-HA) coating and vancomycin (VCM) on methicillin-resistant Pure titanium discs (Ti discs), Ti discs coated with HA (HA discs), and 3% Ag-HA discs developed using a thermal spraying were inoculated with MRSA suspensions containing a mean in vitro 4.3 (SD 0.8) x 106 or 43.0 (SD 8.4) x 105 colony-forming units (CFUs). Immediately after MRSA inoculation, sterile phosphate-buffered saline or VCM (20 µg/ml) was added, and the discs were incubated for 24 hours at 37°C. Viable cell counting, 3D confocal laser scanning microscopy with Airyscan, and scanning electron microscopy were then performed. HA discs and Ag HA discs were implanted subcutaneously in vivo in the dorsum of rats, and MRSA suspensions containing a mean in vivo 7.2 (SD 0.4) x 106 or 72.0 (SD 4.2) x 105 CFUs were inoculated on the discs. VCM was injected subcutaneously daily every 12 hours followed by viable cell counting.Aims
Methods