The purpose of this prospective, randomised study
was to evaluate the clinical and radiological results comparing the
identical cemented or cementless NexGen total knee prostheses implanted
bilaterally in the same patient. Sequential simultaneous bilateral
total knee replacements were performed in 50 patients (100 knees).
There were 39 women and 11 men with a mean age of 58.4 years (51
to 67) who received a cemented prosthesis in one knee and a cementless
prosthesis in the other. The mean follow-up was 13.6 years (13 to
14). At final review, the mean Knee Society scores (96.2 (82 to
100)
We investigated whether the extension gap in total knee replacement (TKR) would be changed when the femoral component was inserted. The extension gap was measured with and without the femoral component in place in 80 patients with varus osteoarthritis undergoing posterior-stabilised TKR. The effect of a post-operative increase in the size of the femoral posterior condyles was also evaluated. The results showed that placement of the femoral component significantly reduced the medial and lateral extension gaps by means of 1.0 mm and 0.9 mm, respectively (p <
0.0001). The extension gap was reduced when a larger femoral component was selected relative to the thickness of the resected posterior condyle. When the post-operative posterior lateral condyle was larger than that pre-operatively, 17 of 41 knees (41%) showed a decrease in the extension gap of >
2.0 mm. When a specially made femoral trial component with a posterior condyle enlarged by 4 mm was tested, the medial and lateral extension gaps decreased further by means of 2.1 mm and 2.8 mm, respectively. If the thickness of the posterior condyle is expected to be larger than that pre-operatively, it should be recognised that the extension gap is likely to be altered. This should be taken into consideration when preparing the extension gap.
Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
We undertook a study in which 138 female patients with a mean age of 71.2 years (51 to 82) received a standard NexGen CR-flex prosthesis in one knee and a gender-specific NexGen CR-flex prosthesis in the other. The mean follow-up period was 3.25 years (3.1 to 3.5). The aspect ratios of the standard and gender-specific prostheses were compared with that of the distal femur. The mean post-operative Knee Society knee scores were 94 (70 to 100) and 93 (70 to 100) points and the function scores were 83 (60 to 100) and 84 (60 to 100) points for the standard implants and the gender-specific designs, respectively. The mean post-operative Western Ontario and McMaster Universities score was 26.4 points (0 to 76). Patient satisfaction, the radiological results and the complication rates were similar in the two groups. In those with a standard prosthesis, the femoral component was closely matched in 80 knees (58.0%), overhung in 14 (10.1%) and undercovered the bone in 44 (31.9%). In those with a gender-specific prosthesis, it was closely matched in 15 knees (10.9%) and undercovered the bone in 123 (89.1%). Since we found no significant differences between the two groups with regard to the clinical and radiological results, patient satisfaction or complication rate, the goal of the design of the gender-specific CR-flex prosthesis to improve the outcome was not achieved in our patients.
The Oxford mobile-bearing unicompartmental knee
replacement (UKR) is an effective and safe treatment for osteoarthritis
of the medial compartment. The results in the lateral compartment
have been disappointing due to a high early rate of dislocation
of the bearing. A series using a newly designed domed tibial component
is reported. The first 50 consecutive domed lateral Oxford UKRs in 50 patients
with a mean follow-up of three years (2.0 to 4.3) were included.
Clinical scores were obtained prospectively and Kaplan-Meier survival
analysis was performed for different endpoints. Radiological variables
related to the position and alignment of the components were measured. One patient died and none was lost to follow-up. The cumulative
incidence of dislocation was 6.2% (95% confidence interval (CI)
2.0 to 17.9) at three years. Survival using revision for any reason
and aseptic revision was 94% (95% CI 82 to 98) and 96% (95% CI 85
to 99) at three years, respectively. Outcome scores, visual analogue
scale for pain and maximum knee flexion showed a significant improvement
(p <
0.001). The mean Oxford knee score was 43 ( Clinical results are excellent and short-term survival has improved
when compared with earlier series. The risk of dislocation remains
higher using a mobile-bearing UKR in the lateral compartment when
compared with the medial compartment. Patients should be informed
about this complication. To avoid dislocations, care must be taken
not to elevate the lateral joint line.
We have investigated the accuracy of placement of the femoral component using imageless navigation in 100 consecutive Birmingham Hip Resurfacings. Pre-operative templating determined the native neck-shaft angle and planned stem-shaft angle of the implant. The latter were verified post-operatively using digital anteroposterior unilateral radiographs of the hip. The mean neck-shaft angle determined before operation was 132.7° (118° to 160°). The mean planned stem-shaft angle was a relative valgus alignment of 9.7° ( Navigation in hip resurfacing may afford the surgeon a reliable and accurate method of placement of the femoral component.
We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p <
0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p <
0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.
We performed a prospective, randomised study to compare the results and rates of complications of primary total knee replacement performed using a quadriceps-sparing technique or a standard arthrotomy in 120 patients who had bilateral total knee replacements carried out under the same anaesthetic. The clinical results, pain scales, surgical and hospital data, post-operative complications and radiological results were compared. No significant differences were found between the two groups with respect to the blood loss, knee score, function score, pain scale, range of movement or radiological findings. In contrast, the operating time (p = 0.0001) and the tourniquet time (p <
0.0001) were significantly longer in the quadriceps-sparing group, as was the rate of complications (p = 0.0468). We therefore recommend the use of a standard arthrotomy with the shortest possible skin incision for total knee replacement.
We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1). The total knee score, pain score, functional score and range of movement were not statistically different (p >
0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing.
We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed. Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.
The computed neck-shaft angle and the size of the femoral component were recorded in 100 consecutive hip resurfacings using imageless computer-navigation and compared with the angle measured before operation and with actual component implanted. The reliability of the registration was further analysed using ten cadaver femora. The mean absolute difference between the measured and navigated neck-shaft angle was 16.3° (0° to 52°). Navigation underestimated the measured neck-shaft angle in 38 patients and the correct implant size in 11. Registration of the cadaver femora tended to overestimate the correct implant size and provided a low level of repeatability in computing the neck-shaft angle. Prudent pre-operative planning is advisable for use in conjunction with imageless navigation since misleading information may be registered intraoperatively, which could lead to inappropriate sizing and positioning of the femoral component in hip resurfacing.
We carried out three total knee replacements with proximal realignment in two patients with severe osteoarthritis of the knee and congenital dislocation of the patella. During the operation, the femur and the tibia were cut according to the recommendations of the manufacturer of the implant. The femoral component was placed in external rotation and the centre of the tibial component aligned in relation to the tibial tuberosity with regard to rotation and translation. After making the bone cuts, the iliotibial tract was detached from Gerdy’s tubercle, the popliteus tendon divided, and the biceps femoris tendon elongated by Z-plasty. After the trial implants were positioned, a proximal re-alignment procedure was performed. One knee had deep infection. There was one dislocated patella which was repositioned, and walking ability was improved in all knees. We believe that the rotational alignment of a prosthesis is as important as the soft-tissue surgery.
Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p <
0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.
We performed a prospective, randomised controlled trial of unicompartmental knee arthroplasty comparing the performance of the Acrobot system with conventional surgery. A total of 27 patients (28 knees) awaiting unicompartmental knee arthroplasty were randomly allocated to have the operation performed conventionally or with the assistance of the Acrobot. The primary outcome measurement was the angle of tibiofemoral alignment in the coronal plane, measured by CT. Other secondary parameters were evaluated and are reported. All of the Acrobot group had tibiofemoral alignment in the coronal plane within 2° of the planned position, while only 40% of the conventional group achieved this level of accuracy. While the operations took longer, no adverse effects were noted, and there was a trend towards improvement in performance with increasing accuracy based on the Western Ontario and McMaster Universities Osteoarthritis Index and American Knee Society scores at six weeks and three months. The Acrobot device allows the surgeon to reproduce a pre-operative plan more reliably than is possible using conventional techniques which may have clinical advantages.
We have compared the biomechanical nature of the reconstruction of the hip in conventional total hip arthroplasty (THA) and surface replacement arthroplasty (SRA) in a randomised study involving 120 patients undergoing unilateral primary hip replacement. The contralateral hip was used as a control. Post-operatively, the femoral offset was significantly increased with THA (mean 5.1 mm; −2.8 to 11.6) and decreased with SRA (mean −3.3 mm; −8.9 to 8.2). Femoral offset was restored within Restoration of the normal proximal femoral anatomy was more precise with SRA. The enhanced stability afforded by the use of a large-diameter femoral head avoided over-lengthening of the limb or increased offset to improve soft-tissue tension as occurs sometimes in THA. In a subgroup of patients with significant pre-operative deformity, restoration of the normal hip anatomy with lower pre-operative femoral offset or significant shortening of the leg was still possible with SRA.
We measured the contact areas and contact stresses at the post-cam mechanism of a posterior-stabilised total knee arthroplasty when a posterior force of 500 N was applied to the Kirschner Performance, Scorpio Superflex, NexGen LPS Flex Fixed, and NexGen LPS Flex Mobile knee systems. Measurements were made at 90°, 120°, and 150° of flexion both in neutral rotation and 10° of internal rotation of the tibial component. Peak contact stresses at 90°, 120°, and 150° were 24.0, 33.9, and 28.8 MPa, respectively, for the Kirschner; 26.0, 32.4, and 22.1 MPa, respectively, for the Scorpio; and 34.1, 31.5, and 32.5 MPa, respectively, for the NexGen LPS Flex Fixed. With an internally rotated tibia, the contact stress increased significantly with all the fixed-bearing arthroplasties but not with the NexGen LPS Flex Mobile arthroplasty. The post-cam design should be modified in order to provide a larger contact area whilst avoiding any impingement and edge loading.