Bone & Joint Research
Vol. 7, Issue 2 | Pages 196 - 204
1 Feb 2018
Krull A Morlock MM Bishop NE

Objectives

Taper junctions between modular hip arthroplasty femoral heads and stems fail by wear or corrosion which can be caused by relative motion at their interface. Increasing the assembly force can reduce relative motion and corrosion but may also damage surrounding tissues. The purpose of this study was to determine the effects of increasing the impaction energy and the stiffness of the impactor tool on the stability of the taper junction and on the forces transmitted through the patient’s surrounding tissues.

Methods

A commercially available impaction tool was modified to assemble components in the laboratory using impactor tips with varying stiffness at different applied energy levels. Springs were mounted below the modular components to represent the patient. The pull-off force of the head from the stem was measured to assess stability, and the displacement of the springs was measured to assess the force transmitted to the patient’s tissues.


Bone & Joint 360
Vol. 7, Issue 5 | Pages 33 - 36
1 Oct 2018


Bone & Joint 360
Vol. 6, Issue 4 | Pages 23 - 25
1 Aug 2017


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1100 - 1105
1 Aug 2018
Howard EL Shepherd KL Cribb G Cool P

Aims

The aim of this study was to validate the Mirels score in predicting pathological fractures in metastatic disease of the lower limb.

Patients and Methods

A total of 62 patients with confirmed metastatic disease met the inclusion criteria. Of the 62 patients, 32 were female and 30 were male. The mean age of patients was 65 years (35 to 89). The primary malignancy originated from the breast in 27 (44%) patients, prostate in 15 (24%) patients, kidney in seven (11%), and lung in four (6%) of patients. One patient (2%) had metastatic carcinoma from the lacrimal gland, two patients (3%) had multiple myeloma, one patient (2%) had lymphoma of bone, and five patients (8%) had metastatic carcinoma of unknown primary. Plain radiographs at the time of initial presentation were scored using Mirels system by the four authors. The radiographic components of the score (anatomical site, size, and radiographic appearance) were scored two weeks apart. Inter- and intraobserver reliability were calculated with Fleiss’ kappa test. Bland-Altman plots were created to compare the variances of the individual components of the score and the total Mirels score.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1286 - 1289
1 Oct 2017
Rajpura A Board TN Siney PD Wynn Jones H Williams S Dabbs L Wroblewski BM

Aims

Our aim in this study was to describe a continuing review of 11 total hip arthroplasties using 22.225 mm Alumina ceramic femoral heads on a Charnley flanged femoral component, articulating against a silane crosslinked polyethylene.

Patients and Methods

Nine patients (11 THAs) were reviewed at a mean of 27.5 years (26 to 28) post-operatively. Outcome was assessed using the d’Aubigne and Postel, and Charnley scores and penetration was recorded on radiographs. In addition, the oxidation of a 29-year-old shelf-aged acetabular component was analysed.


The Bone & Joint Journal
Vol. 100-B, Issue 2 | Pages 212 - 218
1 Feb 2018
Jungbluth P Tanner S Schneppendahl J Grassmann J Wild M Hakimi M Windolf J Laun R

Aims

The aim of this retrospective multicentre study was to evaluate mid-term results of the operative treatment of Monteggia-like lesions and to determine the prognostic factors that influence the clinical and radiological outcome.

Patients and Methods

A total of 46 patients (27 women and 19 men), with a mean age of 57.7 years (18 to 84) who had sustained a Monteggia-like lesion were followed up clinically and radiologically after surgical treatment. The Mayo Modified Wrist Score (MMWS), Mayo Elbow Performance Score (MEPS), Broberg and Morrey Score, and Disabilities of the Arm, Shoulder and Hand (DASH) score were used for evaluation at a mean of 65 months (27 to 111) postoperatively. All ulnar fractures were stabilized using a proximally contoured or precontoured locking compression plate. Mason type I fractures of the radial head were treated conservatively, type II fractures were treated with reconstruction, and type III fractures with arthroplasty. All Morrey type II and III fractures of the coronoid process was stabilized using lag screws.


The Bone & Joint Journal
Vol. 99-B, Issue 4 | Pages 489 - 493
1 Apr 2017
Sadamasu A Yamaguchi S Nakagawa R Kimura S Endo J Akagi R Sasho T

Aims

The purposes of this study were to clarify first, the incidence of peroneal tendon dislocation in patients with a fracture of the talus and second the factors associated with peroneal tendon dislocation.

Patients and Methods

We retrospectively examined 30 patients (30 ankles) with a mean age of 37.5 years, who had undergone internal fixation for a fracture of the talus. Independent examiners assessed for peroneal tendon dislocation using the pre-operative CT images. The medical records were also reviewed for the presence of peroneal tendon dislocation. The associations between the presence of dislocation with the patient characteristics or radiological findings, including age, mechanism of injury, severity of fracture, and fleck sign, were assessed using Fisher’s exact tests.


Objectives

This investigation sought to advance the work published in our prior biomechanical study (Journal of Orthopaedic Research, 2016). We specifically sought to determine whether there are additional easy-to-measure parameters on plain radiographs of the proximal humerus that correlate more strongly with ultimate fracture load, and whether a parameter resembling the Dorr strength/quality characterisation of proximal femurs can be applied to humeri.

Materials and Methods

A total of 33 adult humeri were used from a previous study where we quantified bone mineral density of the proximal humerus using radiographs and dual-energy x-ray absorptiometry (DEXA), and regional mean cortical thickness and cortical index using radiographs. The bones were fractured in a simulated backwards fall with the humeral head loaded at 2 mm/second via a frustum angled at 30° from the long axis of the bone. Correlations were assessed with ultimate fracture load and these new parameters: cortical index expressed in areas (“areal cortical index”) of larger regions of the diaphysis; the canal-to-calcar ratio used analogous to its application in proximal femurs; and the recently described medial cortical ratio.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 865 - 871
1 Jul 2017
Amstutz HC Le Duff MJ Bhaurla SK

Aims

A contact patch to rim (CPR) distance of < 10 mm has been associated with edge-loading and excessive wear. However, not all arthroplasties with a low CPR distance show problems with wear. Therefore, CPR distance may not be the only variable affecting the post-operative metal ion concentrations.

Patients and Methods

We used multiple logistic regression to determine what variables differed between the patients who had high and low cobalt (CoS) and chromium (CrS) serum ion concentrations within a cohort of patients with low (< 10 mm) CPR distances. A total of 56 patients treated with unilateral hip resurfacing arthroplasty (HRA) had CoS and CrS ion studies performed more than one year after surgery. The mean age of the patients at the time of surgery was 51.7 years (29 to 70), with 38 women (68%) and 18 men (32%).


Bone & Joint 360
Vol. 6, Issue 3 | Pages 33 - 35
1 Jun 2017


Bone & Joint Research
Vol. 7, Issue 7 | Pages 468 - 475
1 Jul 2018
He Q Sun H Shu L Zhu Y Xie X Zhan Y Luo C

Objectives

Researchers continue to seek easier ways to evaluate the quality of bone and screen for osteoporosis and osteopenia. Until recently, radiographic images of various parts of the body, except the distal femur, have been reappraised in the light of dual-energy X-ray absorptiometry (DXA) findings. The incidence of osteoporotic fractures around the knee joint in the elderly continues to increase. The aim of this study was to propose two new radiographic parameters of the distal femur for the assessment of bone quality.

Methods

Anteroposterior radiographs of the knee and bone mineral density (BMD) and T-scores from DXA scans of 361 healthy patients were prospectively analyzed. The mean cortical bone thickness (CBTavg) and the distal femoral cortex index (DFCI) were the two parameters that were proposed and measured. Intra- and interobserver reliabilities were assessed. Correlations between the BMD and T-score and these parameters were investigated and their value in the diagnosis of osteoporosis and osteopenia was evaluated.


Bone & Joint 360
Vol. 6, Issue 3 | Pages 41 - 43
1 Jun 2017
Foy MA


Bone & Joint 360
Vol. 6, Issue 6 | Pages 14 - 16
1 Dec 2017


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 853 - 861
1 Jul 2018
Leunig M Hutmacher JE Ricciardi BF Impellizzeri FM Rüdiger HA Naal FD

Aims

The classical longitudinal incision used for the direct anterior approach (DAA) to the hip does not follow the tension lines of the skin and can lead to impaired wound healing and poor cosmesis. The purpose of this retrospective study was to determine the satisfaction with the scar, and functional and radiographic outcomes comparing the classic longitudinal incision with a modified skin crease ‘bikini’ when the DAA is used for total hip arthroplasty (THA).

Patients and Methods

A total of 964 patients (51% female; 59% longitudinal, 41% ‘bikini’) completed a follow-up questionnaire between two and four years postoperatively, including the Oxford Hip Score (OHS), the University of North Carolina ‘4P’ scar scale (UNC4P) and two items for assessing the aesthetic appearance of the scar and symptoms of numbness. The positioning of the components, rates of heterotopic ossification (HO) and rates of revision were assessed.


The Bone & Joint Journal
Vol. 100-B, Issue 1 | Pages 11 - 19
1 Jan 2018
Darrith B Courtney PM Della Valle CJ

Aims

Instability remains a challenging problem in both primary and revision total hip arthroplasty (THA). Dual mobility components confer increased stability, but there are concerns about the unique complications associated with these designs, as well as the long-term survivorship.

Materials and Methods

We performed a systematic review of all English language articles dealing with dual mobility THAs published between 2007 and 2016 in the MEDLINE and Embase electronic databases. A total of 54 articles met inclusion criteria for the final analysis of primary and revision dual mobility THAs and dual mobility THAs used in the treatment of fractures of the femoral neck. We analysed the survivorship and rates of aseptic loosening and of intraprosthetic and extra-articular dislocation.


Bone & Joint 360
Vol. 6, Issue 3 | Pages 2 - 6
1 Jun 2017
Das A Shivji F Ollivere BJ


The Bone & Joint Journal
Vol. 99-B, Issue 9 | Pages 1197 - 1203
1 Sep 2017
Laumonerie P Reina N Ancelin D Delclaux S Tibbo ME Bonnevialle N Mansat P

Aims

Radial head arthroplasty (RHA) may be used in the treatment of non-reconstructable radial head fractures. The aim of this study was to evaluate the mid-term clinical and radiographic results of RHA.

Patients and Methods

Between 2002 and 2014, 77 RHAs were implanted in 54 men and 23 women with either acute injuries (54) or with traumatic sequelae (23) of a fracture of the radial head. Four designs of RHA were used, including the Guepar (Small Bone Innovations (SBi)/Stryker; 36), Evolutive (Aston Medical; 24), rHead RECON (SBi/Stryker; ten) or rHead STANDARD (SBi/Stryker; 7) prostheses. The mean follow-up was 74.0 months (standard deviation (sd) 38.6; 24 to 141). The indication for further surgery, range of movement, mean Mayo Elbow Performance (MEP) score, quick Disabilities of the Arm, Shoulder and Hand (quickDASH) score, osteolysis and positioning of the implant were also assessed according to the design, and acute or delayed use.


Bone & Joint 360
Vol. 6, Issue 4 | Pages 2 - 7
1 Aug 2017
Titchener AG Tambe AA Clark DI


Bone & Joint Research
Vol. 7, Issue 6 | Pages 379 - 387
1 Jun 2018
Hansen L De Raedt S Jørgensen PB Mygind-Klavsen B Kaptein B Stilling M

Objectives

To validate the precision of digitally reconstructed radiograph (DRR) radiostereometric analysis (RSA) and the model-based method (MBM) RSA with respect to benchmark marker-based (MM) RSA for evaluation of kinematics in the native hip joint.

Methods

Seven human cadaveric hemipelves were CT scanned and bone models were segmented. Tantalum beads were placed in the pelvis and proximal femoral bone. RSA recordings of the hips were performed during flexion, adduction and internal rotation. Stereoradiographic recordings were all analyzed with DRR, MBM and MM. Migration results for the MBM and DRR with respect to MM were compared. Precision was assessed as systematic bias (mean difference) and random variation (Pitman’s test for equal variance).


Bone & Joint 360
Vol. 7, Issue 2 | Pages 23 - 25
1 Apr 2018