The routine use of patient reported outcome measures
(PROMs) in evaluating the outcome after arthroplasty by healthcare
organisations reflects a growing recognition of the importance of
patients’ perspectives in improving treatment. Although widely embraced
in the NHS, there are concerns that PROMs are being used beyond
their means due to a poor understanding of their limitations. This paper reviews some of the current challenges in using PROMs
to evaluate total knee arthroplasty. It highlights alternative methods
that have been used to improve the assessment of outcome. Cite this article:
The aims of this retrospective study were to
compare the mid-term outcomes following revision total knee replacement
(TKR) in 76 patients (81 knees) <
55 years of age with those
of a matched group of primary TKRs based on age, BMI, gender and
comorbid conditions. We report the activity levels, functional scores,
rates of revision and complications. Compared with patients undergoing
primary TKR, those undergoing revision TKR had less improvement
in the mean Knee Society function scores (8.14 (–55 to +60) Young patients undergoing revision TKR should be counselled that
they can expect somewhat less improvement and a higher risk of complications
than occur after primary TKR. Cite this article:
We scanned 25 left knees in healthy human subjects
using MRI. Multiplanar reconstruction software was used to take
measurements of the inferior and posterior facets of the femoral
condyles and the trochlea. A ‘basic circle’ can be defined which, in the sagittal plane,
fits the posterior and inferior facets of the lateral condyle, the
posterior facet of the medial condyle and the floor of the groove
of the trochlea. It also approximately fits both condyles in the
coronal plane (inferior facets) and the axial plane (posterior facets).
The circle fitting the inferior facet of the medial condyle in the
sagittal plane was consistently 35% larger than the other circles
and was termed the ‘medial inferior circle’. There were strong correlations
between the radii of the circles, the relative positions of the
centres of the condyles, the width of the condyles, the total knee
width and skeletal measurements including height. There was poor
correlation between the radii of the circles and the position of
the trochlea relative to the condyles. In summary, the condyles are approximately spherical except for
the inferior facet medially, which has a larger radius in the sagittal
plane. The size and position of the condyles are consistent and
change with the size of the person. However, the position of the
trochlea is variable even though its radius is similar to that of
the condyles. This information has implications for understanding
anterior knee pain and for the design of knee replacements. Cite this article:
Instability in flexion after total knee replacement
(TKR) typically occurs as a result of mismatched flexion and extension
gaps. The goals of this study were to identify factors leading to
instability in flexion, the degree of correction, determined radiologically,
required at revision surgery, and the subsequent clinical outcomes.
Between 2000 and 2010, 60 TKRs in 60 patients underwent revision
for instability in flexion associated with well-fixed components.
There were 33 women (55%) and 27 men (45%); their mean age was 65
years (43 to 82). Radiological measurements and the Knee Society
score (KSS) were used to assess outcome after revision surgery.
The mean follow-up was 3.6 years (2 to 9.8). Decreased condylar
offset (p <
0.001), distalisation of the joint line (p <
0.001)
and increased posterior tibial slope (p <
0.001) contributed
to instability in flexion and required correction at revision to regain
stability. The combined mean correction of posterior condylar offset
and joint line resection was 9.5 mm, and a mean of 5° of posterior
tibial slope was removed. At the most recent follow-up, there was
a significant improvement in the mean KSS for the knee and function
(both p <
0.001), no patient reported instability and no patient
underwent further surgery for instability. The following step-wise approach is recommended: reduction of
tibial slope, correction of malalignment, and improvement of condylar
offset. Additional joint line elevation is needed if the above steps
do not equalise the flexion and extension gaps. Cite this article:
We present detailed information about early morbidity
after aseptic revision knee replacement from a nationwide study.
All aseptic revision knee replacements undertaken between 1st October
2009 and 30th September 2011 were analysed using the Danish National
Patient Registry with additional information from the Danish Knee
Arthroplasty Registry. The 1218 revisions involving 1165 patients
were subdivided into total revisions, large partial revisions, partial
revisions and revisions of unicondylar replacements (UKR revisions).
The mean age was 65.0 years (27 to 94) and the median length of
hospital stay was four days (interquartile range: 3 to 5), with
a 90 days re-admission rate of 9.9%,
re-operation rate of 3.5% and mortality rate of 0.2%. The age ranges
of 51 to 55 years (p = 0.018), 76 to 80 years (p <
0.001) and ≥ 81
years (p <
0.001) were related to an increased risk of re-admission.
The age ranges of 76 to 80 years (p = 0.018) and the large partial
revision subgroup (p = 0.073) were related to an increased risk
of re-operation. The ages from 76 to 80 years (p <
0.001), age ≥ 81
years (p <
0.001) and surgical time >
120 min (p <
0.001)
were related to increased length of hospital stay, whereas the use
of a tourniquet (p = 0.008) and surgery in a low volume centre (p
= 0.013) were related to shorter length of stay. In conclusion, we found a similar incidence of early post-operative
morbidity after aseptic knee revisions as has been reported after
primary procedures. This suggests that a length of hospital stay ≤ four
days and discharge home at that time is safe following aseptic knee
revision surgery in Denmark. Cite this article:
The role of arthroscopy in the treatment of soft-tissue
injuries associated with proximal tibial fractures remains debatable.
Our hypothesis was that MRI over-diagnoses clinically relevant associated
soft-tissue injuries. This prospective study involved 50 consecutive
patients who underwent surgical treatment for a split-depression fracture
of the lateral tibial condyle (AO/OTA type B3.1). The mean age of
patients was 50 years (23 to 86) and 27 (54%) were female. All patients
had MRI and arthroscopy. Arthroscopy identified 12 tears of the
lateral meniscus, including eight bucket-handle tears that were
sutured and four that were resected, as well as six tears of the
medial meniscus, of which five were resected. Lateral meniscal injuries
were diagnosed on MRI in four of 12 patients, yielding an overall
sensitivity of 33% (95% confidence interval (CI) 11 to 65). Specificity
was 76% (95% CI 59 to 88), with nine tears diagnosed among 38 menisci
that did not contain a tear. MRI identified medial meniscal injuries
in four of six patients, yielding an overall sensitivity of 67%
(95% CI 24 to 94). Specificity was 66% (95% CI 50 to 79), with 15
tears diagnosed in 44 menisci that did not contain tears. MRI appears to offer only a marginal benefit as the specificity
and sensitivity for diagnosing meniscal injuries are poor in patients
with a fracture. There were fewer arthroscopically-confirmed associated
lesions than reported previously in MRI studies. Cite this article:
Total knee replacement (TKR) is an effective
method of treating end-stage arthritis of the knee. It is not, however,
a procedure without risk due to a number of factors, one of which
is diabetes mellitus. The purpose of this study was to estimate
the general prevalence of diabetes in patients about to undergo
primary TKR and to determine whether diabetes mellitus adversely
affects the outcome. We conducted a systematic review and meta-analysis
according to the Meta-analysis Of Observational Studies in Epidemiology
(MOOSE) guidelines. The Odds Ratio (OR) and mean difference (MD)
were used to represent the estimate of risk of a specific outcome.
Our results showed the prevalence of diabetes mellitus among patients
undergoing TKR was 12.2%. Patients with diabetes mellitus had an increased
risk of deep infection (OR = 1.61, 95% confidence interval (CI),
1.38 to 1.88), deep vein thrombosis (in Asia, OR = 2.57, 95% CI,
1.58 to 4.20), periprosthetic fracture (OR = 1.89, 95% CI, 1.04
to 3.45), aseptic loosening (OR = 9.36, 95% CI, 4.63 to 18.90),
and a poorer Knee Society function subscore (MD = -5.86, 95% CI,
-10.27 to -1.46). Surgeons should advise patients specifically about
these increased risks when obtaining informed consent and be meticulous about
their peri-operative care. Cite this article:
The aim of this study was to find anatomical
landmarks for rotational alignment of the tibial component in total knee
replacement (TKR) in a CT-based study. Pre-operative CT scanning
was performed on 94 South Korean patients (nine men, 85 women, 188
knees) with osteoarthritis of the knee joint prior to TKR. The tibial
anteroposterior (AP) axis was defined as a line perpendicular to
the femoral surgical transepicondylar axis and passing through the centre
of the posterior cruciate ligament (PCL). The angles between the
defined tibial AP axis and anatomical landmarks at various levels
of the tibia were measured. The mean values of the angles between
the defined tibial AP axis and the line connecting the anterior
border of the proximal third of the tibia to the centre of the PCL
was -0.2° (-17 to 14.1, Cite this article:
Although the vast majority of patients that undergo
total knee replacement have satisfactory outcomes with a generally
low complication rate, occasionally a patient will be encountered
that has had multiple failed surgeries, and now reaches a crossroad
as to whether limb salvage will be acceptable or not. Cite this article:
A national, multi-centre study was designed in
which a questionnaire quantifying the degree of patient satisfaction
and residual symptoms in patients following total knee replacement
(TKR) was administered by an independent, blinded third party survey
centre. A total of 90% of patients reported satisfaction with the
overall functioning of their knee, but 66% felt their knee to be
‘normal’, with the reported incidence of residual symptoms and functional
problems ranging from 33% to 54%. Female patients and patients from
low-income households had increased odds of reporting dissatisfaction.
Neither the use of contemporary implant designs (gender-specific,
high-flex, rotating platform) or custom cutting guides (CCG) with
a neutral mechanical axis target improved patient-perceived outcomes.
However, use of a CCG to perform a so-called kinematically aligned
TKR showed a trend towards more patients reporting their knee to
feel ‘normal’ when compared with a so called mechanically aligned
TKR This data shows a degree of dissatisfaction and residual symptoms
following TKR, and that several recent modifications in implant
design and surgical technique have not improved the current situation. Cite this article:
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
Total knee replacement (TKR) smart tibial trials
have load-bearing sensors which will show quantitative compartment
pressure values and femoral-tibial tracking patterns. Without smart
trials, surgeons rely on feel and visual estimation of imbalance
to determine if the knee is optimally balanced. Corrective soft-tissue
releases are performed with minimal feedback as to what and how
much should be released. The smart tibial trials demonstrate graphically
where and how much imbalance is present, so that incremental releases
can be performed. The smart tibial trials now also incorporate accelerometers
which demonstrate the axial alignment. This now allows the surgeon
the option to perform a slight recut of the tibia or femur to provide
soft-tissue balance without performing soft-tissue releases. Using
a smart tibial trial to assist with soft-tissue releases or bone
re-cuts, improved patient outcomes have been demonstrated at one
year in a multicentre study of 135 patients (135 knees). Cite this article:
The use of hinged implants in primary total knee
replacement (TKR) should be restricted to selected indications and mainly
for elderly patients. Potential indications for a rotating hinge
or pure hinge implant in primary TKR include: collateral ligament
insufficiency, severe varus or valgus deformity (>
20°) with necessary
relevant soft-tissue release, relevant bone loss including insertions
of collateral ligaments, gross flexion-extension gap imbalance,
ankylosis, or hyperlaxity. Although data reported in the literature
are inconsistent, clinical results depend on implant design, proper
technical use, and adequate indications. We present our experience
with a specific implant type that we have used for over 30 years
and which has given our elderly patients good mid-term results.
Because revision of implants with long cemented stems can be very
challenging, an effort should be made in the future to use shorter stems
in modular versions of hinged implants. Cite this article:
The aim of this study was to examine the functional
outcome at ten years following lateral closing wedge high tibial osteotomy
for medial compartment osteoarthritis of the knee and to define
pre-operative predictors of survival and determinants of functional
outcome. 164 consecutive patients underwent high tibial osteotomy between
2000 and 2002. A total of 100 patients (100 knees) met the inclusion
criteria and 95 were available for review at ten years. Data were
collected prospectively and included patient demographics, surgical
details, long leg alignment radiographs, Western Ontario and McMaster Universities
osteoarthritis index (WOMAC) and Knee Society scores (KSS) pre-operatively
and at five and ten years follow-up. At ten years, 21 patients had been revised at a mean of five
years. Overall Kaplan–Meier survival was 87% (95% confidence interval
(CI) 81 to 94) and 79% (95% CI 71 to 87) at five and ten years,
respectively. When compared with unrevised patients, those who had
been revised had significantly lower mean pre-operative WOMAC Scores
(47 (21 to 85) This study has shown that improved survival is associated with
age <
55 years, pre-operative WOMAC scores >
45 and, a BMI <
30. In patients over 55 years of age with adequate pre-operative
functional scores, survival can be good and functional outcomes
can be significantly better than their younger counterparts. We
recommend the routine use of pre-operative functional outcome scores
to guide decision-making when considering suitability for high tibial osteotomy. Cite this article:
The optimal timing of total knee replacement
(TKR) in patients with osteoarthritis, in relation to the severity
of disease, remains controversial. This prospective study was performed
to investigate the effect of the severity of osteoarthritis and
other commonly available pre- and post-operative clinical parameters
on the clinical outcome in a consecutive series of cemented TKRs.
A total of 176 patients who underwent unilateral TKR were included
in the study. Their mean age was 68 years (39 to 91), 63 (36%) were
male and 131 knees (74%) were classified as grade 4 on the Kellgren–Lawrence
osteoarthritis scale. A total of 154 patients (87.5%) returned for
clinical review 12 months post-operatively, at which time the outcome
was assessed using the Knee Society score. A low radiological severity of osteoarthritis was not associated
with pain 12 months post-operatively. However, it was significantly
associated with an inferior level of function (p = 0.007), implying
the need for increased focus on all possible reasons for pain in
the knee and the forms of conservative treatment which are available
for patients with lower radiological severity of osteoarthritis. Cite this article:
The relationship between post-operative bone
density and subsequent failure of total knee replacement (TKR) is
not known. This retrospective study aimed to determine the relationship
between bone density and failure, both overall and according to
failure mechanism. All 54 aseptic failures occurring in 50 patients
from 7760 consecutive primary cemented TKRs between 1983 and 2004
were matched with non-failing TKRs, and 47 failures in 44 patients
involved tibial failures with the matching characteristics of age
(65.1 for failed and 69.8 for non-failed), gender (70.2% female), diagnosis
(93.6% OA), date of operation, bilaterality, pre-operative alignment
(0.4 and 0.3 respectively), and body mass index (30.2 and 30.0 respectively).
In each case, the density of bone beneath the tibial component was assessed
at each follow-up interval using standardised, calibrated radiographs.
Failing knees were compared with controls both overall and, as a
subgroup analysis, by failure mechanism. Knees were compared with
controls using univariable linear regression. Significant and continuous elevation in tibial density was found
in knees that eventually failed by medial collapse (p <
0.001)
and progressive radiolucency (p <
0.001) compared with controls,
particularly in the medial region of the tibia. Knees failing due
to ligamentous instability demonstrated an initial decline in density
(p = 0.0152) followed by a non-decreasing density over time (p =
0.034 for equivalence). Non-failing knees reported a decline in
density similar to that reported previously using dual-energy x-ray
absorptiometry (DEXA). Differences between failing and non-failing
knees were observable as early as two months following surgery.
This tool may be used to identify patients at risk of failure following
TKR, but more validation work is needed. Cite this article:
Accurate, reproducible outcome measures are essential
for the evaluation of any orthopaedic procedure, in both clinical
practice and research. Commonly used patient-reported outcome measures (PROMs) have
drawbacks such as ‘floor’ and ‘ceiling’ effects, limitations of
worldwide adaptability and an inability to distinguish pain from
function. They are also unable to measure the true outcome of an
intervention rather than a patient’s perception of that outcome. Performance-based functional outcome tools may address these
problems. It is important that both clinicians and researchers are
aware of these measures when dealing with high-demand patients,
using a new intervention or implant, or testing a new rehabilitation
protocol. This article provides an overview of some of the clinically-validated
performance-based functional outcome tools used in the assessment
of patients undergoing hip and knee surgery. Cite this article:
Patellofemoral complications are common after
total knee replacement (TKR). Leaving the patellar unsurfaced after
TKR may lead to complications such as anterior knee pain, and re-operation
to surface it. Complications after patellar resurfacing include
patellar fracture, aseptic loosening, patellar instability, polyethylene
wear, patellar clunk and osteonecrosis. Historically, patellar complications
account for one of the larger proportions of causes of failure in
TKR, however, with contemporary implant designs, complication rates
have decreased. Most remaining failures relate to patellofemoral
tracking. Understanding the causes of patellofemoral maltracking
is essential to prevent these complications as well as manage them
when they occur. Cite this article:
We report the clinical outcome and findings at
second-look arthroscopy of 216 patients (mean age 25 years (11 to 58))
who underwent anterior cruciate ligament (ACL) reconstruction or
augmentation. There were 73 single-bundle ACL augmentations (44
female, 29 male), 82 double-bundle ACL reconstructions (35 female,
47 male), and 61 single-bundle ACL reconstructions (34 female, 27
male). In 94 of the 216 patients, proprioceptive function of the knee
was evaluated before and 12 months after surgery using the threshold
to detect passive motion test. Second-look arthroscopy showed significantly better synovial
coverage of the graft in the augmentation group (good: 60 (82%),
fair: 10 (14%), poor: 3 (4%)) than in the other groups (p = 0.039).
The mean side-to-side difference measured with a KT-2000 arthrometer
was 0.4 mm (-3.3 to 2.9) in the augmentation group, 0.9 mm (-3.2
to 3.5) in the double-bundle group, and 1.3 mm (-2.7 to 3.9) in
the single-bundle group: the result differed significantly between
the augmentation and single-bundle groups (p = 0 .013). No significant
difference in the Lysholm score or pivot-shift test was seen between
the three groups (p = 0.09 and 0.65, respectively). In patients
with good synovial coverage, three of the four measurements used
revealed significant improvement in proprioceptive function (p = 0.177,
0.020, 0.034, and 0.026). We conclude that ACL augmentation is a reasonable treatment option
for patients with favourable ACL remnants. Cite this article:
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method