The peri-prosthetic tissue response to wear debris
is complex and influenced by various factors including the size, area
and number of particles. We hypothesised that the ‘biologically
active area’ of all metal wear particles may predict the type of
peri-prosthetic tissue response. Peri-prosthetic tissue was sampled from 21 patients undergoing
revision of a small diameter metal-on-metal (MoM) total hip arthroplasty
(THA) for aseptic loosening. An enzymatic protocol was used for
tissue digestion and scanning electron microscope was used to characterise
particles. Equivalent circle diameters and particle areas were calculated.
Histomorphometric analyses were performed on all tissue specimens.
Aspirates of synovial fluid were collected for analysis of the cytokine
profile analysis, and compared with a control group of patients
undergoing primary THA (n = 11) and revision of a failed ceramic-on-polyethylene
arthroplasty (n = 6). The overall distribution of the size and area of the particles
in both lymphocyte and
non-lymphocyte-dominated responses were similar; however, the subgroup
with lymphocyte-dominated peri-prosthetic tissue responses had a
significantly larger total number of particles. 14 cytokines (interleukin (IL)-1ß, IL-2, IL-4, IL-5, IL-6, IL-10,
IL-13, IL-17, interferon (IFN)-γ, and IFN-gamma-inducible protein
10), chemokines (macrophage inflammatory protein (MIP)-1α and MIP-1ß),
and growth factors (granulocyte macrophage colony stimulating factor
(GM-CSF) and platelet derived growth factor) were detected at significantly higher
levels in patients with metal wear debris compared with the control
group. Significantly higher levels for IL-1ß, IL-5, IL-10 and GM-CSF
were found in the subgroup of tissues from failed MoM THAs with
a lymphocyte-dominated peri-prosthetic response compared with those
without this response. These results suggest that the ‘biologically active area’ predicts
the type of
peri-prosthetic tissue response. The cytokines IL-1ß, IL-5, IL-10,
and GM-CSF are associated with lymphocyte-dominated tissue responses
from failed small-diameter MoM THA. Cite this article:
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
Although atlantoaxial rotatory fixation (AARF) is a common cause
of torticollis in children, the diagnosis may be delayed. The condition
is characterised by a lack of rotation at the atlantoaxial joint
which becomes fixed in a rotated and subluxed position. The management of
children with a delayed presentation of this condition is controversial.
This is a retrospective study of a group of such children. Children who were admitted to two institutions between 1988 and
2014 with a diagnosis of AARF were included. We identified 12 children
(four boys, eight girls), with a mean age of 7.3 years (1.5 to 13.4),
in whom the duration of symptoms on presentation was at least four weeks
(four to 39). All were treated with halo traction followed by a
period of cervical immobilisation in a halo vest or a Minerva jacket.
We describe a simple modification to the halo traction that allows
the child to move their head whilst maintaining traction. The mean follow-up
was 59.6 weeks (24 to 156).Aims
Patients and Methods
As the number of younger and more active patients
treated with total knee arthroplasty (TKA) continues to increase,
consideration of better fixation as a means of improving implant
longevity is required. Cemented TKA remains the reference standard
with the largest body of evidence and the longest follow-up to support
its use. However, cementless TKA, may offer the opportunity of a
more bone-sparing procedure with long lasting biological fixation
to the bone. We undertook a review of the literature examining advances
of cementless TKA and the reported results. Cite this article:
The April 2015 Hip &
Pelvis Roundup360 looks at: Goal-directed fluid therapy in hip fracture; Liberal blood transfusion no benefit in the longer term; Repeated measures: increased accuracy or compounded errors?; Peri-acetabular osteotomy safer than perhaps thought?; Obesity and peri-acetabular osteotomy: poor bedfellows; Stress fracture in peri-acetabular osteotomy; Infection and tantalum implants; Highly crosslinked polyethylene really does work
Concerns have been raised that deformation of
acetabular shells may disrupt the assembly process of modular prostheses.
In this study we aimed to examine the effect that the strength of
bone has on the amount of deformation of the acetabular shell. The
hypothesis was that stronger bone would result in greater deformation.
A total of 17 acetabular shells were inserted into the acetabula
of eight cadavers, and deformation was measured using an optical
measuring system. Cores of bone from the femoral head were taken
from each cadaver and compressed using a materials testing machine.
The highest peak modulus and yield stress for each cadaver were used
to represent the strength of the bone and compared with the values
for the deformation and the surgeon’s subjective assessment of the
hardness of the bone. The mean deformation of the shell was 129
µm (3 to 340). No correlation was found between deformation and
either the maximum peak modulus (r² = 0.011, t = 0.426, p = 0.676) or
the yield stress (r² = 0.024, t = 0.614, p = 0.549) of the bone.
Although no correlation was found between the strength of the bone
and deformation, the values for the deformation observed could be
sufficient to disrupt the assembly process of modular acetabular
components. Cite this article:
Modular junctions are ubiquitous in contemporary hip arthroplasty. The head-trunnion junction is implicated in the failure of large diameter metal-on-metal (MoM) hips which are the currently the topic of one the largest legal actions in the history of orthopaedics (estimated costs are stated to exceed $4 billion). Several factors are known to influence the strength of these press-fit modular connections. However, the influence of different head sizes has not previously been investigated. The aim of the study was to establish whether the choice of head size influences the initial strength of the trunnion-head connection. Ti-6Al-4V trunnions (n = 60) and two different sizes of cobalt-chromium (Co-Cr) heads (28 mm and 36 mm; 30 of each size) were used in the study. Three different levels of assembly force were considered: 4 kN; 5 kN; and 6 kN (n = 10 each). The strength of the press-fit connection was subsequently evaluated by measuring the pull-off force required to break the connection. The statistical differences in pull-off force were examined using a Kruskal–Wallis test and two-sample Mann–Whitney U test. Finite element and analytical models were developed to understand the reasons for the experimentally observed differences.Objectives
Materials and Methods
Ankle replacements have improved significantly since the first reported attempt at resurfacing of the talar dome in 1962. We are now at a stage where ankle replacement offers a viable option in the treatment of end-stage ankle arthritis. As the procedure becomes more successful, it is important to reflect and review the current surgical outcomes. This allows us to guide our patients in the treatment of end-stage ankle arthritis. What is the better surgical treatment – arthrodesis or replacement?
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
Revision total hip arthroplasty (THA) is challenging
when there is severe loss of bone in the proximal femur. The purpose
of this study was to evaluate the clinical and radiographic outcomes
of revision THA in patients with severe proximal femoral bone loss
treated with a fluted, tapered, modular femoral component. Between
January 1998 and December 2004, 92 revision THAs were performed
in 92 patients using a single fluted, tapered, modular femoral stem
design. Pre-operative diagnoses included aseptic loosening, infection
and peri-prosthetic fracture. Bone loss was categorised pre-operatively
as Paprosky types III-IV, or Vancouver B3 in patients with a peri-prosthetic
fracture. The mean clinical follow-up was 6.4 years (2 to 12). A
total of 47 patients had peri-operative complications, 27 of whom
required further surgery. However, most of these further operations
involved retention of a well-fixed femoral stem, and 88/92 femoral
components (97%) remained Revision THA in patients with extensive proximal femoral bone
loss using the Link MP fluted, tapered, modular stem led to a high
rate of osseointegration of the stem at mid-term follow-up. Cite this article:
We conducted a case-control study to examine
the merit of silver-coated tumour prostheses. We reviewed 85 patients
with Agluna-treated (silver-coated) tumour implants treated between
2006 and 2011 and matched them with 85 control patients treated
between 2001 and 2011 with identical, but uncoated, tumour prostheses. In all, 106 men and 64 women with a mean age of 42.2 years (18.4
to 90.4) were included in the study. There were 50 primary reconstructions
(29.4%); 79 one-stage revisions (46.5%) and 41 two-stage revisions
for infection (24.1%). The overall post-operative infection rate of the silver-coated
group was 11.8% compared with 22.4% for the control group (p = 0.033,
chi-square test). A total of seven of the ten infected prostheses
in the silver-coated group were treated successfully with debridement,
antibiotics, and implant retention compared with only six of the
19 patients (31.6%) in the control group (p = 0.048, chi-square
test). Three patients in the silver-coated group (3.5%) and 13 controls
(15.3%) had chronic periprosthetic infection (p = 0.009, chi-square
test). The overall success rates in controlling infection by two-stage
revision in the silver-coated group was 85% (17/20) compared with
57.1% (12/21) in the control group (p = 0.05, chi-square test).
The Agluna-treated endoprostheses were associated with a lower rate
of early periprosthetic infection. These silver-treated implants
were particularly useful in two-stage revisions for infection and
in those patients with incidental positive cultures at the time
of implantation of the prosthesis. Debridement with antibiotic treatment and retention of the implant
appeared to be more successful with silver-coated implants. Cite this article:
We report the clinical and radiological outcomes
of a series of contemporary cementless ceramic-on-ceramic total hip
replacements (THRs) at ten years in patients aged ≤ 55 years of
age. Pre- and post-operative activity levels are described. A total
of 120 consecutive ceramic cementless THRs were performed at a single
centre in 110 patients from 1997 to 1999. The mean age of the patients
at operation was 45 years (20 to 55). At ten years, four patients
had died and six were lost to follow-up, comprising ten hips. The
mean post-operative Harris hip score was 94.7 (55 to 100). Radiological
analysis was undertaken in 90 available THRs of the surviving 106
hips at final review: all had evidence of stable bony ingrowth,
with no cases of osteolysis. Wear was undetectable. There were four
revisions. The survival for both components with revision for any
cause as an endpoint was 96.5% (95% confidence interval 94.5 to
98.7). The mean modified University of California, Los Angeles activity
level rose from a mean of 6.4 (4 to 10) pre-operatively to 9.0 (6
to 10) at the ten-year post-operative period. Alumina ceramic-on-ceramic bearings in cementless primary THR
in this series have resulted in good clinical and radiological outcomes
with undetectable rates of wear and excellent function in the demanding
younger patient group at ten years. Cite this article:
The April 2013 Spine Roundup360 looks at: smuggling spinal implants; local bone graft and PLIF; predicting disability with slipped discs; mortality and spinal surgery; spondyloarthropathy; brachytherapy; and fibrin mesh and BMP.
The primary aim of this study was to analyse the position of
the acetabular and femoral components in total hip arthroplasty
undertaken using an anterior surgical approach. In a prospective, single centre study, we used the EOS imaging
system to analyse the position of components following THA performed
via the anterior approach in 102 patients (103 hips) with a mean
age of 64.7 years (Aims
Patients and Methods
An uncemented hemispherical acetabular component
is the mainstay of acetabular revision and gives excellent long-term
results. Occasionally, the degree of acetabular bone loss means that a
hemispherical component will be unstable when sited in the correct
anatomical location or there is minimal bleeding host bone left
for biological fixation. On these occasions an alternative method
of reconstruction has to be used. A major column structural allograft has been shown to restore
the deficient bone stock to some degree, but it needs to be off-loaded
with a reconstruction cage to prevent collapse of the graft. The
use of porous metal augments is a promising method of overcoming
some of the problems associated with structural allograft. If the defect
is large, the augment needs to be protected by a cage to allow ingrowth
to occur. Cup-cage reconstruction is an effective method of treating
chronic pelvic discontinuity and large contained or uncontained
bone defects. This paper presents the indications, surgical techniques and
outcomes of various methods which use acetabular reconstruction
cages for revision total hip arthroplasty. Cite this article:
To employ a simple and fast method to evaluate those patients with neurological deficits and misplaced screws in relatively safe lumbosacral spine, and to determine if it is necessary to undertake revision surgery. A total of 316 patients were treated by fixation of lumbar and lumbosacral transpedicle screws at our institution from January 2011 to December 2012. We designed the criteria for post-operative revision scores of pedicle screw malpositioning (PRSPSM) in the lumbosacral canal. We recommend the revision of the misplaced pedicle screw in patients with PRSPSM = 5′ as early as possible. However, patients with PRSPSM < 5′ need to follow the next consecutive assessment procedures. A total of 15 patients were included according to at least three-stage follow-up.Objectives
Methods
Peri-acetabular tumour resections and their subsequent
reconstruction are among the most challenging procedures in orthopaedic
oncology. Despite the fact that a number of different pelvic endoprostheses
have been introduced, rates of complication remain high and long-term
results are mostly lacking. In this retrospective study, we aimed to evaluate the outcome
of reconstructing a peri-acetabular defect with a pedestal cup endoprosthesis
after a type 2 or type 2/3 internal hemipelvectomy. A total of 19 patients (11M:8F) with a mean age of 48 years (14
to 72) were included, most of whom had been treated for a primary
bone tumour (n = 16) between 2003 and 2009. After a mean follow-up
of 39 months (28 days to 8.7 years) seven patients had died. After
a mean follow-up of 7.9 years (4.3 to 10.5), 12 patients were alive,
of whom 11 were disease-free. Complications occurred in 15 patients.
Three had recurrent dislocations and three experienced aseptic loosening.
There were no mechanical failures. Infection occurred in nine patients,
six of whom required removal of the prosthesis. Two patients underwent
hindquarter amputation for local recurrence. The implant survival rate at five years was 50% for all reasons,
and 61% for non-oncological reasons. The mean Musculoskeletal Tumor
Society score at final follow-up was 49% (13 to 87). Based on these poor results, we advise caution if using the pedestal
cup for reconstruction of a peri-acetabular tumour resection. Cite this article:
The management of spinal deformity in children
with univentricular cardiac pathology poses significant challenges to
the surgical and anaesthetic teams. To date, only posterior instrumented
fusion techniques have been used in these children and these are
associated with a high rate of complications. We reviewed our experience
of both growing rod instrumentation and posterior instrumented fusion
in children with a univentricular circulation. Six children underwent spinal corrective surgery, two with cavopulmonary
shunts and four following completion of a Fontan procedure. Three
underwent growing rod instrumentation, two had a posterior fusion
and one had spinal growth arrest. There were no complications following
surgery, and the children undergoing growing rod instrumentation
were successfully lengthened. We noted a trend for greater blood
loss and haemodynamic instability in those whose surgery was undertaken
following completion of a Fontan procedure. At a median follow-up
of 87.6 months (interquartile range (IQR) 62.9 to 96.5) the median
correction of deformity was 24.2% (64.5° (IQR 46° to 80°) We believe that early surgical intervention with growing rod
instrumentation systems allows staged correction of the spinal deformity
and reduces the haemodynamic insult to these physiologically compromised
children. Due to the haemodynamic changes that occur with the completed
Fontan circulation, the initial scoliosis surgery should ideally
be undertaken when in the cavopulmonary shunt stage. Cite this article:
Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface. Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer. The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no adverse effects identified from either the BONIT-HA coating or the DOTIZE surface treatment.
The aim of this randomised controlled study was
to compare functional and radiological outcomes between modern cemented
and uncemented hydroxyapatite coated stems after one year in patients
treated surgically for a fracture of the femoral neck. A total of
141 patients aged >
65 years were included. Patients were randomised
to be treated with a cemented Exeter stem or an uncemented Bimetric
stem. The patients were reviewed at four and 12 months. The cemented group performed better than the uncemented group
for the Harris hip score (78 In conclusion, our data do not support the use of an uncemented
hydroxyapatite coated stem for the treatment of displaced fractures
of the femoral neck in the elderly. Cite this article: