Results 461 - 480 of 1391
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 2 | Pages 236 - 241
1 Mar 1995
Kim Y Oh J Oh S

We report the incidence of osteolysis in the femur, tibia, and patella of 44 consecutive patients (60 knees) who were followed for more than seven years after cementless knee arthroplasty with a Porous-Coated Anatomic prosthesis. The average age of the patients was 56.5 years (17 to 73); the operative diagnosis was osteoarthritis (33 knees), rheumatoid arthritis (17), tuberculous arthritis (7) and post-traumatic arthritis (3). All patellae were resurfaced. No femoral or tibial component was loose at the final follow-up examination. Thirty patellar components were loose of which six had been revised. Radiographs revealed osteolysis in 90% of the tibial plateaux and in 80% of the 30 intact patellar prostheses. No osteolysis was seen around any femoral component. In 50 knees (83%) the average wear of the polyethylene liner was 2.5 mm in the medial compartment and 1.7 mm in the lateral compartment. Four of 60 knees (6.7%) were revised for complete wear of the polyethylene liner of the tibial component. Fixation of the tibial and patellar components without cement fails to seal the interface between bone and prosthesis and allows the migration of polyethylene particulate debris which causes osteolysis


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 500 - 507
1 Apr 2010
Lu Y Huang C Chang T Ho F Cheng C Huang C

Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R. 2. = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 522 - 529
1 Mar 2021
Nichol T Callaghan J Townsend R Stockley I Hatton PV Le Maitre C Smith TJ Akid R

Aims

The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel coating that is capable of a controlled antibiotic release for cementless hydroxyapatite (HA)-coated titanium orthopaedic prostheses.

Methods

Coatings containing gentamicin at a concentration of 1.25% weight/volume (wt/vol), similar to that found in commercially available antibiotic-loaded bone cement, were prepared and tested in the laboratory for: kinetics of antibiotic release; activity against planktonic and biofilm bacterial cultures; biocompatibility with cultured mammalian cells; and physical bonding to the material (n = 3 in all tests). The sol-gel coatings and controls were then tested in vivo in a small animal healing model (four materials tested; n = 6 per material), and applied to the surface of commercially pure HA-coated titanium rods.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 423 - 429
1 Mar 2021
Diez-Escudero A Hailer NP

Periprosthetic joint infection (PJI) is one of the most dreaded complications after arthroplasty surgery; thus numerous approaches have been undertaken to equip metal surfaces with antibacterial properties. Due to its antimicrobial effects, silver is a promising coating for metallic surfaces, and several types of silver-coated arthroplasty implants are in clinical use today. However, silver can also exert toxic effects on eukaryotic cells both in the immediate vicinity of the coated implants and systemically. In most clinically-used implants, silver coatings are applied on bulk components that are not in direct contact with bone, such as in partial or total long bone arthroplasties used in tumour or complex revision surgery. These implants differ considerably in the coating method, total silver content, and silver release rates. Safety issues, such as the occurrence of argyria, have been a cause for concern, and the efficacy of silver coatings in terms of preventing PJI is also controversial. The application of silver coatings is uncommon on parts of implants intended for cementless fixation in host bone, but this option might be highly desirable since the modification of implant surfaces in order to improve osteoconductivity can also increase bacterial adhesion. Therefore, an optimal silver content that inhibits bacterial colonization while maintaining osteoconductivity is crucial if silver were to be applied as a coating on parts intended for bone contact. This review summarizes the different methods used to apply silver coatings to arthroplasty components, with a focus on the amount and duration of silver release from the different coatings; the available experience with silver-coated implants that are in clinical use today; and future strategies to balance the effects of silver on bacteria and eukaryotic cells, and to develop silver-coated titanium components suitable for bone ingrowth.

Cite this article: Bone Joint J 2021;103-B(3):423–429.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 535 - 537
1 Apr 2007
Evangelista GT Fulkerson E Kummer F Di Cesare PE

During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly recommended that patients with this type of prosthetic head be carefully monitored after a dislocation


Bone & Joint Research
Vol. 2, Issue 8 | Pages 140 - 148
1 Aug 2013
Gauthier L Dinh L Beaulé PE

Objectives. To quantify and compare peri-acetabular bone mineral density (BMD) between a monoblock acetabular component using a metal-on-metal (MoM) bearing and a modular titanium shell with a polyethylene (PE) insert. The secondary outcome was to measure patient-reported clinical function. Methods. A total of 50 patients (25 per group) were randomised to MoM or metal-on-polyethlene (MoP). There were 27 women (11 MoM) and 23 men (14 MoM) with a mean age of 61.6 years (47.7 to 73.2). Measurements of peri-prosthetic acetabular and contralateral hip (covariate) BMD were performed at baseline and at one and two years’ follow-up. The Western Ontario and McMaster Universities osteoarthritis index (WOMAC), University of California, Los Angeles (UCLA) activity score, Harris hip score, and RAND-36 were also completed at these intervals. Results. At two years, only zone 1 showed a loss in BMD (-2.5%) in MoM group compared with a gain in the MoP group (+2.2%). Zone 2 showed loss in both groups (-2.2% for MoM; -3.9% for MoP) and zones 3 and 4 a gain in both groups (+0.1% for MoM; +3.3% for MoP). No other between-group differences were detected. When adjusting for BMD of the contralateral hip, no differences in BMD were observed. The only significant differences in functional scores at two years were higher UCLA activity (7.3 (. sd. 1.2) vs 6.1 (. sd. 1.5); p = 0.01) and RAND-36 physical function (82.1 (. sd. 13.0) vs 64.5 (. sd. 26.4); p = 0.02) for MoM bearings versus MoP. One revision was performed in the MoM group, for aseptic acetabular loosening at 11 months. Conclusions. When controlling for systemic BMD, there were no significant differences between MoM and MoP groups in peri-acetabular BMD. However, increasing reports of adverse tissue reactions with large head MoM THR have restricted the use of the monoblock acetabular component to resurfacing only


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 67 - 69
1 Nov 2013
Brooks PJ

Dislocation is one of the most common causes of patient and surgeon dissatisfaction following hip replacement and to treat it, the causes must first be understood. Patient factors include age greater than 70 years, medical comorbidities, female gender, ligamentous laxity, revision surgery, issues with the abductors, and patient education. Surgeon factors include the annual quantity of procedures and experience, the surgical approach, adequate restoration of femoral offset and leg length, component position, and soft-tissue or bony impingement. Implant factors include the design of the head and neck region, and so-called skirts on longer neck lengths. There should be offset choices available in order to restore soft-tissue tension. Lipped liners aid in gaining stability, yet if improperly placed may result in impingement and dislocation. Late dislocation may result from polyethylene wear, soft-tissue destruction, trochanteric or abductor disruption and weakness, or infection. Understanding the causes of hip dislocation facilitates prevention in a majority of instances. Proper pre-operative planning includes the identification of patients with a high offset in whom inadequate restoration of offset will reduce soft-tissue tension and abductor efficiency. Component position must be accurate to achieve stability without impingement. Finally, patient education cannot be over-emphasised, as most dislocations occur early, and are preventable with proper instructions. Cite this article: Bone Joint J 2013;95-B, Supple A:67–9


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 3 - 10
1 Jul 2020
Sosa BR Niu Y Turajane K Staats K Suhardi V Carli A Fischetti V Bostrom M Yang X

Aims

Current treatments of prosthetic joint infection (PJI) are minimally effective against Staphylococcus aureus biofilm. A murine PJI model of debridement, antibiotics, and implant retention (DAIR) was used to test the hypothesis that PlySs2, a bacteriophage-derived lysin, can target S. aureus biofilm and address the unique challenges presented in this periprosthetic environment.

Methods

The ability of PlySs2 and vancomycin to kill biofilm and colony-forming units (CFUs) on orthopaedic implants were compared using in vitro models. An in vivo murine PJI model of DAIR was used to assess the efficacy of a combination of PlySs2 and vancomycin on periprosthetic bacterial load.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 327 - 332
1 Mar 2009
García-Rey E García-Cimbrelo E Cordero-Ampuero J

We reviewed 111 hemispherical Duraloc series-500 acetabular components with a minimum follow-up of 12 years. The mean clinical and radiological follow-up was 13.4 years (12 to 15). A Profile hydroxyapatite-coated anatomical femoral component was used in each case. Six patients had a late dislocation, for whom the polyethylene liner was exchanged. Each acetabular component was well fixed and all femoral components showed signs of bone ingrowth. The mean rate of femoral head penetration was 0.10 mm/year (0.021 to 0.481). The probability of not developing femoral cortical hypertrophy and proximal osteopenia by 12 years was 80.2% (95% confidence interval, 72.7 to 87.6) and 77.5% (95% confidence interval, 69.7 to 85.2), respectively. Despite these good clinical results, further follow-up is needed to determine whether these prostheses will loosen with time


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1289 - 1296
1 Oct 2020
Amstutz HC Le Duff M

Aims

Hip resurfacing arthroplasty (HRA) is typically indicated for young and active patients. Due to the longevity of arthroplasty, these patients are likely to undergo revision surgery during their lifetime. There is a paucity of information on the long-term outcome of revision surgeries performed after failed HRA. The aim of our study was to provide survivorship data as well as clinical scores after HRA revisions.

Methods

A total of 42 patients (43 hips) were revised after HRA at our centre to a variety of devices, including four HRA and 39 total hip arthroplasties (THAs). In addition to perioperative complications, University of California, Los Angeles (UCLA) hip scores and 12-Item Short-Form Health Survey questionnaire (SF-12) quality of life scores were collected at follow-up visits after the primary HRA and after revision surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 6 | Pages 770 - 776
1 Jun 2010
Sakai T Ohzono K Nishii T Miki H Takao M Sugano N

We compared a modular neck system with a non-modular system in a cementless anatomical total hip replacement (THR). Each group consisted of 74 hips with developmental hip dysplasia. Both groups had the same cementless acetabular component and the same articulation, which consisted of a conventional polyethylene liner and a 28 mm alumina head. The mean follow-up was 14.5 years (13 to 15), at which point there were significant differences in the mean total Harris hip score (modular/non-modular: 98.6 (64 to 100)/93.8 (68 to 100)), the mean range of abduction (32° (15° to 40°)/28 (0° to 40°)), use of a 10° elevated liner (31%/100%), the incidence of osteolysis (27%/79.7%) and the incidence of equal leg lengths (≥ 6 mm, 92%/61%). There was no disassociation or fracture of the modular neck. The modular system reduces the need for an elevated liner, thereby reducing the incidence of osteolysis. It gives a better range of movement and allows the surgeon to make an accurate adjustment of leg length


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1515 - 1521
1 Nov 2010
Clauss M Ilchmann T Zimmermann P Ochsner PE

The aim of this study was to obtain detailed long-term data on the cement-bone interface in patients with cemented stems, implanted using the constrained fixation technique. A total of eight stems were removed together with adjacent bone during post-mortem examinations of patients with well-functioning prostheses. Specimens were cut at four defined levels, contact radiographs were obtained for each level, and slices were prepared for histological analysis. Clinical data, clinical radiographs, contact radiographs and histological samples were examined for signs of loosening and remodelling. The mean radiological follow-up was 9.6 years and all stems were well-fixed, based on clinical and radiological criteria. Contact radiographs revealed an incomplete cement mantle but a complete filling of the medullary canal for all implants. Various amounts of polyethylene particles were evident at the cement-bone interface of seven stems, with no accompanying inflammatory reaction. Cortical atrophy and the formation of an ‘inner cortex’ were confirmed in six of eight stems by contact radiographs and histology, but were only visible on two clinical radiographs. Our results confirm that a complete cement mantle is not essential for the survival of Müller straight stems into the mid term, and support our hypothesis that no benefit to long-term survival can be expected from modern cementing techniques


Bone & Joint Research
Vol. 9, Issue 11 | Pages 761 - 767
1 Nov 2020
Hada M Mizu-uchi H Okazaki K Murakami K Kaneko T Higaki H Nakashima Y

Aims

This study aims to investigate the effects of posterior tibial slope (PTS) on knee kinematics involved in the post-cam mechanism in bi-cruciate stabilized (BCS) total knee arthroplasty (TKA) using computer simulation.

Methods

In total, 11 different PTS (0° to 10°) values were simulated to evaluate the effect of PTS on anterior post-cam contact conditions and knee kinematics in BCS TKA during weight-bearing stair climbing (from 86° to 6° of knee flexion). Knee kinematics were expressed as the lowest points of the medial and lateral femoral condyles on the surface of the tibial insert, and the anteroposterior translation of the femoral component relative to the tibial insert.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 1 | Pages 114 - 115
1 Jan 2005
Dunstan E Tilley S Briggs TWR Cannon SR

We report the case of a 12-year-old boy with polyostotic fibrous dysplasia of the upper femur in whom a massive customised polyethylene prosthesis functioned successfully for more than fifty years


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 360 - 365
1 Feb 2021
Mahendraraj KA Shields MV Grubhofer F Golenbock SW Jawa A

Aims

Existing literature indicates that inferiorly inclined glenoid baseplates following reverse total shoulder arthroplasty (RSA) produce better outcomes compared to superiorly inclined baseplates. We aim to compare clinical outcomes for RSAs with superiorly and neutrally/inferiorly inclined lateralized glenospheres.

Methods

We retrospectively reviewed 154 consecutive patients undergoing RSA between July 2015 and July 2017 by one single-fellowship trained surgeon (AJ). Two raters (KAM and MVS) independently measured glenoid inclination in preoperative and minimum two year follow-up radiographs (anteroposterior/Grashey) using the RSA angle. Inclination was then compared to patient-reported outcomes, range of motion (ROM), and independently assessed degree of scapular notching and staging of heterotopic ossification at two year follow-up.


Bone & Joint Research
Vol. 9, Issue 9 | Pages 593 - 600
1 Sep 2020
Lee J Koh Y Kim PS Kang KW Kwak YH Kang K

Aims

Unicompartmental knee arthroplasty (UKA) has become a popular method of treating knee localized osteoarthritis (OA). Additionally, the posterior cruciate ligament (PCL) is essential to maintaining the physiological kinematics and functions of the knee joint. Considering these factors, the purpose of this study was to investigate the biomechanical effects on PCL-deficient knees in medial UKA.

Methods

Computational simulations of five subject-specific models were performed for intact and PCL-deficient UKA with tibial slopes. Anteroposterior (AP) kinematics and contact stresses of the patellofemoral (PF) joint and the articular cartilage were evaluated under the deep-knee-bend condition.


The Bone & Joint Journal
Vol. 103-B, Issue 2 | Pages 338 - 346
1 Feb 2021
Khow YZ Liow MHL Lee M Chen JY Lo NN Yeo SJ

Aims

This study aimed to identify the tibial component and femoral component coronal angles (TCCAs and FCCAs), which concomitantly are associated with the best outcomes and survivorship in a cohort of fixed-bearing, cemented, medial unicompartmental knee arthroplasties (UKAs). We also investigated the potential two-way interactions between the TCCA and FCCA.

Methods

Prospectively collected registry data involving 264 UKAs from a single institution were analyzed. The TCCAs and FCCAs were measured on postoperative radiographs and absolute angles were analyzed. Clinical assessment at six months, two years, and ten years was undertaken using the Knee Society Knee score (KSKS) and Knee Society Function score (KSFS), the Oxford Knee Score (OKS), the 36-Item Short-Form Health Survey questionnaire (SF-36), and range of motion (ROM). Fulfilment of expectations and satisfaction was also recorded. Implant survivorship was reviewed at a mean follow-up of 14 years (12 to 16). Multivariate regression models included covariates, TCCA, FCCA, and two-way interactions between them. Partial residual graphs were generated to identify angles associated with the best outcomes. Kaplan-Meier analysis was used to compare implant survivorship between groups.


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1128 - 1135
14 Sep 2020
Khoshbin A Haddad FS Ward S O hEireamhoin S Wu J Nherera L Atrey A

Aims

The rate of dislocation when traditional single bearing implants are used in revision total hip arthroplasty (THA) has been reported to be between 8% and 10%. The use of dual mobility bearings can reduce this risk to between 0.5% and 2%. Dual mobility bearings are more expensive, and it is not clear if the additional clinical benefits constitute value for money for the payers. We aimed to estimate the cost-effectiveness of dual mobility compared with single bearings for patients undergoing revision THA.

Methods

We developed a Markov model to estimate the expected cost and benefits of dual mobility compared with single bearing implants in patients undergoing revision THA. The rates of revision and further revision were calculated from the National Joint Registry of England and Wales, while rates of transition from one health state to another were estimated from the literature, and the data were stratified by sex and age. Implant and healthcare costs were estimated from local procurement prices and national tariffs. Quality-adjusted life-years (QALYs) were calculated using published utility estimates for patients undergoing THA.


Bone & Joint Open
Vol. 2, Issue 1 | Pages 33 - 39
14 Jan 2021
McLaughlin JR Lee KR Johnson MA

Aims

We present the clinical and radiological results at a minimum follow-up of 20 years using a second-generation uncemented total hip arthroplasty (THA). These results are compared to our previously published results using a first-generation hip arthroplasty followed for 20 years.

Methods

A total of 62 uncemented THAs in 60 patients were performed between 1993 and 1994. The titanium femoral component used in all cases was a Taperloc with a reduced distal stem. The acetabular component was a fully porous coated threaded hemispheric titanium shell (T-Tap ST). The outcome of every femoral and acetabular component with regard to retention or revision was determined for all 62 THAs. Complete clinical follow-up at a minimum of 20 years was obtained on every living patient. Radiological follow-up was obtained on all but one.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1319 - 1323
3 Oct 2020
Khatun F Gill DF Atrey A Porteous M

Aims

We present the results, in terms of survival, clinical outcome, and radiological appearance at 20 years, in a cohort of 225 cemented Exeter Universal femoral components (Stryker, Newbury, UK) implanted in 207 patients, at a district general hospital.

Methods

All patients in this study had a total hip arthroplasty (THA) using an Exeter Universal femoral component with a cemented (n = 215) or cementless (n = 10) acetabular component. Clinical and radiological data were collected prospectively at one year, five years, and every five years thereafter. Patients lost to radiological and clinical follow-up (five) were cross-referenced with National Joint Registry (NJR) data and general practitioner (GP) records to assess whether they had undergone revision for any reason.