The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 319 - 323
1 Mar 2008
Moonot P Singh PJ Cronin MD Kalairajah YE Kavanagh TG Field RE

Hip resurfacing is a bone-conserving procedure with respect to proximal femoral resection, but there is debate in the literature as to whether the same holds true for the acetabulum. We have investigated whether the Birmingham hip resurfacing conserves acetabular bone.

Between 1998 and 2005, 500 Birmingham hip resurfacings were performed by two surgeons. Between 1996 and 2005 they undertook 700 primary hip replacements, with an uncemented acetabular component. These patients formed the clinical material to compare acetabular component sizing. The Birmingham hip resurfacing group comprised 350 hips in men and 150 hips in women. The uncemented total hip replacement group comprised 236 hips in men and 464 hips in women. Age- and gender-matched analysis of a cohort of patients for the sizes of the acetabular components required for the two types of replacement was also undertaken. Additionally, an analysis of the sizes of the components used by each surgeon was performed.

For age-matched women, the mean outside diameter of the Birmingham hip resurfacing acetabular components was 2.03 mm less than that of the acetabular components in the uncemented total hip replacements (p < 0.0001). In similarly matched men there was no significant difference (p = 0.77). A significant difference was also found between the size of acetabular components used by the two surgeons for Birmingham hip resurfacing for both men (p = 0.0015) and women (p = 0.001). In contrast, no significant difference was found between the size of acetabular components used by the two surgeons for uncemented total hip replacement in either men or women (p = 0.06 and p = 0.14, respectively). This suggests that variations in acetabular preparation also influence acetabular component size in hip resurfacing.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 434 - 440
1 Apr 2009
Dall GF Ohly NE Ballantyne JA Brenkel IJ

We analysed which pre-operative factors could be used to predict the length of in-patient stay following unilateral primary total hip replacement undertaken for osteoarthritis. Data were collected prospectively from 2302 patients undergoing primary total hip replacement over a nine-year period. The relationships between the various pre-operative factors and length of stay were studied separately using either Student’s t-test or Pearson’s correlation, and then subjected to multiple linear regression analysis.

The mean length of stay was 8.1 days (median 7; 3 to 58). After adjusting for the effects of other pre-operative factors, younger age, male gender, higher combined Harris hip function and activity score, higher general health perception dimension of the Short-Form 36 score, and non-steroidal anti-inflammatory drug use were all found to be significantly associated with a reduced length of stay.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 725 - 729
1 Jun 2009
Livesey C Wylde V Descamps S Estela CM Bannister GC Learmonth ID Blom AW

We undertook a randomised controlled trial to compare the outcomes of skin adhesive and staples for skin closure in total hip replacement. The primary outcome was the cosmetic appearance of the scar at three months using a surgeon-rated visual analogue scale. In all, 90 patients were randomised to skin closure using either skin adhesive (n = 45) or staples (n = 45). Data on demographics, surgical details, infection and oozing were collected during the in-patient stay. Further data on complications, patient satisfaction and evaluation of cosmesis were collected at three-month follow-up, and a photograph of the scar was taken. An orthopaedic and a plastic surgeon independently evaluated the cosmetic appearance of the scars from the photographs. No significant difference was found between groups in the cosmetic appearance of scars at three months (p = 0.172), the occurrence of complications (p = 0.3), or patient satisfaction (p = 0.42). Staples were quicker and easier to use than skin adhesive and also less expensive. Skin adhesive and surgical staples are both effective skin closure methods in total hip replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 34 - 38
1 Jan 2011
Charity JAF Tsiridis E Sheeraz A Howell JR Hubble MJW Timperley AJ Gie GA

We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed.

Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 851 - 854
1 Jul 2009
Timperley AJ Whitehouse SL

Recently the National Patient Safety Agency in the United Kingdom published a report entitled “Mitigating surgical risk in patients undergoing hip arthroplasty for fractures of the proximal femur”. A total of 26 deaths had been reported to them when cement was used at hemiarthroplasty between October 2003 and October 2008. This paper considers the evidence for using cement fixation of a hemiarthroplasty in the treatment of hip fractures.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1052 - 1056
1 Aug 2013
Lampropoulou-Adamidou K Georgiades G Vlamis J Hartofilakidis G

We evaluated the outcome of 41 consecutive Charnley low-friction arthroplasties (LFAs) performed by a single surgeon in 28 patients aged ≤ 35 years at operation between 23 and 36 years previously. There were 20 women and eight men with a mean age of 32 years (23 to 35) at surgery. Two patients (three hips) were lost to follow-up at 12 and 17 years post-operatively, respectively, and one patient (one hip) died at 13 years post-operatively. These patients were excluded from the final evaluation. The survival rate of the acetabular components was 92.7% (95% confidence interval (CI) 88.7 to 96.7) at ten years, 67.1% (95% CI 59.75 to 74.45) at 20 years and 53.2% (95% CI 45.3 to 61.1) at 25 years. For the femoral component the survival was 95.1% (95% CI 91.8 to 98.5) at ten years, 77.1% (95% CI 73.9 to 80.3) at 20 years and 68.2% (95% CI 60.7 to 75.8) at 25 years. The results indicate that the Charnley LFA remains a reasonable choice in the treatment of young patients and can serve for comparison with newer techniques and implants.

Cite this article: Bone Joint J 2013;95-B:1052–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 881 - 885
1 Jul 2011
Cobb JP Davda K Ahmad A Harris SJ Masjedi M Hart AJ

Large-head metal-on-metal total hip replacement has a failure rate of almost 8% at five years, three times the revision rate of conventional hip replacement. Unexplained pain remains a feature of this type of arthroplasty.

All designs of the femoral component of large-head metal-on-metal total hip replacements share a unique characteristic: a subtended angle of 120° defining the proportion of a sphere that the head represents. Using MRI, we measured the contact area of the iliopsoas tendon on the femoral head in sagittal reconstruction of 20 hips of patients with symptomatic femoroacetabular impingement. We also measured the articular extent of the femoral head on 40 normal hips and ten with cam-type deformities. Finally, we performed virtual hip resurfacing on normal and cam-type hips, avoiding overhang of the metal rim inferomedially.

The articular surface of the femoral head has a subtended angle of 120° anteriorly and posteriorly, but only 100° medially. Virtual surgery in a normally shaped femoral head showed a 20° skirt of metal protruding medially where iliopsoas articulates.

The excessive extent of the large-diameter femoral components may cause iliopsoas impingement independently of the acetabular component. This may be the cause of postoperative pain with these implants.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 7 - 9
1 Nov 2014
Berend ME Berend KR Lombardi Jr AV

Over the past 30 years there have been many improvements in implant fixation, correction of deformity, improved polyethylene wear, and survival after knee replacement. The work over the last decade has focused on less invasive surgical techniques, multimodal pain management protocols, more rapid functional recovery and reduced length of stay, aiming to minimise the side effects of treatment while maintaining function and implant durability. When combined and standardised these pre-, intra- and post-operative factors have now facilitated outpatient knee replacement procedures for unicompartmental replacement, patella femoral arthroplasty and total knee replacement (TKR).

We have found liposomal bupivacaine, with potential for longer therapeutic action, to be a helpful adjunct and describe our current pain management program. The next step in our multimodal program is to improve the duration of patient satisfaction and reduce cost and length of stay after TKR.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):7–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 708 - 714
1 Jun 2008
McBryde CW Shears E O’Hara JN Pynsent PB

Metal-on-metal hip resurfacing was performed for developmental dysplasia in 96 hips in 85 patients, 78 in women and 18 in men, with a mean age at the time of surgery of 43 years (14 to 65). These cases were matched for age, gender, operating surgeon and date of operation with a group of patients with primary osteoarthritis who had been treated by resurfacing, to provide a control group of 96 hips (93 patients). A clinical and radiological follow-up study was performed. The dysplasia group were followed for a mean of 4.4 years (2.0 to 8.5) and the osteoarthritis group for a mean of 4.5 years (2.2 to 9.4). Of the dysplasia cases, 17 (18%) were classified as Crowe grade III or IV.

There were five (5.2%) revisions in the dysplasia group and none in the osteoarthritic patients. Four of the failures were due to acetabular loosening and the other sustained a fracture of the neck of femur. There was a significant difference in survival between the two groups (p = 0.02). The five-year survival was 96.7% (95% confidence interval 90.0 to 100) for the dysplasia group and 100% (95% confidence interval 100 to 100) for the osteoarthritic group. There was no significant difference in the median Oxford hip score between the two groups at any time during the study.

The medium-term results of metal-on-metal hip resurfacing in all grades of developmental dysplasia are encouraging, although they are significantly worse than in a group of matched patients with osteoarthritis treated in the same manner.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 562 - 569
1 May 2008
Daniel J Pradhan A Pradhan C Ziaee H Moss M Freeman J McMinn DJW

We report a retrospective review of the incidence of venous thromboembolism in 463 consecutive patients who underwent primary total hip arthroplasty (487 procedures). Treatment included both total hip replacement and hip resurfacing, and the patients were managed without anticoagulants. The thromboprophylaxis regimen included an antiplatelet agent, generally aspirin, hypotensive epidural anaesthesia, elastic compression stockings and early mobilisation. In 258 of these procedures (244 patients) performed in 2005 (cohort A) mechanical compression devices were not used, whereas in 229 (219 patients) performed during 2006 (cohort B) bilateral intermittent pneumatic calf compression was used.

All operations were performed through a posterior mini-incision approach. Patients who required anticoagulation for pre-existing medical problems and those undergoing revision arthroplasty were excluded. Doppler ultrasonographic screening for deep-vein thrombosis was performed in all patients between the fourth and sixth post-operative days. All patients were reviewed at a follow-up clinic six to ten weeks after the operation. In addition, reponse to a questionnaire was obtained at the end of 12 weeks post-operatively.

No symptomatic calf or above-knee deep-vein thrombosis or pulmonary embolism occurred. In 25 patients in cohort A (10.2%) and in ten patients in cohort B (4.6%) asymptomatic calf deep-vein thromboses were detected ultrasonographically. This difference was statistically significant (p = 0.03). The regimen followed by cohort B offers the prospect of a low incidence of venous thromboembolism without subjecting patients to the higher risk of bleeding associated with anticoagulant use.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1332 - 1338
1 Oct 2013
Van Der Straeten C Van Quickenborne D De Roest B Calistri A Victor J De Smet K

A retrospective study was conducted to investigate the changes in metal ion levels in a consecutive series of Birmingham Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed 250 BHRs implanted in 232 patients between 1998 and 2001. Implant survival, clinical outcome (Harris hip score), radiographs and serum chromium (Cr) and cobalt (Co) ion levels were assessed.

Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost to follow-up and ten had been revised. The remaining 202 BHRs in 190 patients (136 men and 54 women; mean age at surgery 50.5 years (17 to 76)) were evaluated at a minimum follow-up of ten years (mean 10.8 years (10 to 13.6)). The overall implant survival at 13.2 years was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris hip score was 97.7 (median 100; 65 to 100). Median and mean ion levels were low for unilateral resurfacings (Cr: median 1.3 µg/l, mean 1.95 µg/l (< 0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l (< 0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l, mean 3.46 µg/l (< 0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66 µg/l (< 0.5 to 9.5)). In 80 unilateral BHRs with sequential ion measurements, Cr and Co levels were found to decrease significantly (p < 0.001) from the initial assessment at a median of six years (4 to 8) to the last assessment at a median of 11 years (9 to 13), with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co. Three female patients had a > 2.5 µg/l increase of Co ions, associated with head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there was no significant difference in change of ion levels between genders (Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505; Co, p = 0.370). Higher acetabular component inclination angles correlated with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002). Patients with increased ion levels had lower Harris hip scores (p = 0.038).

In conclusion, in well-functioning BHRs the metal ion levels decreased significantly at ten years. An increase > 2.5 µg/l was associated with poor function.

Cite this article: Bone Joint J 2013;95-B:1332–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 16 - 23
1 Jan 2009
Philippon MJ Briggs KK Yen Y Kuppersmith DA

Over an eight-month period we prospectively enrolled 122 patients who underwent arthroscopic surgery of the hip for femoroacetabular impingement and met the inclusion criteria for this study. Patients with bilateral hip arthroscopy, avascular necrosis and previous hip surgery were excluded. Ten patients refused to participate leaving 112 in the study. There were 62 women and 50 men. The mean age of the patients was 40.6 yrs (95% confidence interval (CI) 37.7 to 43.5). At arthroscopy, 23 patients underwent osteoplasty only for cam impingement, three underwent rim trimming only for pincer impingement, and 86 underwent both procedures for mixed-type impingement. The mean follow-up was 2.3 years (2.0 to 2.9). The mean modified Harris hip score (HHS) improved from 58 to 84 (mean difference = 24 (95% CI 19 to 28)) and the median patient satisfaction was 9 (1 to 10). Ten patients underwent total hip replacement at a mean of 16 months (8 to 26) after arthroscopy.

The predictors of a better outcome were the pre-operative modified HHS (p = 0.018), joint space narrowing ≥ 2 mm (p = 0.005), and repair of labral pathology instead of debridement (p = 0.032).

Hip arthroscopy for femoroacetabular impingement, accompanied by suitable rehabilitation, gives a good short-term outcome and high patient satisfaction.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1299 - 1302
1 Oct 2007
Ramiah RD Ashmore AM Whitley E Bannister GC

We determined the ten-year life expectancy of 5831 patients who had undergone 6653 elective primary total hip replacements at a regional orthopaedic centre between April 1993 and October 2004. Using hospital, general practitioner and the local health authority records, we recorded the dates of death for those who died following surgery.

The mean age at operation was 67 years (13 to 96) with a male:female ratio of 2:3. Of 1154 patients with a ten-year follow-up 340 (29.5%) had died a mean of 5.6 years (0 to 10) after surgery. Using Kaplan-Meier curves, the ten-year survival was 89% in patients under 65 years at surgery, 75% in patients aged between 65 and 74 years, and 51% in patients over 75.

The standardised mortality rates were considerably higher for patients under 45 years, 20% higher for those between 45 and 64 years, and steadily reduced in patients aged 65 and over.

The survival of cemented hip replacement derived from the Swedish Hip Arthroplasty Register Annual Report 2004 exceeds the life expectancy of patients over the age of 60 in our area, suggesting that cemented hip replacement is the procedure of choice in this population.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1128 - 1136
1 Sep 2008
Lemaire R

Several aspects of the management of an orthopaedic surgical patient are not directly related to the surgical technique but are nevertheless essential for a successful outcome. Blood management is one of these. This paper considers the various strategies available for the management of blood loss in patients undergoing orthopaedic and trauma surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 163 - 166
1 Feb 2012
Zahos K Mehendale S Ward AJ Smith EJ Nichols M

We report the use of a 15° face-changing cementless acetabular component in patients undergoing total hip replacement for osteoarthritis secondary to developmental dysplasia of the hip. The rationale behind its design and the surgical technique used for its implantation are described. It is distinctly different from a standard cementless hemispherical component as it is designed to position the bearing surface at the optimal angle of inclination, that is, < 45°, while maximising the cover of the component by host bone.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 8 | Pages 1072 - 1078
1 Aug 2010
Grammatopoulos G Pandit H Glyn-Jones S McLardy-Smith P Gundle R Whitwell D Gill HS Murray DW

Pseudotumours are a rare complication of hip resurfacing. They are thought to be a response to metal debris which may be caused by edge loading due to poor orientation of the acetabular component. Our aim was to determine the optimal acetabular orientation to minimise the risk of pseudotumour formation.

We matched 31 hip resurfacings revised for pseudotumour formation with 58 controls who had a satisfactory outcome from this procedure. The radiographic inclination and anteversion angles of the acetabular component were measured on anteroposterior radiographs of the pelvis using Einzel-Bild-Roentgen-Analyse software. The mean inclination angle (47°, 10° to 81°) and anteversion angle (14°, 4° to 34°) of the pseudotumour cases were the same (p = 0.8, p = 0.2) as the controls, 46° (29° to 60°) and 16° (4° to 30°) respectively, but the variation was greater. Assuming an accuracy of implantation of ± 10° about a target position, the optimal radiographic position was found to be approximately 45° of inclination and 20° of anteversion. The incidence of pseudotumours inside the zone was four times lower (p = 0.007) than outside the zone.

In order to minimise the risk of pseudotumour formation we recommend that surgeons implant the acetabular component at an inclination of 45° (± 10) and anteversion of 20° (± 10) on post-operative radiographs. Because of differences between the radiographic and the operative angles, this may be best achieved by aiming for an inclination of 40° and an anteversion of 25°.


The Bone & Joint Journal
Vol. 96-B, Issue 10 | Pages 1287 - 1289
1 Oct 2014
Nikiphorou E Konan S MacGregor AJ Haddad FS Young A

There has been an in increase in the availability of effective biological agents for the treatment of rheumatoid arthritis as well as a shift towards early diagnosis and management of the inflammatory process. This article explores the impact this may have on the place of orthopaedic surgery in the management of patients with rheumatoid arthritis.

Cite this article: Bone Joint J 2014;96-B:1287–9


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 928 - 934
1 Jul 2009
Palan J Gulati A Andrew JG Murray DW Beard DJ

Balancing service provision and surgical training is a challenging issue that affects all healthcare systems. A multicentre prospective study of 1501 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and the grade of the operating surgeon, and whether there is any difference in outcome if surgeons’ assistants assist with the operation, rather than orthopaedic trainees. The primary outcome measure was the change in the Oxford hip score (OHS) at five years. Secondary outcomes included the rate of revision and dislocation, operating time, and length of hospital stay.

There was no significant difference in ΔOHS or complication rates between operations undertaken by trainers and trainees, or those at which surgeons’ assistants and trainees were the assistant. However, there was a significant difference in the duration of surgery, with a mean reduction of 28 minutes in those in which a surgeons’ assistant was the assistant.

This study provides evidence that total hip replacements can be performed safely and effectively by appropriately trained surgeons in training, and that there are potential benefits of using surgeons’ assistants in orthopaedic surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 28 - 37
1 Jan 2010
Jameson SS Langton DJ Nargol AVF

We present the early clinical and radiological results of Articular Surface Replacement (ASR) resurfacings in 214 hips (192 patients) with a mean follow-up of 43 months (30 to 57). The mean age of the patients was 56 years (28 to 74) and 85 hips (40%) were in 78 women.

The mean Harris hip score improved from 52 (11 to 81) to 95 (27 to 100) at two years and the mean University of California, Los Angeles activity score from 3.9 (1 to 10) to 7.4 (2 to 10) in the same period. Narrowing of the neck (to a maximum of 9%) was noted in 124 of 209 hips (60%). There were 12 revisions (5.6%) involving four (1.9%) early fractures of the femoral neck and two (0.9%) episodes of collapse of the femoral head secondary to avascular necrosis. Six patients (2.8%) had failure related to metal wear debris. The overall survival for our series was 93% (95% confidence interval 80 to 98) and 89% (95% confidence interval 82 to 96) for hips with acetabular components smaller than 56 mm in diameter.

The ASR implant has a lower diametrical clearance and a subhemispherical acetabular component when compared with other more frequently implanted metal-on-metal hip resurfacings. These changes may contribute to the higher failure rate than in other series, compared with other designs. Given our poor results with the small components we are no longer implanting the smaller size.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 8 | Pages 1013 - 1018
1 Aug 2009
Agarwala S Shah S Joshi VR

The use of bisphosphonates in the treatment of avascular necrosis of the femoral head is an encouraging but relatively new option with most published data being derived from small trials with limited follow-up. We present a clinicoradiological analysis of 395 hips with avascular necrosis which were treated with oral alendronate for three years with a mean follow-up of four years (1 to 8).

Our results show an improvement in the clinical function, a reduction in the rate of collapse and a decrease in the requirement for total hip replacement, compared with the findings of other studies in which no treatment was given. This improvement is particularly marked if the treatment is begun in the pre-collapse stages of the disease. Even in Ficat stage-III hips some benefit was obtained from treatment with alendronate by at least a delay in the need for total hip replacement.