Results 21 - 35 of 35
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1506 - 1511
1 Nov 2015
Liddle AD Pandit H Judge A Murray DW

Unicompartmental knee arthroplasty (UKA) has advantages over total knee arthroplasty but national joint registries report a significantly higher revision rate for UKA. As a result, most surgeons are highly selective, offering UKA only to a small proportion (up to 5%) of patients requiring arthroplasty of the knee, and consequently performing few each year. However, surgeons with large UKA practices have the lowest rates of revision. The overall size of the practice is often beyond the surgeon’s control, therefore case volume may only be increased by broadening the indications for surgery, and offering UKA to a greater proportion of patients requiring arthroplasty of the knee.

The aim of this study was to determine the optimal UKA usage (defined as the percentage of knee arthroplasty practice comprised by UKA) to minimise the rate of revision in a sample of 41 986 records from the for National Joint Registry for England and Wales (NJR).

UKA usage has a complex, non-linear relationship with the rate of revision. Acceptable results are achieved with the use of 20% or more. Optimal results are achieved with usage between 40% and 60%. Surgeons with the lowest usage (up to 5%) have the highest rates of revision. With optimal usage, using the most commonly used implant, five-year survival is 96% (95% confidence interval (CI) 94.9 to 96.0), compared with 90% (95% CI 88.4 to 91.6) with low usage (5%) previously considered ideal.

The rate of revision of UKA is highest with low usage, implying the use of narrow, and perhaps inappropriate, indications. The widespread use of broad indications, using appropriate implants, would give patients the advantages of UKA, without the high rate of revision.

Cite this article: Bone Joint J 2015;97-B:1506–11.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1172 - 1179
1 Sep 2008
Wylde V Learmonth I Potter A Bettinson K Lingard E

We compared patient-reported outcomes of the Kinemax fixed- and mobile-bearing total knee replacement in a multi-centre randomised controlled trial. Patients were randomised to the fixed- or the mobile-bearing prosthesis via a sealed envelope method after the bone cuts had been made in the operating theatre. Randomisation was stratified by centre and diagnosis. Patients were assessed pre-operatively and at eight to 12 weeks, one year and two years post-operatively. Validated questionnaires were used which included the Western Ontario MacMasters University, Short-Form 12, Mental Health Index-5, Knee Injury and Osteoarthritis Outcome Score for Knee-Related Quality of Life and Function in Sport and Recreation scales and a validated scale of satisfaction post-operatively. In total, 242 patients (250 knees) with a mean age of 68 years (40 to 80) were recruited from four NHS orthopaedic centres. Of these, 132 patients (54.5%) were women.

No statistically significant differences could be identified in any of the patient-reported outcome scores between patients who received the fixed-bearing or the mobile-bearing knee up to two-years post-operatively.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1339 - 1347
1 Oct 2013
Scott CEH Eaton MJ Nutton RW Wade FA Pankaj P Evans SL

As many as 25% to 40% of unicompartmental knee replacement (UKR) revisions are performed for pain, a possible cause of which is proximal tibial strain. The aim of this study was to examine the effect of UKR implant design and material on cortical and cancellous proximal tibial strain in a synthetic bone model. Composite Sawbone tibiae were implanted with cemented UKR components of different designs, either all-polyethylene or metal-backed. The tibiae were subsequently loaded in 500 N increments to 2500 N, unloading between increments. Cortical surface strain was measured using a digital image correlation technique. Cancellous damage was measured using acoustic emission, an engineering technique that detects sonic waves (‘hits’) produced when damage occurs in material.

Anteromedial cortical surface strain showed significant differences between implants at 1500 N and 2500 N in the proximal 10 mm only (p < 0.001), with relative strain shielding in metal-backed implants. Acoustic emission showed significant differences in cancellous bone damage between implants at all loads (p = 0.001). All-polyethylene implants displayed 16.6 times the total number of cumulative acoustic emission hits as controls. All-polyethylene implants also displayed more hits than controls at all loads (p < 0.001), more than metal-backed implants at loads ≥ 1500 N (p < 0.001), and greater acoustic emission activity on unloading than controls (p = 0.01), reflecting a lack of implant stiffness. All-polyethylene implants were associated with a significant increase in damage at the microscopic level compared with metal-backed implants, even at low loads. All-polyethylene implants should be used with caution in patients who are likely to impose large loads across their knee joint.

Cite this article: Bone Joint J 2013;95-B:1339–47.


Bone & Joint 360
Vol. 5, Issue 3 | Pages 15 - 17
1 Jun 2016


Bone & Joint Research
Vol. 2, Issue 7 | Pages 129 - 131
1 Jul 2013
Wyatt MC Frampton C Horne JG Devane P

Objectives

Our study aimed to examine if a mobile-bearing total knee replacement (TKR) offered an advantage over fixed-bearing designs with respect to rates of secondary resurfacing of the patella in knees in which it was initially left unresurfaced.

Methods

We examined the 11-year report of the New Zealand Joint Registry and identified all primary TKR designs that had been implanted in > 500 knees without primary resurfacing of the patella. We examined how many of these were mobile-bearing, fixed-bearing cruciate-retaining and fixed-bearing posterior-stabilised designs. We assessed the rates of secondary resurfacing of the patella for each group and constructed Kaplan-Meier survival curves.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 636 - 642
1 May 2013
Gøthesen Ø Espehaug B Havelin L Petursson G Lygre S Ellison P Hallan G Furnes O

We evaluated the rates of survival and cause of revision of seven different brands of cemented primary total knee replacement (TKR) in the Norwegian Arthroplasty Register during the years 1994 to 2009. Revision for any cause, including resurfacing of the patella, was the primary endpoint. Specific causes of revision were secondary outcomes.

Three posterior cruciate-retaining (PCR) fixed modular-bearing TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing posterior cruciate-sacrificing TKRs were investigated in a total of 17 782 primary TKRs. The median follow-up for the implants ranged from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from 89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative to the fixed modular-bearing Profix knee (the most frequently used TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision for aseptic tibial loosening was higher in the mobile-bearing LCS Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared with the Profix. These implants (except AGC Universal) also had an increased risk of revision for femoral loosening (RR 2.3 (95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95% CI 1.1 to 11.0), respectively). These results suggest that aseptic loosening is related to design in TKR.

Cite this article: Bone Joint J 2013;95-B:636–42.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1366 - 1371
1 Oct 2013
Muir D Aoina J Hong T Mason R

We performed a retrospective review of a consecutive series of 178 Mobility total ankle replacements (TARs) performed by three surgeons between January 2004 and June 2009, and analysed radiological parameters and clinical outcomes in a subgroup of 129 patients. The mean follow-up was 4 years (2 to 6.3). A total of ten revision procedures (5.6%) were undertaken. The mean Ankle Osteoarthritis Scale (AOS) pain score was 17 (0 to 88) and 86% of patients were clinically improved at follow-up. However, 18 patients (18 TARs, 14%) had a poor outcome with an AOS pain score of > 30. A worse outcome was associated with a pre-operative diagnosis of post-traumatic degenerative arthritis. However, no pre- or post-operative radiological parameters were significantly associated with a poor outcome. Of the patients with persistent pain, eight had predominantly medial-sided pain. Thirty TARs (29%) had a radiolucency in at least one zone.

The outcome of the Mobility TAR at a mean of four years is satisfactory in > 85% of patients. However, there is a significant incidence of persistent pain, particularly on the medial side, for which we were unable to establish a cause.

Cite this article: Bone Joint J 2013;95-B:1366–71.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1500 - 1507
1 Nov 2013
Zaidi R Cro S Gurusamy K Sivanadarajah N Macgregor A Henricson A Goldberg A

We performed a systematic review and meta-analysis of modern total ankle replacements (TARs) to determine the survivorship, outcome, complications, radiological findings and range of movement, in patients with end-stage osteoarthritis (OA) of the ankle who undergo this procedure. We used the methodology of the Cochrane Collaboration, which uses risk of bias profiling to assess the quality of papers in favour of a domain-based approach. Continuous outcome scores were pooled across studies using the generic inverse variance method and the random-effects model was used to incorporate clinical and methodological heterogeneity. We included 58 papers (7942 TARs) with an interobserver reliability (Kappa) for selection, performance, attrition, detection and reporting bias of between 0.83 and 0.98. The overall survivorship was 89% at ten years with an annual failure rate of 1.2% (95% confidence interval (CI) 0.7 to 1.6). The mean American Orthopaedic Foot and Ankle Society score changed from 40 (95% CI 36 to 43) pre-operatively to 80 (95% CI 76 to 84) at a mean follow-up of 8.2 years (7 to 10) (p < 0.01). Radiolucencies were identified in up to 23% of TARs after a mean of 4.4 years (2.3 to 9.6). The mean total range of movement improved from 23° (95% CI 19 to 26) to 34° (95% CI 26 to 41) (p = 0.01).

Our study demonstrates that TAR has a positive impact on patients’ lives, with benefits lasting ten years, as judged by improvement in pain and function, as well as improved gait and increased range of movement. However, the quality of evidence is weak and fraught with biases and high quality randomised controlled trials are required to compare TAR with other forms of treatment such as fusion.

Cite this article: Bone Joint J 2013;95-B:1500–7.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 436 - 444
1 Apr 2013
Scott CEH Nutton RW Biant LC

The lateral compartment is predominantly affected in approximately 10% of patients with osteoarthritis of the knee. The anatomy, kinematics and loading during movement differ considerably between medial and lateral compartments of the knee. This in the main explains the relative protection of the lateral compartment compared with the medial compartment in the development of osteoarthritis. The aetiology of lateral compartment osteoarthritis can be idiopathic, usually affecting the femur, or secondary to trauma commonly affecting the tibia. Surgical management of lateral compartment osteoarthritis can include osteotomy, unicompartmental knee replacement and total knee replacement. This review discusses the biomechanics, pathogenesis and development of lateral compartment osteoarthritis and its management.

Cite this article: Bone Joint J 2013;95-B:436–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1641 - 1648
1 Dec 2012
Baker PN Jameson SS Deehan DJ Gregg PJ Porter M Tucker K

Current analysis of unicondylar knee replacements (UKRs) by national registries is based on the pooled results of medial and lateral implants. Consequently, little is known about the differential performance of medial and lateral replacements and the influence of each implant type within these pooled analyses. Using data from the National Joint Registry for England and Wales (NJR) we aimed to determine the proportion of UKRs implanted on the lateral side of the knee, and their survival and reason for failure compared with medial UKRs. By combining information on the side of operation with component details held on the NJR, we were able to determine implant laterality (medial versus lateral) for 32 847 of the 35 624 unicondylar replacements (92%) registered before December 2010. Of these, 2052 (6%) were inserted on the lateral side of the knee. The rates of survival at five years were 93.1% (95% confidence interval (CI) 92.7 to 93.5) for medial and 93.0% (95% CI 91.1 to 94.9) for lateral UKRs (p = 0.49). The rates of failure remained equivalent after adjusting for patient age, gender, American Society of Anesthesiologists (ASA) grade, indication for surgery and implant design using Cox’s proportional hazards method (hazard ratio for lateral relative to medial replacement = 0.88 (95% CI 0.69 to 1.13); p = 0.32). Aseptic loosening/lysis and unexplained pain were the main reasons for revision in both groups, although the reasons did vary depending on whether a mobile- or a fixed-bearing design was used. At a maximum of eight years the mid-term survival rates of medial and lateral UKRs are similar.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1058 - 1066
1 Aug 2012
Baker PN Deehan DJ Lees D Jameson S Avery PJ Gregg PJ Reed MR

Patient-reported outcome measures (PROMs) are increasingly being used to assess functional outcome and patient satisfaction. They provide a framework for comparisons between surgical units, and individual surgeons for benchmarking and financial remuneration. Better performance may bring the reward of more customers as patients and commissioners seek out high performers for their elective procedures. Using National Joint Registry (NJR) data linked to PROMs we identified 22 691 primary total knee replacements (TKRs) undertaken for osteoarthritis in England and Wales between August 2008 and February 2011, and identified the surgical factors that influenced the improvements in the Oxford knee score (OKS) and EuroQol-5D (EQ-5D) assessment using multiple regression analysis. After correction for patient factors the only surgical factors that influenced PROMs were implant brand and hospital type (both p < 0.001). However, the effects of surgical factors upon the PROMs were modest compared with patient factors. For both the OKS and the EQ-5D the most important factors influencing the improvement in PROMs were the corresponding pre-operative score and the patient’s general health status. Despite having only a small effect on PROMs, this study has shown that both implant brand and hospital type do influence reported subjective functional scores following TKR. In the current climate of financial austerity, proposed performance-based remuneration and wider patient choice, it would seem unwise to ignore these effects and the influence of a range of additional patient factors.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 479 - 483
1 Apr 2011
Chang C Lai K Yang C Lan S

Between April 2004 and July 2007, we performed 241 primary total knee replacements in 204 patients using the e.motion posterior cruciate-retaining, multidirectional mobile-bearing prosthesis. Of these, 100 were carried out using an image-free navigation system, and the remaining 141 with the conventional technique. We conducted a retrospective study from the prospectively collected data of these patients to assess the early results of this new mobile-bearing design.

At a mean follow-up of 49 months (32 to 71), 18 knees (7.5%) had mechanical complications of which 13 required revision. Three of these had a peri-prosthetic fracture, and were removed from the study. The indication for revision in the remaining ten was loosening of the femoral component in two, tibiofemoral dislocation in three, disassociation of the polyethylene liner in four, and a broken polyethyene liner in one. There were eight further mechanically unstable knees which presented with recurrent disassociation of the polyethylene liner. There was no significant difference in the incidence of mechanical instability between the navigation-assisted procedures (8 of 99, 8.1%) and the conventionally implanted knees (10 of 139, 7.2%).

In our view, the relatively high rate of mechanical complications and revision within 30 months precludes the further use of new design of knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1457 - 1461
1 Nov 2007
Han HS Kang S Yoon KS

We have examined the results obtained with 72 NexGen legacy posterior stabilised-flex fixed total knee replacements in 47 patients implanted by a single surgeon between March 2003 and September 2004.

Aseptic loosening of the femoral component was found in 27 (38%) of the replacements at a mean follow-up of 32 months (30 to 48) and 15 knees (21%) required revision at a mean of 23 months (11 to 45). We compared the radiologically-loose and revised knees with those which had remained well-fixed to identify the factors which had contributed to this high rate of aseptic loosening.

Post-operatively, the mean maximum flexion was 136° (110° to 140°) in the loosened group and 125° (95° to 140°) in the well-fixed group (independent t-test, p = 0.022). Squatting, kneeling, or sitting cross-legged could be achieved by 23 (85%) of the loosened knees, but only 22 (49%) of the well-fixed knees (chi-squared test, p = 0.001). The loosened femoral components were found to migrate into a more flexed position, but no migration was detected in the well-fixed group.

These implants allowed a high degree of flexion, but showed a marked rate of early loosening of the femoral component, which was associated with weight-bearing in maximum flexion.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 44 - 48
1 Jan 2006
Keene G Simpson D Kalairajah Y

Twenty patients underwent simultaneous bilateral medial unicompartmental knee arthroplasty. Pre-operative hip-knee-ankle alignment and valgus stress radiographs were used to plan the desired post-operative alignment of the limb in accordance with established principles for unicompartmental arthroplasty. In each patient the planned alignment was the same for both knees. Overall, the mean planned post-operative alignment was to 2.3° of varus (0° to 5°).

The side and starting order of surgery were randomised, using conventional instrumentation for one knee and computer-assisted surgery for the opposite side.

The mean variation between the pre-operative plan and the achieved correction in the navigated and the non-navigated limb was 0.9° (sd 1.1; 0° to 4°) and 2.8° (sd 1.4; 1° to 7°), respectively. Using the Wilcoxon signed rank test, we found the difference in variation statistically significant (p < 0.001).

Assessment of lower limb alignment in the non-navigated group revealed that 12 (60%) were within ± 2° of the pre-operative plan, compared to 17 (87%) of the navigated cases.

Computer-assisted surgery significantly improves the post-operative alignment of medial unicompartmental knee arthroplasty compared to conventional techniques in patients undergoing bilateral simultaneous arthroplasty. Improved alignment after arthroplasty is associated with better function and increased longevity.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1016 - 1021
1 Aug 2006
Delport HP Banks SA De Schepper J Bellemans J

Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the in vivo kinematics of this new group of implants. We investigated 31 patients who had undergone a total knee replacement with a similar prosthetic design but with three different options: fixed-bearing posterior cruciate ligament-retaining, fixed-bearing posterior-stabilised and mobile-bearing posterior-stabilised. To do this we used a three-dimensional to two-dimensional model registration technique. Both the fixed- and mobile-bearing posterior-stabilised configurations used the same femoral component. We found that fixed-bearing posterior stabilised and mobile-bearing posterior-stabilised knee replacements demonstrated similar kinematic patterns, with consistent femoral roll-back during flexion. Mobile-bearing posterior-stabilised knee replacements demonstrated greater and more natural internal rotation of the tibia during flexion than fixed-bearing posterior-stabilised designs. Such rotation occurred at the interface between the insert and tibial tray for mobile-bearing posterior-stabilised designs. However, for fixed-bearing posterior-stabilised designs, rotation occurred at the proximal surface of the bearing. Posterior cruciate ligament-retaining knee replacements demonstrated paradoxical sliding forward of the femur.

We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.