Aims. The aim of this study was to evaluate the effects of using a
portable, accelerometer-based surgical navigation system (KneeAlign2)
in total knee arthroplasty (TKA) on the alignment of the femoral
component, and blood loss. Patients and Methods. A total of 241 consecutive patients with primary osteoarthritis
of the knee were enrolled in this prospective, randomised controlled
study. There were 207 women and 34 men. The mean age of the patients
was 74.0 years (57 to 89). The KneeAlign2 system was used for distal
femoral resection in 121 patients (KA2 group) and a conventional intramedullary
femoral guide was used in 120 patients (IM group). Results. One patient (0.8%) in the KA2 group and 19 in the IM group had
an alignment which was
>
3° away from the neutral
Aims. To present our experience of using a combination of intra-articular
osteotomy and external fixation to treat different deformities of
the knee. Patients and Methods. A total of six patients with a mean age of 26.5 years (15 to
50) with an abnormal hemi-joint line convergence angle (HJLCA) and
mechanical axis deviation (MAD) were included. Elevation of a tibial
hemiplateau or femoral condylar advancement was performed and limb
lengthening with correction of residual deformity using a circular
or monolateral Ilizarov frame. Results. At a mean follow-up of 2.8 years (1.5 to 4.1), the mean HJLCA
improved from 15.6° (10° to 23°) pre-operatively to 0.4° (0° to
2°). The mean MAD improved from 70.0 mm (20.1 to 118.5) pre-operatively
to 9.1 mm (3 to 15). The mean tibiofemoral angle improved from 31.0°
(8° to 54°) pre-operatively to 4.9° (2° to 8°). The mean limb-length discrepancy
decreased from 6.3 cm (2.9 to 13.6) pre-operatively to 1.1 cm (0
to 5). All osteotomies and distraction zones healed without complications. Conclusion. The use of intra-articular corrective osteotomies combined with
the Ilizarov technique allowed correction of deformities of the
knee joint with satisfactory HJLCA and overall
Substantial healthcare resources have been devoted
to computer navigation and patient-specific instrumentation systems
that improve the reproducibility with which neutral mechanical alignment
can be achieved following total knee replacement (TKR). This choice of
alignment is based on the long-held tenet that the alignment of
the limb post-operatively should be within 3° of a neutral mechanical
axis. Several recent studies have demonstrated no significant difference
in survivorship when comparing well aligned versus malaligned TKRs.
Our aim was to review the anatomical alignment of the knee, the
historical and contemporary data on a neutral
In this study we randomised 140 patients who
were due to undergo primary total knee arthroplasty (TKA) to have the
procedure performed using either patient-specific cutting guides
(PSCG) or conventional instrumentation (CI). . The primary outcome measure was the
Deformity of the proximal femur in fibrous dysplasia
leads to deviation of the
The requirement for release of collateral ligaments to achieve a stable, balanced total knee replacement has been reported to arise in about 50% to 100% of procedures. This wide range reflects a lack of standardised quantitative indicators to determine the necessity for a release. Using recent advances in computerised navigation, we describe two navigational predictors which provide quantitative measures that can be used to identify the need for release. The first was the ability to restore the
We present the results of the surgical correction of lower-limb deformities caused by metabolic bone disease. Our series consisted of 17 patients with a diagnosis of hypophosphataemic rickets and two with renal osteodystrophy; their mean age was 25.6 years (14 to 57). In all, 43 lower-limb segments (27 femora and 16 tibiae) were osteotomised and the deformity corrected using a monolateral external fixator. The segment was then stabilised with locked intramedullary nailing. In addition, six femora in three patients were subsequently lengthened by distraction osteogenesis. The mean follow-up was 60 months (18 to 120). The frontal alignment parameters (the
We retrospectively reviewed the records of 1150
computer-assisted total knee replacements and analysed the clinical
and radiological outcomes of 45 knees that had arthritis with a
pre-operative recurvatum deformity. The mean pre-operative hyperextension
deformity of 11° (6° to 15°), as measured by navigation at the start
of the operation, improved to a mean flexion deformity of 3.1° (0°
to 7°) post-operatively. A total of 41 knees (91%) were managed
using inserts ≤ 12.5 mm thick, and none had mediolateral laxity
>
2 mm from a
Survivors of infantile meningococcal septicaemia often develop progressive skeletal deformity as a result of physeal damage at many sites, particularly in the lower limb. Distal tibial physeal arrest typically occurs with sparing of the distal fibular physis leading to a rapidly progressive varus deformity. There have been reports of isolated cases of this deformity, but to our knowledge there have been no papers which specifically describe the development of the deformity and the options for treatment. Surgery to correct this deformity is complex because of the patient’s age, previous scarring and the multiplanar nature of the deformity. The surgical goal is to restore leg-length equality and the
Minimally invasive total knee replacement (MIS-TKR)
has been reported to have better early recovery than conventional
TKR. Quadriceps-sparing (QS) TKR is the least invasive MIS procedure,
but it is technically demanding with higher reported rates of complications
and outliers. This study was designed to compare the early clinical
and radiological outcomes of TKR performed by an experienced surgeon
using the QS approach with or without navigational assistance (NA),
or using a mini-medial parapatellar (MP) approach. In all, 100 patients
completed a minimum two-year follow-up: 30 in the NA-QS group, 35
in the QS group, and 35 in the MP group. There were no significant
differences in clinical outcome in terms of ability to perform a
straight-leg raise at 24 hours (p = 0.700), knee score (p = 0.952),
functional score (p = 0.229) and range of movement (p = 0.732) among
the groups. The number of outliers for all three radiological parameters
of
Intra-articular resection of bone with soft-tissue balancing and total knee replacement (TKR) has been described for the treatment of patients with severe osteoarthritis of the knee associated with an ipsilateral malunited femoral fracture. However, the extent to which deformity in the sagittal plane can be corrected has not been addressed. We treated 12 patients with severe arthritis of the knee and an extra-articular malunion of the femur by TKR with intra-articular resection of bone and soft-tissue balancing. The femora had a mean varus deformity of 16° (8° to 23°) in the coronal plane. There were seven recurvatum deformities with a mean angulation of 11° (6° to 15°) and five antecurvatum deformities with a mean angulation of 12° (6° to 15°). The mean follow-up was 93 months (30 to 155). The median Knee Society knee and function scores improved from 18.7 (0 to 49) and 24.5 (10 to 50) points pre-operatively to 93 (83 to 100) and 90 (70 to 100) points at the time of the last follow-up, respectively. The mean
Total knee arthroplasty (TKA) may provoke ankle symptoms. The aim of this study was to validate the impact of the preoperative mechanical tibiofemoral angle (mTFA), the talar tilt (TT) on ankle symptoms after TKA, and assess changes in the range of motion (ROM) of the subtalar joint, foot posture, and ankle laxity. Patients who underwent TKA from September 2020 to September 2021 were prospectively included. Inclusion criteria were primary end-stage osteoarthritis (Kellgren-Lawrence stage IV) of the knee. Exclusion criteria were missed follow-up visit, post-traumatic pathologies of the foot, and neurological disorders. Radiological angles measured included the mTFA, hindfoot alignment view angle, and TT. The Foot Function Index (FFI) score was assessed. Gait analyses were conducted to measure mediolateral changes of the gait line and ankle laxity was tested using an ankle arthrometer. All parameters were acquired one week pre- and three months postoperatively.Aims
Methods
This study aimed to evaluate if total knee arthroplasty (TKA) femoral components aligned in either mechanical alignment (MA) or kinematic alignment (KA) are more biomimetic concerning trochlear sulcus orientation and restoration of trochlear height. Bone surfaces from 1,012 CT scans of non-arthritic femora were segmented using a modelling and analytics system. TKA femoral components (Triathlon; Stryker) were virtually implanted in both MA and KA. Trochlear sulcus orientation was assessed by measuring the distal trochlear sulcus angle (DTSA) in native femora and in KA and MA prosthetic femoral components. Trochlear anatomy restoration was evaluated by measuring the differences in medial, lateral, and sulcus trochlear height between native femora and KA and MA prosthetic femoral components.Aims
Methods
The aim of mechanical alignment in total knee arthroplasty is to align all knees into a fixed neutral position, even though not all knees are the same. As a result, mechanical alignment often alters a patient’s constitutional alignment and joint line obliquity, resulting in soft-tissue imbalance. This annotation provides an overview of how the Coronal Plane Alignment of the Knee (CPAK) classification can be used to predict imbalance with mechanical alignment, and then offers practical guidance for bone balancing, minimizing the need for soft-tissue releases. Cite this article:
Robotic arm-assisted surgery offers accurate and reproducible guidance in component positioning and assessment of soft-tissue tensioning during knee arthroplasty, but the feasibility and early outcomes when using this technology for revision surgery remain unknown. The objective of this study was to compare the outcomes of robotic arm-assisted revision of unicompartmental knee arthroplasty (UKA) to total knee arthroplasty (TKA) versus primary robotic arm-assisted TKA at short-term follow-up. This prospective study included 16 patients undergoing robotic arm-assisted revision of UKA to TKA versus 35 matched patients receiving robotic arm-assisted primary TKA. In all study patients, the following data were recorded: operating time, polyethylene liner size, change in haemoglobin concentration (g/dl), length of inpatient stay, postoperative complications, and hip-knee-ankle (HKA) alignment. All procedures were performed using the principles of functional alignment. At most recent follow-up, range of motion (ROM), Forgotten Joint Score (FJS), and Oxford Knee Score (OKS) were collected. Mean follow-up time was 21 months (6 to 36).Aims
Methods
The understanding of rotational alignment of the distal femur is essential in total knee replacement to ensure that there is correct placement of the femoral component. Many reference axes have been described, but there is still disagreement about their value and mutual angular relationship. Our aim was to validate a geometrically-defined reference axis against which the surface-derived axes could be compared in the axial plane. A total of 12 cadaver specimens underwent CT after rigid fixation of optical tracking devices to the femur and the tibia. Three-dimensional reconstructions were made to determine the anatomical surface points and geometrical references. The spatial relationships between the femur and tibia in full extension and in 90° of flexion were examined by an optical infrared tracking system. After co-ordinate transformation of the described anatomical points and geometrical references, the projection of the relevant axes in the axial plane of the femur were mathematically achieved. Inter- and intra-observer variability in the three-dimensional CT reconstructions revealed angular errors ranging from 0.16° to 1.15° for all axes except for the trochlear axis which had an interobserver error of 2°. With the knees in full extension, the femoral transverse axis, connecting the centres of the best matching spheres of the femoral condyles, almost coincided with the tibial transverse axis (mean difference −0.8°, . sd. 2.05). At 90° of flexion, this femoral transverse axis was orthogonal to the tibial
Recently, gender-specific designs of total knee replacement have been developed to accommodate anatomical differences between males and females. We examined a group of male and female distal femora matched for age and height, to determine if there was a difference in the aspect ratio (mediolateral distance versus anteroposterior distance) and the height of the anterior flange between the genders. The Hamann-Todd Collection provided 1207 skeletally mature cadaver femora. The femoral length, the anteroposterior height, height of the lateral and medial flanges and the mediolateral width were measured in all the specimens. The
Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear. A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay.Aims
Methods
Accurate identification of the ankle joint centre is critical for estimating tibial coronal alignment in total knee arthroplasty (TKA). The purpose of the current study was to leverage artificial intelligence (AI) to determine the accuracy and effect of using different radiological anatomical landmarks to quantify mechanical alignment in relation to a traditionally defined radiological ankle centre. Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A sub-cohort of 250 radiographs were annotated for landmarks relevant to knee alignment and used to train a deep learning (U-Net) workflow for angle calculation on the entire database. The radiological ankle centre was defined as the midpoint of the superior talus edge/tibial plafond. Knee alignment (hip-knee-ankle angle) was compared against 1) midpoint of the most prominent malleoli points, 2) midpoint of the soft-tissue overlying malleoli, and 3) midpoint of the soft-tissue sulcus above the malleoli.Aims
Methods
As advancements in total knee arthroplasty progress at an exciting pace, two areas are of special interest, as they directly impact implant design and surgical decision making. Knee morphometry considers the three-dimensional shape of the articulating surfaces within the knee joint, and knee phenotyping provides the ability to categorize alignment into practical groupings that can be used in both clinical and research settings. This annotation discusses the details of these concepts, and the ways in which they are helping us better understand the individual subtleties of each patient’s knee. Cite this article: