Advertisement for orthosearch.org.uk
Results 21 - 40 of 138
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 63 - 66
1 Nov 2013
Rodriguez JA Cooper HJ

Large ceramic femoral heads offer several advantages that are potentially advantageous to patients undergoing both primary and revision total hip replacement. Many high-quality studies have demonstrated the benefit of large femoral heads in reducing post-operative instability. Ceramic femoral heads may also offer an advantage in reducing polyethylene wear that has been reported in vitro and is starting to become clinically apparent in mid-term clinical outcome studies. Additionally, the risk of taper corrosion at a ceramic femoral head–neck junction is clearly lower than when using a metal femoral head. With improvements in the material properties of both modern ceramic femoral heads and polyethylene acetabular liners that have reduced the risk of mechanical complications, large ceramic heads have gained popularity in recent years. Cite this article: Bone Joint J 2013;95-B, Supple A:63–6


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 38 - 43
1 Mar 2024
Buckner BC Urban ND Cahoy KM Lyden ER Deans CF Garvin KL

Aims

Oxidized zirconium (OxZi) and highly cross-linked polyethylene (HXLPE) were developed to minimize wear and risk of osteolysis in total hip arthroplasty (THA). However, retrieval studies have shown that scratched femoral heads may lead to runaway wear, and few reports of long-term results have been published. The purpose of this investigation is to report minimum ten-year wear rates and clinical outcomes of THA with OxZi femoral heads on HXLPE, and to compare them with a retrospective control group of cobalt chrome (CoCr) or ceramic heads on HXLPE.

Methods

From 2003 to 2006, 108 THAs were performed on 96 patients using an OxZi head with a HXLPE liner with minimum ten-year follow-up. Harris Hip Scores (HHS) were collected preoperatively and at the most recent follow-up (mean 13.3 years). Linear and volumetric liner wear was measured on radiographs of 85 hips with a minimum ten-year follow-up (mean 14.5 years). This was compared to a retrospective control group of 45 THAs using ceramic or CoCr heads from October 1999 to February 2005, with a minimum of ten years’ follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 125 - 132
1 May 2024
Carender CN Perry KI Sierra RJ Trousdale RT Berry DJ Abdel MP

Aims

Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components.

Methods

We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.


Bone & Joint 360
Vol. 12, Issue 3 | Pages 13 - 15
1 Jun 2023

The June 2023 Hip & Pelvis Roundup360 looks at: Machine learning to identify surgical candidates for hip and knee arthroplasty: a viable option?; Poor outcome after debridement and implant retention; Can you cement polyethylene liners into well-fixed acetabular shells in hip revision?; Revision stem in primary arthroplasties: the Exeter 44/0 125 mm stem; Depression and anxiety: could they be linked to infection?; Does where you live affect your outcomes after hip and knee arthroplasties?; Racial disparities in outcomes after total hip arthroplasty and total knee arthroplasty are substantially mediated by socioeconomic disadvantage both in black and white patients.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 565 - 572
1 Jun 2024
Resl M Becker L Steinbrück A Wu Y Perka C

Aims

This study compares the re-revision rate and mortality following septic and aseptic revision hip arthroplasty (rTHA) in registry data, and compares the outcomes to previously reported data.

Methods

This is an observational cohort study using data from the German Arthroplasty Registry (EPRD). A total of 17,842 rTHAs were included, and the rates and cumulative incidence of hip re-revision and mortality following septic and aseptic rTHA were analyzed with seven-year follow-up. The Kaplan-Meier estimates were used to determine the re-revision rate and cumulative probability of mortality following rTHA.


Bone & Joint Open
Vol. 4, Issue 2 | Pages 53 - 61
1 Feb 2023
Faraj S de Windt TS van Hooff ML van Hellemondt GG Spruit M

Aims

The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA).

Methods

Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 825 - 831
3 Oct 2024
Afghanyar Y Afghanyar B Loweg L Drees P Gercek E Dargel J Rehbein P Kutzner KP

Aims

Limited implant survival due to aseptic cup loosening is most commonly responsible for revision total hip arthroplasty (THA). Advances in implant designs and materials have been crucial in addressing those challenges. Vitamin E-infused highly cross-linked polyethylene (VEPE) promises strong wear resistance, high oxidative stability, and superior mechanical strength. Although VEPE monoblock cups have shown good mid-term performance and excellent wear patterns, long-term results remain unclear. This study evaluated migration and wear patterns and clinical and radiological outcomes at a minimum of ten years’ follow-up.

Methods

This prospective observational study investigated 101 cases of primary THA over a mean duration of 129 months (120 to 149). At last follow-up, 57 cases with complete clinical and radiological outcomes were evaluated. In all cases, the acetabular component comprised an uncemented titanium particle-coated VEPE monoblock cup. Patients were assessed clinically and radiologically using the Harris Hip Score, visual analogue scale (pain and satisfaction), and an anteroposterior radiograph. Cup migration and polyethylene wear were measured using Einzel-Bild-Röntgen-Analyze software. All complications and associated treatments were documented until final follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1206 - 1215
1 Nov 2024
Fontalis A Buchalter D Mancino F Shen T Sculco PK Mayman D Haddad FS Vigdorchik J

Understanding spinopelvic mechanics is important for the success of total hip arthroplasty (THA). Despite significant advancements in appreciating spinopelvic balance, numerous challenges remain. It is crucial to recognize the individual variability and postoperative changes in spinopelvic parameters and their consequential impact on prosthetic component positioning to mitigate the risk of dislocation and enhance postoperative outcomes. This review describes the integration of advanced diagnostic approaches, enhanced technology, implant considerations, and surgical planning, all tailored to the unique anatomy and biomechanics of each patient. It underscores the importance of accurately predicting postoperative spinopelvic mechanics, selecting suitable imaging techniques, establishing a consistent nomenclature for spinopelvic stiffness, and considering implant-specific strategies. Furthermore, it highlights the potential of artificial intelligence to personalize care.

Cite this article: Bone Joint J 2024;106-B(11):1206–1215.


Bone & Joint Open
Vol. 5, Issue 6 | Pages 514 - 523
24 Jun 2024
Fishley W Nandra R Carluke I Partington PF Reed MR Kramer DJ Wilson MJ Hubble MJW Howell JR Whitehouse SL Petheram TG Kassam AM

Aims

In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and modular (MDM) dual-mobility polyethylene bearings match those of Birmingham hip MoM components. If the acetabular component is satisfactorily positioned, well integrated into the bone, and has no surface damage, this presents the opportunity for revision with exchange of the metal head for ADM/MDM polyethylene bearings without removal of the acetabular component.

Methods

Between 2012 and 2020, across two centres, 94 patients underwent revision of Birmingham MoM hip arthroplasties or resurfacings. Mean age was 65.5 years (33 to 87). In 53 patients (56.4%), the acetabular component was retained and dual-mobility bearings were used (DM); in 41 (43.6%) the acetabulum was revised (AR). Patients underwent follow-up of minimum two-years (mean 4.6 (2.1 to 8.5) years).


The Bone & Joint Journal
Vol. 106-B, Issue 3 Supple A | Pages 81 - 88
1 Mar 2024
Lustig S Cotte M Foissey C Asirvatham RD Servien E Batailler C

Aims

The benefit of a dual-mobility acetabular component (DMC) for primary total hip arthroplasties (THAs) is controversial. This study aimed to compare the dislocation and complication rates when using a DMC compared to single-mobility (SM) acetabular component in primary elective THA using data collected at a single centre, and compare the revision rates and survival outcomes in these two groups.

Methods

Between 2010 and 2019, 2,075 primary THAs using either a cementless DM or SM acetabular component were included. Indications for DMC were patients aged older than 70 years or with high risk of dislocation. All other patients received a SM acetabular component. Exclusion criteria were cemented implants, patients treated for femoral neck fracture, and follow-up of less than one year. In total, 1,940 THAs were analyzed: 1,149 DMC (59.2%) and 791 SM (40.8%). The mean age was 73 years (SD 9.2) in the DMC group and 57 years (SD 12) in the SM group. Complications and revisions have been analyzed retrospectively.


The Bone & Joint Journal
Vol. 106-B, Issue 6 | Pages 555 - 564
1 Jun 2024
Leal J Holland CT Cochrane NH Seyler TM Jiranek WA Wellman SS Bolognesi MP Ryan SP

Aims

This study aims to assess the relationship between history of pseudotumour formation secondary to metal-on-metal (MoM) implants and periprosthetic joint infection (PJI) rate, as well as establish ESR and CRP thresholds that are suggestive of infection in these patients. We hypothesized that patients with a pseudotumour were at increased risk of infection.

Methods

A total of 1,171 total hip arthroplasty (THA) patients with MoM articulations from August 2000 to March 2014 were retrospectively identified. Of those, 328 patients underwent metal artefact reduction sequence MRI and had minimum two years’ clinical follow-up, and met our inclusion criteria. Data collected included demographic details, surgical indication, laterality, implants used, history of pseudotumour, and their corresponding preoperative ESR (mm/hr) and CRP (mg/dl) levels. Multivariate logistic regression modelling was used to evaluate PJI and history of pseudotumour, and receiver operating characteristic curves were created to assess the diagnostic capabilities of ESR and CRP to determine the presence of infection in patients undergoing revision surgery.


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 833 - 843
1 Jul 2022
Kayani B Baawa-Ameyaw J Fontalis A Tahmassebi J Wardle N Middleton R Stephen A Hutchinson J Haddad FS

Aims

This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized zirconium (OxZr) versus cobalt-chrome (CoCr) femoral heads with ultra-high molecular weight polyethylene (UHMWPE) and highly cross-linked polyethylene (XLPE) liners in total hip arthroplasty (THA).

Methods

Patients undergoing primary THA were recruited from four institutions and prospectively allocated to the following treatment groups: Group A, CoCr femoral head with XLPE liner; Group B, OxZr femoral head with XLPE liner; and Group C, OxZr femoral head with UHMWPE liner. All study patients and assessors recording outcomes were blinded to the treatment groups. The outcomes of 262 study patients were analyzed at ten years’ follow-up.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 1 - 2
1 May 2024
Berry DJ Haddad FS


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 844 - 851
1 Jul 2022
Rogmark C Nåtman J Jobory A Hailer NP Cnudde P

Aims

Patients with femoral neck fractures (FNFs) treated with total hip arthroplasty (THA) have an almost ten-fold increased risk of dislocation compared to patients undergoing elective THA. The surgical approach influences the risk of dislocation. To date, the influence of differing head sizes and dual-mobility components (DMCs) on the risk of dislocation has not been well studied.

Methods

In an observational cohort study on 8,031 FNF patients with THA between January 2005 and December 2014, Swedish Arthroplasty Register data were linked with the National Patient Register, recording the total dislocation rates at one year and revision rates at three years after surgery. The cumulative incidence of events was estimated using the Kaplan-Meier method. Cox multivariable regression models were fitted to calculate adjusted hazard ratios (HRs) with 95% confidence intervals (CIs) for the risk of dislocation, revision, or mortality, stratified by surgical approach.


Bone & Joint 360
Vol. 11, Issue 2 | Pages 15 - 18
1 Apr 2022


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 464 - 471
1 Apr 2022
Veerman K Raessens J Telgt D Smulders K Goosen JHM

Aims

Debridement, antibiotics, and implant retention (DAIR) is a widely accepted form of surgical treatment for patients with an early periprosthetic joint infection (PJI) after primary arthroplasty. The outcome of DAIR after revision arthroplasty, however, has not been reported. The aim of this study was to report the success rate of DAIR after revision arthroplasty with a follow-up of two years.

Methods

This retrospective study, conducted at the Sint Maartenskliniek, Nijmegen, the Netherlands, included 88 patients who underwent DAIR within 90 days of revision total hip or total knee arthroplasty between 2012 and 2019. Details of the surgical procedures and PJI were collected. Univariate analysis and a subgroup analysis of the culture-positive group were performed. Kaplan-Meier survivorship curves were constructed.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 47 - 53
1 May 2024
Jones SA Parker J Horner M

Aims

The aims of this study were to determine the success of a reconstruction algorithm used in major acetabular bone loss, and to further define the indications for custom-made implants in major acetabular bone loss.

Methods

We reviewed a consecutive series of Paprosky type III acetabular defects treated according to a reconstruction algorithm. IIIA defects were planned to use a superior augment and hemispherical acetabular component. IIIB defects were planned to receive either a hemispherical acetabular component plus augments, a cup-cage reconstruction, or a custom-made implant. We used national digital health records and registry reports to identify any reoperation or re-revision procedure and Oxford Hip Score (OHS) for patient-reported outcomes. Implant survival was determined via Kaplan-Meier analysis.


The Bone & Joint Journal
Vol. 106-B, Issue 5 Supple B | Pages 105 - 111
1 May 2024
Apinyankul R Hong C Hwang KL Burket Koltsov JC Amanatullah DF Huddleston JI Maloney WJ Goodman SB

Aims

Instability is a common indication for revision total hip arthroplasty (THA). However, even after the initial revision, some patients continue to have recurrent dislocation. The aim of this study was to assess the risk for recurrent dislocation after revision THA for instability.

Methods

Between 2009 and 2019, 163 patients underwent revision THA for instability at Stanford University Medical Center. Of these, 33 (20.2%) required re-revision due to recurrent dislocation. Cox proportional hazard models, with death and re-revision surgery for periprosthetic infection as competing events, were used to analyze the risk factors, including the size and alignment of the components. Paired t-tests or Wilcoxon signed-rank tests were used to assess the outcome using the Veterans RAND 12 (VR-12) physical and VR-12 mental scores, the Harris Hip Score (HHS) pain and function, and the Hip disability and Osteoarthritis Outcome score for Joint Replacement (HOOS, JR).


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 535 - 537
1 Apr 2007
Evangelista GT Fulkerson E Kummer F Di Cesare PE

During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly recommended that patients with this type of prosthetic head be carefully monitored after a dislocation


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 326 - 332
1 Mar 2013
García-Rey E García-Cimbrelo E Cruz-Pardos A

Between 1999 and 2001, 90 patients underwent total hip replacement using the same uncemented acetabular and femoral components with a 28 mm metallic femoral head but with prospective randomisation of the acetabular liner to either Durasul highly cross-linked polyethylene or nitrogen-sterilised Sulene polyethylene. We assessed 83 patients at a minimum follow-up of ten years. Linear penetration of the femoral head was estimated at six weeks, six and 12 months and annually thereafter, using the Dorr method, given the non-spherical shape of the acetabular component. There was no loosening of any component; only one hip in the Sulene group showed proximal femoral osteolysis. The mean penetration of the femoral head at six weeks was 0.08 mm (0.02 to 0.15) for the Durasul group and 0.16 mm (0.05 to 0.28) for the Sulene group (p = 0.001). The mean yearly linear penetration was 64.8% lower for the Durasul group at 0.05 mm/year (. sd. 0.035) for the Sulene group and 0.02 mm/year (. sd. 0.016) for the Durasul (p < 0.001). Mean linear femoral head penetration at ten years was 61% less in the Durasul than Sulene group. Highly cross-linked polyethylene gives excellent results at ten years. Cite this article: Bone Joint J 2013;95-B:326–32