Advertisement for orthosearch.org.uk
Results 21 - 40 of 69
Results per page:
The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims

The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection.

Methods

We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 543 - 550
1 May 2023
Abel F Avrumova F Goldman SN Abjornson C Lebl DR

Aims

The aim of this study was to assess the accuracy of pedicle screw placement, as well as intraoperative factors, radiation exposure, and complication rates in adult patients with degenerative disorders of the thoracic and lumbar spines who have undergone robotic-navigated spinal surgery using a contemporary system.

Methods

The authors reviewed the prospectively collected data on 196 adult patients who had pedicle screws implanted with robot-navigated assistance (RNA) using the Mazor X Stealth system between June 2019 and March 2022. Pedicle screws were implanted by one experienced spinal surgeon after completion of a learning period. The accuracy of pedicle screw placement was determined using intraoperative 3D fluoroscopy.


Bone & Joint Research
Vol. 12, Issue 4 | Pages 245 - 255
3 Apr 2023
Ryu S So J Ha Y Kuh S Chin D Kim K Cho Y Kim K

Aims

To determine the major risk factors for unplanned reoperations (UROs) following corrective surgery for adult spinal deformity (ASD) and their interactions, using machine learning-based prediction algorithms and game theory.

Methods

Patients who underwent surgery for ASD, with a minimum of two-year follow-up, were retrospectively reviewed. In total, 210 patients were included and randomly allocated into training (70% of the sample size) and test (the remaining 30%) sets to develop the machine learning algorithm. Risk factors were included in the analysis, along with clinical characteristics and parameters acquired through diagnostic radiology.


Bone & Joint Open
Vol. 5, Issue 8 | Pages 644 - 651
7 Aug 2024
Hald JT Knudsen UK Petersen MM Lindberg-Larsen M El-Galaly AB Odgaard A

Aims

The aim of this study was to perform a systematic review and bias evaluation of the current literature to create an overview of risk factors for re-revision following revision total knee arthroplasty (rTKA).

Methods

A systematic search of MEDLINE and Embase was completed in accordance with the Preferred Reporting Items for Systematic reviews and Meta-Analyses (PRISMA) guidelines. The studies were required to include a population of index rTKAs. Primary or secondary outcomes had to be re-revision. The association between preoperative factors and the effect on the risk for re-revision was also required to be reported by the studies.


Bone & Joint Open
Vol. 6, Issue 1 | Pages 82 - 92
14 Jan 2025
Ranieri R Borroni M Delle Rose G Conti M Garofalo R Castagna A

Aims

The aim of this study was to report long-term clinical outcomes of a modern convertible metal-backed glenoid (MBG) in total shoulder arthroplasty (TSA).

Methods

After a minimum of 15 years, a previously studied cohort of 35 patients who received a modern convertible MBG during the period 1996 to 2005 was contacted for clinical and radiological follow-up. At last follow-up, patients were evaluated radiologically and clinically according to the Constant Score, Simple Shoulder Test, and visual analogue scale for pain. Complications and revisions were recorded, and survival analysis was performed.


Bone & Joint Open
Vol. 4, Issue 3 | Pages 146 - 157
7 Mar 2023
Camilleri-Brennan J James S McDaid C Adamson J Jones K O'Carroll G Akhter Z Eltayeb M Sharma H

Aims

Chronic osteomyelitis (COM) of the lower limb in adults can be surgically managed by either limb reconstruction or amputation. This scoping review aims to map the outcomes used in studies surgically managing COM in order to aid future development of a core outcome set.

Methods

A total of 11 databases were searched. A subset of studies published between 1 October 2020 and 1 January 2011 from a larger review mapping research on limb reconstruction and limb amputation for the management of lower limb COM were eligible. All outcomes were extracted and recorded verbatim. Outcomes were grouped and categorized as per the revised Williamson and Clarke taxonomy.


Bone & Joint Open
Vol. 5, Issue 4 | Pages 294 - 303
11 Apr 2024
Smolle MA Fischerauer SF Vukic I Leitner L Puchwein P Widhalm H Leithner A Sadoghi P

Aims

Patients with proximal femoral fractures (PFFs) are often multimorbid, thus unplanned readmissions following surgery are common. We therefore aimed to analyze 30-day and one-year readmission rates, reasons for, and factors associated with, readmission risk in a cohort of patients with surgically treated PFFs across Austria.

Methods

Data from 11,270 patients with PFFs, treated surgically (osteosyntheses, n = 6,435; endoprostheses, n = 4,835) at Austrian hospitals within a one-year period (January to December 2021) was retrieved from the Leistungsorientierte Krankenanstaltenfinanzierung (Achievement-Oriented Hospital Financing). The 30-day and one-year readmission rates were reported. Readmission risk for any complication, as well as general medicine-, internal medicine-, and surgery/injury-associated complications, and factors associated with readmissions, were investigated.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 108 - 112
1 Jan 2009
Chandrasekar CR Grimer RJ Carter SR Tillman RM Abudu A Buckley L

Endoprosthetic replacement of the proximal femur may be required to treat primary bone tumours or destructive metastases either with impending or established pathological fracture. Modular prostheses are available off the shelf and can be adapted to most reconstructive situations for this purpose. We have assessed the clinical and functional outcome of using the METS (Stanmore Implants Worldwide) modular tumour prosthesis to reconstruct the proximal femur in 100 consecutive patients between 2001 and 2006. We compared the results with the published series for patients managed with modular and custom-made endoprosthetic replacements for the same conditions. There were 52 males and 48 females with a mean age of 56.3 years (16 to 84) and a mean follow-up of 24.6 months (0 to 60). In 65 patients the procedure was undertaken for metastases, in 25 for a primary bone tumour, and in ten for other malignant conditions. A total of 46 patients presented with a pathological fracture, and 19 presented with failed fixation of a previous pathological fracture. The overall patient survival was 63.6% at one year and 23.1% at five years, and was significantly better for patients with a primary bone tumour than for those with metastatic tumour (82.3% vs 53.3%, respectively at one year (p = 0.003)). There were six early dislocations of which five could be treated by closed reduction. No patient needed revision surgery for dislocation. Revision surgery was required by six (6%) patients, five for pain caused by acetabular wear and one for tumour progression. Amputation was needed in four patients for local recurrence or infection. The estimated five-year implant survival with revision as the endpoint was 90.7%. The mean Toronto Extremity Salvage score was 61% (51% to 95%). The implant survival and complications resulting from the use of the modular system were comparable to the published series of both custom-made and other modular proximal femoral implants. We conclude that at intermediate follow-up the modular tumour prosthesis for proximal femur replacement provides versatility, a low incidence of implant-related complications and acceptable function for patients with metastatic tumours, pathological fractures and failed fixation of the proximal femur. It also functions as well as a custom-made endoprosthetic replacement


The Bone & Joint Journal
Vol. 104-B, Issue 10 | Pages 1148 - 1155
1 Oct 2022
Watts AC Hamoodi Z McDaid C Hewitt C

Aims

Arthroplasties of the elbow, including total elbow arthroplasty, radial head arthroplasty, distal humeral hemiarthroplasty, and radiocapitellar arthroplasty, are rarely undertaken. This scoping review aims to outline the current research in this area to inform the development of future research.

Methods

A scoping review was undertaken adhering to the Joanna Briggs Institute guidelines using Medline, Embase, CENTRAL, and trial registries, limited to studies published between 1 January 1990 and 7 February 2021. Endnote software was used for screening and selection, and included randomized trials, non-randomized controlled trials, prospective and retrospective cohort studies, case-control studies, analytical cross-sectional studies, and case series of ten or more patients reporting the clinical outcomes of elbow arthroplasty. The results are presented as the number of types of studies, sample size, length of follow-up, clinical outcome domains and instruments used, sources of funding, and a narrative review.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims

The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes.

Methods

A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.


Bone & Joint 360
Vol. 10, Issue 1 | Pages 31 - 33
1 Feb 2021


The Bone & Joint Journal
Vol. 102-B, Issue 9 | Pages 1113 - 1121
14 Sep 2020
Nantha Kumar N Kunutsor SK Fernandez MA Dominguez E Parsons N Costa ML Whitehouse MR

Aims

We conducted a systematic review and meta-analysis to compare the mortality, morbidity, and functional outcomes of cemented versus uncemented hemiarthroplasty in the treatment of intracapsular hip fractures, analyzing contemporary and non-contemporary implants separately.

Methods

PubMed, Medline, EMBASE, CINAHL, and Cochrane Library were searched to 2 February 2020 for randomized controlled trials (RCTs) comparing the primary outcome, mortality, and secondary outcomes of function, quality of life, reoperation, postoperative complications, perioperative outcomes, pain, and length of hospital stay. Relative risks (RRs) and mean differences (with 95% confidence intervals (CIs)) were used as summary association measures.


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 11 - 16
1 Jan 2020
Parker MJ Cawley S

Aims

Debate continues about whether it is better to use a cemented or uncemented hemiarthroplasty to treat a displaced intracapsular fracture of the hip. The aim of this study was to attempt to resolve this issue for contemporary prostheses.

Methods

A total of 400 patients with a displaced intracapsular fracture of the hip were randomized to receive either a cemented polished tapered stem hemiarthroplasty or an uncemented Furlong hydroxyapatite-coated hemiarthroplasty. Follow-up was conducted by a nurse blinded to the implant at set intervals for up to one year from surgery.


Bone & Joint 360
Vol. 9, Issue 3 | Pages 5 - 7
1 Jun 2020
Lebel DE Rocos B


Bone & Joint Research
Vol. 9, Issue 9 | Pages 534 - 542
1 Sep 2020
Varga P Inzana JA Fletcher JWA Hofmann-Fliri L Runer A Südkamp NP Windolf M

Aims

Fixation of osteoporotic proximal humerus fractures remains challenging even with state-of-the-art locking plates. Despite the demonstrated biomechanical benefit of screw tip augmentation with bone cement, the clinical findings have remained unclear, potentially as the optimal augmentation combinations are unknown. The aim of this study was to systematically evaluate the biomechanical benefits of the augmentation options in a humeral locking plate using finite element analysis (FEA).

Methods

A total of 64 cement augmentation configurations were analyzed using six screws of a locking plate to virtually fix unstable three-part fractures in 24 low-density proximal humerus models under three physiological loading cases (4,608 simulations). The biomechanical benefit of augmentation was evaluated through an established FEA methodology using the average peri-screw bone strain as a validated predictor of cyclic cut-out failure.


Bone & Joint 360
Vol. 8, Issue 5 | Pages 33 - 35
1 Oct 2019


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1362 - 1369
1 Nov 2019
Giannicola G Calella P Bigazzi P Mantovani A Spinello P Cinotti G

Aims

The aim of this study was to analyze the results of two radiocapitellar prostheses in a large case series followed prospectively, with medium-term follow-up.

Patients and Methods

A total of 31 patients with a mean age of 54 years (27 to 73) were analyzed; nine had primary osteoarthritis (OA) and 17 had post-traumatic OA, three had capitellar osteonecrosis, and two had a fracture. Overall, 17 Lateral Resurfacing Elbow (LRE) and 14 Uni-Elbow Radio-Capitellum Implant (UNI-E) arthroplasties were performed. Pre- and postoperative assessment involved the Mayo Elbow Performance Score (MEPS), the Quick Disabilities of the Arm, Shoulder and Hand (Q-DASH) score, and the modified American Shoulder Elbow Surgeons (m-ASES) score.


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 627 - 634
1 Jun 2019
King JJ Dalton SS Gulotta LV Wright TW Schoch BS

Aims

Acromial fractures following reverse shoulder arthroplasty (RSA) have a wide range of incidences in reported case series. This study evaluates their incidence following RSA by systematically reviewing the current literature.

Materials and Methods

A systematic review using the search terms “reverse shoulder”, “reverse total shoulder”, or “inverted shoulder” was performed using PubMed, Web of Science, and Cochrane databases between 1 January 2010 and 31 March 2018. Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines were used. Studies were included if they reported on RSA outcomes and the incidence rate of acromial and/or scapular spine fractures. The rate of these fractures was evaluated for primary RSA, revision RSA, RSA indications, and RSA implant design.


Bone & Joint 360
Vol. 7, Issue 1 | Pages 32 - 35
1 Feb 2018