Tibial nonunion represents a spectrum of conditions
which are challenging to treat, and optimal management remains unclear
despite its high rate of incidence. We present 44 consecutive patients
with 46 stiff tibial nonunions, treated with hexapod external fixators
and distraction to achieve union and gradual deformity correction.
There were 31 men and 13 women with a mean age of 35 years (18 to
68) and a mean follow-up of 12 months (6 to 40). No tibial osteotomies
or bone graft procedures were performed. Bony union was achieved
after the initial surgery in 41 (89.1%) tibias. Four persistent
nonunions united after repeat treatment with closed hexapod distraction,
resulting in bony union in 45 (97.8%) patients. The mean time to
union was 23 weeks (11 to 49). Leg-length was restored to within
1 cm of the contralateral side in all tibias. Mechanical alignment
was restored to within 5° of normal in 42 (91.3%) tibias. Closed
distraction of stiff tibial nonunions can predictably lead to union
without further surgery or bone graft. In addition to generating
the required distraction to achieve union, hexapod circular external
fixators can accurately correct concurrent deformities and limb-length
discrepancies. Cite this article:
The purpose of this study was to compare the results and complications of tibial lengthening over an intramedullary nail with treatment using the traditional Ilizarov method. In this matched case study, 16 adult patients underwent 19 tibial lengthening over nails (LON) procedures. For the matched case group, 17 patients who underwent 19 Ilizarov tibial lengthenings were retrospectively matched to the LON group.Objectives
Methods
Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator. A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.Aims
Patients and Methods
The aims of this study were to establish whether composite fixation
(rail-plate) decreases fixator time and related problems in the
management of patients with infected nonunion of tibia with a segmental
defect, without compromising the anatomical and functional outcomes
achieved using the classical Ilizarov technique. We also wished
to study the acceptability of this technique using patient-based
objective criteria. Between January 2012 and January 2015, 14 consecutive patients
were treated for an infected nonunion of the tibia with a gap and
were included in the study. During stage one, a radical debridement
of bone and soft tissue was undertaken with the introduction of
an antibiotic-loaded cement spacer. At the second stage, the tibia
was stabilized using a long lateral locked plate and a six-pin monorail
fixator on its anteromedial surface. A corticotomy was performed
at the appropriate level. During the third stage, i.e. at the end
of the distraction phase, the transported fragment was aligned and
fixed to the plate with two to four screws. An iliac crest autograft
was added to the docking site and the fixator was removed. Functional
outcome was assessed using the Association for the Study and Application
of Methods of Ilizarov (ASAMI) criteria. Patient-reported outcomes
were assessed using the Musculoskeletal Tumor Society (MSTS) score.Aims
Patients and Methods
Computer hexapod assisted orthopaedic surgery (CHAOS), is a method
to achieve the intra-operative correction of long bone deformities
using a hexapod external fixator before definitive internal fixation
with minimally invasive stabilisation techniques. The aims of this study were to determine the reliability of this
method in a consecutive case series of patients undergoing femoral
deformity correction, with a minimum six-month follow-up, to assess
the complications and to define the ideal group of patients for
whom this treatment is appropriate. The medical records and radiographs of all patients who underwent
CHAOS for femoral deformity at our institution between 2005 and
2011 were retrospectively reviewed. Records were available for all
55 consecutive procedures undertaken in 49 patients with a mean
age of 35.6 years (10.9 to 75.3) at the time of surgery.Aims
Patients and Methods
To present our experience of using a combination of intra-articular
osteotomy and external fixation to treat different deformities of
the knee. A total of six patients with a mean age of 26.5 years (15 to
50) with an abnormal hemi-joint line convergence angle (HJLCA) and
mechanical axis deviation (MAD) were included. Elevation of a tibial
hemiplateau or femoral condylar advancement was performed and limb
lengthening with correction of residual deformity using a circular
or monolateral Ilizarov frame.Aims
Patients and Methods
Many authors have reported a shorter treatment time when using trifocal bone transport (TFT) rather than bifocal bone transport (BFT) in the management of long segmental tibial bone defects. However, the difference in the incidence of additional procedures, the true complications, and the final results have not been investigated. A total of 86 consecutive patients with a long tibial bone defect (≥ 8 cm), who were treated between January 2008 and January 2015, were retrospectively reviewed. A total of 45 were treated by BFT and 41 by TFT. The median age of the 45 patients in the BFT group was 43 years (interquartile range (IQR) 23 to 54).Aims
Patients and Methods
A retrospective study was performed in 18 patients
with achondroplasia, who underwent bilateral humeral lengthening
between 2001 and 2013, using monorail external fixators. The mean
age was ten years (six to 15) and the mean follow-up was 40 months
(12 to 104). The mean disabilities of the arm, shoulder and hand (DASH) score
fell from 32.3 (20 to 40)
pre-operatively to 9.4 (6 to 14) post-operatively (p = 0.037). A
mean lengthening of 60% (40% to 95%) was required to reach the goal
of independent perineal hygiene. One patient developed early consolidation,
and fractures occurred in the regenerate bone of four humeri in
three patients. There were three transient radial nerve palsies. Humeral lengthening increases the independence of people with
achondroplasia and is not just a cosmetic procedure. Cite this article:
Anatomical atlases document classical safe corridors for the
placement of transosseous fine wires through the calcaneum during
circular frame external fixation. During this process, the posterior
tibial neurovascular bundle (PTNVB) is placed at risk, though this
has not been previously quantified. We describe a cadaveric study
to investigate a safe technique for posterolateral to anteromedial
fine wire insertion through the body of the calcaneum. A total of 20 embalmed cadaveric lower limbs were divided into
two groups. Wires were inserted using two possible insertion points
and at varying angles. In Group A, wires were inserted one-third
along a line between the point of the heel and the tip of the lateral
malleolus while in Group B, wires were inserted halfway along this
line. Standard dissection techniques identified the structures at
risk and the distance of wires from neurovascular structures was measured.
The results from 19 limbs were subject to analysis.Aims
Materials and Methods
We wished to examine the effectiveness of tibial lengthening
using a two ring Ilizarov frame in skeletally immature patients.
This is a potentially biomechanically unstable construct which risks
the loss of axial control. We retrospectively reviewed a consecutive series of 24 boys and
26 girls, with a mean age of 8.6 years (4 to 14), who underwent
52 tibial lengthening procedures with a mean follow-up of 4.3 years
(4.0 to 16.9). Tibial alignment was measured before and after treatment
using joint orientation lines from the knee and a calculation of
the oblique plane axis.Aims
Patients and Methods
We have carried out a retrospective review of 20 patients with segmental fractures of the tibia who had been treated by circular external fixation. We describe the heterogeneity of these fractures, their association with multiple injuries and the need for multilevel stability with the least compromise of the biology of the fracture segments. The assessment of outcome included union, complications, the measurement of the functional IOWA knee and ankle scores and the general health status (Short-form 36). The mean time to union was 21.7 weeks (12.8 to 31), with no difference being observed between proximal and distal levels of fracture. Complications were encountered in four patients. Two had nonunion at the distal level, one a wire-related infection which required further surgery and another shortening of 15 mm with 8° of valgus which was clinically insignificant. The functional scores for the knee and ankle were good to excellent, but the physical component score of the short-form 36 was lower than the population norm. This may be explained by the presence of multiple injuries affecting the overall score.
Anatomical atlases document safe corridors for placement of wires when using fine-wire circular external fixation. The furthest posterolateral corridor described in the distal tibia is through the fibula. This limits the crossing angle and stability of the frame. In this paper we describe a new, safe Retro-Fibular Wire corridor, which provides greater crossing angles and increased stability. In a cadaver study, 20 formalin-treated legs were divided into two groups. Wires were inserted into the distal quarter of the tibia using two possible corridors and standard techniques of dissection identified the distance of the wires from neurovascular structures. In both groups the posterior tibial neurovascular bundle was avoided. In group A the peroneal artery was at risk. In group B this injury was avoided. Comparison of the groups showed a significant difference (p <
0.001). We recommend the Retro-Fibular wire technique whereby wires are inserted into the tibia mid-way between the posteromedial border of the fibula and the tendo Achillis, at 30° to 45° to the sagittal plane, and introduced from a posterolateral to an anteromedial position. Subsequently, when using this technique in 30 patients, we have had no neurovascular complications or problems relating to tethering of the peroneal tendons.
The last decade has seen a considerable increase
in the use of in total ankle arthroplasty (TAA) to treat patients
with end-stage arthritis of the ankle. However, the longevity of
the implants is still far from that of total knee and hip arthroplasties. The aim of this review is to outline a diagnostic and treatment
algorithm for the painful TAA to be used when considering revision
surgery. Cite this article:
Most problems encountered in complex revision
total knee arthroplasty can be managed with the wide range of implant
systems currently available. Modular metaphyseal sleeves, metallic
augments and cones provide stability even with significant bone
loss. Hinged designs substitute for significant ligamentous deficiencies.
Catastrophic failure that precludes successful reconstruction can
be encountered. The alternatives to arthroplasty in such drastic
situations include knee arthrodesis, resection arthroplasty and
amputation. The relative indications for the selection of these
alternatives are recurrent deep infection, immunocompromised host,
and extensive non-reconstructible bone or soft-tissue defects.
We report our experience of the use of callus distraction with a monolateral fixator for the treatment of acquired radial club-hand deformity after osteomyelitis. Between 1994 and 2004, 13 patients with a mean age of eight years (4 to 15) were treated by callus distraction with a monolateral fixator after a preliminary period of at least four weeks in a corrective short-arm cast. All patients achieved bony union and were satisfied with the functional and cosmetic outcome. There were no major complications, but three patients required cancellous bone grafting at the docking site for delayed union. Local treatment and oral antibiotics were required for pin-site infection in six patients. There were no deep infection or recurrence of osteomyelitis.
The aim of this study was to report the outcome following primary
fixation or a staged protocol for type C fractures of the tibial
plafond. We studied all patients who sustained a complex intra-articular
fracture (AO type C) of the distal tibia over an 11-year period.
The primary short-term outcome was infection. The primary long-term
outcome was the Foot and Ankle Outcome Score (FAOS).Aims
Patients and Methods
We undertook a retrospective analysis of 306
procedures on 233 patients, with a mean age of 12 years (1 to 21),
in order to evaluate the use of somatosensory evoked potential (SSEP)
monitoring for the early detection of nerve compromise during external
fixation procedures for limb lengthening and correction of deformity.
Significant SSEP changes were identified during 58 procedures (19%).
In 32 instances (10.5%) the changes were transient, and resolved
once the surgical cause had been removed. The remaining 26 (8.5%)
were analysed in two groups, depending on whether or not corrective
action had been performed in response to critical changes in the
SSEP recordings. In 16 cases in which no corrective action was taken,
13 (81.2%, 4.2% overall) developed a post-operative neurological
deficit, six of which were permanent and seven temporary, persisting
for five to 18 months. In the ten procedures in which corrective
action was taken, four patients (40%, 1.3% overall) had a temporary
(one to eight months) post-operative neuropathy and six had no deficit. After appropriate intervention in response to SSEP changes, the
incidence and severity of neurological deficits were significantly
reduced, with no cases of permanent neuropathy. SSEP monitoring
showed 100% sensitivity and 91% specificity for the detection of
nerve injury during external fixation. It is an excellent diagnostic
technique for identifying nerve lesions when they are still highly
reversible.