We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
A total of 219 hips in 192 patients aged between
18 and 65 years were randomised to 28-mm metal-on-metal uncemented
total hip replacements (THRs, 107 hips) or hybrid hip resurfacing
(HR, 112 hips). At a mean follow-up of eight years (6.6 to 9.3)
there was no significant difference between the THR and HR groups
regarding rate of revision (4.0% (4 of 99) Cite this article:
Dysplasia of the hip, hypotonia, osteopenia,
ligamentous laxity, and mental retardation increase the complexity
of performing and managing patients with Down syndrome who require
total hip replacement (THR). We identified 14 patients (six males,
eight females, 21 hips) with Down syndrome and degenerative disease
of the hip who underwent THR, with a minimum follow-up of two years
from 1969 to 2009. In seven patients, bilateral THRs were performed
while the rest had unilateral THRs. The mean clinical follow-up
was 5.8 years (standard deviation ( Cite this article:
Metaphyseal bone loss is common with revision
total knee replacement (RTKR). Using the Anderson Orthopaedic Research
Institute (AORI) classification, type 2-B and type 3 defects usually
require large metal blocks, bulk structural allograft or highly
porous metal cones. Tibial and femoral trabecular metal metaphyseal
cones are a unique solution for large bone defects. These cones
substitute for bone loss, improve metaphyseal fixation, help correct
malalignment, restore the joint line and may permit use of a shorter
stem. The technique for insertion involves sculpturing of the remaining
bone with a high speed burr and rasp, followed by press-fit of the
cone into the metaphysis. The fixation and osteoconductive properties
of the porous cone outer surface allow ingrowth and encourage long-term
biological fixation. The revision knee component is then cemented
into the porous cone inner surface, which provides superior fixation
compared with cementing into native but deficient metaphyseal bone.
The advantages of the cone compared with allograft include: technical
ease, biological fixation, no resorption, and possibly a lower risk
of infection. The disadvantages include: difficult extraction and
relatively short-term follow-up. Several studies using cones report
promising short-term results for the reconstruction of large bone
defects in RTKR. Cite this article:
Radiological evidence of post-traumatic osteoarthritis
(PTOA) after fracture of the tibial plateau is common but end-stage arthritis
which requires total knee arthroplasty is much rarer. The aim of this study was to examine the indications for, and
outcomes of, total knee arthroplasty after fracture of the tibial
plateau and to compare this with an age and gender-matched cohort
of TKAs carried out for primary osteoarthritis. Between 1997 and 2011, 31 consecutive patients (23 women, eight
men) with a mean age of 65 years (40 to 89) underwent TKA at a mean
of 24 months (2 to 124) after a fracture of the tibial plateau.
Of these, 24 had undergone ORIF and seven had been treated non-operatively.
Patients were assessed pre-operatively and at 6, 12 and >
60 months
using the Short Form-12, Oxford Knee Score and a patient satisfaction
score. Patients with instability or nonunion needed total knee arthroplasty
earlier (14 and 13.3 months post-injury) than those with intra-articular
malunion (50 months, p <
0.001). Primary cruciate-retaining implants
were used in 27 (87%) patients. Complication rates were higher in
the PTOA cohort and included wound complications (13% Total knee arthroplasty undertaken after fracture of the tibial
plateau has a higher rate of complications than that undertaken
for primary osteoarthritis, but patient-reported outcomes and satisfaction
are comparable. Cite this article:
The October 2014 Spine Roundup360 looks at: microdiscectomy is not exactly a hands-down winner; lumbar spinal stenosis unpicked; Wallis implant helpful in lumbosacral decompression; multidisciplinary rehabilitation is good for back pain; and understanding the sciatic stretch test.
A retrospective review was performed of patients
undergoing primary cementless total knee replacement (TKR) using
porous tantalum performed by a group of surgical trainees. Clinical
and radiological follow-up involved 79 females and 26 males encompassing
115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up
was 7 years (2 to 11). Tibial and patellar components were porous
tantalum monoblock implants, and femoral components were posterior
stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological
assessments were made for implant positioning, alignment, radiolucencies,
lysis, and loosening. There was 95.7% survival of implants. There
was no radiological evidence of loosening and no osteolysis found.
No revisions were performed for aseptic loosening. Average tibial
component alignment was 1.4° of varus (4°of valgus to 9° varus),
and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean
femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral
alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar
tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient
satisfaction with improvement in pain was 91%. Cementless TKR incorporating
porous tantalum yielded good clinical and radiological outcomes
at a mean of follow-up of seven-years. Cite this article:
Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability. Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°,
We describe the early results of a prospective study of 100 total ankle replacements (96 patients) at a single centre using the Mobility Total Ankle Replacement. At final review, six patients had died and five ankles (5%) had been revised, two by fusion and three by exchange of components. All remaining patients were reviewed at a minimum of three years. The mean follow-up was 43 months (4 to 63). The three-year survival was 97% (95% confidence interval (CI) 91 to 99). The four-year survival was 93.6% (95% CI 84.7 to 97.4). The portion of bony interface that was visible on plain radiograph was divided into 15 zones and a radiolucent line or osteolytic cavity was seen in one zone in 14 ankles. It was not seen in more than one zone. In five ankles it was >
10 mm in width. This study suggests that the early outcome of ankle replacement is comparable to that of other total joint replacements.
We report the clinical and radiographic outcome of a consecutive series of 138 hydroxyapatite-coated total knee replacements with a mean follow-up of 11 years (10 to 13). The patients were entered into a prospective study and all living patients (76 knees) were evaluated. The Hospital for Special Surgery knee score was obtained for comparison with the pre-operative situation. No patient was lost to follow-up. Radiographic assessment revealed no loosening. Seven prostheses have been revised, giving a cumulative survival rate of 93% at 13 years. We believe this to be the longest follow-up report available for an hydroxyapatite-coated knee replacement and the first for this design of Insall-Burstein II knee.
We developed a method of applying vibration to the impaction bone grafting process and assessed its effect on the mechanical properties of the impacted graft. Washed morsellised bovine femoral heads were impacted into shear test rings. A range of frequencies of vibration was tested, as measured using an accelerometer housed in a vibration chamber. Each shear test was repeated at four different normal loads to generate stress-strain curves. The Mohr-Coulomb failure envelope from which shear strength and interlocking values are derived was plotted for each test. The experiments were repeated with the addition of blood in order to replicate a saturated environment. Graft impacted with the addition of vibration at all frequencies showed improved shear strength when compared with impaction without vibration, with 60 Hz giving the largest effect. Under saturated conditions the addition of vibration was detrimental to the shear strength of the aggregate. The civil-engineering principles of particulate settlement and interlocking also apply to impaction bone grafting. Although previous studies have shown that vibration may be beneficial in impaction bone grafting on the femoral side, our study suggests that the same is not true in acetabular revision.
The presacral retroperitoneal approach for axial lumbar interbody fusion (presacral ALIF) is not widely reported, particularly with regard to the mid-term outcome. This prospective study describes the clinical outcomes, complications and rates of fusion at a follow-up of two years for 26 patients who underwent this minimally invasive technique along with further stabilisation using pedicle screws. The fusion was single-level at the L5-S1 spinal segment in 17 patients and two-level at L4–5 and L5-S1 in the other nine. The visual analogue scale for pain and Oswestry Disability Index scores were recorded pre-operatively and during the 24-month study period. The evaluation of fusion was by thin-cut CT scans at six and 12 months, and flexion-extension plain radiographs at six, 12 and 24 months. Significant reductions in pain and disability occurred as early as three weeks postoperatively and were maintained. Fusion was achieved in 22 of 24 patients (92%) at 12 months and in 23 patients (96%) at 24 months. One patient (4%) with a pseudarthrosis underwent successful revision by augmentation of the posterolateral fusion mass through a standard open midline approach. There were no severe adverse events associated with presacral ALIF, which in this series demonstrated clinical outcomes and fusion rates comparable with those of reports of other methods of interbody fusion.
The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.
We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed.
The use of impaction bone grafting during revision arthroplasty of the hip in the presence of cortical defects has a high risk of post-operative fracture. Our laboratory study addressed the effect of extramedullary augmentation and length of femoral stem on the initial stability of the prosthesis and the risk of fracture. Cortical defects in plastic femora were repaired using either surgical mesh without extramedullary augmentation, mesh with a strut graft or mesh with a plate. After bone impaction, standard or long-stem Exeter prostheses were inserted, which were tested by cyclical loading while measuring defect strain and migration of the stem. Compared with standard stems without extramedullary augmentation, defect strains were 31% lower with longer stems, 43% lower with a plate and 50% lower with a strut graft. Combining extramedullary augmentation with a long stem showed little additional benefit (p = 0.67). The type of repair did not affect the initial stability. Our results support the use of impaction bone grafting and extramedullary augmentation of diaphyseal defects after mesh containment.
We report the five year outcomes of a two-stage
approach for infected total hip replacement. This is a single-surgeon
experience at a tertiary centre where the more straightforward cases
are treated using single-stage exchange. This study highlights the
vital role of the multidisciplinary team in managing these cases. A total of 125 patients (51 male, 74 female) with a mean age
of 68 years (42 to 78) were reviewed prospectively. Functional status
was assessed using the Harris hip score (HHS). The mean HHS improved
from 38 (6 to 78.5) pre-operatively to 81.2 (33 to 98) post-operatively.
Staphylococcus species were isolated in 85 patients (68%). The rate of control of infection was 96% at five years. In all,
19 patients died during the period of the study. This represented
a one year mortality of 0.8% and an overall mortality of 15.2% at
five years. No patients were lost to follow-up. We report excellent control of infection in a series of complex
patients and infections using a two-stage revision protocol supported
by a multidisciplinary approach. The reason for the high rate of
mortality in these patients is not known. Cite this article:
The aim of this experimental study on New Zealand’s white rabbits
was to investigate the transplantation of autogenous growth plate
cells in order to treat the injured growth plate. They were assessed
in terms of measurements of radiological tibial varus and histological
characteristics. An experimental model of plate growth medial partial resection
of the tibia in 14 New Zealand white rabbits was created. During
this surgical procedure the plate growth cells were collected and
cultured. While the second surgery was being performed, the autologous
cultured growth plate cells were grafted at the right tibia, whereas
the left tibia was used as a control group. Objectives
Methods
An experimental sheep model was used for impaction allografting of 12 hemiarthroplasty femoral components placed into two equal-sized groups. In group 1, a 50:50 mixture of ApaPore hydroxyapatite bone-graft substitute and allograft was used. In group 2, ApaPore and allograft were mixed in a 90:10 ratio. Both groups were killed at six months. Ground reaction force results demonstrated no significant differences (p >
0.05) between the two groups at 8, 16 and 24 weeks post-operatively, and all animals remained active. The mean bone turnover rates were significantly greater in group 1, at 0.00206 mm/day, compared to group 2 at 0.0013 mm/day (p <
0.05). The results for the area of new bone formation demonstrated no significant differences (p >
0.05) between the two groups. No significant differences were found between the two groups in thickness of the cement mantle (p >
0.05) and percentage ApaPore-bone contact (p >
0.05). The results of this animal study demonstrated that a mixture of ApaPore allograft in a 90:10 ratio was comparable to using a 50:50 mixture.
We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed. Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.
The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score.Objectives
Methods