Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series. A total of 610 Birmingham Hip Resurfacing arthroplasties were performed in 532 patients with a mean age of 51.8 years (16.5 to 81.6). They were followed for between two and eight years; 107 patients (120 hips) had been followed up for more than five years. Two patients were lost to follow-up. At a minimum of five years’ follow-up, 79 of 85 hips (93%) had an excellent or good outcome according to the Harris hip score. The mean Oxford hip score was 16.1 points ( Considering that these patients are young and active these results are good, and support the use of resurfacing. Further study is needed to address the early failures, particularly those related to fracture and metal debris.
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
We reviewed the long-term radiological outcome,
complications and revision operations in 19 children with quadriplegic
cerebral palsy and hip dysplasia who underwent combined peri-iliac
osteotomy and femoral varus derotation osteotomy. They had a mean
age of 7.5 years (1.6 to 10.9) and comprised 22 hip dislocations
and subluxations. We also studied the outcome for the contralateral
hip. At a mean follow-up of 11.7 years (10 to 15.1) the Melbourne
cerebral palsy (CP) hip classification was grade 2 in 16 hips, grade
3 in five, and grade 5 in one. There were five complications seen
in four hips (21%, four patients), including one dislocation, one
subluxation, one coxa vara with adduction deformity, one subtrochanteric
fracture and one infection. A recurrent soft-tissue contracture occurred
in five hips and ten required revision surgery. In pre-adolescent children with quadriplegic cerebral palsy good
long-term outcomes can be achieved after reconstruction of the hip;
regular follow-up is required.
Cite this article:
Posterolateral rotatory instability is the most common type of symptomatic chronic instability of the elbow. In this condition the forearm complex rotates externally in relation to the humerus, causing posterior subluxation or dislocation of the radial head. The lateral ligament complex, radial head and coronoid process are important constraints to posterolateral rotatory instability, and their disruption is involved in the pathogenesis of this condition. The diagnosis relies on a high index of clinical suspicion, active and passive apprehension tests, and examination under anaesthesia. Surgical treatment has given consistently successful results. Open reconstruction of the lateral ligaments with a tendon graft has been the procedure of choice, with arthroscopic techniques emerging as a potential alternative.
Between 1987 and 2006 we performed a modified Thompson’s quadricepsplasty on 40 fracture-related stiff knees and followed the patients for a mean of 7.9 years (2 to 11.1). The factors affecting the final gain of movement were investigated. A total of 15 knees required lengthening of the rectus femoris. The mean flexion gain was 70.2° (42.3° to 112.5°). According to Judet’s criteria, the results were excellent in 30 knees, good in seven, and fair in three. The range of movement which was achieved intra-operatively was related to the gain of knee flexion on univariate analysis. Five patients had complications: deep infection in one, recurrent patellar dislocation in one, and rupture of the extensor mechanism in three. This modified technique gives satisfactory results. Achieving maximum knee flexion intra-operatively seems to be the most important factor in enhancing the outcome in patients with stiffness of the knee following fracture.
The Kent hip is a distally-locked femoral stem which was developed to address severe proximal bone loss, severe bony deformity and peri-prosthetic fracture. We reviewed the results of 145 consecutive Kent hips implanted into 141 patients between 1987 and 2000. The indications for implantation were aseptic loosening (75 hips), septic loosening (two), peri-prosthetic and prosthetic fracture (37), severe bony deformity (24), and fracture through a proximal femoral metastasis (seven). The median time to full weight-bearing after surgery was two days and the mean length of follow-up was 5.1 years (2 to 15). Further revisions were required for 13 femoral stems. With removal of the stem for any reason as an end-point, the cumulative survival at five, ten and 15 years was 93%, 89% and 77%, respectively. In patients aged ≥ 70 years, the cumulative survival at 15 years was 92%, compared with 68% in those aged <
70 years. Because of these findings, we recommend the use of interlocking stems in patients aged ≥ 70 years, particularly in those with a peri-prosthetic fracture, for whom alternative methods are limited. Outcome scores and survival data, compared with other systems, indicate that the Kent hip should be used with caution in younger patients.
The June 2012 Shoulder &
Elbow Roundup360 looks at: reverse shoulder replacement; torn rotator cuffs and platelet-rich fibrin; rotator cuff repair; frozen shoulder; whether an arthroscopic rotator cuff repair actually heals; the torn rotator cuff’s effect on activities of daily living; subacromial impingement; how to improve the reliability of the Constant-Murley score; and failure of the Neer modification of an open Bankart procedure.
We retrospectively evaluated 42 hips which had undergone acetabular reconstruction using the Kerboull acetabular reinforcement device between September 1994 and December 1998. We used autogenous bone chips from the ilium and ceramic particle morsellised grafts, even in large acetabular bone defects, in the early stages of the study. Thereafter, femoral head allograft was used as bulk graft in patients with large acetabular defects. Ceramic blocks and the patients’ contralateral femoral head were also used as bulk graft. The mean follow-up period was 8.7 years (4.3 to 12). Survivorship analysis was performed using radiological failure of the acetabular component, irrespective of whether it was revised, or not, as the end-point. The survival rate of the morsellised graft group (25 hips) and the bulk graft group (17 hips) at ten years was 53% (95% confidence interval (CI) 42.5% to 63.5%) and 82% (95% CI 72.4% to 91.6%), respectively. The mid-term results of revision total hip replacement with the Kerboull device were better when bulk graft was used in any size of bone defect.
There are no recent studies comparing cable with
wire for the fixation of osteotomies or fractures in total hip replacement
(THR). Our objective was to evaluate the five-year clinical and
radiological outcomes and complication rates of the two techniques.
We undertook a review including all primary and revision THRs performed
in one hospital between 1996 and 2005 using cable or wire fixation.
Clinical and radiological evaluation was performed five years post-operatively.
Cables were used in 51 THRs and wires in 126, and of these, 36 THRs
with cable (71%) and 101 with wire (80%) were evaluated at follow-up.
The five-year radiographs available for 33 cable and 91 wire THRs
revealed rates of breakage of fixation of 12 of 33 (36%) and 42
of 91 (46%), respectively. With cable there was a significantly
higher risk of metal debris (68% In conclusion, we found a higher incidence of complications and
a trend towards increased infection and foreign-body reaction with
the use of cables.
We describe the results of 81 consecutive revision
total hip replacements with impaction grafting in 79 patients using
a collared polished chrome–cobalt stem, customised in length according
to the extent of distal bone loss. Our hypothesis was that the features
of this stem would reduce the rate of femoral fracture and subsidence
of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative
fracture or significant subsidence occurred. Only one patient suffered
a post-operative diaphyseal fracture, which was associated with
a fall. All but one femur showed incorporation of the graft. No
revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using
further femoral revision as the endpoint, was 100% (95% confidence
interval (CI) 95.9 to 100), and at nine years using re-operation
for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem
individually adapted to the extent of bone loss and with a collar
may reduce subsidence and the rate of fracture while maintaining
the durability of the fixation.
Dislocation of the shoulder may occur during
seizures in epileptics and other patients who have convulsions. Following
the initial injury, recurrent instability is common owing to a tendency
to develop large bony abnormalities of the humeral head and glenoid
and a susceptibility to further seizures. Assessment is difficult
and diagnosis may be missed, resulting in chronic locked dislocations
with protracted morbidity. Many patients have medical comorbidities,
and successful treatment requires a multidisciplinary approach addressing
the underlying seizure disorder in addition to the shoulder pathology.
The use of bony augmentation procedures may have improved the outcomes
after surgical intervention, but currently there is no evidence-based
consensus to guide treatment. This review outlines the epidemiology
and pathoanatomy of seizure-related instability, summarising the
currently-favoured options for treatment, and their results.
Periprosthetic fracture of the femur is an uncommon complication after total hip replacement, but appears to be increasing. We undertook a nationwide observational study to determine the risk factors for failure after treatment of these fractures, examining patient- and implant-related factors, the classification of the fractures and the outcome. Between 1979 and 2000, 1049 periprosthetic fractures of the femur were reported to the Swedish National Hip Arthroplasty Register. Of these, 245 had a further operation after failure of their initial management. Data were collected from the Register and hospital records. The material was analysed by the use of Poisson regression models. It was found that the risk of failure of treatment was reduced for Vancouver type B2 injuries (p = 0.0053) if revision of the implant was undertaken (p = 0.0033) or revision and open reduction and internal fixation (p = 0.0039) were performed. Fractures classified as Vancouver type B1 had a significantly higher risk of failure (p = 0.0001). The strongest negative factor was the use of a single plate for fixation (p = 0.001). The most common reasons for failure in this group were loosening of the femoral prosthesis, nonunion and re-fracture. It is probable that many fractures classified as Vancouver type B1 (n = 304), were in reality type B2 fractures with a loose stem which were not recognised. Plate fixation was inadequate in these cases. The difficulty in separating type B1 from type B2 fractures suggests that the prosthesis should be considered as loose until proven otherwise.
We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up. Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain. Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.
Recent events have highlighted the importance
of implant design for survival and wear-related complications following
metal-on-metal hip resurfacing arthroplasty. The mid-term survival
of the most widely used implant, the Birmingham Hip Resurfacing
(BHR), has been described by its designers. The aim of this study
was to report the ten-year survival and patient-reported functional
outcome of the BHR from an independent centre. In this cohort of 554 patients (646 BHRs) with a mean age of
51.9 years (16.5 to 81.5) followed for a mean of eight years (1
to 12), the survival and patient-reported functional outcome depended
on gender and the size of the implant. In female hips (n = 267)
the ten-year survival was 74% (95% confidence interval (CI) 83 to
91), the ten-year revision rate for pseudotumour was 7%, the mean
Oxford hip score (OHS) was 43 ( This study supports the ongoing use of resurfacing in young active
men, who are a subgroup of patients who tend to have problems with
conventional THR. In contrast, the results in women have been poor
and we do not recommend metal-on-metal resurfacing in women. Continuous
follow-up is recommended because of the increasing incidence of
pseudotumour with the passage of time.
Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated We found that long-term alumina wear in association with a loose
acetabular component could be categorised into three groups. Of
20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight
‘severe wear’, which was characterised by a change in the physical
shape of the bearing and a loss of volume. This suggests that the
wear in alumina–alumina bearings in association with a loose acetabular
component may be variable in pattern, and may explain, in part,
why the wear of a ceramic head
The painful subluxed or dislocated hip in adults
with cerebral palsy presents a challenging problem. Prosthetic dislocation
and heterotopic ossification are particular concerns. We present
the first reported series of 19 such patients (20 hips) treated
with hip resurfacing and proximal femoral osteotomy. The pre-operative
Gross Motor Function Classification System (GMFCS) was level V in
13 (68%) patients, level IV in three (16%), level III in one (5%) and
level II in two (11%). The mean age at operation was 37 years (13
to 57). The mean follow-up was 8.0 years (2.7 to 11.6), and 16 of the
18 (89%) contactable patients or their carers felt that the surgery
had been worthwhile. Pain was relieved in 16 of the 18 surviving
hips (89%) at the last follow-up, and the GMFCS level had improved
in seven (37%) patients. There were two (10%) early dislocations;
three hips (15%) required revision of femoral fixation, and two
hips (10%) required revision, for late traumatic fracture of the
femoral neck and extra-articular impingement, respectively. Hence
there were significant surgical complications in a total of seven
hips (35%). No hips required revision for instability, and there
were no cases of heterotopic ossification. We recommend hip resurfacing with proximal femoral osteotomy
for the treatment of the painful subluxed or dislocated hip in patients
with cerebral palsy.
A 16-year-old professional female ballet student sustained a plantar flexion-inversion injury to her left ankle while dancing. Clinical examination and MRI suggested subluxation of the tibiotalar joint. However, accurate diagnosis was hampered by a transient palsy of the common peroneal nerve. It was subsequently established that she had also sustained a dislocation of her calcaneocuboid joint, a rare injury, which was successfully stabilised by using a hamstring graft. The presentation and management of this rare condition are discussed.
There has been a substantial increase in the
number of hip and knee prostheses implanted in recent years, with
a consequent increase in the number of revisions required. Total
femur replacement (TFR) following destruction of the entire femur,
usually after several previous revision operations, is a rare procedure
but is the only way of avoiding amputation. Intramedullary femur
replacement (IFR) with preservation of the femoral diaphysis is
a modification of TFR. Between 1999 and 2010, 27 patients with non-oncological
conditions underwent surgery in our department with either IFR (n
= 15) or TFR (n = 12) and were included in this study retrospectively.
The aim of the study was to assess the indications, complications
and outcomes of IFR and TFR in revision cases. The mean follow-up
period was 31.3 months (6 to 90). Complications developed in 37%
of cases, 33% in the IFR group and 4% in the TFR group. Despite
a trend towards a slightly better functional outcome compared with
TFR, the indication for intramedullary femur replacement should
be established on a very strict basis in view of the procedure’s
much higher complication rate.
We have assessed the long-term results of 292 cemented total hip replacements which were performed for developmental dysplasia of the hip in 206 patients. The mean age of the patients at operation was 42.6 years (15.9 to 79.5) and most (202) were women. The severity of dysplasia was graded according to both the Crowe and the Hartofilakidis classifications. A 22.25-mm Charnley head was always used and the acetabular components were inserted with cement into the true acetabulum. Bone grafting of the acetabulum, using the patient’s own femoral head, was performed on 48 occasions. At a mean follow-up of 15.7 years (2.2 to 31.2) the overall survival of the acetabular component was 78%. The main cause of revision was aseptic loosening (88.3%). The rate of survival at 20 years based on the Hartofilakidis classification was 76% in the dysplastic, 55% in the low-dislocation and 12% in the high-dislocation groups and on the Crowe classification, 72.7% for group I, 70.7% for group II, 36.7% for group III and 15.6% for group IV. There was no statistical correlation between bone grafting of the acetabulum and survival of the acetabular component. This study has shown a higher rate of failure of the acetabular component with increasing severity of hip dysplasia.
This prospective study compares the outcome of
157 hydroxyapatite (HA)-coated tibial components with 164 cemented
components in the ROCC Rotating Platform total knee replacement
in 291 patients. The mean follow-up was 7.6 years (5.2 to 11). There
were two revisions for loosening: one for an HA-coated and one for
a cemented tibial component. Radiological evaluation demonstrated
no radiolucent lines with the HA-coated femoral components. A total
of three HA-coated tibial components exhibited radiolucent lines
at three months post-operatively and these disappeared after three
further months of protected weight-bearing. With HA-coated components
the operating time was shorter (p <
0.006) and the radiological
assessment of the tibial interface was more stable (p <
0.01).
Using revision for aseptic loosening of the tibial component as
the end point, the survival rates at nine years was identical for
both groups at 99.1%. Our results suggest that HA-coated components perform at least
as well as the same design with cemented components and compare
favourably with those of series describing cemented or porous-coated
knee replacements, suggesting that fixation of both components with
hydroxyapatite is a reliable option in primary total knee replacement.