Metal-on-metal bearings are being increasingly used in young patients. The potential adverse effects of systemic metal ion elevation are the subject of ongoing investigation. The purpose of this study was to investigate whether cobalt and chromium ions cross the placenta of pregnant women with a metal-on-metal hip resurfacing and reach the developing fetus. Whole blood levels were estimated using high-resolution inductively-coupled plasma mass spectrometry. Our findings showed that cobalt and chromium are able to cross the placenta in the study patients with metal-on-metal hip resurfacings and in control subjects without any metal implants. In the study group the mean concentrations of cobalt and chromium in the maternal blood were 1.39 μg/l (0.55 to 2.55) and 1.28 μg/l (0.52 to 2.39), respectively. The mean umbilical cord blood concentrations of cobalt and chromium were comparatively lower, at 0.839 μg/l (0.42 to 1.75) and 0.378 μg/l (0.14 to 1.03), respectively, and this difference was significant with respect to chromium (p <
0.05). In the control group, the mean concentrations of cobalt and chromium in the maternal blood were 0.341 μg/l (0.18 to 0.54) and 0.199 μg/l (0.12 to 0.33), and in the umbilical cord blood they were 0.336 μg/l (0.17 to 0.5) and 0.194 μg/l (0.11 to 0.56), respectively. The differences between the maternal and umbilical cord blood levels in the controls were marginal, and not statistically significant (p >
0.05). The mean cord blood level of cobalt in the study patients was significantly greater than that in the control group (p <
0.01). Although the mean umbilical cord blood chromium level was nearly twice as high in the study patients (0.378 μg/l) as in the controls (0.1934 μg/l), this difference was not statistically significant. (p >
0.05) The transplacental transfer rate was in excess of 95% in the controls for both metals, but only 29% for chromium and 60% for cobalt in study patients, suggesting that the placenta exerts a modulatory effect on the rate of metal ion transfer.
Oxidised zirconium (OxZi) has been developed
as an alternative bearing surface for femoral heads in total hip arthroplasty
(THA). This study has investigated polyethylene wear, functional
outcomes and complications, comparing OxZi and cobalt–chrome (CoCr)
as part of a three-arm, multicentre randomised controlled trial.
Patients undergoing THA from four institutions were prospectively
randomised into three groups. Group A received a CoCr femoral head
and highly cross-linked polyethylene (XLPE) liner; Group B received
an OxZi femoral head and XLPE liner; Group C received an OxZi femoral
head and ultra-high molecular weight polyethylene (UHMWPE) liner.
At five years, 368 patients had no statistically significant differences
in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western
Ontario and McMaster Universities Osteoarthritis Index (p = 0.847),
pain scores
(p = 0.458) or complications. The mean rate of linear wear was 0.028
mm/year (standard deviation ( Cite this article:
The August 2013 Hip &
Pelvis Roundup360 looks at: are we getting it right first time?; tantalum augments in revision hip surgery; lower wear in dual mobility?; changing faces changes outcomes; synovial fluid aspiration in MOM hips; taper disease: the new epidemic of hip surgery; the super-obese and THR; and whether well fixed stems can remain in infected hips
The April 2012 Hip &
Pelvis Roundup360 looks at osteoporotic hip fractures, retrotrochanteric pain, fibrin adhesive and reattachment of articular cartilage, autologous bone marrow mononuclear cells and avascular necrosis, bearing surfaces, stability after THR, digital templating, pelvic tilt after THR, custom-made sockets for DDH, and dogs and THR
We evaluated the clinical results and complications
after extra-articular resection of the distal femur and/or proximal
tibia and reconstruction with a tumour endoprosthesis (MUTARS) in
59 patients (mean age 33 years (11 to 74)) with malignant bone or
soft-tissue tumours. According to a Kaplan–Meier analysis, limb
survival was 76% (95% confidence interval (CI) 64.1 to 88.5) after
a mean follow-up of 4.7 years (one month to 17 years). Peri-prosthetic infection
was the most common indication for subsequent amputation (eight
patients). Survival of the prosthesis without revision was 48% (95%
CI 34.8 to 62.0) at two years and 25% (95% CI 11.1 to 39.9) at five years
post-operatively. Failure of the prosthesis was due to deep infection
in 22 patients (37%), aseptic loosening in ten patients (17%), and
peri-prosthetic fracture in six patients (10%). Wear of the bearings
made a minor revision necessary in 12 patients (20%). The mean Musculoskeletal
Tumor Society score was 23 (10 to 29). An extensor lag >
10° was
noted in ten patients (17%). These results suggest that limb salvage after extra-articular
resection with a tumour prosthesis can achieve good functional results
in most patients, although the rates of complications and subsequent
amputation are higher than in patients treated with intra-articular
resection. Cite this article:
With the established success of the National
Joint Registry and the emergence of a range of new national initiatives for
the capture of electronic data in the National Health Service, orthopaedic
surgery in the United Kingdom has found itself thrust to the forefront
of an information revolution. In this review we consider the benefits
and threats that this revolution poses, and how orthopaedic surgeons
should marshal their resources to ensure that this is a force for
good.
The aim of this study was to assess the effect
of frictional torque and bending moment on fretting corrosion at
the taper interface of a modular femoral component and to investigate
whether different combinations of material also had an effect. The
combinations we examined were 1) cobalt–chromium (CoCr) heads on
CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic
heads on CoCr stems. In test 1 increasing torque was imposed by offsetting the stem
in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and
8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14
Nm and 18 Nm. In test 2 we investigated the effect of increasing the bending
moment by offsetting the application of axial load from the midline
in the mediolateral plane. Increments of offset equivalent to head
+ 0 mm, head + 7 mm and head + 14 mm were used. Significantly higher currents and amplitudes were seen with increasing
torque for all combinations of material. However, Ti stems showed
the highest corrosion currents. Increased bending moments associated
with using larger offset heads produced more corrosion: Ti stems
generally performed worse than CoCr stems. Using ceramic heads did
not prevent corrosion, but reduced it significantly in all loading
configurations. Cite this article:
The purpose of this study was twofold: first,
to determine whether the five-year results of hip resurfacing arthroplasty
(HRA) in Canada justified the continued use of HRA; and second,
to identify whether greater refinement of patient selection was
warranted. This was a retrospective cohort study that involved a review
of 2773 HRAs performed between January 2001 and December 2008 at
11 Canadian centres. Cox’s proportional hazards models were used
to analyse the predictors of failure of HRA. Kaplan–Meier survival
analysis was performed to predict the cumulative survival rate at
five years. The factors analysed included age, gender, body mass
index, pre-operative hip pathology, surgeon’s experience, surgical
approach, implant sizes and implant types. The most common modes
of failure were also analysed. The 2773 HRAs were undertaken in 2450 patients: 2127 in men and
646 in women. The mean age at operation was 50.5 years ( The failure rates of HRA at five years justify the ongoing use
of this technique in men. Female gender is an independent predictor
of failure, and a higher failure rate at five years in women leads
the authors to recommend this technique only in exceptional circumstances
for women. Cite this article:
Ceramic-on-metal (CoM) is a relatively new bearing
combination for total hip arthroplasty (THA) with few reported outcomes.
A total of 287 CoM THAs were carried out in 271 patients (mean age
55.6 years (20 to 77), 150 THAs in female patients, 137 in male)
under the care of a single surgeon between October 2007 and October
2009. With the issues surrounding metal-on-metal bearings the decision
was taken to review these patients between March and November 2011,
at a mean follow-up of 34 months (23 to 45) and to record pain,
outcome scores, radiological analysis and blood ion levels. The
mean Oxford Hip Score was 19.2 (12 to 53), 254 patients with 268
hips (95%) had mild/very mild/no pain, the mean angle of inclination
of the acetabular component was 44.8o (28o to
63o), 82 stems (29%) had evidence of radiolucent lines
of >
1 mm in at least one Gruen zone and the median levels of cobalt
and chromium ions in the blood were 0.83 μg/L (0.24 μg/L to 27.56 μg/L)
and 0.78 μg/L (0.21 μg/L to 8.84 μg/L), respectively. The five-year
survival rate is 96.9% (95% confidence interval 94.7% to 99%). Due to the presence of radiolucent lines and the higher than
expected levels of metal ions in the blood, we would not recommend
the use of CoM THA without further long-term follow-up. We plan
to monitor all these patients regularly. Cite this article:
Cementless acetabular fixation has demonstrated superior long-term durability in total hip replacement, but most series have studied implants with porous metal surfaces. We retrospectively evaluated the results of 100 consecutive patients undergoing total hip replacement where a non-porous Allofit component was used for primary press-fit fixation. This implant is titanium alloy, grit-blasted, with a macrostructure of forged teeth and has a biradial shape. A total of 81 patients (82 hips) were evaluated at final follow-up at a mean of 10.1 years (8.9 to 11.9). The Harris Hip Score improved from a mean 53 points (23 to 73) pre-operatively to a mean of 96 points (78 to 100) at final review. The osseointegration of all acetabular components was radiologically evaluated with no evidence of loosening. The survival rate with revision of the component as the endpoint was 97.5% (95% confidence interval 94 to 100) after 11.9 years. Radiolucency was found in one DeLee-Charnley zone in four acetabular components. None of the implants required revision for aseptic loosening. Two patients were treated for infection, one requiring a two-stage revision of the implant. One femoral stem was revised for osteolysis due to the production of metal wear debris, but the acetabular shell did not require revision. This study demonstrates that a non-porous titanium acetabular component with adjunct surface fixation offers an alternative to standard porous-coated implants.
Implant-associated infection is a major source
of morbidity in orthopaedic surgery. There has been extensive research
into the development of materials that prevent biofilm formation,
and hence, reduce the risk of infection. Silver nanoparticle technology
is receiving much interest in the field of orthopaedics for its
antimicrobial properties, and the results of studies to date are
encouraging. Antimicrobial effects have been seen when silver nanoparticles are
used in trauma implants, tumour prostheses, bone cement, and also
when combined with hydroxyapatite coatings. Although there are promising
results with Cite this article:
A high radiographic inclination angle (RI) contributes
to accelerated wear and has been associated with dislocation after
total hip arthroplasty (THA). With freehand positioning of the acetabular
component there is a lack of accuracy, with a trend towards a high
radiographic inclination angle. The aim of this study was to investigate
whether the use of a digital protractor to measure the operative
inclination angle (OI) could improve the positioning of the acetabular
component in relation to a ‘safe zone’. We measured the radiographic inclination angles of 200 consecutive
uncemented primary THAs. In the first 100 the component was introduced
freehand and in the second 100 a digital protractor was used to
measure the operative inclination angle. The mean difference between the operative and the radiographic
inclination angles
(∆RI–OI) in the second cohort was 12.3° (3.8° to 19.8°). There was
a strong correlation between the circumference of the hip and ∆RI–OI.
The number of RI outliers was significantly reduced in the protractor
group (p = 0.002). Adjusting the OI, using a digital protractor and taking into
account the circumference of the patient’s hip, improves the RI
significantly (p <
0.001) and does not require additional operating
time. Cite this article:
Dislocation remains among the most common complications
of, and reasons for, revision of both primary and revision total
hip replacements (THR). Hence, there is great interest in maximising
stability to prevent this complication. Head size has been recognised
to have a strong influence on the risk of dislocation post-operatively.
As femoral head size increases, stability is augmented, secondary
to an increase in impingement-free range of movement. Larger head
sizes also greatly increase the ‘jump distance’ required for the
head to dislocate in an appropriately positioned cup. Level-one
studies support the use of larger diameter heads as they decrease
the risk of dislocation following primary and revision THR. Highly cross-linked
polyethylene has allowed us to increase femoral head size, without
a marked increase in wear. However, the thin polyethylene liners
necessary to accommodate larger heads may increase the risk of liner
fracture and larger heads have also been implicated in causing soft-tissue
impingement resulting in groin pain. Larger diameter heads also
impart larger forces on the femoral trunnion, which may contribute
to corrosion, metal release, and adverse local tissue reactions.
Alternative large bearings including large ceramic heads and dual
mobility bearings may mitigate some of these risks, and several
of these devices have been used with clinical success. Cite this article:
We report serum metal ion level data in patients with unilateral and bilateral hip resurfacing over a ten-year period. In these patients there is an increase in both cobalt and chromium levels above the accepted reference ranges during the first 18 months after operation. Metal ion levels remain elevated, but decline slowly for up to five years. However, the levels then appear to start rising again in some patients up to the ten-year mark. There was no significant difference in cobalt or chromium levels between men and women. These findings appear to differ from much of the current literature. The clinical significance of a raised metal ion level remains under investigation.
There is little evidence on the cost effectiveness
of different brands of hip prostheses. We compared lifetime cost effectiveness
of frequently used brands within types of prosthesis including cemented
(Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite
Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed)
and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy).
We used data from three linked English national databases to estimate
the lifetime risk of revision, quality-adjusted life years (QALYs)
and cost. For women with osteoarthritis aged 70 years, the Exeter V40 Elite
Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0
QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision
risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision
risk, 9.22 QALYs), the most commonly used brand, and assuming a
willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is
most cost effective, with an incremental net monetary benefit of £876.
Differences in cost effectiveness between the hybrid CPT Trilogy
and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc
Exceed were small, and a cautious interpretation is required, given
the limitations of the available information. However, it is unlikely that cemented brands are among the most
cost effective. Similar patterns of results were observed for men
and other ages. The gain in quality of life after total hip arthroplasty,
rather than the risk of revision, was the main driver of cost effectiveness. Cite this article:
Total knee arthroplasty (TKA) is an established
and successful procedure. However, the design of prostheses continues
to be modified in an attempt to optimise the functional outcome
of the patient. The aim of this study was to determine if patient outcome after
TKA was influenced by the design of the prosthesis used. A total of 212 patients (mean age 69; 43 to 92; 131 female (62%),
81 male (32%)) were enrolled in a single centre double-blind trial
and randomised to receive either a Kinemax (group 1) or a Triathlon
(group 2) TKA. Patients were assessed pre-operatively, at six weeks, six months,
one year and three years after surgery. The outcome assessments
used were the Oxford Knee Score; range of movement; pain numerical
rating scales; lower limb power output; timed functional assessment
battery and a satisfaction survey. Data were assessed incorporating
change over all assessment time points, using repeated measures
analysis of variance longitudinal mixed models. Implant group 2
showed a significantly greater range of movement (p = 0.009), greater
lower limb power output (p = 0.026) and reduced report of ‘worst
daily pain’ (p = 0.003) over the three years of follow-up. Differences
in Oxford Knee Score (p = 0.09), report of ‘average daily pain’
(p = 0.57) and timed functional performance tasks (p = 0.23) did
not reach statistical significance. Satisfaction with outcome was
significantly better in group 2 (p = 0.001). These results suggest that patient outcome after TKA can be influenced
by the prosthesis used. Cite this article:
We systematically reviewed the peer-reviewed literature to relate the survival of hybrid metal-on-metal hip resurfacing arthroplasty devices to a National Institute of Clinical Excellence (NICE) benchmark for choosing a primary total hip replacement, which is a survival rate of 90% at a follow-up of ten years. A total of 29 articles (10 621 resurfaced hips) met the inclusion criteria. The mean follow-up ranged from 0.6 to 10.5 years and the survival of the implant ranged from 84% to 100%. Of the 10 621 hips, 370 were revised (3.5%), with aseptic loosening as the most frequent mode of failure. None of the hip resurfacing arthroplasty implants used to date met the full ten-year NICE benchmark of survival. A total of 13 studies showed satisfactory survival compared with the three-year NICE benchmark.
Degenerative problems of the hip in patients
with childhood and adult onset neuromuscular disorders can be challenging
to treat. Many orthopaedic surgeons are reluctant to recommend total
hip replacement (THR) for patients with underlying neuromuscular
disorders due to the perceived increased risks of dislocation, implant loosening,
and lack of information about the functional outcomes and potential
benefits of these procedures in these patients. Modular femoral
components and alternative bearings which facilitate the use of
large femoral heads, constrained acetabular components and perhaps
more importantly, a better understanding about the complications
and outcomes of THR in the patient with neuromuscular disorders,
make this option viable. This paper will review the current literature
and our experience with THR in the more frequently encountered neuromuscular
disorders. Cite this article: