Aseptic loosening is seen in a significant proportion of cemented total hip replacements (THR). In a small subgroup of patients who suffer early loosening polyethylene debris is unlikely to be responsible. We recently reported one case of allergic contact dermatitis to N,N-dimethylparatoluidine (DMT), an accelerator used in bone cement. We have therefore investigated this using skin-patch tests to a variety of substances including metals, polyethylene and the separated individual components of Simplex cement. We studied 70 patients, 15 with aseptic loosening less than two years after THR, 25 with satisfactory long-term cemented fixation, five with infected loosening of cemented THRs and 25 awaiting hip arthroplasty. We found seven positive reactions to DMT, all of them in patients with the rapid onset of aseptic loosening. Allergy to DMT is recognised in the dental profession in respect of the ‘denture sore mouth’ syndrome, and could also be an occupational hazard since some industrial glues contain DMT. Our results suggest the need for awareness of possible previous dental or occupational exposure to the constituents of bone cement. We recommend the use of skin-patch testing in high-risk cases.
We describe a modified lateral approach to the hip which exploits the functional continuity of gluteus medius and vastus lateralis and their dense crescentic attachment to the greater trochanter. The gluteus medius is not incised or split, but is detached and mobilised with gluteus minimus as one unit. This facilitates reattachment of the glutei and helps to preserve abductor function.
The Kinematic Stabilizer is a posterior-cruciate-substituting design of total knee replacement. We have reviewed 109 primary total knee replacements in 95 patients at a mean follow-up time of 12.7 years (10 to 14). We used survival analysis with failure defined as revision of the implant. This gave a cumulative survival rate of 95% (95% CI± 5%) at ten years and 87% (± 10%) at 13 years. These results from an independent centre confirm the value of an established design of cemented total knee replacement and question the wisdom of the introduction of modifications and new designs without properly controlled trials.
We report a prospective, stratified study of 60 PCA-cups and 60 RM-polyethylene cups which have been followed for a median time of 90 months, with annual radiography. The radiological migration of cups was measured by the computer-assisted EBRA method. A number of threshold migration rates from 1 mm in the first year to 1 mm in five years have been assessed and related to clinically determined revision rates. A total of 28 cups showed a total migration of 1 mm or more within the first two years; 13 of these cups have required revision and been exchanged. The survival curves of cups which had previously shown early migration were considerably different from those without early migration. For cups with a migration of less than 1 mm within the first two years the mean survival at 96 months was 0.96 ± 0.02; for migrating cups, it was 0.63 ± 0.11 (log-rank test, p = 0.0001; chi-square value = 39.4). Early migration is a good predictor for late loosening of hip sockets.
Any operation induces an elevation in the level of serum C-reactive protein (CRP). After hip and knee arthroplasty the maximal values are seen on the second and third postoperative days, after which the CRP decreases rapidly. There is no difference between patients with cemented or uncemented prostheses. Major postoperative complications may cause a further increase in CRP levels at one and two weeks.
Postoperative deep-vein thrombosis (DVT) is believed to be rare in Asians. We studied 88 consecutive patients in Malaysia who had operations for fracture of the proximal femur or for total hip or knee replacement. No patient had prophylaxis against DVT; bilateral ascending venography was performed between six and ten days after operation. A total of 55 patients (62.5%) showed venographic evidence of DVT. The prevalence was greatest after total knee replacement (76.5%), less after total hip replacement (64.3%) and smallest in the fracture group (50%). One patient developed symptomatic pulmonary embolism. In contrast to other reports from Asia, we found an incidence of postoperative DVT which is similar to that reported in Western populations. This suggests that the present practice of withholding routine prophylaxis against thromboembolism in Asian patients undergoing high-risk orthopaedic procedures should be reconsidered.
We investigated the effect of a fibrinolytic inhibitor, tranexamic acid, on blood loss and blood transfusion in knee arthroplasty by a randomised, double-blind study of 86 patients. A dose of 10 mg/kg body-weight of either tranexamic acid or placebo was given intravenously shortly before the release of the tourniquet, and repeated three hours later. The mean total blood loss was 730 ± 280 ml in the tranexamic acid group as against 1410 ± 480 ml in the placebo group (p <
0.001). Both the number of patients receiving blood transfusion and the number of blood units transfused were reduced to one-third in the treated group, and mean postoperative Hb concentrations were significantly higher after prophylaxis. The number of thromboembolic complications was the same in both groups. Tranexamic acid should be given prophylactically in order to be effective.
Total hip and total knee arthroplasty both provide a considerable improvement in quality of life, but there is no evidence to suggest that one is more successful than the other. We studied 72 patients in a prospective trial before and after total hip or total knee replacement. We recorded scores for disability and distress derived from the Harris hip score and the British Orthopaedic Association knee assessment score, and used them to generate quality of life (QoL) scores using the Rosser Index Matrix immediately before and at one year after surgery. The patients awaiting knee replacement had significantly lower QoL scores than those awaiting hip replacement (p = 0.011). The QoL scores at one year were high and almost identical for both groups (p = 0.46). Further analysis showed that gender and weight were not significant predictors of improvement of QoL scores, but age (p = 0.03) and whether the hip or knee was replaced (p = 0.006) were significant factors.
A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis.
Despite the use of ultraclean air, there are still cases of infection in total joint arthroplasty. One possible route by which bacteria may enter the wound is indirectly by contamination of instruments during skin preparation and draping. We found that bacterial air counts were 4.4 times higher during preparation and draping for hip or knee arthroplasty using an unscrubbed, ungowned leg holder than during the operation itself. With the leg holder scrubbed and gowned during preparation and draping, the air counts were reduced but were still 2.4 fold greater than intraoperatively. On some occasions, the air counts during preparation and draping exceeded the standards for ultraclean air irrespective of the attire of the leg holder. We recommend that the leg is held by a scrubbed and gowned member of the team. More importantly, we consider that instrument packs should be opened only after skin preparation and draping have been completed.
In a prospective trial we studied 176 consecutive patients having a primary total hip arthroplasty to compare the quality of life before and after operation in non-obese and obese groups. We used a modified Harris hip score and the Rosser Index Matrix to generate these scores, and found that the median preoperative quality of life score for both groups was similar. During review, both groups showed a corresponding marked improvement in quality of life scores at one and three years. There was no statistically significant difference in the improvement in scores between the non-obese and obese groups. It appears that relative body-weight on its own does not influence the benefit derived from primary total hip arthroplasty.
We report a prospective study of gait and tibial component migration in 45 patients with osteoarthritis treated by total knee arthroplasty (TKA). Migration was measured over two years using roentgen stereophotogrammetry. We used the previously established threshold of 200 μm migration in the second postoperative year to distinguish two groups: a risk group of 15 patients and a stable group of 28 patients. We performed gait analysis before operation and at six months and at two years after TKA. On all three occasions we found significant differences between the two groups in the mean sagittal plane moments of the knee joint. The risk group walked with higher peak flexion moments than the stable group. The two groups were not discriminated by any clinical or radiological criteria or other gait characteristics. The relationship which we have found between gait with increased flexion moments and risk of tibial component loosening warrants further study as regards the aetiology of prosthetic loosening and possible methods of influencing its incidence.
We studied wear in the ultra-high-molecular-weight polyethylene offset bore socket in 54 hips which had had Charnley low-friction arthroplasty. At an average follow-up of 8.1 years, the mean penetration rate was 0.04 mm per year. Correlation between the depth of socket penetration and the incidence of socket migration was confirmed, but socket migration occurred with lower penetration than had been previously reported.