Bioengineering reasons for increased wear and failure of metal-on-metal (MoM) bearings in hip prostheses have been described. Low wear occurs in MoM hips when the centre of the femoral head is concentric with the centre of the acetabular component and the implants are correctly positioned. Translational or rotational malpositioning of the components can lead to the contact-patch of the femoral component being displaced to the rim of the acetabular component, resulting in a ten- to 100-fold increase in wear and metal ion levels. This may cause adverse tissue reactions, loosening of components and failure of the prosthesis.
In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
Metal-on-metal resurfacing of the hip (MoMHR)
has enjoyed a resurgence in the last decade, but is now again in question
as a routine option for osteoarthritis of the hip. Proponents of
hip resurfacing suggest that its survival is superior to that of
conventional hip replacement (THR), and that hip resurfacing is
less invasive, is easier to revise than THR, and provides superior
functional outcomes. Our argument serves to illustrate that none
of these proposed advantages have been realised and new and unanticipated
serious complications, such as pseudotumors, have been associated
with the procedure. As such, we feel that the routine use of MoMHR
is not justified. Cite this article:
Ciprofloxacin hydrochloride-loaded microspheres were prepared by a spray-drying method using pectin and chitosan. The effects of different polymers and drug ratios were investigated. The most appropriate carriers were selected by The drug was released rapidly from the pectin carrier but this was more sustained in the chitosan formulation. Chitosan microspheres loaded with ciprofloxacin hydrochloride were more effective for the treatment of osteomyelitis than equivalent intramuscular antibiotics.
Trabecular metal (TM) augments are a relatively
new option for reconstructing segmental bone loss during acetabular
revision. We studied 34 failed hip replacements in 34 patients that
were revised between October 2003 and March 2010 using a TM acetabular
shell and one or two augments. The mean age of the patients at the
time of surgery was 69.3 years (46 to 86) and the mean follow-up
was 64.5 months (27 to 107). In all, 18 patients had a minor column
defect, 14 had a major column defect, and two were associated with
pelvic discontinuity. The hip centre of rotation was restored in
27 patients (79.4%). The Oxford hip score increased from a mean
of 15.4 points (6 to 25) before revision to a mean of 37.7 (29 to
47) at the final follow-up. There were three aseptic loosenings
of the construct, two of them in the patients with pelvic discontinuity.
One septic loosening also occurred in a patient who had previously
had an infected hip replacement. The augments remained stable in
two of the failed hips. Whenever there was a loose acetabular component
in contact with a stable augment, progressive metal debris shedding
was evident on the serial radiographs. Complications included another
deep infection treated without revision surgery. Good clinical and
radiological results can be expected for bone-deficient acetabula
treated by a TM cup and augment, but for pelvic discontinuities
this might not be a reliable option. Cite this article:
We report the clinical and radiographic outcomes
of 208 consecutive femoral revision arthroplasties performed in 202
patients (119 women, 83 men) between March 1991 and December 2007
using the X-change Femoral Revision System, fresh-frozen morcellised
allograft and a cemented polished Exeter stem. All patients were
followed prospectively. The mean age of the patients at revision
was 65 years (30 to 86). At final review in December 2013 a total
of 130 patients with 135 reconstructions (64.9%) were alive and
had a non re-revised femoral component after a mean follow-up of
10.6 years (4.7 to 20.9). One patient was lost to follow-up at six
years, and their data were included up to this point.
Re-operation for any reason was performed in 33 hips (15.9%), in
13 of which the femoral component was re-revised (6.3%). The mean
pre-operative Harris hip score was 52 (19 to 95) (n = 73) and improved
to 80 (22 to 100) (n = 161) by the last follow-up. Kaplan–Meier
survival with femoral re-revision for any reason as the endpoint
was 94.9% (95% confidence intervals (CI) 90.2 to 97.4) at ten years;
with femoral re-revision for aseptic loosening as the endpoint it was
99.4% (95% CI 95.7 to 99.9); with femoral re-operation for any reason
as the endpoint it was 84.5% (95% CI 78.3 to 89.1); and with subsidence ≥ 5
mm it was 87.3% (95% CI 80.5 to 91.8). Femoral revision with the
use of impaction allograft bone grafting and a cemented polished
stem results in a satisfying survival rate at a mean of ten years’ follow-up. Cite this article:
To review the current best surgical practice and detail a multi-disciplinary
approach that could further reduce joint replacement infection. Review of relevant literature indexed in PubMed.Objectives
Methods
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
It has been suggested that the wear of ultra-high molecular weight polyethylene (UHMWPE) in total hip replacement is substantially reduced when the femoral head is ceramic rather than metal. However, studies of alumina and zirconia ceramic femoral heads on the penetration of an UHMWPE liner The purpose of this study was to examine the surface characteristics of 30 alumina and 24 zirconia ceramic femoral heads and to identify any phase transformation in the zirconia heads. We also studied the penetration rate of alumina and zirconia heads into contemporary UHMWPE liners. The alumina heads had been implanted for a mean of 11.3 years (8.1 to 16.2) and zirconia heads for a mean of 9.8 years (7.5 to 15). The mean surface roughness values of the explanted alumina heads (Ra 40.12 nm and Rpm 578.34 nm) were similar to those for the explanted zirconia heads (Ra 36.21 nm and Rpm 607.34 nm). The mean value of the monoclinic phase of two control zirconia heads was 1% (0.8% to 1.5%) and 1.2% (0.9% to 1.3%), respectively. The mean value of the monoclinic phase of 24 explanted zirconia heads was 7.3% (1% to 26%). In the alumina group, the mean linear penetration rate of the UMWPE liner was 0.10 mm/yr (0.09 to 0.12) in hips with low Ra and Rpm values (13.22 nm and 85.91 nm, respectively). The mean linear penetration rate of the UHMWPE liner was 0.13 mm/yr (0.07 to 0.23) in hips with high Ra and Rpm values (198.72 nm and 1329 nm, respectively). This difference was significant (p = 0.041). In the zirconia head group, the mean linear penetration rate of the UHMWPE liner was 0.09 mm/yr (0.07 to 0.14) in hips with low Ra and Rpm values (12.78 nm and 92.99 nm, respectively). The mean linear penetration rate of the UHMWPE liner was 0.12 mm/yr (0.08 to 0.22) in hips with high Ra and Rpm values (199.21 nm and 1381 nm, respectively). This difference was significant (p = 0.039). The explanted zirconia heads which had a minimal phase transformation had similar surface roughness and a similar penetration rate of UHMWPE liner as the explanted alumina head.
We investigated the incidence of soft-tissue
lesions after small head metal-on-metal total hip replacement (MoM THR).
Between December 1993 and May 1999, 149 patients (195 hips) underwent
primary cementless MoM THR. During the follow-up period, three patients (five THRs) died
and eight patients (14 THRs) were lost to follow-up. We requested
that all patients undergo CT evaluation. After exclusion of five
patients (six THRs) who had undergone a revision procedure, and
22 (28 THRs) who were unwilling to take part in this study, 111
patients (142 THRs) were evaluated. There were 63 men (88 THRs)
and 48 women (54 THRs) with a mean age of 45.7 years (37 to 56)
at the time of surgery. The mean follow-up was 15.4 years (13 to
19). A soft-tissue lesion was defined as an abnormal peri-prosthetic
collection of fluid, solid lesion or asymmetrical soft-tissue mass. At final follow-up, soft-tissue lesions were found in relation
to 28 THRs (19.7%), including 25 solid and three cystic lesions.
They were found in 20 men and eight women; 26 lesions were asymptomatic
and two were symptomatic. The mean maximal diameter of the soft-tissue
lesion was 42.3 mm (17 to 135). The relatively high rate of soft-tissue lesions
observed with small head MoM THR remains a concern. Cite this article:
Impacted morsellised allografts have been used successfully to address the problem of poor bone stock in revision surgery. However, there are concerns about the transmission of pathogens, the high cost and the shortage of supply of donor bone. Bone-graft extenders, such as tricalcium phosphate (TCP) and hydroxyapatite (HA), have been developed to minimise the use of donor bone. In a human cadaver model we have evaluated the surgical and mechanical feasibility of a TCP/HA bone-graft extender during impaction grafting revision surgery. A TCP/HA allograft mix increased the risk of producing a fissure in the femur during the impaction procedure, but provided a higher initial mechanical stability when compared with bone graft alone. The implications of the use of this type of graft extender in impaction grafting revision surgery are discussed.
This review examines the future of total hip arthroplasty, aiming to avoid past mistakes
Cancellous allograft bone chips are commonly
used in the reconstruction of defects in bone after removal of benign tumours.
We investigated the MRI features of grafted bone chips and their
change over time, and compared them with those with recurrent tumour.
We retrospectively reviewed 66 post-operative MRIs from 34 patients
who had undergone curettage and grafting with cancellous bone chips
to fill the defect after excision of a tumour. All grafts showed
consistent features at least six months after grafting: homogeneous
intermediate or low signal intensities with or without scattered
hyperintense foci (speckled hyperintensities) on T1 images; high
signal intensities with scattered hypointense foci (speckled hypointensities)
on T2 images, and peripheral rim enhancement with or without central
heterogeneous enhancements on enhanced images. Incorporation of
the graft occurred from the periphery to the centre, and was completed
within three years. Recurrent lesions consistently showed the same signal
intensities as those of pre-operative MRIs of the primary lesions.
There were four misdiagnoses, three of which were chondroid tumours. We identified typical MRI features and clarified the incorporation
process of grafted cancellous allograft bone chips. The most important
characteristics of recurrent tumours were that they showed the same
signal intensities as the primary tumours. It might sometimes be
difficult to differentiate grafted cancellous allograft bone chips
from a recurrent chondroid tumour. Cite this article:
We studied the effects of hyperbaric oxygen (HBO) and zoledronic acid (ZA) on posterior lumbar fusion using a validated animal model. A total of 40 New Zealand white rabbits underwent posterior lumbar fusion at L5–6 with autogenous iliac bone grafting. They were divided randomly into four groups as follows: group 1, control; group 2, HBO (2.4 atm for two hours daily); group 3, local ZA (20 μg of ZA mixed with bone graft); and group 4, combined HBO and local ZA. All the animals were killed six weeks after surgery and the fusion segments were subjected to radiological analysis, manual palpation, biomechanical testing and histological examination. Five rabbits died within two weeks of operation. Thus, 35 rabbits (eight in group 1 and nine in groups 2, 3 and 4) completed the study. The rates of fusion in groups 3 and 4 (p = 0.015) were higher than in group 1 (p <
0.001) in terms of radiological analysis and in group 4 was higher than in group 1 with regard to manual palpation (p = 0.015). We found a statistically significant difference in the biomechanical analysis between groups 1 and 4 (p = 0.024). Histological examination also showed a statistically significant difference between groups 1 and 4 (p = 0.036). Our results suggest that local ZA combined with HBO may improve the success rate in posterior lumbar spinal fusion.
We review the history and literature of hip resurfacing arthroplasty. Resurfacing and the science behind it continues to evolve. Recent results, particularly from the national arthroplasty registers, have spread disquiet among both surgeons and patients. A hip resurfacing arthroplasty is not a total hip replacement, but should perhaps be seen as a means of delaying it. The time when hip resurfacing is offered to a patient may be different from that for a total hip replacement. The same logic can apply to the timing of revision surgery. Consequently, the comparison of resurfacing with total hip replacement may be a false one. Nevertheless, the need for innovative solutions for young arthroplasty patients is clear. Total hip replacement can be usefully delayed in many of these patients by the use of hip resurfacing arthroplasty.
Components from 73 failed knee replacements (TKRs) consisting of rotating-platform, mobile-bearing and fixed-bearing implants were examined to assess the patterns of wear. The patterns were divided into low-grade (burnishing, abrasion and cold flow) and high-grade (scratching, pitting/metal embedding and delamination) to assess the severity of the wear of polyethylene. The rotating-platform group had a higher incidence of low-grade wear on the upper surface compared with the fixed-bearing group. By contrast, high-grade wear comprising scratching, pitting and third-body embedding was seen on the lower surface. Linear regression analysis showed a significant correlation of the wear scores between the upper and lower surfaces of the tibial insert (R2 = 0.29, p = 0.04) for the rotating-platform group, but no significant correlation was found for the fixed-bearing counterpart. This suggests that high-grade wear patterns on the upper surface are reduced with the rotating-platform design. However, the incidence of burnishing, pitting/third-body embedding and scratching wear patterns on the lower surface was higher compared with that in the fixed-bearing knee.
Our study aimed to examine not only the incidence but also the
impact of noise from two types of total hip replacement articulations:
ceramic-on-ceramic and ceramic-on-polyethylene. We performed a case-controlled study comparing subjective and
objective questionnaire scores of patients receiving a ceramic-on-ceramic
or a ceramic-on-polyethylene total hip replacement by a single surgeon.Objectives
Methods
The objective of this study was to compare the elution characteristics,
antimicrobial activity and mechanical properties of antibiotic-loaded
bone cement (ALBC) loaded with powdered antibiotic, powdered antibiotic
with inert filler (xylitol), or liquid antibiotic, particularly focusing
on vancomycin and amphotericin B. Cement specimens loaded with 2 g of vancomycin or amphotericin
B powder (powder group), 2 g of antibiotic powder and 2 g of xylitol
(xylitol group) or 12 ml of antibiotic solution containing 2 g of
antibiotic (liquid group) were tested.Objectives
Methods
We report 17 patients (20 hips) in whom metal-on-metal resurfacing had been performed and who presented with various symptoms and a soft-tissue mass which we termed a pseudotumour. Each patient underwent plain radiography and in some, CT, MRI and ultrasonography were also performed. In addition, histological examination of available samples was undertaken. All the patients were women and their presentation was variable. The most common symptom was discomfort in the region of the hip. Other symptoms included spontaneous dislocation, nerve palsy, a noticeable mass or a rash. The common histological features were extensive necrosis and lymphocytic infiltration. To date, 13 of the 20 hips have required revision to a conventional hip replacement. Two are awaiting revision. We estimate that approximately 1% of patients who have a metal-on-metal resurfacing develop a pseudotumour within five years. The cause is unknown and is probably multifactorial. There may be a toxic reaction to an excess of particulate metal wear debris or a hypersensitivity reaction to a normal amount of metal debris. We are concerned that with time the incidence of these pseudotumours may increase. Further investigation is required to define their cause.
Wear of metal-on-metal bearings causes elevated levels of cobalt and chromium in blood and body fluids. Metal-on-metal bearings have two distinct wear phases. In the early phase, the wear rate is high. Later, it decreases and the bearing enters a steady-state phase. It is expected that as the wear rates decline, the level of cobalt detected in plasma will also decrease. We studied the baseline and exercise-related cobalt rise in 21 patients (13 men and eight women) with a mean age of 54 years (38 to 80) who had undergone successful hip resurfacing at a mean of 44 months (10 to 96) earlier. Our results showed that circulating baseline cobalt levels were not significantly correlated with the time since implantation (r = 0.08, p = 0.650). By contrast, the exercise-related cobalt rise was directly correlated with the inclination angle of the acetabular component (r = 0.47, p = 0.032) and inversely correlated with the time since implantation (r = −0.5, p = 0.020). Inclination of the acetabular component should be kept less than 40° to decrease the production of wear debris.