Mechanical wear and corrosion at the head-stem junction of total hip arthroplasties (THAs) (trunnionosis) have been implicated in their early revision, most commonly in metal-on-metal (MOM) hips. We can isolate the role of the head-stem junction as the predominant source of metal release by investigating non-MOM hips; this can help to identify clinically significant volumes of material loss and corrosion from these surfaces. In this study we examined a series of 94 retrieved metal-on-polyethylene (MOP) hips for evidence of corrosion and material loss at the taper junction using a well published visual grading method and an established roundness-measuring machine protocol. Hips were retrieved from 74 male and 20 female patients with a median age of 57 years (30 to 76) and a median time to revision of 215 months (2 to 324). The reasons for revision were loosening of both the acetabular component and the stem (n = 29), loosening of the acetabular component (n = 58) and infection (n = 7). No adverse tissue reactions were reported by the revision surgeons.Objectives
Methods
The purpose of this retrospective study was to evaluate the minimum
five-year outcome of revision total hip arthroplasty (THA) using
the Kerboull acetabular reinforcement device (KARD) in patients
with Paprosky type III acetabular defects and destruction of the
inferior margin of the acetabulum. We identified 36 patients (37 hips) who underwent revision THA
under these circumstances using the KARD, fresh frozen allograft
femoral heads, and reconstruction of the inferior margin of the
acetabulum. The Merle d’Aubigné system was used for clinical assessment.
Serial anteroposterior pelvic radiographs were used to assess migration
of the acetabular component.Aims
Patients and Methods
We aimed to retrospectively assess the accuracy and safety of
CT navigated pedicle screws and to compare accuracy in the cervical
and thoracic spine (C2-T8) with (COMB) and without (POST) prior
anterior surgery (anterior cervical discectomy or corpectomy and
fusion with ventral plating: ACDF/ACCF). A total of 592 pedicle screws, which were used in 107 consecutively
operated patients (210 COMB, 382 POST), were analysed. The accuracy
of positioning was determined according to the classification of
Gertzbein and Robbins on post-operative CT scans.Aims
Patients and Methods
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
Unicompartmental knee arthroplasty (UKA) is a demanding procedure, with tibial component subsidence or pain from high tibial strain being potential causes of revision. The optimal position in terms of load transfer has not been documented for lateral UKA. Our aim was to determine the effect of tibial component position on proximal tibial strain. A total of 16 composite tibias were implanted with an Oxford Domed Lateral Partial Knee implant using cutting guides to define tibial slope and resection depth. Four implant positions were assessed: standard (5° posterior slope); 10° posterior slope; 5° reverse tibial slope; and 4 mm increased tibial resection. Using an electrodynamic axial-torsional materials testing machine (Instron 5565), a compressive load of 1.5 kN was applied at 60 N/s on a meniscal bearing via a matching femoral component. Tibial strain beneath the implant was measured using a calibrated Digital Image Correlation system.Objectives
Methods
The aim of this study was to evaluate the clinical and radiological
outcome of using an anatomical short-stem shoulder prosthesis to
treat primary osteoarthritis of the glenohumeral joint. A total of 66 patients (67 shoulders) with a mean age of 76 years
(63 to 92) were available for clinical and radiological follow-up
at two different timepoints (T1, mean 2.6 years, Aims
Patients and Methods
Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
Controversy about the use of an anatomical total shoulder arthroplasty
(aTSA) in young arthritic patients relates to which is the ideal
form of fixation for the glenoid component: cemented or cementless.
This study aimed to evaluate implant survival of aTSA when used
in patients aged < 60 years with primary glenohumeral osteoarthritis (OA),
and to compare the survival of cemented all-polyethylene and cementless
metal-backed glenoid components. A total of 69 consecutive aTSAs were performed in 67 patients
aged < 60 years with primary glenohumeral OA. Their mean age
at the time of surgery was 54 years (35 to 60). Of these aTSAs,
46 were undertaken using a cemented polyethylene component and 23
were undertaken using a cementless metal-backed component. The age, gender,
preoperative function, mobility, premorbid glenoid erosion, and
length of follow-up were comparable in the two groups. The patients
were reviewed clinically and radiographically at a mean of 10.3
years (5 to 12, Aims
Materials and Methods
We carried out a further study of the long-term results of the
cemented Exeter femoral component in patients under the age of 40
with a mean follow-up of 13.6 years (10 to 20). We reviewed our original cohort of 104 cemented Exeter stems
in 78 consecutive patients with a mean age of 31 years (16 to 39).
Only one patient was lost to radiological follow-up.Aims
Patients and Methods
Adolescent idiopathic scoliosis (AIS) is a complex
3D deformity of the spine. Its prevalence is between 2% and 3% in the
general population, with almost 10% of patients requiring some form
of treatment and up to 0.1% undergoing surgery. The cosmetic aspect
of the deformity is the biggest concern to the patient and is often
accompanied by psychosocial distress. In addition, severe curves
can cause cardiopulmonary distress. With proven benefits from surgery,
the aims of treatment are to improve the cosmetic and functional
outcomes. Obtaining correction in the coronal plane is not the only
important endpoint anymore. With better understanding of spinal
biomechanics and the long-term effects of multiplanar imbalance,
we now know that sagittal balance is equally, if not more, important.
Better correction of deformities has also been facilitated by an
improvement in the design of implants and a better understanding
of metallurgy. Understanding the unique character of each deformity
is important. In addition, using the most appropriate implant and
applying all the principles of correction in a bespoke manner is important
to achieve optimum correction. In this article, we review the current concepts in AIS surgery. Cite this article:
The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces. In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.Objectives
Methods
We compared the incidence of pseudotumours after
large head metal-on-metal (MoM) total hip arthroplasty (THA) with
that after conventional metal-on-polyethylene (MoP) THA and assessed
the predisposing factors to pseudotumour formation. From a previous randomised controlled trial which compared large
head (38 mm to 60 mm) cementless MoM THA with conventional head
(28 mm) cementless MoP THA, 93 patients (96 THAs: 41 MoM (21 males,
20 females, mean age of 64 years, standard deviation ( The incidence of pseudotumours, measured using a standardised
CT protocol was 22 (53.7%) after MoM THA and 12 (21.8%) after MoP
THA. Women with a MoM THA were more likely to develop a pseudotumour
than those with a MoP THA (15 Contrary to popular belief, pseudotumours occur frequently around
MoP THAs. Women with a MoM THA and an elevated cobalt level are
at greatest risk. In this study, pseudotumours had no effect on
the functional outcome after either large head MoM or conventional
MoP THA. Cite this article:
The increased The proximal penetration of the femoral head five years post-operatively
was measured for 15 patients using radiostereometric analysis (RSA).Aims
Patients and Methods
We present the ten-year data of a cohort of patients, aged between
18 and 65 years (mean age 52.7 years; 19 to 64), who underwent total
hip arthroplasty. Patients were randomised to be treated with a
cobalt-chrome (CoCr) femoral head with an ultra-high molecular weight
polyethylene (UHMWPE), highly cross-linked polyethylene (XLPE) or
ceramic-on-ceramic (CoC) bearing surface. A total of 102 hips (91 patients) were randomised into the three
groups. At ten years, 97 hips were available for radiological and
functional follow-up. Two hips (two patients) had been revised (one
with deep infection and one for periprosthetic fracture) and three
were lost to follow-up. Radiological analysis was performed using
a validated digital assessment programme to give linear, directional
and volumetric wear of the two polyethylene groups.Aims
Patients and Methods
Patient-specific (PS) implantation surgical technology has been introduced in recent years and a gradual increase in the associated number of surgical cases has been observed. PS technology uses a patient’s own geometry in designing a medical device to provide minimal bone resection with improvement in the prosthetic bone coverage. However, whether PS unicompartmental knee arthroplasty (UKA) provides a better biomechanical effect than standard off-the-shelf prostheses for UKA has not yet been determined, and still remains controversial in both biomechanical and clinical fields. Therefore, the aim of this study was to compare the biomechanical effect between PS and standard off-the-shelf prostheses for UKA. The contact stresses on the polyethylene (PE) insert, articular cartilage and lateral meniscus were evaluated in PS and standard off-the-shelf prostheses for UKA using a validated finite element model. Gait cycle loading was applied to evaluate the biomechanical effect in the PS and standard UKAs.Objectives
Methods