This study tests the biomechanical properties of adjacent locked
plate constructs in a femur model using Sawbones. Previous studies
have described biomechanical behaviour related to inter-device distances.
We hypothesise that a smaller lateral inter-plate distance will
result in a biomechanically stronger construct, and that addition
of an anterior plate will increase the overall strength of the construct. Sawbones were plated laterally with two large-fragment locking
compression plates with inter-plate distances of 10 mm or 1 mm.
Small-fragment locking compression plates of 7-hole, 9-hole, and
11-hole sizes were placed anteriorly to span the inter-plate distance.
Four-point bend loading was applied, and the moment required to
displace the constructs by 10 mm was recorded.Objectives
Methods
Stems may improve fixation and stability of components
during revision total knee replacement. However, the choice between
cemented and cementless stems is not a clear one. Cemented stems
offer several advantages in terms of versatility, mechanical stability, surgical
technique and clinical outcome over their cementless counterpart. Cite this article:
The December 2015 Children’s orthopaedics Roundup360 looks at: Paediatric femoral fractures: a single incision nailing?; Lateral condylar fractures: open or percutaneous?; Forearm refracture: the risks; Tibial spine fractures; The child’s knee in MRI; The mechanics of SUFE; Idiopathic clubfoot
Sagittal alignment of the lumbosacral spine, and specifically pelvic incidence (PI), has been implicated in the development of spine pathology, but generally ignored with regards to diseases of the hip. We aimed to determine if increased PI is correlated with higher rates of hip osteoarthritis (HOA). The effect of PI on the development of knee osteoarthritis (KOA) was used as a negative control. We studied 400 well-preserved cadaveric skeletons ranging from 50 to 79 years of age at death. Each specimen’s OA of the hip and knee were graded using a previously described method. PI was measured from standardised lateral photographs of reconstructed pelvises. Multiple regression analysis was performed to determine the relationship between age and PI with HOA and KOA.Objectives
Methods
Instability after total knee replacement (TKR)
accounts for 10% to 22% of revision procedures. All patients who
present for evaluation of instability require a thorough history to
be taken and physical examination, as well as appropriate imaging.
Deep periprosthetic infection must be ruled out by laboratory testing
and an aspiration of the knee must be carried out. The three main
categories of instability include flexion instability, extension instability
(symmetric and asymmetric), and genu recurvatum. Most recently,
the aetiologies contributing to, and surgical manoeuvres required
to correct, flexion instability have been elucidated. While implant
design and patient-related factors may certainly contribute to the
aetiology, surgical technique is also a significant factor in all
forms of post-operative instability. Cite this article:
One method of femoral head preservation following avascular necrosis
(AVN) is core decompression and insertion of a tantalum rod. However,
there may be a high failure rate associated with this procedure.
The purpose of this study was to document the clinical and radiological
outcomes following total hip arthroplasty (THA) subsequent to failed
tantalum rod insertion. A total of 37 failed tantalum rods requiring total hip arthroplasty
were identified from a prospective database. There were 21 hips
in 21 patients (12 men and nine women, mean age 37 years, 18 to
53) meeting minimum two year clinical and radiographic follow-up
whose THAs were carried out between November 2002 and April 2013
(mean time between tantalum rod implantation and conversion to a
THA was 26 months, 6 to 72). These were matched by age and gender
to individuals (12 men, nine women, mean age 40 years, 18 to 58)
receiving THA for AVN without prior tantalum rod insertion.Aims
Patients and Methods
Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
The accurate reconstruction of hip anatomy and
biomechanics is thought to be important in achieveing good clinical
outcomes following total hip arthroplasty (THA). To this end some
newer hip designs have introduced further modularity into the design
of the femoral component such that neckshaft angle and anteversion,
which can be adjusted intra-operatively. The clinical effect of
this increased modularity is unknown. We have investigated the changes
in these anatomical parameters following conventional THA with a
prosthesis of predetermined neck–shaft angle and assessed the effect
of changes in the hip anatomy on clinical outcomes. In total, 44 patients (mean age 65.3 years (standard deviation
( The mean pre-operative neck–shaft angle was significantly increased
by 2.8° from 128° ( Cite this article:
The lateral compartment is predominantly affected
in approximately 10% of patients with osteoarthritis of the knee. The
anatomy, kinematics and loading during movement differ considerably
between medial and lateral compartments of the knee. This in the
main explains the relative protection of the lateral compartment
compared with the medial compartment in the development of osteoarthritis.
The aetiology of lateral compartment osteoarthritis can be idiopathic,
usually affecting the femur, or secondary to trauma commonly affecting
the tibia. Surgical management of lateral compartment osteoarthritis
can include osteotomy, unicompartmental knee replacement and total
knee replacement. This review discusses the biomechanics, pathogenesis
and development of lateral compartment osteoarthritis and its management. Cite this article:
We investigated a new intramedullary locking
nail that allows the distal interlocking screws to be locked to
the nail. We compared fixation using this new implant with fixation
using either a conventional nail or a locking plate in a laboratory
simulation of an osteoporotic fracture of the distal femur. A total
of 15 human cadaver femora were used to simulate an AO 33-A3 fracture
pattern. Paired specimens compared fixation using either a locking
or non-locking retrograde nail, and using either a locking retrograde
nail or a locking plate. The constructs underwent cyclical loading
to simulate single-leg stance up to 125 000 cycles. Axial and torsional
stiffness and displacement, cycles to failure and modes of failure
were recorded for each specimen. When compared with locking plate
constructs, locking nail constructs had significantly longer mean
fatigue life (75 800 cycles ( The new locking retrograde femoral nail showed better stiffness
and fatigue life than locking plates, and superior fatigue life
to non-locking nails, which may be advantageous in elderly patients. Cite this article:
The June 2014 Oncology Roundup360 looks at: Infection still a problem in endoprosthetic reconstruction; massive allografts not as successful as we perhaps think; curopsy for aneurysmal bone cysts?; lengthening prosthesis: days are numbered; new WHO classification in brief; proximal tumours and fluid levels: bad news; infection is predictable in orthopaedic oncology; psychosocial support key in oncological outcomes.
Mechanical alignment has been a fundamental tenet of total knee arthroplasty (TKA) since modern knee replacement surgery was developed in the 1970s. The objective of mechanical alignment was to infer the greatest biomechanical advantage to the implant to prevent early loosening and failure. Over the last 40 years a great deal of innovation in TKA technology has been focusing on how to more accurately achieve mechanical alignment. Recently the concept of mechanical alignment has been challenged, and other alignment philosophies are being explored with the intention of trying to improve patient outcomes following TKA. This article examines the evolution of the mechanical alignment concept and whether there are any viable alternatives.
In skeletally immature patients, resection of
bone tumours and reconstruction of the lower limb often results
in leg-length discrepancy. The Stanmore non-invasive extendible
endoprosthesis, which uses electromagnetic induction, allows post-operative
lengthening without anaesthesia. Between 2002 and 2009, 55 children
with a mean age of 11.4 years (5 to 16) underwent reconstruction
with this prosthesis; ten patients (18.2%) died of disseminated
disease and one child underwent amputation due to infection. We
reviewed 44 patients after a mean follow-up of 41.2 months (22 to
104). The mean Musculoskeletal Tumor Society score was 24.7 (8 to
30) and the Toronto Extremity Salvage score was 92.3% (55.2% to
99.0%). There was no local recurrence of tumour. Complications developed
in 16 patients (29.1%) and ten (18.2%) underwent revision. The mean length gained per patient was 38.6 mm (3.5 to 161.5),
requiring a mean of 11.3 extensions (1 to 40), and ten component
exchanges were performed in nine patients (16.4%) after attaining
the maximum lengthening capacity of the implant. There were 11 patients
(20%) who were skeletally mature at follow-up, ten of whom had equal
leg lengths and nine had a full range of movement of the hip and
knee. This is the largest reported series using non-invasive extendible
endoprostheses after excision of primary bone tumours in skeletally
immature patients. The technique produces a good functional outcome,
with prevention of limb-length discrepancy at skeletal maturity.
Osteochondritis Dissecans (OCD) is a condition
for which the aetiology remains unknown. It affects subchondral bone
and secondarily its overlying cartilage and is mostly found in the
knee. It can occur in adults, but is generally identified when growth
remains, when it is referred to as juvenile OCD. As the condition
progresses, the affected subchondral bone separates from adjacent
healthy bone, and can lead to demarcation and separation of its associated
articular cartilage. Any symptoms which arise relate to the stage
of the disease. Early disease without separation of the lesion results
in pain. Separation of the lesion leads to mechanical symptoms and
swelling and, in advanced cases, the formation of loose bodies. Early identification of OCD is essential as untreated OCD can
lead to the premature degeneration of the joint, whereas appropriate
treatment can halt the disease process and lead to healing. Establishing
the stability of the lesion is a key part of providing the correct
treatment. Stable lesions, particularly in juvenile patients, have
greater propensity to heal with non-surgical treatment, whereas
unstable or displaced lesions usually require surgical management. This article discusses the aetiology, clinical presentation and
prognosis of OCD in the knee. It presents an algorithm for treatment,
which aims to promote healing of native hyaline cartilage and to
ensure joint congruity. Take home message: Although there is no clear consensus as to
the best treatment of OCD, every attempt should be made to retain
the osteochondral fragment when possible as, with a careful surgical
technique, there is potential for healing even in chronic lesions Cite this article:
We retrospectively reviewed the records of 1150
computer-assisted total knee replacements and analysed the clinical
and radiological outcomes of 45 knees that had arthritis with a
pre-operative recurvatum deformity. The mean pre-operative hyperextension
deformity of 11° (6° to 15°), as measured by navigation at the start
of the operation, improved to a mean flexion deformity of 3.1° (0°
to 7°) post-operatively. A total of 41 knees (91%) were managed
using inserts ≤ 12.5 mm thick, and none had mediolateral laxity
>
2 mm from a mechanical axis of 0° at the end of the surgery. At
a mean follow-up of 26.4 months (13 to 48) there was significant
improvement in the mean Knee Society, Oxford knee and Western Ontario
and McMaster Universities Osteoarthritis Index scores compared with
the pre-operative values. The mean knee flexion improved from 105°
(80° to 125°) pre-operatively to 131° (120° to 145°), and none of
the limbs had recurrent recurvatum. These early results show that total knee replacement using computer
navigation and an algorithmic approach for arthritic knees with
a recurvatum deformity can give excellent radiological and functional
outcomes without recurrent deformity.
This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for peri-prosthetic fracture after
21 years, and one patient (one hip) was lost to follow-up. Using
aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular
revision but none needed femoral revision, demonstrating 100% survival
(95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral
component gives excellent long-term results. Cite this article:
End caps are intended to prevent nail migration
(push-out) in elastic stable intramedullary nailing. The aim of
this study was to investigate the force at failure with and without
end caps, and whether different insertion angles of nails and end caps
would alter that force at failure. Simulated oblique fractures of the diaphysis were created in
15 artificial paediatric femurs. Titanium Elastic Nails with end
caps were inserted at angles of 45°, 55° and 65° in five specimens
for each angle to create three study groups. Biomechanical testing
was performed with axial compression until failure. An identical
fracture was created in four small adult cadaveric femurs harvested
from two donors (both female, aged 81 and 85 years, height 149 cm and
156 cm, respectively). All femurs were tested without and subsequently
with end caps inserted at 45°. In the artificial femurs, maximum force was not significantly
different between the three groups (p = 0.613). Push-out force was
significantly higher in the cadaveric specimens with the use of
end caps by an up to sixfold load increase (830 N, standard deviation
(SD) 280 These results indicate that the nail and end cap insertion angle
can be varied within 20° without altering construct stability and
that the risk of elastic stable intramedullary nailing push–out
can be effectively reduced by the use of end caps. Cite this article:
The aim of this study was to compare the maximum
laxity conferred by the cruciate-retaining (CR) and posterior-stabilised
(PS) Triathlon single-radius total knee arthroplasty (TKA) for anterior
drawer, varus–valgus opening and rotation in eight cadaver knees
through a defined arc of flexion (0º to 110º). The null hypothesis
was that the limits of laxity of CR- and PS-TKAs are not significantly
different. The investigation was undertaken in eight loaded cadaver knees
undergoing subjective stress testing using a measurement rig. Firstly
the native knee was tested prior to preparation for CR-TKA and subsequently
for PS-TKA implantation. Surgical navigation was used to track maximal
displacements/rotations at 0º, 30º, 60º, 90º and 110° of flexion.
Mixed-effects modelling was used to define the behaviour of the
TKAs. The laxity measured for the CR- and PS-TKAs revealed no statistically
significant differences over the studied flexion arc for the two
versions of TKA. Compared with the native knee both TKAs exhibited
slightly increased anterior drawer and decreased varus-valgus and
internal-external roational laxities. We believe further study is required
to define the clinical states for which the additional constraint
offered by a PS-TKA implant may be beneficial. Cite this article:
Fixed flexion deformities are common in osteoarthritic
knees that are indicated for total knee arthroplasty. The lack of
full extension at the knee results in a greater force of quadriceps
contracture and energy expenditure. It also results in slower walking
velocity and abnormal gait mechanics, overloading the contralateral
limb. Residual flexion contractures after TKA have been associated
with poorer functional scores and outcomes. Although some flexion contractures may resolve with time after
surgery, a substantial percentage will become permanent. Therefore,
it is essential to correct fixed flexion deformities at the time
of TKA, and be vigilant in the post-operative course to maintain
the correction. Surgical techniques to address pre-operative flexion contractures
include: adequate bone resection, ligament releases, removal of
posterior osteophytes, and posterior capsular releases. Post-operatively,
extension can be maintained with focused physiotherapy, a specially
modified continuous passive motion machine, a contralateral heel
lift, and splinting.
The aim of this study was to report the incidence
of arthrofibrosis of the knee and identify risk factors for its development
following a fracture of the tibial plateau. We carried out a retrospective
review of 186 patients (114 male, 72 female) with a fracture of
the tibial plateau who underwent open reduction and internal fixation.
Their mean age was 46.4 years (19 to 83) and the mean follow-up
was16.0 months (6 to 80). A total of 27 patients (14.5%) developed arthrofibrosis requiring
a further intervention. Using multivariate regression analysis,
the use of a provisional external fixator (odds ratio (OR) 4.63,
95% confidence interval (CI) 1.26 to 17.7, p = 0.021) was significantly
associated with the development of arthrofibrosis. Similarly, the
use of a continuous passive movement (CPM) machine was associated
with significantly less development of arthrofibrosis (OR = 0.32,
95% CI 0.11 to 0.83, p = 0.024). The effect of time in an external
fixator was found to be significant, with each extra day of external
fixation increasing the odds of requiring manipulation under anaesthesia
(MUA) or quadricepsplasty by 10% (OR = 1.10, p = 0.030). High-energy
fracture, surgical approach, infection and use of tobacco were not
associated with the development of arthrofibrosis. Patients with
a successful MUA had significantly less time to MUA (mean 2.9 months; Based our results, CPM following operative fixation for a fracture
of the tibial plateau may reduce the risk of the development of
arthrofibrosis, particularly in patients who also undergo prolonged
provisional external fixation. Cite this article: