We evaluated the efficacy of Cite this article:
If a surgeon is faced with altered lesser trochanter
anatomy when revising the femoral component in revision total hip
replacement, a peri-prosthetic fracture, or Paprosky type IIIb or
type IV femoral bone loss, a modular tapered stem offers the advantages
of accurately controlling femoral version and length. The splines
of the taper allow rotational control, and improve the fit in femoral
canals with diaphyseal bone loss. In general, two centimetres of diaphyseal
contact is all that is needed to gain stability with modular tapered
stems. By allowing the proximal body trial to rotate on a well-fixed
distal segment during trial reduction, appropriate anteversion can
be obtained in order to improve intra-operative stability, and decrease
the dislocation risk. However, modular stems should not be used for
all femoral revisions, as implant fracture and corrosion at modular
junctions can still occur. Cite this article:
There is a North Atlantic divide, with cementless
femoral stems being used more frequently in the USA and cemented
stems being used more frequently in many countries in Europe. This
is primarily because different cemented stems have been used on
different sides of the Atlantic and the results of the cemented
stems in the US have often been poor, whereas the results of the
stems used in Europe have been good. In the National registers in
Europe, cemented stems have tended to achieve better results than
cementless. Cite this article:
The aim of this study was to determine the comorbid risk factors for failure in young patients who undergo fixation of a displaced fracture of the femoral neck. We identified from a prospective database all such patients ≤ 60 years of age treated with reduction and internal fixation. The main outcome measures were union, failure of fixation, nonunion and the development of avascular necrosis. There were 122 patients in the study. Union occurred in 83 patients (68%) at a mean follow-up of 58 months (18 to 155). Complications occurred in 39 patients (32%) at a mean of 11 months (0.5 to 39). The rate of nonunion was 7.4% (n = 9) and of avascular necrosis was 11.5% (n = 14). Failures were more common in patients over 40 years of age (p = 0.03). Univariate analysis identified that delay in time to fixation (>
24 hours), alcohol excess and pre-existing renal, liver or respiratory disease were all predictive of failure (all p <
0.05). Of these, alcohol excess, renal disease and respiratory disease were most predictive of failure on multivariate analysis. Younger patients with fractures of the femoral neck should be carefully evaluated for comorbidities that increase the risk of failure after reduction and fixation. In patients with a history of alcohol abuse, renal or respiratory disease, arthroplasty should be considered as an alternative treatment.
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods
Total hip replacement (THR) after acetabular
fracture presents unique challenges to the orthopaedic surgeon.
The majority of patients can be treated with a standard THR, resulting
in a very reasonable outcome. Technical challenges however include
infection, residual pelvic deformity, acetabular bone loss with
ununited fractures, osteonecrosis of bone fragments, retained metalwork,
heterotopic ossification, dealing with the sciatic nerve, and the
difficulties of obtaining long-term acetabular component fixation.
Indications for an acute THR include young patients with both femoral
head and acetabular involvement with severe comminution that cannot
be reconstructed, and the elderly, with severe bony comminution.
The outcomes of THR for established post-traumatic arthritis include
excellent pain relief and functional improvements. The use of modern
implants and alternative bearing surfaces should improve outcomes
further. Cite this article:
Fractures of the proximal femur are one of the
greatest challenges facing the medical community, constituting a
heavy socioeconomic burden worldwide. Controversy exists regarding
the optimal treatment for patients with unstable trochanteric proximal
femoral fractures. The recognised treatment alternatives are extramedullary
fixation usually with a sliding hip screw and intramedullary fixation
with a cephalomedullary nail. Current evidence suggests that best
results and lowest complication rates occur using a sliding hip screw.
Complications in these difficult fractures are relatively common
regardless of type of treatment. We believe that a novel device,
the X-Bolt dynamic plating system, may offer superior fixation over
a sliding hip screw with lower reoperation risk and better function.
We therefore propose to investigate the clinical effectiveness of
the X-bolt dynamic plating system compared with standard sliding
hip screw fixation within the framework of a the larger WHiTE (Warwick
Hip Trauma Evaluation) Comprehensive Cohort Study. Cite this article:
Total hip replacement for developmental hip dysplasia
is challenging. The anatomical deformities on the acetabular and
femoral sides are difficult to predict. The Crowe classification
is usually used to describe these cases – however, it is not a very
helpful tool for pre-operative planning. Small acetabular components,
acetabular augments, and modular femoral components should be available
for all cases. Regardless of the Crowe classification, the surgeon must
be prepared to perform a femoral osteotomy for shortening, or to
correct rotation, and/or angulation. Cite this article:
The August 2013 Research Roundup360 looks at: passive smoking and bone substitutes; platelet-rich plasma and osteogenesis; plantar fasciitis and platelet-rich plasma: a match made in heaven?; MRSA decolonisation decreases infection rates; gums, bisphosphonates and orthopaedics; PRAISE and partner violence; blunt impact and post-traumatic OA; and IDEAL research and implants
Hip arthrodesis remains a viable surgical technique
in well selected patients, typically the young manual labourer with
isolated unilateral hip disease. Despite this, its popularity with
patients and surgeons has decreased due to the evolution of hip
replacement, and is seldom chosen by young adult patients today.
The surgeon is more likely to encounter a patient who requests conversion
to total hip replacement (THR). The most common indications are
a painful pseudarthrosis, back pain, ipsilateral knee pain or contralateral
hip pain. Occasionally the patient will request conversion because
of difficulty with activities of daily living, body image and perceived
cosmesis. The technique of conversion and a discussion of the results
are presented. Cite this article:
We retrospectively reviewed 40 hips in 36 patients who had undergone acetabular reconstruction using a titanium Kerboull-type acetabular reinforcement device with bone allografts between May 2001 and April 2006. Impacted bone allografts were used for the management of American Academy of Orthopaedic Surgeons Type II defects in 17 hips, and bulk bone allografts together with impacted allografts were used for the management of Type III defects in 23 hips. A total of five hips showed radiological failure at a mean follow-up of 6.7 years (4.5 to 9.3), two of which were infected. The mean pre-operative Merle d’Aubigné score was 10 (5 to 15) This clinical study indicates that revision total hip replacement using the Kerboull-type acetabular reinforcement device with bone allografts yielded satisfactory mid-term results.
Although the importance of lateral femoral wall
integrity is increasingly being recognised in the treatment of intertrochanteric
fracture, little attention has been put on the development of a
secondary post-operative fracture of the lateral wall. Patients
with post-operative fractures of the lateral wall were reported
to have high rates of re-operation and complication. To date, no
predictors of post-operative lateral wall fracture have been reported.
In this study, we investigated the reliability of lateral wall thickness
as a predictor of lateral wall fracture after dynamic hip screw
(DHS) implantation. A total of 208 patients with AO/OTA 31-A1 and -A2 classified
intertrochanteric fractures who received internal fixation with
a DHS between January 2003 and May 2012 were reviewed. There were
103 men and 150 women with a mean age at operation of
78 years (33 to 94). The mean follow-up was 23 months (6 to 83).
The right side was affected in 97 patients and the left side in
111. Clinical information including age, gender, side, fracture
classification, tip–apex distance, follow-up time, lateral wall
thickness and outcome were recorded and used in the statistical
analysis. Fracture classification and lateral wall thickness significantly
contributed to post-operative lateral wall fracture (both p <
0.001). The lateral wall thickness threshold value for risk of developing
a secondary lateral wall fracture was found to be 20.5 mm. To our knowledge, this is the first study to investigate the
risk factors of post-operative lateral wall fracture in intertrochanteric
fracture. We found that lateral wall thickness was a reliable predictor
of post-operative lateral wall fracture and conclude that intertrochanteric
fractures with a lateral wall thickness <
20.5 mm should not
be treated with DHS alone. Cite this article:
We aimed to further evaluate the biomechanical characteristics
of two locking screws Synthetic tubular bone models representing normal bone density
and osteoporotic bone density were used. Artificial fracture gaps
of 1 cm were created in each specimen before fixation with one of
two constructs: 1) two locking screws using a five-hole locking
compression plate (LCP) plate; or 2) three non-locking screws with
a seven-hole LCP plate across each side of the fracture gap. The
stiffness, maximum displacement, mode of failure and number of cycles
to failure were recorded under progressive cyclic torsional and
eccentric axial loading.Objectives
Methods
While an increasing amount of arthroplasty articles
report comorbidity measures, none have been validated for outcomes.
In this study, we compared commonly used International Classification
of Diseases-based comorbidity measures with re-operation rates after
total hip replacement (THR). Scores used included the Charlson,
the Royal College of Surgeons Charlson, and the Elixhauser comorbidity
score. We identified a nationwide cohort of 134 423 THRs from the
Swedish Hip Arthroplasty Register. Re-operations were registered
post-operatively for up to 12 years. The hazard ratio was estimated
by Cox’s proportional hazards regression, and we used C-statistics
to assess each measure’s ability to predict re-operation. Confounding
variables were age, gender, type of implant fixation, hospital category,
hospital implant volume and year of surgery. In the first two years only the Elixhauser score showed any significant
relationship with increased risk of re-operation, with increased
scores for both one to two and three or more comorbidities. However,
the predictive C-statistic in this period for the Elixhauser score
was poor (0.52). None of the measures proved to be of any value between
two and 12 years. They might be of value in large cohort or registry
studies, but not for the individual patient. Cite this article:
We reviewed 91 patients (103 feet) who underwent
a Ludloff osteotomy combined with additional procedures. According
to the combined procedures performed, patients were divided into
Group I (31 feet; first web space release), Group II (35 feet; Akin
osteotomy and trans-articular release), or Group III (37 feet; Akin
osteotomy, supplementary axial Kirschner (K-) wire fixation, and
trans-articular release). Each group was then further subdivided
into severe and moderate deformities. The mean hallux valgus angle correction of Group II was significantly
greater than that of Group I (p = 0.001). The mean intermetatarsal
angle correction of Group III was significantly greater than that
of Group II (p <
0.001). In severe deformities, post-operative
incongruity of the first metatarsophalangeal joint was least common
in Group I (p = 0.026). Akin osteotomy significantly increased correction
of the hallux valgus angle, while a supplementary K-wire significantly
reduced the later loss of intermetatarsal angle correction. First
web space release can be recommended for severe deformity. Additionally,
K-wire fixation (odds ratio (OR) 5.05 (95% confidence interval (CI) 1.21
to 24.39); p = 0.032) and the pre-operative hallux valgus angle
(OR 2.20 (95% CI 1.11 to 4.73); p = 0.001) were shown to be factors
affecting recurrence of hallux valgus after Ludloff osteotomy. Cite this article:
We report the effectiveness of revision of total
elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows
in 52 patients were treated with re-cementing of a total elbow replacement
into part or all of the existing cement mantle or into the debrided
host-bone interface, without the use of structural bone augmentation
or a custom prosthesis. The original implant revision was still
Options for the treatment of subcapital femoral
neck fractures basically fall into two categories: internal fixation
or arthroplasty (either hemiarthroplasty or total hip arthroplasty).
Historically, the treatment option has been driven by a diagnosis-related approach
(non-displaced neck fractures versus displaced neck fractures).
More recently, the traditional paradigm has changed. Instead of
a diagnosis-related approach, it has become more of a patient-related
approach. Treatment options take in to consideration the patient’s age,
functional demands, and individual risk profile. A simple algorithm
can be helpful in terms of directing the treatment. Non-displaced
fractures, regardless of age of the patient, should be treated with
closed reduction and internal fixation. For displaced femoral neck fractures,
the treatment differs depending on the age of the patient. The younger
patient should be treated with urgent ORIF with the goal of an anatomic
reduction. For displaced femoral neck fractures in the elderly,
cognitive function should be determined. For those who are cognitively
functioning, total hip arthroplasty appears to be the best option.
In the cognitively dysfunctional, a bipolar hemiarthroplasty or
a total hip arthroplasty with use of larger heads (32 mm or 36 mm)
and/or constrained sockets are a viable option.
The popularity of cementless total hip replacement
(THR) has surpassed cemented THR in England and Wales. This retrospective
cohort study records survival time to revision following primary
cementless THR with the most common combination (accounting for
almost a third of all cementless THRs), and explores risk factors independently
associated with failure, using data from the National Joint Registry
for England and Wales. Patients with osteoarthritis who had a DePuy
Corail/Pinnacle THR implanted between the establishment of the registry
in 2003 and 31 December 2010 were included within analyses. There
were 35 386 procedures. Cox proportional hazard models were used
to analyse the extent to which the risk of revision was related
to patient, surgeon and implant covariates. The overall rate of
revision at five years was 2.4% (99% confidence interval 2.02 to
2.79). In the final adjusted model, we found that the risk of revision
was significantly higher in patients receiving metal-on-metal (MoM:
hazard ratio (HR) 1.93, p <
0.001) and ceramic-on-ceramic bearings
(CoC: HR 1.55, p = 0.003) compared with the best performing bearing
(metal-on-polyethylene). The risk of revision was also greater for
smaller femoral stems (sizes 8 to 10: HR 1.82, p <
0.001) compared
with mid-range sizes. In a secondary analysis of only patients where body
mass index (BMI) data were available (n = 17 166), BMI ≥ 30 kg/m2 significantly
increased the risk of revision (HR 1.55, p = 0.002). The influence
of the bearing on the risk of revision remained significant (MoM:
HR 2.19, p <
0.001; CoC: HR 2.09,
p = 0.001). The risk of revision was independent of age, gender,
head size and offset, shell, liner and stem type, and surgeon characteristics. We found significant differences in failure between bearing surfaces
and femoral stem size after adjustment for a range of covariates
in a large cohort of single-brand cementless THRs. In this study
of procedures performed since 2003, hard bearings had significantly
higher rates of revision, but we found no evidence that head size
had an effect. Patient characteristics, such as BMI and American
Society of Anesthesiologists grade, also influence the survival
of cementless components. Cite this article:
The purpose of this paper is to discuss the risk
factors, prevention strategies, classification, and treatment of
intra-operative femur fractures sustained during primary and revision
total hip arthroplasty.
A moderator and panel of five experts led an
interactive session in discussing five challenging and interesting patient
case presentations involving surgery of the hip. The hip pathologies
reviewed included failed open reduction internal fixation of subcapital
femoral neck fracture, bilateral hip disease, evaluation of pain
after metal-on-metal hip arthroplasty, avascular necrosis, aseptic
loosening secondary to osteolysis and polyethylene wear, and management
of ceramic femoral head fracture.