It is accepted that resurfacing hip replacement
preserves the bone mineral density (BMD) of the femur better than total
hip replacement (THR). However, no studies have investigated any
possible difference on the acetabular side. Between April 2007 and March 2009, 39 patients were randomised
into two groups to receive either a resurfacing or a THR and were
followed for two years. One patient’s resurfacing subsequently failed,
leaving 19 patients in each group. Resurfaced replacements maintained proximal femoral BMD and,
compared with THR, had an increased bone mineral density in Gruen
zones 2, 3, 6, and particularly zone 7, with a gain of 7.5% (95%
confidence interval (CI) 2.6 to 12.5) compared with a loss of 14.6%
(95% CI 7.6 to 21.6). Resurfacing replacements maintained the BMD
of the medial femoral neck and increased that in the lateral zones
between 12.8% (95% CI 4.3 to 21.4) and 25.9% (95% CI 7.1 to 44.6). On the acetabular side, BMD was similar in every zone at each
point in time. The mean BMD of all acetabular regions in the resurfaced
group was reduced to 96.2% (95% CI 93.7 to 98.6) and for the total
hip replacement group to 97.6% (95% CI 93.7 to 101.5) (p = 0.4863).
A mean total loss of 3.7% (95% CI 1.0 to 6.5) and 4.9% (95% CI 0.8
to 9.0) of BMD was found above the
Instability is a common cause of failure after total hip arthroplasty. A novel reverse total hip has been developed, with a femoral cup and acetabular ball, creating enhanced mechanical stability. The purpose of this study was to assess the implant fixation using radiostereometric analysis (RSA), and the clinical safety and efficacy of this novel design. Patients with end-stage osteoarthritis were enrolled in a prospective cohort at a single centre. The cohort consisted of 11 females and 11 males with mean age of 70.6 years (SD 3.5) and BMI of 31.0 kg/m2 (SD 5.7). Implant fixation was evaluated using RSA as well as Western Ontario and McMaster Universities Osteoarthritis Index, Harris Hip Score, Oxford Hip Score, Hip disability and Osteoarthritis Outcome Score, 38-item Short Form survey, and EuroQol five-dimension health questionnaire scores at two-year follow-up. At least one acetabular screw was used in all cases. RSA markers were inserted into the innominate bone and proximal femur with imaging at six weeks (baseline) and six, 12, and 24 months. Independent-samples Aims
Methods
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods
The use of a porous metal shell supported by two augments with the ‘footing’ technique is one solution to manage Paprosky IIIB acetabular defects in revision total hip arthroplasty. The aim of this study was to assess the medium-term implant survival and radiological and clinical outcomes of this technique. We undertook a retrospective, two-centre series of 39 hips in 39 patients (15 male, 24 female) treated with the ‘footing’ technique for Paprosky IIIB acetabular defects between 2007 and 2020. The median age at the time of surgery was 64.4 years (interquartile range (IQR) 54.4 to 71.0). The median follow-up was 3.9 years (IQR 3.1 to 7.0).Aims
Methods
We present a series of 30 uncemented total hip replacements performed between June 1985 and January 2002 with a mean follow-up of seven years (5 to 20) in 27 patients who had previously undergone a valgus intertrochanteric osteotomy. No further osteotomy was undertaken to enable hip replacement. We used a number of uncemented modular or monoblock femoral
Pelvic discontinuity is a challenging acetabular defect without a consensus on surgical management. Cup-cage reconstruction is an increasingly used treatment strategy. The present study evaluated implant survival, clinical and radiological outcomes, and complications associated with the cup-cage construct. We included 53 cup-cage construct (51 patients) implants used for hip revision procedures for pelvic discontinuity between January 2003 and January 2022 in this retrospective review. Mean age at surgery was 71.8 years (50.0 to 92.0; SD 10.3), 43/53 (81.1%) were female, and mean follow-up was 6.4 years (0.02 to 20.0; SD 4.6). Patients were implanted with a Trabecular Metal Revision Shell with either a ZCA cage (n = 12) or a TMARS cage (n = 40, all Zimmer Biomet). Pelvic discontinuity was diagnosed on preoperative radiographs and/or intraoperatively. Kaplan-Meier survival analysis was performed, with failure defined as revision of the cup-cage reconstruction.Aims
Methods
Between April 1992 and July 2005, 310 posterior lip augmentation devices were used for the treatment of recurrent dislocation of the hip in 307 patients who had received primary total hip replacements (THRs) using Charnley/Charnley Elite components with a cemented acetabulum. The mean number of dislocations before stabilisation with the device was five (1 to 16) with a mean time to this intervention from the first dislocation of 3.8 years (0 days to 22.5 years). The mean age of the patients at this reconstruction was 75.4 years (39 to 96). A retrospective clinical and radiological review was carried out at a mean follow-up of six years and nine months (4.4 months to 13 years and 7 months). Of the 307 patients, 53 had died at the time of the latest review, with a functioning THR and with the posterior lip augmentation device in situ. There were four revisions (1.3%), one for pain, two for deep infection and one for loosening of the
Three-dimensional surface models of the normal hemipelvis derived from volumetric CT data on 42 patients were used to determine the radius, depth and orientation of the native acetabulum. A sphere fitted to the lunate surface and a plane matched to the acetabular rim were used to calculate the radius, depth and anatomical orientation of the acetabulum. For the 22 females the mean acetabular abduction, anteversion, radius and normalised depth were 57.1° (50.7° to 66.8°), 24.1° (14.0° to 33.3°), 25 mm (21.7 to 30.3) and 0.79 mm (0.56 to 1.04), respectively. The same parameters for the 20 males were 55.5° (47.7° to 65.9°), 19.3° (8.5° to 32.3°), 26.7 mm (24.5 to 28.7) and 0.85 mm (0.65 to 0.99), respectively. The orientation of the native acetabulum did not match the safe zone for
In metal-on-polyethylene (MoP) total hip arthroplasty (THA), large metal femoral heads have been used to increase stability and reduce the risk of dislocation. The increased size of the femoral head can, however, lead to increased taper corrosion, with the release of metal ions and adverse reactions. The aim of this study was to investigate the relationship between the size of the femoral head and the levels of metal ions in the blood in these patients. A total of 96 patients were enrolled at two centres and randomized to undergo MoP THA using either a 32 mm metal head or a femoral head of between 36 mm and 44 mm in size, being the largest possible to fit the thinnest available polyethylene insert. The levels of metal ions and patient-reported outcome measures (Oxford Hip Score, University of California, Los Angeles Activity Scale) were recorded at two and five years postoperatively.Aims
Methods
For rare cases when a tumour infiltrates into the hip joint, extra-articular resection is required to obtain a safe margin. Endoprosthetic reconstruction following tumour resection can effectively ensure local control and improve postoperative function. However, maximizing bone preservation without compromising surgical margin remains a challenge for surgeons due to the complexity of the procedure. The purpose of the current study was to report clinical outcomes of patients who underwent extra-articular resection of the hip joint using a custom-made osteotomy guide and 3D-printed endoprosthesis. We reviewed 15 patients over a five-year period (January 2017 to December 2022) who had undergone extra-articular resection of the hip joint due to malignant tumour using a custom-made osteotomy guide and 3D-printed endoprosthesis. Each of the 15 patients had a single lesion, with six originating from the acetabulum side and nine from the proximal femur. All patients had their posterior column preserved according to the surgical plan.Aims
Methods
Aims. The purpose of this retrospective study was to evaluate the minimum
five-year outcome of revision total hip arthroplasty (THA) using
the Kerboull acetabular reinforcement device (KARD) in patients
with Paprosky type III acetabular defects and destruction of the
inferior margin of the acetabulum. Patients and Methods. We identified 36 patients (37 hips) who underwent revision THA
under these circumstances using the KARD, fresh frozen allograft
femoral heads, and reconstruction of the inferior margin of the
acetabulum. The Merle d’Aubigné system was used for clinical assessment.
Serial anteroposterior pelvic radiographs were used to assess migration
of the
Aims. The classical longitudinal incision used for the direct anterior
approach (DAA) to the hip does not follow the tension lines of the
skin and can lead to impaired wound healing and poor cosmesis. The
purpose of this retrospective study was to determine the satisfaction
with the scar, and functional and radiographic outcomes comparing
the classic longitudinal incision with a modified skin crease ‘bikini’
when the DAA is used for total hip arthroplasty (THA). Patients and Methods. A total of 964 patients (51% female; 59% longitudinal, 41% ‘bikini’)
completed a follow-up questionnaire between two and four years postoperatively,
including the Oxford Hip Score (OHS), the University of North Carolina
‘4P’ scar scale (UNC4P) and two items for assessing the aesthetic
appearance of the scar and symptoms of numbness. The positioning
of the components, rates of heterotopic ossification (HO) and rates
of revision were assessed. Results. The mean OHS was similar in both groups (p = 0.41). The mean
UNC4P total score was slightly better (p = 0.01) and the proportion
of patients who were very satisfied with the cosmetic aspects of
the scar was higher in the ‘bikini’ group (p < 0.001). The proportion
of patients reporting numbness in the scar was higher in the longitudinal
group (14.5% vs 7.5%, respectively, p < 0.001).
The abduction angle of the
The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate. A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.Aims
Methods
Precise implant positioning, tailored to individual spinopelvic biomechanics and phenotype, is paramount for stability in total hip arthroplasty (THA). Despite a few studies on instability prediction, there is a notable gap in research utilizing artificial intelligence (AI). The objective of our pilot study was to evaluate the feasibility of developing an AI algorithm tailored to individual spinopelvic mechanics and patient phenotype for predicting impingement. This international, multicentre prospective cohort study across two centres encompassed 157 adults undergoing primary robotic arm-assisted THA. Impingement during specific flexion and extension stances was identified using the virtual range of motion (ROM) tool of the robotic software. The primary AI model, the Light Gradient-Boosting Machine (LGBM), used tabular data to predict impingement presence, direction (flexion or extension), and type. A secondary model integrating tabular data with plain anteroposterior pelvis radiographs was evaluated to assess for any potential enhancement in prediction accuracy.Aims
Methods
Large bone defects resulting from osteolysis, fractures, osteomyelitis, or metastases pose significant challenges in acetabular reconstruction for total hip arthroplasty. This study aimed to evaluate the survival and radiological outcomes of an acetabular reconstruction technique in patients at high risk of reconstruction failure (i.e. periprosthetic joint infection (PJI), poor bone stock, immunosuppressed patients), referred to as Hip Reconstruction In Situ with Screws and Cement (HiRISC). This involves a polyethylene liner embedded in cement-filled bone defects reinforced with screws and/or plates for enhanced fixation. A retrospective chart review of 59 consecutive acetabular reconstructions was performed by four surgeons in a single institution from 18 October 2018 to 5 January 2023. Cases were classified based on the Paprosky classification, excluding type 1 cases (n = 26) and including types 2 or 3 for analysis (n = 33). Radiological loosening was evaluated by an orthopaedic surgeon who was not the operating surgeon, by comparing the immediate postoperative radiographs with the ones at latest follow-up. Mean follow-up was 557 days (SD 441; 31 to 1,707).Aims
Methods
Dual-mobility (DM) components are increasingly used to prevent and treat dislocation after total hip arthroplasty (THA). Intraprosthetic dissociation (IPD) is a rare complication of DM that is believed to have decreased with contemporary implants. This study aimed to report incidence, treatment, and outcomes of contemporary DM IPD. A total of 1,453 DM components were implanted at a single academic institution between January 2010 and December 2021: 695 in primary and 758 in revision THA. Of these, 49 presented with a dislocation of the large DM head and five presented with an IPD. At the time of closed reduction of the large DM dislocation, six additional IPDs occurred. The mean age was 64 years (SD 9.6), 54.5% were female (n = 6), and mean follow-up was 4.2 years (SD 1.8). Of the 11 IPDs, seven had a history of instability, five had abductor insufficiency, four had prior lumbar fusion, and two were conversions for failed fracture management.Aims
Methods
Aims. Few reconstructive techniques are available for patients requiring
complex acetabular revisions such as those involving Paprosky type
2C, 3A and 3B deficiencies and pelvic discontinuity. Our aim was
to describe the development of the patient specific Triflange acetabular
component for use in these patients, the surgical technique and
mid-term results. We include a description of the pre-operative
CT scanning, the construction of a model, operative planning, and
surgical technique. All implants were coated with porous plasma
spray and hydroxyapatite if desired. Patients and Methods. A multicentre, retrospective review of 95 complex acetabular
reconstructions in 94 patients was performed. A total of 61 (64.2%)
were female. The mean age of the patients was 66 (38 to 85). The
mean body mass index was 29 kg/m. 2. (18 to 51). Outcome
was reported using the Harris Hip Score (HHS), complications, failures
and survival. Results. The mean follow-up was 3.5 years (1 to 11). The mean HHS improved
from 46 (15 to 90)
pre-operatively to 75 (14 to 100). A total of 21 hips (22%) had
at least one complication with some having more than one; including
dislocation (6%), infection (6%), and femoral complications (2%).
The implant was subsequently removed in five hips (5%), only one
for suspected aseptic loosening. Conclusion. The Triflange patient specific
Total femoral arthroplasty (TFA) is a rare procedure used in cases of significant femoral bone loss, commonly from cancer, infection, and trauma. Low patient numbers have resulted in limited published work on long-term outcomes, and even less regarding TFA undertaken for non-oncological indications. The aim of this study was to evaluate the long-term clinical outcomes of all TFAs in our unit. Data were collected retrospectively from a large tertiary referral revision arthroplasty unit’s database. Inclusion criteria included all patients who underwent TFA in our unit. Preoperative demographics, operative factors, and short- and long-term outcomes were collected for analysis. Outcome was defined using the Musculoskeletal Infection Society (MSIS) outcome reporting tool.Aims
Methods
Ultra-high-molecular-weight polyethylene sterilised in the absence of air and highly cross-linked polyethylene have been used to avoid osteolysis and loosening in total hip replacement. Our prospective randomised study has assessed the results using two different polyethylenes associated with the same prosthetic design. We assessed 45 Allofit