Stems improve the mechanical stability of tibial
components in total knee replacement (TKR), but come at a cost of stress
shielding along their length. Their advantages include resistance
to shear, reduced tibial lift-off and increased stability by reducing
micromotion. Longer stems may have disadvantages including stress
shielding along the length of the stem with associated reduction
in bone density and a theoretical risk of subsidence and loosening, peri-prosthetic
fracture and end-of-stem pain. These features make long stems unattractive
in the primary TKR setting, but often desirable in revision surgery
with bone loss and instability. In the revision scenario, stems
are beneficial in order to convey structural stability to the construct
and protect the reconstruction of bony defects. Cemented and uncemented
long stemmed implants have different roles depending on the nature
of the bone loss involved. This review discusses the biomechanics of the design of tibial
components and stems to inform the selection of the component and
the technique of implantation.
The Oxford hip and knee scores (OHS and OKS)
are validated patient-reported outcome measures used in patients undergoing
total hip replacement (THR), hip resurfacing (HR), total knee replacement
(TKR) and unicompartmental knee replacement (UKR). We analysed the
absolute OHS and OKS and change in scores following THR, HR, TKR,
and UKR performed at one specialist centre. All patients undergoing
and completing at least one Oxford score were eligible for inclusion
in the study which included 27 950 OHS and 19 750 OKS in 13 682
patients. Data were analysed using non-linear quantile regression.
The median absolute Oxford scores for THR, HR, TKR and UKR were
pre-operative 68.8% (15.0/48), 58.3% (20.0/48), 66.7% (16.0/48),
60.4% (19.0/48) respectively: and post-operative asymptote was 14.6%
(41.0/48), 5.8% (45.2/48), 31.2% (33.0/48), 29.2% (34.0/48). The
median asymptotic change from the pre-operative score for THR, HR,
TKR and UKR were 47.9% (23.0/48), 47.9% (23.0/48), 33.3% (16.0/48)
and 32.4% (15.5/48), respectively. The median time at which no further
appreciable change in score was achieved post-operatively was 0.7
years for THR, 1.1 years for HR, 0.9 years for TKR and 1.1 years
for UKR. The curves produced from this analysis could be used to educate
patients, and to audit the performance of a surgeon and an institution.
The time to achieve a stable improvement in outcome varied between
different types of joint replacement, which may have implications
for the timing of post-operative review. Cite this article:
Aims. The extended wait that most patients are now experiencing for hip and
We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article:
We report ten-year clinical and radiological
follow-up data for the Sigma Press Fit Condylar total knee replacement system
(Sigma PFC TKR). Between October 1998 and October 1999 a total of
235 consecutive PFC Sigma TKRs were carried out in 203 patients.
Patients were seen at a specialist nurse-led clinic seven to ten
days before admission and at six and 18 months, three, five and
eight to ten years after surgery. Data were recorded prospectively
at each clinic visit. Radiographs were obtained at the five- and
eight- to ten-year follow-up appointments. Of the 203 patients,
147 (171 knees) were alive at ten years and 12 were lost to follow-up.
A total of eight knees (3.4%) were revised, five for infection and
three to change the polyethylene insert. The survival at ten years
with an endpoint of revision for any reason was 95.9%, and with
an endpoint of revision for aseptic failure was 98.7%. The mean
American Knee Society Score (AKSS) was 79 (10 to 99) at eight to
ten years, compared with 31 (2 to 62) pre-operatively. Of 109 knee
with radiographs reviewed, 47 knees had radiolucent lines but none
showed evidence of loosening. Cite this article:
We have studied prospectively the outcome of wound discharge in patients after arthroplasty of the hip and knee. Over a period of 3.5 years 530 primary arthroplasties were carried out in one hospital. Postoperative wound infections developed in 82. At a mean follow-up of two years a comparison was made between these patients and 82 with healthy wounds, in terms of symptoms and signs of deep infection. There was an incidence of 1.1% of early deep infection, within six weeks in all cases. The rate of ‘superficial’ infection was 17.3% in the hips, 10.5% in the knees and 14.3% in total. At a mean follow-up of 26 months, there were no significant differences between the patients with infected wounds and a matched group of patients with healthy wounds in terms of the ESR, level of C-reactive protein, white cell count and radiological scores, but clinical scores were significantly worse in the patients with infected knees (p <
0.05). The length of stay was also significantly longer in this group (mean 14.6 days in the healthy wound group, 19 days in the problem group; p <
0.005). There was, however, no convincing evidence that these wound infections led to deep infection and early revision in the early to medium follow-up period. A larger and longer prospective trial would be necessary to shed more light on this potential problem.
Smart trials are total knee tibial trial liners
with load bearing and alignment sensors that will graphically show quantitative
compartment load-bearing forces and component track patterns. These
values will demonstrate asymmetrical ligament balancing and misalignments
with the medial retinaculum temporarily closed. Currently surgeons
use feel and visual estimation of imbalance to assess soft-tissue
balancing and tracking with the medial retinaculum open, which results
in lower medial compartment loads and a wider anteroposterior tibial
tracking pattern. The sensor trial will aid the total knee replacement
surgeon in performing soft-tissue balancing by providing quantitative
visual feedback of changes in forces while performing the releases
incrementally. Initial experience using a smart tibial trial is
presented.
A series of 19 patients with severe rheumatoid arthritis had replacement of both hips and both knees. They are reviewed after a mean follow-up of 27 months. The preferred programme is to replace hips before knees. Pain was relieved in all the patients and function was improved in all but two. Severe rheumatoid disease and prolonged immobilisation before the operations were not contra-indications to a successful outcome, but the presence or the development of cervical myelopathy combined with gross upper limb deformity militated against a good result.
The December 2023 Hip & Pelvis Roundup. 360. looks at: Early hip fracture surgery is safe for patients on direct oral anticoagulants; Time to return to work by occupational class after total hip or
Peri-prosthetic patellar fracture following resurfacing
as part of total knee replacement (TKR) is an infrequent yet challenging
complication. This case-control study was performed to identify
clinical, radiological and surgical factors that increase the risk
of developing a spontaneous patellar fracture after TKR. Patellar
fractures were identified in 74 patients (88 knees) from a series
of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding
those with a previous history of extensor mechanism realignment
or a clear traumatic event, a metal-backed patella, any uncemented
component or subsequent infection, the remaining 64 fractures were
compared with a matched group of TKRs with an excellent outcome
defined by the Knee Society score. The mean age of patients with
a fracture was 70 years (51 to 81) at the time of TKR. Patellar
fractures were detected at a mean of 13.4 months (2 to 84) after
surgery. The incidence of patellar fracture was found to be strongly
associated with the number of previous knee operations, greater
pre-operative mechanical malalignment, smaller post-operative patellar
tendon length, thinner post-resection patellar thickness, and a
lower post-operative Insall-Salvati ratio. An understanding of the risk factors associated with spontaneous
patellar fracture following TKR provides a valuable insight into
prevention of this challenging complication.
The April 2024 Knee Roundup. 360. looks at: Challenging the status quo: re-evaluating the impact of obesity on unicompartmental knee arthroplasty outcomes; Timing matters: the link between ACL reconstruction delays and cartilage damage; Custom fit or off the shelf: evaluating patient outcomes in tailored versus standard
Arthrodesis of the knee is sometimes needed for failed total knee replacement, but fusion can be difficult to obtain. We describe a method of arthrodesis that uses the simple, inexpensive, Portsmouth external fixator. Bony union was obtained in all six patients treated with this technique. These results are compared with those obtained by other methods of arthrodesis.
Following total hip arthroplasty (THA) and total knee arthroplasty (TKR) only the ‘visible’ measured blood loss is usually known. This underestimates the ‘true’ total loss, as some loss is ‘hidden’. Correct management of blood loss should take hidden loss into account. We studied 101 THAs and 101 TKAs (with re-infusion of drained blood). Following THA, the mean total loss was 1510 ml and the hidden loss 471 ml (26%). Following TKA, the mean total loss was 1498 ml. The hidden loss was 765 ml (49%). Obesity made no difference with either operation. THA involves a small hidden loss, the total loss being 1.3 times that measured. However, following TKA, there may be substantial hidden blood loss due to bleeding into the tissues and residual blood in the joint. The true total loss can be determined by doubling the measured loss.
We carried out 60 revision procedures for failed porous coated anatomic total knee replacements in 54 patients, which were divided into two groups. The 14 knees in group I had a well-fixed femoral component at surgery which was retained, and in the 46 knees in group II both tibial and femoral components were loose and were revised using a variety of implants. Our review comprised clinical and radiological assessment. A total of 13 knees required a second revision. Six (42%) in group I failed very early (mean 2.1 years) when compared with seven (15%) in group II (mean 6.8 years). Failure was due to wear of the polyethylene insert by the abraded, retained femoral component (crude odds ratio 4.07; 95% CI 1.07 to 15.5). We recommend a complete change of primary bearing surfaces at the time of revision of an uncemented total knee replacement in order to prevent early wear of polyethylene.
The optimal timing of total knee replacement
(TKR) in patients with osteoarthritis, in relation to the severity
of disease, remains controversial. This prospective study was performed
to investigate the effect of the severity of osteoarthritis and
other commonly available pre- and post-operative clinical parameters
on the clinical outcome in a consecutive series of cemented TKRs.
A total of 176 patients who underwent unilateral TKR were included
in the study. Their mean age was 68 years (39 to 91), 63 (36%) were
male and 131 knees (74%) were classified as grade 4 on the Kellgren–Lawrence
osteoarthritis scale. A total of 154 patients (87.5%) returned for
clinical review 12 months post-operatively, at which time the outcome
was assessed using the Knee Society score. A low radiological severity of osteoarthritis was not associated
with pain 12 months post-operatively. However, it was significantly
associated with an inferior level of function (p = 0.007), implying
the need for increased focus on all possible reasons for pain in
the knee and the forms of conservative treatment which are available
for patients with lower radiological severity of osteoarthritis. Cite this article: